Health o meter 600KL Midlevel Electronic Scales Instruction Manual
- June 1, 2024
- Health o meter
Table of Contents
- Health o meter 600KL Midlevel Electronic Scales
- FCC STATEMENT
- SPECIFICATIONS
- CERTIFICATIONS / CONNECTIVITY / DISPOSAL
- ASSEMBLY INSTRUCTIONS
- SET-UP INSTRUCTIONS
- OPERATING INSTRUCTIONS
- MAINTENANCE AND CLEANING
- CALIBRATION
- TROUBLESHOOTING
- WARRANTY
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Health o meter 600KL Midlevel Electronic Scales
PELSTAR, LLC 9500 West 55th St. McCook, IL 60525-7110 USA
- www.homscales.com
- Health o meter® Professional 600KL Health o meter® Professional 600KG © Pelstar, LLC 2021
Thank you for purchasing this Health o meter® Professional scale. Please read this manual carefully and keep it for easy reference or training.
Note: This scale has been factory calibrated, and does not require calibration before use.
- Please register your scale for warranty coverage at: www.homscales.com
- For User Instructions updates and revisions please go to: www.homscales.com
CAUTIONS AND WARNINGS
INTENDED USE
This Health o meter® Professional platform scale is intended to be used in a
professional medical environment by trained medical staff. This product was
designed to weigh patients who are safely positioned and standing on the
platform. The intended use of the built-in height rod is to measure patient
height. Do not modify the product or use it for anything other than its
intended purpose.
To prevent patient/caregiver injury or damage to your scale, please follow the instructions in this user manual very carefully.
- Do not exceed the weight capacity specified for this scale.
- For accurate weighing, this scale must be placed on a flat, stable surface.
- For accurate weighing, verify before each use the proper operation according to the procedure described in this manual.
- Do not use it in the presence of flammable or explosive materials.
- If the scale becomes damaged, it should not be operated until properly serviced.
- Do not wheel or pull the scale down stairs, doing so may damage internal parts.
- Ensure that the AC adapter and scale do not come in contact with liquids, excessive temperature, or excessive humidity.
Patient/caregiver safety
- This scale is designed for static weighing of patients only. No scale should be used for patient transfer.
- To prevent patient injury, the patient must be attended throughout the entire weighing event. Caregivers should ensure that the patient is stable and provide support as needed when getting onto and off the scale.
- Patients should wear socks or other form of lightweight foot covering during the weighing event.
When operating the scale with the AC adapter
- Operate this scale exclusively with the AC adapter that is provided (ADPT50). Use of an unspecified adapter will void the warranty and can pose a serious safety hazard.
- Before using this scale, inspect the AC adapter cord for cracking / fraying, or for broken/bent plug prongs.
- Before using this scale, ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation.
- Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source.
- Unplug the AC adapter, and carefully store both the adapter and adapter cord before moving the scale.
- Operating this scale at voltages and frequencies other than specified can damage the equipment and will void the warranty.
When operating the scale with batteries
- If the scale’s battery icon indicates that the batteries are low replace the batteries or switch to the AC adapter as soon as possible.
- When replacing/inserting batteries, be sure to use all new batteries.
FCC STATEMENT
This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 93/42/EEC (EN 55011 Class A and EN 60601-1-2). These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. The equipment generates, uses, and may radiate radio frequency energy and, if not installed and used by these instructions, may cause harmful interference to other devices in the vicinity. There is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to eliminate the interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment to an outlet on a circuit different from that to which the other device(s) is connected.
- Consult Customer Service or a field service technician for help.
SPECIFICATIONS
600KL/600KG Scale Specifications
Health o meter® Professional model 600KL/600KG uses sophisticated microprocessor technology. This precision instrument is designed to provide accurate, reliable, and repeatable weight measurements and features that make the weighing process simple, fast, and convenient. The scale is set up to determine the weight of a stationary patient. The weight can be displayed in pounds or kilograms. The 600KL/600KG features a swivel display which allows for viewing from either side of the scale plus wheels for easy mobility. The scale can be operated with a power adapter or 6-C cell batteries (not included), depending on use.
