GOSSEN METRAWATT Z502S PRO-AB Leakage Current Measuring Adapter Instruction Manual
- June 12, 2024
- GOSSEN METRAWATT
Table of Contents
- Scope of delivery
- Safety Precautions
- General
- Preparing for Measurement
- Performing Measurement
- Self-Test Function for Testing the Adapter
- Measuring Instrument Requirements
- Remedying Malfunctions
- Characteristic Values
- Return and Environmentally Sound Disposal
- Repair and Replacement Parts Service Calibration Center* and Rental
- Product Support
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Operating Instructions
PRO-AB
Leakage Current Measuring Adapter
3-349-621-15 4/5.23
Scope of delivery
1 Leakage current measuring adapter
1 Set of operating instructions
Safety Precautions
The leakage current measuring adapter is a measuring device in accordance with
IEC 62353 (VDE 0750, part 1). It may only be used for its intended purpose!
Permissible values for continuous leakage and patient auxiliary current must
be taken into consideration. For safety reasons, the adapter must be checked
for correct functioning before and after measurements (e.g. by means of the
self-test function included in the test instrument, or perhaps by measuring
ohmic input and output impedance).
Input impedance
With open output
- When fuse is OK: 1020 3%
- When fuse if blown: infinite
Output impedance
Range: 10:1
- With open input: 10.1 k
- With closed input: 10 k
Range: 1:1
- With open input: 11.0 k
- With closed input: 10 k
Opening of Adapter / Repair
The adapter may be opened only by authorized service personnel to ensure the
safe and correct operation of the equipment and to keep the warranty valid.
Even original spare parts may be installed only by authorized service
personnel. In case the adapter was opened by unauthorized personnel, no
warranty regarding personal safety, measurement accuracy, conformity with
applicable safety measures or any consequential damage is granted by the
manufacturer.
General
In order to measure contact voltage in accordance with DIN VDE 0107, part 10,
as well as continuous leakage and patient auxiliary current per IEC 62353 (VDE
0750, part 1) / IEC 601-1 / EN 60 6011:2006 (Medical electrical equipment
General requirements for basic safety), the PRO-AB leakage current measuring
adapter must be used as an accessory for the PROFITEST MXTRA or SECULIFE IP
test instrument.
When performing leakage current measurements, the adapter is plugged into the
measurement inputs at the left-hand side of the PROFITEST MXTRA or SECULIFE IP
(2-pole current clamp input and probe input) as shown in figure 1. As
specified in the standards listed above, current values of up to 10 mA may be
measured with this adapter. In order to be able to fully cover this measuring
range using the measurement input provided on the test instrument (2-pole
current clamp input), the measuring instrument is equipped with range
switching between transformation ratios of 10:1 and 1:1. In the 10:1 range,
voltage dividing takes place at the same ratio. The internal measurement
resistor (1.0 k) is protected by an upstream device fuse in order to prevent
overloading (see “Characteristic Values” on page 8. The adapter’s frequency
response is determined by a low-pass filter (10 k / 0.015 µF) with a cut-off
frequency of 1 kHz (20 dB attenuation).
Preparing for Measurement
Before performing measurements, the adapter’s measurement outputs must be plugged into the measurement inputs at the left-hand side of the PROFITEST MXTRA or SECULIFE IP (2-pole current clamp input and probe input).
PROFITEST MXTRA
Either of the leakage current measuring adapter’s inputs is connected to
reference earth (e.g. safe earth electrode / equipotential bonding) via a
measurement cable. The metallic housing (accessible part) of the device under
test is contacted with a test probe or alligator clip which is connected to
the other input by means of a second measurement cable.
Performing Measurement
Measurement is performed with the rotary switch in the “EXTRA” position and
with “EXTRA” measurements selected as an option in the submenu.
The leakage current measuring adapter is equipped with a 1 k measurement
resistor and a low-pass filter with a cut-off frequency of 1 kHz. The
measuring range is selected by means of a switch for the 10:1 and 1:1 ranges.
The filter function is always active and cannot be deactivated with the
current device design.
After completing the test setup as described in section 4, measurement can be
started at the test instrument.
Attention!
The test plug should be located in the storage slot during leakage current
measurement. Under no circumstances may the test plug be connected with any
system components (including PE / ground potential).
