Step 1: Preparation
Remove the protective foil from the kit.

Step 2: Sample Processing

  1. Add up to 16 samples to the Lysis/Binding Buffer in columns 1 and 7 of the provided plate.
  2. Follow the VRKit miQron protocol parameters for each step:

Step Name| Column| Volume| Time (sec)| Mixing Speed| Dry Time| Magnet Capture Time (sec)
---|---|---|---|---|---|---
Lysis / Binding| 1 & 7| 800| 300| 5| 0| 180

Step 3: Elution and Completion

  1. Place the plate into the miQron, ensuring the label faces outward.
  2. Insert two combs and select the VRKit protocol.
  3. Press “Run” to start the extraction process.
  4. When the program is complete, remove the plate from miQron and discard the combs.
  5. Columns 5 and 11 will contain the purified RNA elution.

FAQ:

  • Q: How do I update the protocol on miQron?
    A: To import the updated protocol, press the Scan Protocol icon from the Run menu and scan the QR code provided.


  • Q: Can I process fewer than 16 samples?
    A: Yes, you can process fewer samples by adjusting the volume accordingly in the Lysis/Binding Buffer.


  • Q: What should I do if there is an error during the extraction process?
    A: If you encounter any errors, refer to the troubleshooting section in the user manual or contact customer support for assistance.

Overview Description

VRKit miQron protocol parameters

GALENVS-magnet-Viral-RNA-Extraction-Kit-fig- \(2\)

Using Instruction

  1. Remove the protective foil.
  2. Add up to 16 samples to Lysis/Binding Buffer (columns 1 and 7).
  3. Place plate into the miQron, taking care that the label is facing outward.
  4. Insert two combs.GALENVS-magnet-Viral-RNA-Extraction-Kit-fig- \(6\)
  5. Select the VRKit protocol and press When the program is complete, remove the plate from miQron and discard the combs.

Columns 5 and 11 contain the purified RNA elution.

Protocol Update

To import the updated protocol into the miQron, press the Scan Protocol icon from the Run menu (protocol list view window). Use the scanner on the QR code below.

GALENVS-magnet-Viral-RNA-Extraction-Kit-fig- \(10\)

VRKIT PROTOCOL V1.0

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals