METRAHIT IM XTRA Short-Circuited Coil Tester
- Insulation resistance measurement up to 3.1 GΩ with interference voltage detection, test voltages:50, 100, 250, 500 and 1000 V per DIN EN 61557-2 / VDE 0413-2
- DAR: dielectric absorption rate, PI: polarization index
- 4-wire milliohm measurement (Kelvin connection), 200 mA or 1 A measuring current for precise measurement of extremely small resistances with a resolution of 1 μ
- 2-wire Rlo measurement with 200 mA test current per DIN EN 61557-4 / VDE 0413-4
- Short-circuited coil test with 1000 V and optional COIL adapter
- Multifunctional measuring instrument (V, A, Ω, F, Hz, %, RPM, °C/°F)
- TRMS AC / AC+DC measurement for current/voltage value up to 10/100 kHz
- Low-pass filter can be activated, 1 kHz (-3 dB) in the V AC, AC+DC range
- Direct current measurement, 10 nA to 1 A
- Current measurement with clamp sensors – transformation ratio can be adjusted with CLIP from 1:1 to 1:1000 and is taken into account in the amperage display
- Capacitance measurement
- Precision temperature measurement – °C, and °F for RTD and TC-K sensors
- Diode measurement (IK = 1 mA, UFlow up to 5.1 V) and continuity testing
- Acoustic signals
- Acquisition of min./max. values, DATA Hold
- Data logger thanks to integrated memory module and real-time clock, individual measurements as well
- Push/print function transfers measured values to software by pressing a key
- Programmable sequences for test routines
- Color graphic display
- Modular supply power: standard quick-change rechargeable lithium battery, optional WPC module for inductive charging and mains module with USB port, change without interrupting the measuring circuit thanks to touch protected module socket
- Automatic blocking sockets for the current input
- Remote probe with START (ISO) and STORE keys
- Housing with IP52 protection, dust and splash protected, replaceable rubber holster
- Interfaces: Bluetooth or WLAN integrated, USB with optional mains module
- IZYTRONIQ windows software for documentation, preparation of test reports and graphic evaluation of measurements
Applications
The METRAHIT IM XTRA and the METRAHIT IM E-DRIVE are portable, extremely
rugged multimeters designed for use in the field. They’re suitable for
maintenance, service and diagnosis at electric machines, drive units and
systems, for example in automotive, energy and automation applications.
METRAHIT IM XTRA and METRAHIT IM E-DRIVE multimeters are all-inone
instruments: insulation tester, milliohmmeter, short-circuited coil tester and
universal
multimeter. They’re ideal for safety testing and diagnosis at electric and
hybrid vehicles, as wellas all types of electric machines.
The METRAHIT IM XTRA and the METRAHIT IM E-DRIVE make it possi- ble to
test coils for short-circuits within an inductance range of 10 μH to 50 mH (at
100 Hz) in combination with the optional COILAdapter 50mH. This range
corresponds to motors in accordance with DIN standards with power ratings of
roughly 15 kVA to 80 MVA. A universal adapter for motors with medium power
ratings is in preparation.
Features
Insulation Resistance Measurement with Interference Voltage Detection
Insulation resistance measurement with test voltages of 50 to 1000 V. If
interference voltage of greater than 15 V AC or 25 V DC is detected during
insulation measurement, an error message appears briefly at the display, after
which automatic switching to TRMSAC+DC voltage measurement at 1 MΩ takes
place.
Polarization Index (PI):
When test voltage is applied, insulation resistance is measured after one
minute and after ten minutes. The polarization index is the ratio which
results from the two measured values. In the case of electric drive units, a
value of at least 2 indicates intact insulation and a value of greater than 4
indicates very good insulation.
Absorption Index (DAR)
Practically speaking, the absorption index test is a quick polarization index
measurement. The ISO values measured after 30 and 60 seconds are used to
generate a ratio.
