GOSSEN METRAWATT DIN VDE 0100-600 Electric Tester Instruction Manual
- June 12, 2024
- GOSSEN METRAWATT
Table of Contents
Short-form Operating InstructionsP
ROFITEST MASTER IQ Series
DIN VDE 0100-600, DIN VDE 0105-100, OVE E 8101, NIN / NIV,
IEC 60364-6, EN 50110-1
DIN VDE 0100-600 Electric Tester
Read the detailed operating instructions, available at
www.gossenmetrawatt.com.
The short-form instructions are no substitute for the detailed instructions!
Download Center
https://www.gmc- instruments.de/services/download-center/?q=&doctype=product_manual&lang=de- DE
Safety Instructions
Observe this documentation, in particular all included safety information, in
order to protect yourself and others from injury, and to prevent damage to
the instrument.
These condensed operating instructions and the detailed operating instructions
should be made available to all users.
General
- Tests/measurements may only be performed by a qualified electrician, or under the supervision and direction of a qualified electrician. The user must be instructed by a qualified electrician concerning performance and evaluation of the tests/measurements.
- Observe the five safety rules in accordance with DIN VDE 0105-100:2015-10, VDE 0105-100:2015-10, Operation of electrical installations – Part 100: General requirements (1. Shut down entirely. 2. Secure against restart. 3. Assure absence of voltage at all poles. 4. Ground and short circuit. 5. Cover neighbouring live components, or cordon them off.)
- Observe and comply with all safety regulations which are applicable for your work environment.
- Wear suitable and appropriate personal protective equipment (PPE) whenever working with the instrument.
- The functioning of active medical devices (e.g. pacemakers, defibrillators) and passive medical devices may be affected by voltages, currents and electromagnetic fields generated by the instrument and the health of their users may be impaired. Implement orresponding protective measures in consultation with the manufacturer of the medical device and your physician. If any potential risk cannot be ruled out, do not use the instrument.
Accessories
- Use only the specified accessories (included in the scope of delivery or listed as options) with the instrument.
- Carefully and completely read and adhere to the product documentation for optional accessories. Retain these documents for future reference.
Handling
- Use the instrument in undamaged condition only. Inspect the instrument before use. Pay particular attention to damage, interrupted insulation or kinked cables. Damaged components must be replaced immediately.
- Accessories and cables may only be used as long as they’re fully intact. Inspect accessories and all cables before use. Pay particular attention to damage, interrupted insulation or kinked cables.
- If the instrument or its accessories don’t function flawlessly, permanently remove the instrument/accessories from operation and secure them against inadvertent use.
- If the instrument or accessories are damaged during use, for example if they’re dropped, permanently remove the instrument/accessories from operation and secure them against inadvertent use.
- The instrument and the accessories may only be used for the tests/measurements described in the documentation for the instrument.
- The integrated voltage measuring function and mains check may not be used to test systems or system components for the absence of voltage. Testing for the absence of voltage is only permissible with a suitable voltage tester or voltage measuring system which fulfills the requirements specified in DIN EN 61243.
Operating conditions
- Do not use the instrument and its accessories after long periods of storage under unfavorable conditions (e.g. humidity, dust or extreme temperature).
- Do not use the instrument and its accessories after extraordinary stressing due to transport.
- The instrument must not be exposed to direct sunlight.
- Only use the instrument and its accessories within the limits of the specified technical data and conditions (ambient conditions, IP protection code, measuring category etc.).
- Do not use the instrument in potentially explosive atmospheres.
Rechargeable batteries
- If the charger is used, only the rechargeable battery pack (Z502H) may be fitted in the instrument.
- Do not use the instrument while the rechargeable battery pack (Z502H) is being charged.
- Do not use the instrument if the battery compartment cover has been removed. Touch contact with dangerous voltages is otherwise possible.
- The rechargeable battery pack (Z502H) may only be charged in undamaged condition. Inspect the rechargeable battery pack (Z502H) before use. Pay particular attention to leaky and damaged batteries.
Fuses
- The instrument is equipped with fuses. The instrument may only be used as long as the fuses are in flawless condition. Defective fuses must be replaced. See detailed operating instructions
Measurement cables and establishing contacts
- Plugging in the measurement cables must not necessitate any undue force.
- Never touch conductive ends (e.g. of test probes).
- Fully unroll all measurement cables before starting a test/measurement. Never perform a test/measurement with the measurement cable rolled up.
- Avoid short-circuits due to incorrectly connected measurement cables.
- Ensure that alligator clips, test probes or Kelvin probes make good contact.