Specifications
Capacity and
resolution
| 600KL: 660 lb x 0.2 lb (300 kg x 0.1 kg)
600KG: 300 kg x 0.1 kg
---|---
Power Requirements:
| Adapter Model UES06W0CP-060050SPA or UE15WCP-060050SPA
(Health o meter Professional Model ADPT50) (INCLUDED), Input: 100-240V – 50/60Hz, 500mA
Output: 6.0V 0.5A
or 6 C-cell 1.5V batteries (not included)
Environmental:
| Operating Temperature: 10°C to 40 °C (50°F to 104°F) Storage Temperature:
0°c to 50°C (32ºF to 122ºF) Maximum Humidity: 85% RH in use or storage
Platform Dimensions:| 14 ¼” x 14 ¼” x 2 5/8” (36cm x 36cm x 6.8cm)
Assembled Dimensions:| 14 ¾” x 24 1/8” x 54 3/8”
(37.5cm x 61.3cm x 138cm)
Definition of Symbols
The scale surface, along with the height rod, (if provided) are the Type B applied parts.
CERTIFICATIONS / CONNECTIVITY / DISPOSAL
Certification Descriptions
All Health o meter® Professional scales are manufactured in FDA-registered factories.
Scale Connectivity Information:
To enable reliable transmission of weight, height, and Body Mass Index (BMI)
data, this scale is designed to connect to a computer, monitor, or other
electronic device via USB or Pelstar® wireless technology. This connectivity
helps minimize a source of potential medical errors caused by incorrectly
copying and then recording patient data. For more information regarding how to
connect the scale to other electronic devices, please contact Health o meter®
Professional Scales Technical Support, available during normal business hours
at 1-800-638-3722.
Note: If this product was purchased as a “BT” model, refer to the Wireless Communication Set Up Instructions included with the packaging.
Scale Disposal
Do not dispose of this device in domestic waste. This Health o meter®
Professional scale must be disposed of properly as electronic waste. Follow
the national, regional, or local regulations that apply to you for disposal.
For scale assistance or telephone support, please call Health o meter®
Professional Customer Service (available on normal business days) at (+1)
800-815-6615 or (+1)
708-377-0600.
ASSEMBLY INSTRUCTIONS
Before Assembly
Each 600KL/600KG Heavy Duty Eye Level Digital Scale with Integral Digital Height Rod scale is shipped disassembled in one carton. Carefully inspect the carton for shipping damage before unpacking. If damage is found, contact your shipper or a Health o meter® Professional representative immediately at 1-800-815-6615. Claims must be filed with the shipper as soon as possible after receipt of the package. The following information details what you will find inside the main carton as you unpack the parts for assembly. To prevent scratching any components, carefully remove each assembly from the carton and unwrap the packing materials. Set the carton aside for storage. To avoid damaging the scale parts when unpacking, do not use a box cutter, knife, scissors, or any sharp object to open the protective inner packaging.
Carton
- Large brown box containing:
- (1) Scale platform
- (1) Long brown box containing:
- (1) Pillar
- Small brown box containing:
- (1) Small white box containing the AC adapter (ADPT 50)
- (1) Plastic bag containing:
- (1) 5mm Allen wrench
- (4) 5mm hex screws (Allen screws)
- Medium brown box containing:
- (1) Display assembly
- (1) User Instructions
Tools Required
- 5mm Allen Wrench (included), Phillips Screwdriver (not included)
Before assembly, remove all brown cardboard in and around the pillar, base, and display head.
- Carefully remove the contents of the carton and place each part on a flat, level, and dry surface. Place the scale platform on its side and align the pillar perpendicular to the platform. The front of the display head on the pillar should be facing forward and above the scale platform.
- Locate the load cell cable connector at the bottom of the pillar. Feed the cable through the opening at the base of the platform.
- Insert the pillar into the pillar base on the platform. Insert the 4 hex screws on the underside of the platform and tighten using the supplied 5mm Allen wrench.
- Plug the load cell cable connector into the load cell plug in the platform base. The base of the scale has been designed to provide wire management for the load cell cable. Route the load cell cable as shown.
- After plugging in the load cell cable, slide the section of plastic tubing surrounding the cable upwards.
- Tuck the cable and tubing into the protective guide on the underside of the scale platform. This holds the cable in place and protects it from being damaged.
- Set the base/pillar assembly upright. Obtain the display head module. Plug the cable labeled “SCALE” into the SCALE jack in the display module. Plug the cable labeled “COM” into the COM jack in the display module.