METRAHIT IM XTRA & METRAHIT IM E-DRIVE
Isolation Tester, Milliohmmeter, TRMS Multimeter, Short-Circuited Coil
Tester
Kelvin Connection for 4-Wire Measurement (4-L) (milliohm measurement)
The 4-wire measurement compensates for influences resulting from cable and
contact resistances which must not be neglected when measuring very small
resistances. Measuring current can be set to 200 mA or 1 A. In this way, even
extremely small contact resistances can be measured, for example at welded
and riveted joints and on aircraft outer skins (lightning protection and wick
test), or equipotential bonding is measured in accordance with UN ECE R100 in
hybrid and electric vehicles.
2-Wire Rlo Measurement with 200 mA Test Current per EN 61557 / VDE 0413
Low-resistance measurement per EN 61557-4 / VDE 0413, part 4, for earth,
protective and equipotential bonding conductors.
RMS Value with Distorted Waveform
The utilized measuring method allows for waveform-independent TRMS measurement
of periodic quantities (AC) and pulsating quantities (AC and DC) for voltage
and current at up to 100 kHz.
Activatable Filter for V AC Measurement
A 1 kHz low-pass filter can be activated if required, for example when
measuring cables with parasitic external signals. The input signal is checked
by a voltage comparator for dangerous voltages as long as the low-pass filter
is activated, and these are indicated at the display if applicable.
Diode Testing with Constant Current Ik = 1 mA
Testing of the polarity of diodes and checking for short-circuits and
interruptions in electrical circuits. The test voltage source makes it
possible to measure LEDs and reference diodes up to 4.5 V, e.g. also white
LEDs.
Fast Acoustic Continuity Test Ik = 1 mA
Testing for short-circuiting or interruption in the switch position. The
threshold value for acoustic signaling can be set to 1, 10, 20, 30, 40 or 90
Ω.
Automatic/Manual Measuring Range Selection
Measured quantities are selected with the rotary switch.
The measuring range can be automatically matched to the mesured value, or
selected manually for quick, repetitive measurements.
Color Graphic Display
A high-resolution transmissive 3½” TFT color graphic display with320 x 480
dots is used for measured values and menu navigationThe display is easily
readable from all directions, as well as undedifficult lighting conditions
(controllable with light sensor). Graphic representation permits user-friendly
menu navigation including help texts.
Analog Bar Graph for Quick Trend Displays
The bar graph (with additional negative axis range for zero-frequency
quantities) permits faster detection of measured value changes as compared
with digital value displays.
Display Resolution
High resolution with 30,000 digits and a basic accuracy of 0.15%.
Automatic Storage of Measured Values
The DATA HOLD function automates the storage of measured values after they
have settled in. A patented process assures that random values are not saved
to memory in the case of rapidly changing measured quantities, but rather the
actual measured value. The stored measured value is displayed as a digital
value. The bar graph continuously indicates the momentary measured value.
Overload Protection
Overload protection safeguards the instrument in all measuring functions for
up to 1000 V. Voltages of greater than 1000 V and currents of greater than 1 A
are indicated acoustically. FUSE appears at the display if the fuse for the
current or mΩ measurement input blows.
Battery Charge Level – Power Saving Circuit
The battery charge level is accurately indicated in the graphic display.
The device is switched off automatically if the measured value remains
unchanged for a period of between 10 and 59 minutes (adjustable), if none of
the controls are activated during this time and continuous operation is not
activated.
**Automatic Blocking Sockets (ABS) ***
All current ranges are implemented via a single connector jack which prevents
any possibility of operator error. The automatic blocking sockets prevent
incorrect connection of the measurement cables, as well as selection of the
wrong measured quantity. Danger to the user, the instrument and the device
under test resulting from operator error is
thus ruled out.
- Patented (patent no. EP 1801 598 and US 7,439,725)
Housing and Protective Cover for Harsh Conditions
- New housing design
- Separate fuse compartment
- Quick-change rechargeable battery
The instrument is protected against damage in the event of impacts or dropping by means of a soft rubber cover with tilt stand. The rubber material also assures that the instrument doesn’t wander if it’s set up on a vibrating surface
Data Interfaces
The instrument can be remote configured and momentary and saved measurement
data can be read out via a bidirectional wireless interface, Bluetooth or
WLAN, or the USB port at the optional mains module. IZYTRONIQ software is
required to this end. Interface protocol and device driver software for
LabVIEW (National Instruments™)
are available upon request.