First of all, the rotary selector switch at the test instrument is set to the
“EXTRA” measuring function. The “TYPE” softkey is then pressed after which a
selection menu appears, by means of which the “Leakage Current Measurement”
sub-function is selected with the cursor keys and selection is acknowledged by
pressing the enter key. The menu screen for leakage current measurement then
appears. The measurement can be started or stopped by pressing the “START”
button. Leakage current measurement is a long-term measurement, i.e. it
continues until it is stopped by the user. The momentary measured value is
display continuously during measurement.
In order to perform a measurement, the self-test must be deactivated in the
menu (set the “TEST ON/OFF” function key to “OFF”).
Always start with the large measuring range (10:1), unless there’s no doubt
that small measured values can be expected, in which case the small measuring
range can be used (1:1). The measuring range must be selected at the measuring
adapter, as well as in the menu using the corresponding function key
(“RANGE”). It must be assured that the range settings at the adapter and at
the test instrument are always identical, in order to prevent any distortion
of measurement results.
Depending on the magnitude of the measured values, the range setting can, or
must (in the case of overranging), be manually corrected at the measuring
adapter and the test instrument.
Individual limit values can be adjusted after pressing the “Limits” function
key. Exceeded limit values are indicated by the red limit value LED at the
test instrument.
The menu includes a self-test function for testing the adapter, which can be
selected as needed via the parameter settings. For testing as well as for
performing measurements, the adapter is plugged into the sensor sockets at the
PROFITEST MXTRA or SECULIFE IP test instrument. In addition to this, the two
test probes at the 2-pole adapter equipped with test plug must be inserted
into the measurement sockets at the leakage current measuring adapter.
Polarity is irrelevant. The adapter’s status is displayed by means of an
appropriate message after the test routine has been completed (“OK”, “Blown
Fuse” or “Interruption”).
A detailed description of the test routing is included in section 6.
Self-Test Function for Testing the Adapter
A self-test which checks the adapter’s essential functions is included in the
test instrument, in order to be able to assure that the leakage current
measuring adapter is functioning correctly. During this self-test, the
integrated fuse (see “Characteristic Values” on page 8) is tested for
continuity and the measurement resistor is checked for a correct value of 1.0
k within the permissible tolerance range. Two additional tests are then
executed which check response characteristics, and in particular test for a
correct divider ratio after switchover.
The self-test is executed automatically at the test instrument, i.e. the
individual tests described above are executed sequentially and the results are
displayed after testing has been completed.
6.1 Executing the Test Function at the Instrument
The performance test for the leakage current measuring adapter executed at the
test instrument is conducted in accordance with the steps described below:
-
Plug the leakage current measuring adapter into the test instrument as shown in figure 1.
-
Use the 2-pole adapter at the test instrument (test plug insert), and plug the “PE test probe” and the “L test probe” into the two input jacks at the leakage current measuring adapter (polarity is irrelevant).
-
Within the selected measuring function (EXTRA / IL): Activate the test function by pressing the function key (TEST ON / OFF) apop-up window appears containing the message “Test 10:1”, after which the range selector switch at the leakage current measuring adapter must be set to the “10:1” position (if this is not already the case).
-
Start the test sequence by acknowledging with the “START” key at the test instrument. The test sequence is then executed automatically.
-
After about 1 second, apop-up window appears containing the message “Test 1:1”. The range selector switch at the leakage current measuring adapter must then be set to the “1:1” position, and testing is continued automatically after once again acknowledging with the “START” key (in this case in order to acknowledge that the range has been switched). In the event that an error should occur at the leakage current measuring adapter, this is also indicated by means of an appropriate pop-up window (“Blown Fuse” → replace the fuse and repeat self-testing / “Test Error” → faulty test execution or defective, unusable leakage current measuring adapter).
-
After the second test phase has been completed (after having switched to the 1:1 range), another pop-up window appears after about 1 second in which the test results are displayed (“Test OK” → test passed, leakage current measuring adapter can be used / “Test Error” → faulty test execution or defective, unusable leakage current measuring adapter, repair or replacement required).
Measuring Instrument Requirements
At the moment, this leakage current measuring adapter can only be used in combination with the PROFITEST MXTRA and SECULIFE IP from Gossen Metrawatt GmbH.