Data security
- Always create a backup copy of your measurement data.
- Observe and comply with the applicable national data protection regulations. Use the corresponding functions provided by the instrument such as access protection, as well as other appropriate measures.
Applications
Please read this important information!
Intended Use / Use for Intended Purpose
The measuring and test instruments in the PROFITEST MASTER IQ series include
- PROFITEST MXTRA IQ (M534M) *
- PROFITEST MTECH+ IQ (M534L) *
- PROFITEST MPRO IQ (M534C) *
- Article number on the rating plate (test instrument only); order numbers (instrument with standard scope of delivery or extended accessories) see data sheet.
The test instruments are used to test the effectiveness of protective measures
at stationary electrical installationsper DIN VDE 0100-600, DIN VDE 0105-100,
OVE8101, NIN / NIV and other country-specific standards. They also allow for
testing of electric charging stations in accordance with EN 61851-1 (DIN VDE
0122-1), and earth measurements. The test instruments includes pre-defined
test sequences for easy use; optionally, user-defined test sequences can be
programmed. The test instruments are specially suited for tests during
installation, initial start-up, and for recurrent tests and troubleshooting
in electrical installations.
The range of application of the test instruments covers all alternating and
3-phase current systems with nominal voltages of 230 V / 400 V (300 V / 500 V)
with nominal frequencies of 16⅔ Hz / 50 Hz / 60 Hz / 200 Hz / 400 Hz.
A system structure is created in the test instrument and the meas- ured values
are assigned to the objects. Completed tests and measured values can be saved
and documented in a measurement and test report.
Safety of the operator, as well as that of the test instrument, is only
assured when it’s used for its intended purpose.
Use for Other than Intended Purpose
Using the test instrument for any purposes other than those described in these
condensed operating instructions, or in the test instrument’s operating
instructions, is contrary to use for intended purpose.
Liability and Guarantee
Gossen Metrawatt GmbH assumes no liability for property damage, personal
injury or consequential damage resulting from improper or incorrect use of the
product, in particular due to failure to observe the product documentation.
Furthermore, all guarantee claims are rendered null and void in such cases.
Nor does Gossen Metrawatt GmbH assume any liability for data loss.
Opening the Instrument / Repairs
The test instrument may only be opened by authorized, trained personnel in
order to ensure flawless, safe operation and to assure that the guarantee
isn’t rendered null and void. Even origi- nal replacement parts may only be
installed by authorized, trained personnel. Unauthorized modification of the
test instrument is prohibited. If it can be ascertained that the test
instrument has been opened by unauthorized personnel, no guarantee claims can
be honored by the manufacturer with regard to personal safety, measuring
accuracy, compliance with applicable safety measures or any consequential
damages. If the guarantee seal is damaged or removed, all guarantee claims are
rendered null and void.
Scope of Delivery
Standard equipment of the PROFITEST MASTER IQ series:
1 Test instrument
1 Earthing contact plug insert, country-specific (PRO-SCHUKO /
GTZ3228000R0001)
1 2-pole measuring adapter and 1 cable for expansion into 3-pole adapter
(PRO-A3-II / Z501O)
2 Alligator clips
1 USB cable
1 Shoulder strap
1 Compact rechargeable battery pack (Z502H)
1Charger (Z502R)
1 DAkkS calibration certificate
1 Condensed operating instructions* (this document)
1 Software IZYTRONIQ BUSINESS Starter (for PROFITEST MPRO IQ) or IZYTRONIQ
CLOUD (for PROFITEST MTECH+ IQ and PROFITEST MXTRA IQ)**
- Detailed operating instructions available on the Internet for download from www.gossenmetrawatt.com
** Download from the Internet; registration certificate in the scope of delivery
Symbols on the Instrument
| Warning concerning a point of danger (attention, observe documentation!)
---|---
| Protection category II device
| Charging socket for extra-low direct voltage (for charger Z502R)
| The instrument and its batteries may not be disposed of with household
trash. Further information is included in the operating instructions.
****| EU conformity marking
| If the guarantee seal is damaged or removed, all guarantee claims are
rendered null and void.
| Special technical knowledge of qualified personnel is required for
electrical installation or repair.
| CAT III 600 V CAT IV 300 V Measuring
category
Technical Data (excerpt)
The complete technical data and characteristic values (display ranges, resolution, uncertainties etc.) can be found in the complete operating instructions.