- Place the display module on top of the pillar and align the tab on the front of the pillar with the slot at the bottom of the display module.
- Remove the screw from the back of the display module, place the module on top of the pillar, and use a Phillips screwdriver to replace and tighten the screw securing the display head to the pillar.
SET-UP INSTRUCTIONS
Set-Up
- Remove the protective film from the display window on the display head.
- Ensure the scale base is on a stable, level, and flat surface and that the sides of the base are clear of any obstacles.
- Utilizing the leveling bubble located at the top of the scale platform, level the scale by adjusting the leveling feet at each corner of the underside of the base.
Powering the Scale
-
For AC operation, the scale offers two areas to attach the AC adapter. The AC adapter can be plugged into the base of the scale and the power source. The AC adapter can also be plugged into the “DCIN” plug on the back of the display module and then into the power source.
Note: The Auto Off function is only active when the scale is being used without the AC adapter. -
For battery operation, see Battery Installation below.
Battery Installation
- Place the scale on its side. The battery holder is located in the center of the bottom of the base.
- Slide the battery cover in the direction of the “open” arrow.
- Place the 6 C-cell batteries into the battery holder. Note the orientation that is shown and ensure the batteries are placed in the proper direction. Health o meter® Professional recommends using alkaline batteries. Do not use zinc-carbon batteries.
- Replace the battery cover.
Note: When powering the scale via batteries, the AC adapter must be completely disconnected from the scale.
Warning: If the scale will not be used for some time, remove all batteries to avoid a safety hazard. Disposal of batteries should be done according to the national, regional, or local regulations which apply to your disposal of batteries.
OPERATING INSTRUCTIONS
Using the Keypad
KEY | FUNCTION |
---|---|
ON/OFF | Power scale ON and OFF. |
ZERO | Zeroes the scale before weighing. |
HOLD/RELEASE
| Holds the value of the weighted object on the display until the button is
pressed again to clear the value.
UNIT| 600KL Only: Toggles between kilograms or pounds.
▲| Scrolling method for entering weight when in Calibration mode.
WEIGHT/HEIGHT/ BMI| Toggles the display between the weight reading,
height reading, and corresponding Body Mass Index (BMI) reading.
▼
| Scrolling method for entering weight when in Calibration mode and when in
the Program Mode.
RESET| Functions to unlock height and BMI display.
SEND| Used to “lock” the height and BMI display and to send weight,
height, and BMI data.
CAUTION: To prevent patient injury, the patient must be attended to throughout the entire event.
Weighing a Patient
-
Press the ON/OFF button to power the scale on.
-
Wait until ‘0.0’ and ‘Zero’ appear on the left side of the display.
-
Ask the patient to step on the scale. Ensure they are positioned directly in the center of the platform. The scale will display the patient’s weight.
NOTE: The weight will change until the patient’s weight is stable. Once the patient is standing still, the scale will beep three times, and the word “Lock” will appear on the display. The weight may be held on the display by pressing the HOLD/RELEASE button. To release, press the HOLD/RELEASE button again. -
If desired, press the SEND button to send the weight information to the scale USB port.
NOTE: The scale will always return to the settings and units last used.
Height Measurement Instructions
- WARNING: Use caution when sliding the height rod up and down. The rod should move smoothly, do not move the rod in a jerking motion. Fold the headpiece down when not in use.
- Note: A patient needs to be a minimum of 23-5/8” (60cm) in height for the height to be accurately measured.
-
Press the WEIGHT/HEIGHT/BMI button once. The scale will display “—-” until the height rod or headpiece is moved. With the headpiece in the folded-down position, slide the headpiece up or down to the estimated height of the patient. Then raise the headpiece into its horizontal position.
Note: For patients under 44”, slide the headpiece down the pillar.
Note: When moving the headpiece, ensure the height rod and headpiece are conjoined. -
Ask the patient to step on the scale platform. Position the patient facing the display head. Ensure that the patient’s posture is upright and that the patient’s head is straight and level. Push the headpiece down until it gently rests on the patient’s head. The display will show the patient’s height. Press the RESET button to revert to the weight display.
Using the Scale for Height Measurement Only
- Press the ON/OFF button to turn the scale on.