Remedying Malfunctions
In the event that the self-test conducted with the test instrument of the manual inspection of the leakage current measuring adapter reveals a defect, either the internal fuse must be replaced or the adapter must be sent to the service department for repair, depending upon the type of defect. A blown fuse can be replaced by the user. Proceed as follows in the event that this should become necessary:
- Remove the adapter from the test instrument and disconnect the measurement cables from the adapter.
- Remove the screw-on lid from fuse holder at the adapter.
- Carefully remove the fuse and insert a new one (use correct type only).
Attention!
Use the specified fuse type only (see “Characteristic Values” on page 8).
- After replacing the fuse, the leakage current measuring adapter must be tested for correct functioning as described above.
Characteristic Values
Input
Input current | 0 to 10 mA |
---|---|
Input impedance | 1020 Ω ±3% (±5% according to EN60601-1) |
Measuring impedance | 1000 Ω ±3% (±5% according to EN60601-1) |
Nominal insulation voltage | 250 V |
Fuse | GMC-I type G fuse |
5 x 20 mm, T 50 mA / 250 V
(article no. 3-578-292-01), DIN 41571-3
Output
10:1 range | Output voltage: 0 to 1 V (0.1 V/mA) |
---|---|
1:1 range | Output voltage: 0 to 10 V (1 V/mA) |
Output impedance | 10 kΩ |
Frequency Response
Frequency response per | EN 60601-1 1 |
---|---|
Low-pass filter with cut-off frequency of | kHz |
Temperature
Ambient temperature | 0 to +40 °C |
---|---|
Storage temperature | –25 to +70 °C |
Ambient Conditions
Operating temperature | 0 to + 40 °C |
---|---|
Storage temperature | –25 to + 70 °C |
Relative humidity | Max. 75%, no condensation allowed |
Elevation | Max. 2000 m |
Electromagnetic Compatibility (EMC)
Interference emission | EN 61326-1:2006, class B |
---|---|
Interference immunity | EN 61326-1:2006 |
Mechanical Design
Dimensions | W x H x D: 80 x 27 x 71 mm |
---|---|
Weight | Approx. 100 g |
Protection | Housing: IP 40 per EN 60529 /DIN VDE 0470, part 1 |
Return and Environmentally Sound Disposal
The instrument is a category 9 product (monitoring and control instrument) in
accordance with ElektroG (German electrical and electronic device law). This
device is subject to the RoHS directive. Furthermore, we make reference to the
fact that the current status in this regard can be accessed on the Internet at
www.gossenmetrawatt.com by entering the
search term WEEE.
In accordance with WEEE 2012/19/EU and ElektroG, we identify our electrical
and electronic devices with the symbol in accordance with DIN EN 50419 which
is shown at the right. Devices identified with this symbol may not be disposed
of with the trash. Please contact our service department regarding the return
of old devices (see address in section 11).
Repair and Replacement Parts Service Calibration Center* and Rental
Instrument Service
If required please contact:
GMC-I Service GmbH
Service-Center
Beuthener Str. 41
90471 Nürnberg, Germany
Phone: +49 911 817718-0
Fax: +49 911 817718-253
E-mail service@gossenmetrawatt.com
www.gmci-service.com
This address is only valid in Germany. Please contact our representatives or
subsidiaries for service in other countries.
- DAkkS Calibration laboratory for measured electrical quantities, D-K-15080-01-01, accredited in accordance with DIN EN ISO/IEC 17025:2005
Accredited quantities: direct voltage, direct current value, direct current resistance, alternating voltage, alternating current value, AC active power, AC apparent power, DC power, capacitance, frequency, temperature
Product Support
If required please contact:
Gossen Metrawatt GmbH
Product Support Hotline
Phone: +49 911 8602-0
Fax: +49 911 8602-709
E-mail: support@gossenmetrawatt.com
Edited in Germany · Subject to change, errors excepted · PDF version available
on the Internet
All trademarks, registered trademarks, logos, product names, and company names
are the property of their respective owners
Gossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg, Germany
Phone +49 911 8602-0
Fax +49 911 8602-669
E-mail info@gossenmetrawatt.com
www.gossenmetrawatt.com
References
- Servicepartner für Kalibrierungen, Reparaturen, Ersatzteile, Mietgeräte | GOSSEN METRAWATT GMC-I Service GmbH
- GOSSEN METRAWATT, GMC-Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>