GOSSEN METRAWATT Metriso Xtra Resistance Tester Instruction Manual
- June 12, 2024
- GOSSEN METRAWATT
Table of Contents
- Switching on battery test
- Brightness Contrast
- Parameter
- Measurement
- Connection
- Test Resistance
- RINS/RLO k Ω
- RISO/RINS
- RISO/RINS → AUTO 1 / AUTO 3
- RLO (I > 200 mA !)
- Sockets
- Switching off Rechargeable batteries Fuses
- Subject to change without notice
- Customers Support
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
GOSSEN METRAWATT Metriso Xtra Resistance Tester

Switching on battery test


Brightness Contrast
Request help


Parameter
select confirm

Measurement
-
select

-
start

Connection


Test Resistance



RINS/RLO k Ω
Interference voltage control

The following measurements are possible on electrically dead devices.
External voltage disables the measurement!
RISO/RINS

RISO/RINS → AUTO 1 / AUTO 3

kΩ

RLO (I > 200 mA !)

Sockets


Switching off Rechargeable batteries Fuses

Subject to change without notice
Technical Data
| Measuring Range
---|---
RISO / RINS| U = 50, 100, 250, 500, 1000 V
RISO / RINS| Uvar = 50 … 1000 V
RISO / RINS| Uramp (U ◢ ) = 100 … 1000 V
R| 10 … 10 kW
RLO| 0,01 … 10 Ω (I ³ 200 mA !)
U| 0 … 1000 V
| Limit Values
---|---
Limit RISO / RINS| 50 k Ω @ UISO/UINS = 50 V
| 100 k Ω @ UISO/UINS = 100 V
| 500 k Ω @ UISO/UINS = 250 V
| 1 M Ω @ UISO/UINS = 500 V
| 1 M Ω @ UISO/UINS = 1000 V & @ Uvar
Limit PI| 1.0 / 1.1 / 1.5 / 2.0 / 3.0 / 4.0
Limit DAR| 1.25 / 1.6
Limit RLO| 2 W
Customers Support
Telefon+49 911 8602-0
Telefax +49 911 8602-669
E-Mail info@gossenmetrawatt.com
www.gossenmetrawatt.com

References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>