- Wait until ‘0.0’ and ‘Zero’ appear on the left side of the display.
- Press the WEIGHT/HEIGHT/BMI button once. The scale will display the position of the height rod.
- Ask the patient to step on the scale platform. Adjust the height rod so that the headpiece gently rests on the patient’s head. Read the height.
- Press the RESET button to revert to the weight display.
To Obtain BMI
- Press the ON/OFF button to power on the scale.
- Ask the patient to step on the scale platform and obtain the patient’s weight. Press theWEIGHT/HEIGHT/BMI button once. The scale will now display the position of the height rod. While the patient remains on the platform, adjust the height rod so that the headpiece gently rests on the patient’s head. Press the WEIGHT/HEIGHT/BMI button again. The display will show the patient’s BMI.
- If desired, press the SEND button to send the weight, height, and BMI information to the scale USB port.
- Press the RESET button to revert to the weight display.
Locking the Weight Measuring Unit in KG (600KL only)
Caution: Please make sure you carefully and precisely follow the directions
below as you will be in the PROGRAM MODE of your scale. Any modification
outside of what is listed below may greatly affect the functionality of your
scale.
- Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the ZERO button once, the UNIT button twice, and then the ZERO button once again. The scale will display “CAL”.
- Press the DOWN once and “PROG” will show on the display.
- Press the SEND button and OP1-0 or OP1-1 will be displayed.
- Press the DOWN once and the display will read “OP2-0” for the KG lock function, then press the UNIT button to enable KG lock. Your display should read “OP2-1”.
- Press the DOWN until the display reads “DONE”
- Press the SEND button, the scale will read “START”, then read “0.0 Kg”
You have successfully locked your scale into KG Only.
Deactivating KG Lock (600KL only)
- Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the ZERO button once, the UNIT button twice, and then the ZERO button once again. The scale will display “CAL”.
- Press the DOWN once and “PROG” will show on the display.
- Press the SEND button and OP1-0 or OP1-1 will be displayed.
- Press the DOWN once and the display will read “OP2-1” for the KG lock function, then press the UNIT button to disable the KG lock. Your display should read “OP2-0”.
- Press the DOWN until the display reads “DONE”
- Press the SEND button, the scale will read “START”, then read “0.0”
You have successfully deactivated the KG lock on your scale.
MAINTENANCE AND CLEANING
General
This section provides instructions for maintenance, cleaning, calibration, and
troubleshooting. Maintenance operations other than those described in this
section should be performed by qualified service personnel.
- This scale contains electronic parts. Use it carefully. Do not drop, kick, or jump on it.
- Do not store chemicals, heaters, or radiators.
- Do not store the scale on its side.
Before first use and after periods of non-use, check the scale for proper operation. If the scale does not operate correctly, refer to qualified service personnel.
- Check the overall appearance of the total scale for any obvious damage, abuse, etc.
- Inspect the condition of the AC adapter cord and prongs, and replace them if needed.
Cleaning and Disinfecting
Proper care and cleaning are essential to ensure a long life of accurate and effective weighing.
Caution: Disconnect the scale from the AC adapter power source before cleaning the unit.
- Health o meter® Professional recommends using one of the following solutions on a soft cloth or disposable wipe:
- mild soap and water solution
- 70% isopropyl alcohol
- solution with 1-5% hydrogen peroxide concentration After cleaning/disinfecting, wipe with a cloth dampened with water and then with a clean dry cloth. To avoid residue build-up or damp surfaces, ensure the screen and scale parts are completely dry after cleaning.
- Never use rough or abrasive materials to clean the scale, as these will damage the scale’s finish.
- Do not submerge the scale in water or any other liquid.
- Do not pour or spray fluids directly on the scale.
Replacement Parts
Parts may be available if a piece of the scale needs to be replaced. Contact
Health or Meter Professional Scales Customer Service at
1-800-815-6615 to inquire about the
availability of these replacement parts.
Part # | Description |
---|---|
ADPT50 | POWER ADAPTER |
CALIBRATION
This scale has been factory-calibrated and does not require calibration before use. If required, the scale can be calibrated following the procedure below.
Note: Only weights certified and traceable to national standards are to be used for calibration procedures.
