GALENVS magenti Viral RNA Extraction Kit User Guide

June 3, 2024
GALENVS

Viral RNA Extraction Kit
Quick Start Guide

magenti Viral RNA Extraction Kit

Lysis / Binding Buμer (800μl)
Columns 1 & 7
Wash #1 Buμer (600μl)
Columns 2 & 8μ
Wash #2 Buμer (600μl)
Columns 3 & 9
Elution Buμer (100μl)
Columns 5 & 11

200μl sample added to Lysis/ Binding Buffer| 90μl extracted and purified RNA elution| 200μl sample added to Lysis/
Binding Buffer| 90μl extracted and purified RNA elution
---|---|---|---

VRKit miQron protocol parameters

Step Name| Column| Volume (p1)| Time
(sec)| Mixing Speed
(1-10)| Dry Time
(sec)| Magnet Capture
Time (sec)
---|---|---|---|---|---|---
Lysis / Binding| 1 & 7| 800| 300| 5| 0| 180
Wash #1| 2 & 8| 600| 60| 5| 0| 120
Wash #2| 3 & 9| 600| 60| 5| 60| 120
Elution| 5 & 11| 90| 60| 5| 0| 90
Discard comb| 2& 8| 600| 0| 5| 0| 0
1. Remove the protective foil.| 2. Add up to 16 samples to Lysis/Binding Buffer (columns 1 and 7).| 3. Place plate into the miQron, taking care that the label is facing outward.
---|---|---
| |
4. Insert two combs.| 5. Select the VRKit protocol and press| Columns 5 and 11 contain the purified RNA elution.
| When program is complete, remove plate from miQron and discard combs.|