Voluntary Manufacturer’s Guarantee
36 months for materials and workmanship.
1 year for calibration.
DAkkS calibration certificate
The METRAHIT IM XTRA multimeter is furnished with a DAkkS cali-bration
certificate, which is also recognized internationally (EA, ILAC).
After the user-specified calibration interval has elapsed (recommended
interval: 1 to 3 years),the multimeter can be inexpensively recalibrated in
our own DAkkS calibration
laboratory.
Overview of Included Features
Function| METRAHIT IM XTRA (BT)
METRAHIT IM E-DRIVE (BT)
---|---
VDc (RI = 9 MS2)| •
Vpic / Hz TRMS (Ri = 9 MΩ)|
VAc+Dc TRMS (Ri = 9 MΩ)’|
VAC+DC TRMS (Ri =1 MΩ) Risc, range (interference voltage)| •
Hz OW| … 300 kHz
VAc. AC+DC bandwidth| 100 kHz
ADC, AC, AC+DC / Hz TRMS| 10 nA…1 A
Fuse| 1 A/1000 V-30 kA
Current sensor transformation ratio| 1 mV :1 610 6100 61000 mA
Hz (A AC)| … 300 kHz
Insulation resistance RISO: test voltages| 50 • 100 .250 .500 .1000 V
Short-circuited coil test (1 kV) with COIL adapter| Opfion
Duty cycle measurement as %
=| •
Speed measurement in RPM| •
Resistance Rlo with 200 mA per EN 61557 / VDE 0413| •
Milliohm with 4-wire method, mΩ with 200 mA| •
Milliohm with 4-wire method, mΩ with 1 A pulse| •
Fuse| 1 A/1000 V-30 kA
Resistance Ω| •
Continuity| •
Diode … 5.1v “| •
Temperature: °C/°F TC type K and Pt100/1000 2| •
Capacitance| •
Min-Max / data hold| •
Test sequence| 20 steps
64 MBit memory3| •
Bluetooth interface| METRAHIT IM XTRA BT
METRAHIT IM E-DRIVE BT
WIFI interface| Option
3.5″ ITT color graphic display| •
2-key Remote probe: start/stop and store| •
Quick-change battery with USB charging| •
Mains module with electrical isolation and USB charger| Option
WPC quick change battery for inductive charging| Option
Protection| IP 52
Measuring category| 1000 V CAT III, 600 V CAT IV
- Due to the system, the DC component indicated in the smallest measuring range (300 mV) has an offset. For a precise measurement of the DC component, please select measuring function VDC.
- with optional temperature sensors
- For 300,000 measured values, sampling rate adjustable from 0.1 seconds to 9hours
Standard Equipment (depending on Device Variant)
1 METRAHIT IM XTRA or METRAHIT IM E-DRIVE multimeter with rubber holster
1 Remote probe with start/stop and store/send functions
1 Type KS17-2 cable set: 1 pair of safety measurement cables with 4 mm test
tip red/black
1 Type KC4 Kelvin clip, 1 pair (only METRAHIT IM XTRA)
1 Type KC&S Kelvin clip and Kelvin probe (only METRAHIT IM EDRIVE)
1 Quick change, rechargeable lithium polymer battery with micro USB charging
socket
1 USB mains power pack (5 V DC, 2 A) with cable and micro USB charging plug
1 DAkkS calibration certificate
1 Hard case for the multimeter and accessories
1 Condensed operating instructions, German/English
– Comprehensive operating instructions in German and English available on the
Internet for download at
www.gossenmetrawatt.com
1 Card with registration key for the software
Overview Supply
Accessories | Type | Article No. | M273S | M274S |
---|---|---|---|---|
METRAHIT IM XTRA | M273D | |||
METRAHIT IM E-DRIVE BT | M274B | X | ||
Lilon module & USB charger | Z270A+ | X | X | |
Mains module with electrical isolation and USB charger | 0 | 0 | ||
Remote probe | Z270S | Z270S | X | X |
Cable set | KS17-2 | GTY362003 P0002 | X | X |
1 pair Kelvin clips | KC4 | Z227A | X | 0 |
1 pair Kelvin probes | KC27 | Z227B | 0 | 0 |
1 Kelvin clip & 1 Kelvin probe | KC&S | Z227C | — | X |
Hard case | HC40 | Z270K | X | X |
COIL Adapter 10 pH „, 50 mH | COIL Adapter 50mH | Z270F | 0 | 0 |
COIL Adapter 10 pH „, 500 mH | COIL Adapter XTRA | Z270M | 0 | 0 |
Adapter cable 4 mm male to 6 mm female | AK-4M16F | Z110L | 0 | 0 |
IZYTRONIO Business Starter Licence | S101S & Z956A | S101S & Z956A | X | X |
Legend
X = Standard
O = Option
— = not possible, not provided for
Characteristic Values
- 15 … 45 … 65 Hz … 100 kHz sinusoidal. For influence see page 5.
- At 0° … + 40 °C
- Display of up to max. 5.1 V, “OL” in excess of 5.1 V.
- Applies to measurements at film capacitors during battery operation
- Lowest measurable frequency for sinusoidal measuring signals symmetrical to the zero point
- Overload capacity of the voltage measurement input: power limiting: frequency x voltage max. 6 x 106 V x Hz at > 100 V
- Overload capacity of the current measurement input: See current measuring ranges for maximum current values.
- Input sensitivity, sinusoidal signal: 10% to 100% of the voltage or current measuring range, restriction in mV measuring range: 30% rdg. The voltage measuring ranges with max. 10 kHz apply in the A measuring range.
- Plus sensor deviation
- With ZERO function active
- Accuracy applies as from 1 % of MR; due to the TRMS converter, values < 50 digits are suppressed in the zero point.
- 10 minute cool-down period
- Required signal range: 30% to 100% of the voltage measuring range
- The integrated FF1A/1000 V fuse blows in the event of overloading
- The integrated FF0.315A/1000 V fuse blows in the event of overloading
- For measuring range 30 mΩ and 300 mΩ with function TComp active
Insulation Measurement
Measuring Range| Resolution| Nominal Voltage UISO| Intrinsic Uncer- tainty at
Reference Conditions
±(% rdg. + d)
---|---|---|---
3 … 1000 V ≅1| | Ri = 1MW| 3 + 3
300 kW| 0.1 kΩ| 50/100/250/500 V| 2 + 10
3 MW| 1 kΩ| 50/100/250/500/1000 V| 2 + 10
30 MW| 10 kΩ| 50/100/250/500/1000 V| 2 + 10
300 MW| 100 kΩ| 50/100/250/500/1000 V| 5 + 10
3000 MW| 1 MΩ| 250/500/1000 V| 5 + 10
1 TRMS interference voltage measurement (V AC+DC) with 1 MΩ input resistance, frequency response width: > 65 … 500 Hz, accuracy: 3% + 30 digits
Measuring Function| Nom. Voltage UN| Open- Circuit Voltage Uo Max.| Nom. Cur-
rent IN| Short- Circuit Cur- rent Ik| Acoustic Signal for| Overload Capacity
---|---|---|---|---|---|---
Value| Time
Uint./MW@UISO| —| —| —| —| U > 1000 V| 1000 V| Cont.
MW@UISO| 50
100| 1.2x
UIso| 1.0 mA| < 1.4 mA| U > 1000 V| 1000 V| 10 s
250
500 V
1000 V| 1.12x
UIso
Short-Circuited Coil Test (only with optional COIL adapter)
Measuring Range| Resolution| Nominal Voltage UISO| Intrinsic Uncer- tainty at
Reference Conditions ±(% rdg. + d)
---|---|---|---
0.3 … 1000 V 1| | Ri = 1MW| 3 + 30 > 100 digits
10.0 30.9 μs| 0.1 [μs]| 1000 V| 10 + 5 digits
31 … 250 μs| 1 [μs]
1 TRMS interference voltage measurement (V AC+DC) with 1 M input resistance,
frequency response width: > 65 … 500 Hz, accuracy: 3% + 30 digits
2) The time value may vary for differnt COIL adapters by up to 10 %. This has
no influence whatsoever if you perform the measurements with the same COIL
adapter and compare them with each other.
Inductance measuring ranges of optional COIL adapters:
- COIL adapter XTRA (Z270M): 10 μH up to 5 H
- COIL adapter 50mH (Z270F): 10 μH up to 50 mH
Internal Clock
Time format
Resolution
Accuracy
Temperature
influence| DD.MM.YYYY hh:mm:ss
0.1 s (measured values time stamp)
±1 minute per month
50 ppm/K
---|---
Reference Conditions
Ambient temperature
Relative humidity
Measured quantity
frequency
Measured quantity waveform Supply voltage| +23 °C 2K
40% …… 75%
45 Hz …. 65 Hz
Sinusoidal
4.0 V ±0.1 V
---|---
Influencing Quantities and Influence Error
Influencing Quantity| Sphere of Influence| Measured Quantity /
Measuring Range 1| Influence Error (…% rdg. + … d) / 10 K
---|---|---|---
Temperature| 0 °C … +21 °C and
+25 °C … +40 °C| V | 0.2 + 5
V≅| 0.4 + 5
300 W … 3 MW| 0.5 + 5
30 MW| 1 + 5
mA/A | 0.5 + 5
mA/A≅| 0.8 + 5
30 nF … 300 mF| 2 + 5
Hz| 0.2 + 5
°C/°F (Pt100/Pt1000)| 0.5 + 5
1 With zero balancing
Frequency Influence for VAC VAC+DC Voltage Ranges
| Deviation 1
---|---
Frequency Range| 300 mV range ± (… % rdg. + … d)| 3 V, 30 V, 300 V range 2 ±
(… % rdg. + … d)| 1000 V range 2 ± (… % rdg.)
15 Hz … 45 Hz| 2 + 30| 2 + 30| 2 + 30
65 Hz … 1 kHz| 0.5 + 30| 0.5 + 30| 1 + 30
1 kHz … 10 kHz| 2 + 30| 1.5 + 30| 10 + 30
10 kHz … 20 kHz| 3 + 30| 1.5 + 30| —
20 kHz … 50 kHz| 3 + 30| 5 + 30| —
50 kHz … 100 kHz| 10 + 30| 10 + 30| —
-
For sinusoidal input signals > 10% to 100% of the range (mV range: as of 30% of range, at 1% to 10% of the range: f < 50 kHz, intrinsic error increased by 0.2% of
the upper range limit. -
Overload capacity of the voltage measurement input: power limiting: frequency x voltage max. 6 x 106 V x Hz at > 100 V
Frequency Influence for IAC / IAC+DC Current Measuring Ranges
| Influence Error 1
---|---
Frequency Range| 300 mA to 300 mA ± (… % rdg. + … digits)| 1 A range ±
(… % rdg. + … digits)
15 Hz … 45 Hz| 2 + 30| 2 + 30
65 Hz … 1 kHz| 1 + 30| 1 + 30
1 kHz … 2 kHz| 1 + 30| 1 + 30
2 kHz … 5kHz| 1 + 30| 3 + 30
5 kHz … 10 kHz| 5 + 30| 5 + 30
1 For sinusoidal input signals > 10% to 100% of the range.
Influencing Quantity| Sphere of Influence| Measured Quantity / Measuring
Range| Influence Error 1
---|---|---|---
Crest Factor CF| 1 … 3| V ∼, A∼| ± 1% rdg.
3 … 5| ± 3% rdg.
1 Except for sinusoidal waveform
Influencing Quantity | Sphere of Influence | Measured Quantity | Influence Error |
---|---|---|---|
Relative Atmospheric Humidity | 75% 3 days instrument off | V, A, W, Ω F, Hz, | |
°C | 1 x intrinsic uncertainty | ||
Battery Voltage | ditto | Included in intrinsic uncertainty | |
Influencing Quantity | Sphere of Influence | Measured Qty. / Measuring Range |
Damping
---|---|---|---
Common Mode Interference Voltage
| Interference quantity max. 1000 V∼| V ****| > 90 dB
Interference quantity max. 1000 V∼ 50 Hz … 60 Hz, sinusoidal
| 3 V∼| > 90 dB
30, 300 V∼| > 150 dB
1000 V| > 150 dB
Series Mode Interference Voltage
| Interference quantity: V ∼ , respective nominal value of the measuring
range,
max. 1000 V ∼, 50 Hz … 60 Hz sinu- soidal| V ****| > 50 dB
Interference quantity max. 1000 V| V∼| > 50 dB
Response Time (after manual range selection)
Measured Quantity / Measuring Range| Digital Display Response Time| Measured
Quantity Jump Function
---|---|---
V , V∼
A , A∼| 1.5 s| From 0 to 80% of upper range limit value
300 W … 3 MW| 2s| From ¥ to 50% of upper range limit value
30 MW, MW@UISO| Max. 5 s
Continuity| < 50 ms
°C (Pt 100)| Max. 3 s
| 1.5 s
30 nF … 300 mF| Max. 5 s| From 0 to 50% of upper range limit value
10 Hz| 1.5 s
Fuse
Current measuring
ranges
& 4L mW measuring
ranges| F1: FF 1 A/1000 V AC/DC, 6.3 x 32 mm
Fuse with breaking capacity of 30 kA at 1000 V AC/DC, protects the current
measurement
input in the 300 mA to 1 A ranges
---|---
2L mW measuring ranges| F2: FF 0,315 A/1000 V
6.3 x 32 mm
Fuse with breaking capacity of 30 kA at 1000 V AC/DC
Display
TFT color graphic display (55 x 36 mm) with analog and digital display
including unit of measure, type of current and various special functions
Background Illumination
Activated background illumination can be regulated by means of a light sensor.
Analog Bar Graph
Scaling | Linear |
---|---|
Polarity display | With automatic switching |
Measuring rate | 40 measurements per second and display refresh |
Digital Measured Value Display
Resolution /char. height
Number of places| 320 x 480 dots, 12 mm
31,000 / 3100
4¾-place in the V, A, Hz and W measuring functions, depending on parameter
setting
---|---
Overflow display| “OL” is displayed for ³ 31,000 digits or ³ 3100 digits
Polarity display| “–” (minus sign) is displayed if plus pole is connected to
“⊥”
Measuring rate| 10 and 40 measurements per second with the Min-Max function
except for the capacitance, frequency and duty cycle measuring functions
Refresh Rate| 2 times per second, every 500 ms
Electrical Safety
Protection category | II per EN 61010-1:2010/VDE 04111:2011 |
---|---|
Measuring category | CAT III |
CAT IV
Nominal Voltage| 1000 V
600 V
Pollution degree| 2
Test voltage| 7.4 kV~ per EN 61010-1:2011/VDE 0411-1:2011
Power Supply
Battery module | 3.7 V, 4000 mAh, LiPo (approx. 25% self-discharge per year) |
---|---|
Service life | Approx. 20 hours (without MWISO mea- surement / RLo / R 4-wire) |
Battery indicator | Battery capacity display via battery symbol , querying of |
momentary charge level via menu function
Power OFF function| The multimeter is switched off automatically:
– When battery voltage drops to below approx. 3.6 V
– If none of the keys or the rotary switch are activated for an adjustable
duration (10 to 59 min.) and the multimeter is not in the continuous operation
mode
Rechargeable battery modules can only be recharged externally.
Measuring Function| Nominal Voltage UN| Resistance of the DUT| Service Life in
Hours| Number of Possible Mea- surements with Nominal Current per VDE 0413
---|---|---|---|---
V ****| | | 20 1|
V∼| | | 15 1|
RINS| 100 V| 1 MΩ| 5|
100 V| 100 kΩ| | 300
500 V| 500 kΩ| | 60
1000 V| 2 MΩ| | 20
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2013 class B
Interference immunity EN 61326-1:2013
Short-term measured value deviation of up to 10% may occur during
electromagnetic interference thus reducing the specified operating quality.
Ambient Conditions
Accuracy range
Operating temperatures
(Storage temperature
with batteries)| 0 °C to +40 °C
-10 °C … +50 °C
-20 °C … +50 °C with rubber holster
---|---
Storage temperatures| -25 °C … +70 °C (without battery)
Relative humidity| 40 to 75%, no condensation allowed
Elevation| To 2000 m
Deployment| Indoors, except within specified ambient conditions
Data Interface
Type | Bluetooth 4.2 |
---|---|
Frequency band | 2.402 … 2.480 GHz |
Transmitting power Functions | max. 91 mW |
– Query measuring functions and parameters
– Query momentary measurement data
Internal Measured Value Storage
Housing Dimensions| Impact resistant plastic (ABS) 235 x 105 x 56 mm (without
rubber holster)
---|---
Weight Protection| Approx. 0.7 kg with battery module Housing: IP 52 (pressure
equalization by means of the housing) Excerpt from table on the meaning of IP
Codes
IP XY (1st digit X)| Protection against foreign object entry| IP XY (2nd digit
Y)| Protection against the penetration of water
---|---|---|---
0| not protected| 0| not protected
1| ³ 50.0 mm dia.| 1| vertically falling drops
2| ³ 12.5 mm dia.| 2| vertically falling drops with enclosure tilted 15°
3| ³ 2.5 mm dia.| 3| spraying water
4| ³ 1.0 mm dia.| 4| splashing water
5| dust protected| 5| water jets
Applicable Regulations and Standards
DIN EN 61010–
VDE 0411-1| Safety requirements for electrical equipment for measurement,
control and laboratory use – Part 1: General requirements
---|---
DIN EN 61326-1
VDE 0843-20-1| Electrical equipment for measurement, control and laboratory
use
– EMC requirements – Part 1: General requirements
DIN EN 60529
VDE 0470-1| Test instruments and test procedures – degrees of protection
provided by enclosures (IP code)
DIN EN 61557-1
VDE 0413-1| Electrical safety in low voltage distribution systems up to 1000 V
a.c. and 1500 V d.c. – Equipment for testing, measuring or moni- toring of
protective measures
Part 1: General requirements
DIN EN 61557-2
VDE 0413-2| Part 2: Insulation resistance
DIN EN 61557-4
VDE 0413-4| Part 4: Resistance of earth connection and equipotential bonding
METRAHIT IM XTRA
with Accessory COIL Adapter 50mH (Z270F)
METRAHIT IM XTRA
with Accessory COIL Adapter XTRA (Z270M)
METRAHITIM XTRA & METRAHITIM E-DRIVE Isolation Tester, Milliohmmeter, TRMS
Multimeter, Short-Circuited Coil Tester
Order Information
Designation | Type | Article Number |
---|
multimeter, milliohm-meter and isolation resistance tester (COIL Ready) with
graphic display, Bluetooth, and software IZYTRONIQ Business Starter. R-ISO up
to 1kV & mW @ 200 mA 2-wire & mW @ 200 mA 4-wire & mW @ 1 A 4-wire, delivery
content comprises multimeter (M273D), push-button probe, cable set, kelvin-
clips, hard case,rechargeable lithium battery, USB wall supply, calibration
certificate, and SW licence.| METRAHIT IM XTRA BT| M273S
All-in-One Tester for electric machines, multimeter, milliohm-meter and
isolation resis- tance tester (COIL Ready) with graphic display, Bluetooth,
and software IZYTRONIQ Business Starter. R-ISO up to 1kV & mW @ 200 mA 2-wire
& mW @ 200 mA 4-wire & mW @ 1 A 4-wire, delivery content comprises multimeter
(M274B), push-button probe, cable set, each one kelvin-clip and kelvin-probe,
hard case, rechargeable lith-ium battery, USB wall supply, calibration
certificate, and SW licence.| METRAHIT IM E-DRIVE BT| M274S
Accessory cables and adapters
Cable set (1 pair of measurement cables) 1.2 m, with VDE-GS mark, 600 V CAT IV
1 A 1, 1000 V CAT III 1 A 1 1000 V CAT II 16 A 2| KS17-2| GTY3620034P0002
Cable set with 2 mm diameter steel tips and 120 cm cable, 1000 V / CAT III|
KS17-S| Z110H
Adapter cable 4 mm male to 6 mm female for the charging plug of hybrid and
electric vehicles| AK-4M/6F| Z110L
Cable set including Remote probe, clamps and US test probes (1000 V CAT II /
III 20 A)| KS-NTS| Z110W
Alligator clips (1 pair) for KS17-2 1000 V CAT III 16 A| KY95-3| Z110J
Current clamp sensor, 10 mA ¼ 100 A, 1 mV/10 mA, clamp opening: 15 mm dia.|
WZ12B| Z219B
Kelvin clips (1 set of 2 ea.) for 4-pole connection of low-resistance DUTs,
cable length: 150 cm| KC4| Z227A
Kelvin probes (1 set of 2 ea.) with double steel tips for 4-pole connection of
low-resistance DUTs| KC27| Z227B
Set including 1 Kelvin clip and 1 Kelvin probe, as well as 2 stainless steel
tips for 4-wire measurement, 120 cm cable length with 4 mm angle plugs| KC&S|
Z227C
Rechargeable lithium polymer battery M27x 14,8 Wh| Z270A| Z270A
Rechargeable lithium polymer battery M27x 14,8 Wh| Z270G| Z270G
USB+Power Module M27x| Z270E| Z270E
Charger M27x| Z270L| Z270L
Coil adapter for interturn short circuit detection at inductivities from 10 μH
to 50 mH| COIL Adapter 50mH| Z270F
Coil adapter for interturn short circuit detection at inductivities from 10 μH
to 5 H| COIL Adapter XTRA| Z270M
Probe with keys| Z270S| Z270S
AC/DC current clamp sensor, 5 mA … 30 A, 100 mV/A| CP30| Z201B
AC/DC current clamp sensor, 0.5 … 30 A, 5 … 300 A, 10 mV/A, 1 mV/A| CP330|
Z202B
AC/DC current clamp sensor, 0.5 … 100 A, 5 … 1000 A, 10 mV/A, 1 mV/A| CP1100|
Z203B
AC/DC current clamp sensor, 0.5 … 125 A, 5 … 1250 A, 10 mV/A, 1 mV/A| CP1800|
Z204A
Accessories for temperature measurement with resistance thermometer
Pt100 temperature sensor for surface and immersion measurements, –40 ¼ +600
°C| Z3409| GTZ3409000R0001
Pt1000 temperature sensor for measurement in gases and liquids,
–50 ¼ +220 °C (for servicing household appliances)| TF220| Z102A
Pt100 oven sensor, –50 ¼ +550 °C| TF550| GTZ3408000R0001
Protection and transport accessories
Hard case with foam insert and compartments for 1 METRAHIT IM XTRA or METRAHIT
IME-DRIVE and 2 batteries, as well as 2 universal compartments for
accessories.| HC40| Z270K
Replacement fuses
Fuse F1 for current measuring ranges FF1 A/1000 V AC/DC (10 pcs.)| FF1 A/1000
V AC/DC| Z109O
Fuse F2 for milliohm measuring ranges FF0,315 A/1000 V AC/DC (10 pcs.)|
FF0,315 A/1000 V AC/DC| Z109P
- With plugged on safety caps
- Without plugged on safety caps
For additional information regarding accessories please refer to:
- Measuring Instruments and Testers catalog
- www.gossenmetrawatt.com
© GMC-I Messtechnik GmbH
Edited in Germany, Subject to change without notice / Errors excepted.
All trademarks, registered trademarks, logos, product names, and company names
are property of their respective owners.
GMC-I Messtechnik GmbH
Südwestpark 15
90449 Nürnberg • Germany
Phone +49 911 8602-111
Fax +49 911 8602-777
E-Mail: info@gossenmetrawatt.com
www.gossenmetrawatt.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>