KL Model Calibration
KL model scales can be calibrated in pounds or kilograms. The default
calibration weight unit is pounds. If you prefer to calibrate in kilograms you
must first follow this procedure to change the calibration weight unit.
Caution: Please make sure you carefully and precisely follow the directions
below as you will be in the PROGRAM MODE of your scale. Any modification
outside of what is listed below may greatly affect the functionality of your
scale.
To Change the Calibration Weight Mode to KG
- Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the ZERO button once, the UNIT button twice, and then the ZERO button once again. The scale will display “CAL”.
- Press the DOWN once and “PROG” will show on the display.
- Press the SEND button and OP1-0 or OP1-1 will be displayed.
- Press the UNIT button to toggle between OP1-0 and OP1-1. Select OP1-0 to set your calibration weight in pounds. Select OP1-1 to set your calibration weight to kilograms.
- After making your selection, press the DOWN until the display reads “DONE”
- Press the SEND button, the scale will read “START”, followed by a version code, and then return to zero.
- Turn the scale off. Then begin the calibration procedure below.
KG Model Calibration
KG models can only be calibrated in kilograms. Follow the calibration
procedure below.
Calibration Procedure for KL and KG Models
The calibration weight must be no less than 60 lb (28 kg) and no more than 300
lb /150 kg. For optimum accuracy, Health o meter Professional Scales
recommends calibrating at 300 lb / 150 kg.
OPERATOR ACTION | DISPLAY |
---|
Turn on the scale and while “START’ is displayed, press the following keys in this sequence:
ZERO, UNIT, UNIT, ZERO
Note: You must press these keys while “START” is displayed before the display shows zero.
| ****
“CAL”
Press SEND. “Load” and “300.0” or “150.0” will appear on the display.|
“Load/300.0” or “Load/150.0”
Press SEND to calibrate with 300 lb or 150 kg. To calibrate with a
different amount, press BMI. Use the arrows to change the flashing digit,
and BMI to move to the next digit until you reach the required weight (no
less than 60lb and no more than 300lb). When the required calibration weight
is reached, press SEND.
“Clear” will appear on the display.
| ****
“CLEAR”
Clear the platform of any weight and press SEND. Dashes will appear while
the scale calibrates to zero.| “———–”
Wait without touching or shaking the scale until the zero calibration process
is finished. The display will then prompt for the calibration weight to be
loaded.| “Put XXX.X”
Load the platform with the required weight and press SEND. Dashes will
appear on the display as the scale calibrates.| “———–”
When calibration is complete “DONE/CLEAR/XXX” will appear on the display.|
“DONE/CLEAR/XXX”
Remove all weight from the platform. When all weight is cleared the scale will
reboot and “StArt” will appear on the display followed by the software version
and then return to “0.0”.| ****
“StArt”; numbers; “0.0”
TROUBLESHOOTING
Refer to the following instructions to check and correct any failure before contacting Customer Service at 1-800-815-6615.
SYMPTOM | POSSIBLE CAUSE | CORRECTIVE ACTION |
---|---|---|
The scale does not turn on | 1. Dead battery |
2. Faulty electrical outlet
3. Bad power supply (Check the LED light at the back of the display module to determine if the adapter cord is working)
| 1. Replace batteries
2. Use a different outlet
3. Replace adapter
Questionable weight or the scale does not zero| 1. External object interfering with the scale
2. The display did not show ‘0.0’ before weighing
3. Scale is not placed on a level floor
4. The scale is out of calibration
| 1. Remove interfering objects from the scale
2. Ask the patient to get off the scale, zero the scale and begin the weighing process again
3. Place the scale on a level floor and begin the weighing process again
4. Check weight with a known weight value
The display shows “OVER”| The load on the scale exceeds the capacity (660
lb/300 kg)| Remove the excess weight and use the scale according to its limits
The display shows the “LO Bat” message| The batteries are depleted| Replace
batteries according to instructions
Calibration Errors
ERROR NO. | DEFINITION | CHECK/REPLACE |
---|---|---|
Err 1 | Mathematical mistake; division by zero. Cannot calculate the | |
calibration factor | Will show when trying to calibrate with no calibration |
weight
Err 2| Weight loaded too low. Cannot calibrate| While in the calibration
procedure add more weight in the load mode
Err 3| Unstable weight. Cannot calibrate| Check the load cells’
mechanical surroundings and see nothing touches them and that the cables are
well-connected
Err 5| Calibration disabled| On the back of the display, locate the
silver extended screw and plastic tab. Use a nut driver or pliers to remove
the tab and gain access to the push button underneath. Press and release one
time to enable proper calibration. When finished, toggle the calibration
switch to enable
Err 6| High saturation state (high A/D)| The load cell cable is not
connected correctly. Check the connection, cables, and mechanical connection.