Accuracy | 0 … + 40 °C |
---|---|
Operation | –5 … + 50 °C |
Storage | –20 … + 60 °C (without rechargeable batteries) |
Relative humidity | Max. 75%, no condensation allowed |
Elevation | Max. 2000 m |
Nominal Ranges of Use
Voltage UN| 120 V (108 … 132 V)
230 V (196 … 253 V)
400 V (340 … 440 V)
---|---
Frequency fN| 16 ⅔ Hz (15.4 … 18 Hz)
50 Hz (49.5 … 50.5 Hz)
60 Hz (59.4 … 60.6 Hz)
200 Hz (190 … 210 Hz)
400 Hz (380 … 420 Hz)
Overall voltage range| 65 … 550 V
Overall frequency range| 15.4…420Hz
Waveform| Sinusoidal
Temperature range| 0 °C … + 40 °C
Supply voltage| 8 … 12 V
Line impedance angle| Corresponds to cos =1 … 0,95
Probe resistance| < 50 k
Overload Capacity
RISO | 1200 V continuous |
---|---|
UL-PE, UL-N | 600 V continuous |
RCD RE RF | 440 V continuous |
ZL-PE, ZL-N | 550 V (Limits the number of measurements and pause duration. If |
overload occurs, the
instrument is switched off by means of a thermostatic switch.)
RLO| Electronic protection prevents switching onif interference voltage is
present
Protection with| FF 3.15 A 10 s,
Fine-Wire Fuse| Fuses blow at > 5 A
Electrical Safety
Protection class | II per EN 61010-1 |
---|---|
Nominal Voltage | 230/400 V (300/500 V) |
Test voltage | 3.7 kV, 50 Hz |
Measuring category | CAT III 600 V and CAT IV 300 V |
Pollution degree | 2 |
Fuses | 1 fuse link ea. |
L and N terminals | FF 3.15 A/500G 6.3 x 32 mm |
Mechanical Design
Dimensions | W × L × D = 260 × 330 × 90 mm |
---|---|
Weight | approx. 2.7 kg with rechargeable batteries |
Type of protection | Housing: IP40; test probes: IP40 |
per EN 60529
Data Interfaces
Type | USB for PC connection |
---|---|
Type | RS-232 for barcode and RFID readers |
Scope of Functions
PROFITEST … (article number) | MPRO IQ (M534C) |
---|
MTECH+ IQ (M534L)
MXTRA IQ (M534M)
Testing of residual current devices (RCDs)| |
---|---|---
UB measurement without tripping the RCCB| √| √| √
Tripping time measurement| √| √| √
Measurement of tripping current IF| √| √| √
Selective, SRCDs, PRCDs, type G/R| √| √| √
AC/DC sensitive RCDs, types B and B+| –| √| √
DC sensitive RDC-DDs and RCMBs| –| √| √
Testing of insulation monitoring devices (IMDs)| –| –| √
Testing Residual Current Monitoring Devices (RCMs)| –| –| √
Testing for N-PE reversal| √| √| √
Measurements of loop impedance ZL-PE / ZL-N
Fuse table for systems without RCDs| √| √| √
Without tripping the RCD, fuse table| –| √| √
15 mA measurement1)| √| √| √
Earth resistance RE (mains operation) I/U measuring method (2/3-wire measuring
method via
measuring adapter: 2-wire/2-wire + probe)| √| √| √
Earthing resistance RE (battery operation)
3 or 4-wire measuring method via PRO-RE adapter| √| –| √
Soil resistivity E (battery operation) (4-wire measuring method via PRO-RE
adapter)| √| –| √
Selective earthing resistance RE (mains operation) with 2-pole adapter, probe,
earth electrode and current clamp sensor (3-wire measuring method)| √| √| √
Selective earthing resistance RE (battery operation)
with probe, earth electrode and current clamp sensor
(4-wire measuring method via PRO-RE adapter and curclamp sensor)| √| –| √
Earth loop resistance RESCHL (battery operation) with 2 clamps (clip-on
current transformer direct and clipon current transformer via PRO-RE/2
adapter)| √| –| √
Measurement of equipotential bonding RLO automatic polarity reversal| √| √| √
Insulation resistance RISO
Variable or rising test voltage (ramp)| √| √| √
oltage UL-N / UL-PE / UN-PE / f| √| √| √
Special measurements
Current measurement with clamp IL, IAMP| √| √| √
Rotating Field Direction| √| √| √
Earth leakage resistance RE(ISO| √| √| √
Voltage drop (U)| √| √| √
Standing-surface insulation ZST| √| √| √
Meter start-up (kWh test)| √| √| √
Leakage current with PRO-AB adapter (IL)| –| –| √
Residual voltage test (Ures)| –| –| √
Intelligent ramp (ta + I)| –| –| √
Electric vehicles at charging stations (IEC 61851-1)| –| √| √
Documentation of fault simulations at PRCDs with the PROFITEST PRCD adapter|
–| –| √
Features
Selectable user interface language 2)| √| √| √
Memory (database for up to 50,000 objects)| √| √| √
Automatic test sequences function| √| √| √
RS 232 port for RFID/barcode reader| √| √| √
USB port for data transmission| √| √| √
PC database and report generating software IZYTRONIQ 3)| √| √| √
Measuring category: CAT III 600V / CAT IV 300 V| √| √| √
DAkkS calibration certificate| √| √| √
-
The so-called live measurement is only advisable if there is no bias current within the system. Only suitable for motor protection switches with small nominal current values.
As of firmware 3.4.4: 15 mA test current, only applies if RCD is set with IΔN = 30 mA; otherwise test current = ½ · IΔN of the pre-set RCDs. -
currently available languages D, GB, I, F, E, P, NL, S, N, FIN, CZ, PL
-
IZYTRONIQ BUSINESS Starter for PROFITEST MPRO IQ or IZYTRONIQ CLOUD for PROFITEST MTECH+ IQ and PROFITEST MXTRA IQ
Sockets
Initial Start-Up
Attention!
Remove the protective films from both sensor surfaces (finger contacts) of the
test plug.
This is the only way to safely detect contact voltages.
Insert the enclosed,pre-charged compact rechargeable battery pack (Z502H):
➭ Loosen the screw on the battery compartment cover (back of instrument).
➭ Remove the battery compartment cover.
➭ Insert the compact rechargeable battery pack.
➭ Replace the battery compartment cover.
➭ Screw the battery compartment cover into place.
urther information on the power supply can be ound in the operating
instructions.
- Switching on,
- Connection &
- rech. batterytest
Settings
- Brightness
- Contrast
- Request help
Parameter:
- 3 select
- confirm
- take over
Measurement
Measurement
- select
- start
- trip RCD
see operating instructions ![GOSSEN METRAWATT DIN VDE 0100-600 Electric Tester
- Measurement](https://manuals.plus/wp-content/uploads/2023/07/GOSSEN- METRAWATT-DIN-VDE-0100-600-Electric-Tester-Measurement.jpg)
- Bias magnetization with DC only possible with PROFITEST MXTRA IQ and PROFITEST MTECH+ IQ
up to firmware 3.4.4:| for fN < 45 Hz => UN < 253 V| as of firmware 3.6.0:|
for fN < 45 Hz => UN < 500 V
---|---|---|---
*** as of firmware 3.4.4: 15 mA test current, only applies if RCD is set with IΔN = 30 mA; otherwise test current = ½ · IΔN of the pre-set RCDsBias magnetization with DC only possible with PROFITEST MXTRA IQ and PROFITEST MTECH+ IQBias magnetization with DC only possible with PROFITEST MXTRA IQ and PROFITEST MTECH+ IQ
he following measurements are only possible on voltage-free devices. Interference voltage disables the measurement!
Switching of
Contact & Support
+49 911 8602-0
info@gossenmetrawatt.com /
support@gossenmetrawatt.com
Monday to Thursday: 8 a.m. to 4 p.m.
Friday: 8 a.m. to 2 p.m.
Service, repairs, spare parts and calibrations
+49 911 817718-0
service@gossenmetrawatt.com
www.gmci-service.com
CE Declaration
The instrument fulfils all requirements of applicable EU directives and
national regulations. We confirm this with the CE mark. The CE declaration is
available upon request.
A calibration certificate is included with the instrument.
© Gossen Metrawatt GmbH
Prepared in Germany · Subject to
change without notice / Errors excepted · A PDF
version is available on the Internet All trademarks, registered trademarks,
logos, product names, and company names are the property of their respective
owners.
Gossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg • Germany
Telefon / Phone / Téléphone / Téléfono / Telefono / Telefooon / Telefon +49
911 8602-0
Telefax / Fax +49 911 8602-669
E-Mail / e-mail / e-mailem
info@gossenmetrawatt.com
www.gossenmetrawatt.com
Documents / Resources
|
GOSSEN METRAWATT DIN VDE 0100-600 Electric
Tester
[pdf] Instruction Manual
DIN VDE 0100-600, DIN VDE 0105-100, OVE E 8101, IEC 60364-6, EN 50110-1, DIN
VDE 0100-600 Electric Tester, DIN VDE 0100-600, Electric Tester, Tester
---|---
References