Err 7| Low saturation state (low A/D)| See Err 6
HEIGHT ROD TROUBLESHOOTING
SYMPTOM | POSSIBLE CAUSE | CORRECTIVE ACTION |
---|---|---|
Questionable height | 1. Improper connection |
2. The display shows “ __ ”
3. Height is out of tolerance
| 1. Check scale base connection, pillar connection, and display head connection – ensure cables are correctly connected
2. Check scale base connection and ensure cables are correctly connected to the display head
3. Check the 3 known height standards.
(Dimensions are in cm.) Lowest position – 60.5 +/- 0.8
Neutral position – 112.5 +/- 0.8
Highest position – 222.4 +/- 0.8
After checking the 3 known height standards and the connection – determine if the
scale needs to be replaced.
The height rod is difficult to move or immobile| 1. The pillar may still contain the cardboard within the pillar.
2. Pillar may be damaged.
| 1. Look inside the groove of the pillar to ensure the brown cardboard from shipping has been removed.
2. Contact Customer Service at 1-800-815- 6615.
WARRANTY
Limited Warranty
What does the Warranty Cover?
This Health o meter® Professional scale is warranted from the date of purchase
against defects of materials or in workmanship for two (2) years. If the
product fails to function properly, return the product, freight prepaid, and
properly packed to Pelstar, LLC (see “To Get Warranty Service”, below, for
instructions). If the manufacturer determines that a defect in material or
workmanship exists, the customer’s sole remedy will be the replacement of the
scale at no charge. Replacement will be made with a new or remanufactured
product or component. If the product is no longer available, a replacement may
be made with a similar product of equal or greater value. All replaced parts
are covered only for the original warranty period.
Who is Covered?
The original purchaser of the product must have proof of purchase to receive
warranty service. Please save your invoice or receipt. Pelstar dealers or
retail stores selling Pelstar products do not have the right to alter, modify,
or in any way change the terms and conditions of this warranty.
What is Excluded?
Your warranty does not cover normal wear of parts or damage resulting from
any of the following: negligent use or misuse of the product, use on
improper voltage or current, use contrary to the operating instructions, abuse
including tampering, damage in transit, or unauthorized repair or
alternations. Further, the warranty does not cover natural disasters, such as
fire, flood, hurricanes, and tornadoes. This warranty gives you specific legal
rights, and you may also have other rights that vary from country to country,
state to state, province to province, or jurisdiction to jurisdiction. To get
Warranty Service make sure you keep your sales receipt or document showing
proof of purchase. Call (+1)
800-638-3722 or (+1)
708-377-0600 to receive a return
authorization (RA) number, which must be included on the return label. Attach
your proof of purchase to your defective product along with your name,
address, daytime telephone number, and description of the problem. Carefully
package the product and send it with shipping and insurance prepaid to
Pelstar, LLC
- Attention R/A#
- Return Department
- 9500 West 55th Street
- McCook, IL 60525
Extended Warranty Available
This product is eligible for the Health o meter® Professional ScaleSurance
Extended Warranty Program. ScaleSurance extends the warranty period for an
additional two years. This extension to the standard Limited Warranty can be
purchased with new products or for a facility’s existing product before its
current warranty has expired. To learn more, visit
www.homscales.com/scalesurance/ or
contact your medical supply distributor.
- Not available in all countries.
- PELSTAR, LLC 9500 West 55th St. McCook, IL 60525-7110 USA 1-800-638-3722 or 1-708-377-0600 PLEASE REGISTER YOUR SCALE FOR WARRANTY COVERAGE AT: www.homscales.com
- Health o meter® is a registered trademark of Sunbeam Products, Inc. used under license. Health o meter® Professional products are manufactured, designed, and owned by Pelstar, LLC. We reserve the right to improve, enhance, or modify Health o meter® Professional product features or specifications without notice.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>