Features
- 3 pole AC inlet IEC320-C14, Class I power unit
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Extremely low leakage current
- No load power consumption<0.15W
- Energy efficiency level VI and meet CoC Version 5
- 30~+70°C wide range working temperature
- Protections: Short circuit / Overload/ Over voltage/Over temperature
- LED indicator for power on
- Lifetime > 85 K hours
- Various DC plug quick adapter accessory available (Plug kit sold separately, please refer to :https://www.meanwell.com/upload/pdf/DC plug.pdf)
- • 3 years warranty
Applications
- Mobile clinical workstation
- Oral irrigator
- Portable hemodialysis machine
- Breath Machine
- Medical computer monitor
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
Description
GSM160A is a highly reliable, 160W desktop style single-output green medical
adaptor series. This product is equipped with a 3-pin (with FG) standard
IEC320-C14 power plug, adopting the input range from 80VAC to 264VAC. The
entire series supplies different output voltages between 12VDC and 48VDC that
can satisfy the demands for various kinds of medical electrical devices. The
circuitry design meets the international medical standards (2*MOPP), having an
ultra low leakage current (<100µA), fitting the medical devices in direct
electrical contact with the patients.
With the efficiency up to 94% and the extremely low no-load power consumption
below 0.15W, GSM160A is compliant with USA EISA 2007/DOE, Canada NRCan,
Australia and New Zealand MEPS, EU ErP, and meet Code of Conduct (CoC) Version
5. The supreme feature allows the adaptor to save the energy when it is either
under the operating mode or the standby mode. The entire series utilizes the
94V-0 flame retardant plastic case. GSM160A is approved with the international
medical safety certificates.
Model Encoding
SPECIFICATION
ORDER NO.| GSM160A12-R7B| GSM160A15-R7B| GSM160A20-R7B|
GSM160A24-R7B| GSM160A48-R7B
---|---|---|---|---|---
OUTPUT| SAFETY MODEL NO.| GSM160A12| GSM160A15| GSM160A20|
GSM160A24| GSM160A48
DC VOLTAGE| Note.2| 12V| 15V| 20V| 24V| 48V
RATED CURRENT| 11.5A| 9.6A| 8A| 6.67A| 3.34A
CURRENT RANGE| 0 ~ 11.5A| 0 ~ 9.6A| 0 ~ 8A| 0 ~ 6.67A| 0 ~ 3.34A
RATED POWER (max.)| 138W| 144W| 160W| 160W| 160W
RIPPLE & NOISE (max.) **Note.3| 80mVp-p| 100mVp-p| 100mVp-p| 120mVp-p|
150mVp-p
VOLTAGE TOLERANCE Note.4| ±5.0%| ±5.0%| ±4.0%| ±3.0%| ±3.0%
LINE REGULATION Note.5| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
LOAD REGULATION| ±5.0%| ±5.0%| ±4.0%| ±3.0%| ±3.0%
SETUP, RISE TIME| Note.6| 2000ms, 50ms / 230VAC 2500ms, 50ms
/ 115VAC at full load
HOLD UP TIME (Typ.)| 24ms / 230VAC 24ms / 115VAC at full load
INPUT| VOLTAGE RANGE Note.7| 80 ~ 264VAC
FREQUENCY RANGE| 47 ~ 63Hz
POWER FACTOR (Typ.)| 12V,15V:PF>0.93 / 230VAC 20V,24V,48V:PF>0.94 /
230VAC| PF>0.98 / 115VAC at full load|
EFFICIENCY (Typ.)| 90%| 91%| 92.5%| 93%| 94%
AC CURRENT (Typ.)| 1.85A / 115VAC 1A / 230VAC
INRUSH CURRENT (Typ.)| Cold start 90A / 115VAC 110A / 230VAC
LEAKAGE CURRENT(max.)| Earth leakage current < 175μA/264VAC , Touch
current <90μA/264VAC
PROTECTION| OVERLOAD| 105 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| 105 ~ 135% rated output voltage
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE| Shut down o/p voltage, re-power on to recover
ENVIRONMENT| WORKING TEMP.| -30 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03% / ℃ (0~40℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, Z axes
OPERATING ALTITUDE Note.8| 3000 meters
SAFETY & **EMC (Note 10)| SAFETY STANDARDS| IEC60601-1, TUV
BS EN/EN60601-1, ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22.2 No.
60601-1:14 – Edition 3, EAC TP TC 004 approved
ISOLATION LEVEL| Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP
WITHSTAND VOLTAGE Note.9| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22 CAN
ICES-3(B)/NMB-3(B)| Class B
Radiated emission| BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22 CAN
ICES-3(B)/NMB-3(B)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN60601-1-2, BS EN/EN61204-3
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz
)Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 3, 1KV/Line-Line , 2KV/Line-FG
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25
periods,100% interruptions 250 periods
OTHERS| MTBF| 2200.0K hrs min. Telcordia SR-332 (Bellcore) ;
239.4K hrs min.| MIL-HDBK-217F (25℃)|
DIMENSION| 1757235mm (LWH)
PACKING| 0.66Kg; 20pcs/14.2Kg/0.95CUFT
CONNECTOR| PLUG| See page 4~5 ; Other type available by customer
requested
CABLE| See page 4~5 ; Other type available by customer requested
NOTE| 1. All parameters are specified at 230VAC input, rated load, 25℃
70% RH ambient.2. DC voltage: The output voltage set at point measure by plug
terminal & 50% load.3. Ripple & noise are measured at 20MHz by using a 12
twisted pair terminated with a 0.1μf & 47μf capacitor.4. Tolerance: includes
set up tolerance, line regulation, load regulation.5. Line regulation is
measured from low line to high line at rated load.6. Length of set up time is
measured at first cold start. Turning ON/OFF the power supply may lead to
increase of the set up time.7. Derating may be needed under low input
voltages. Pleas check the derating curve for more details.8. The ambient
temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with
fan models for operating altitude higher than 2000m(6500ft).9. Optional for
1.5KVAC with BF rated.10. The power supply is considered as an independent
unit, but the final equipment still need to re-confirm that the whole system
complies with the EMC directives. For guidance on how to perform these EMC
tests, please refer to “EMI testing of component power supplies.”(as available
on http://www.meanwell.com)※ Product Liability Disclaimer For detailed
information, please refer to
https://[www.meanwell.com/serviceDisclaimer.aspx](http://www.meanwell.com/serviceDisclaimer.aspx)
| Arrow.| com.| File
Name:GSM160A-SPEC ____2022-03-18
Overview
◎ DC plug changeable through:
(1) Customization of the standard part with an optional DC plug according to
the table (MOQ applicable)
(2) Quick adapter accessory (sold separately without MOQ)
Please refer to below table and online selection guide:
https://www.meanwell.com/upload/pdf/DC plug.pdf
Example quick adapter accessory:
Optional DC plug: (Available in customized cable or quick adapter)
Tuning Fork Style | Type No. | A | B | C | Quick Adapter Accessory |
---|---|---|---|---|---|
OD | ID | L |
| P1J| 5.5| 2.1| 11.0| Available
for 24 ~ 48V(Current rating: 7.5A max.)
P1M| 5.5| 2.5| 11.0
Min. DIN 4 Pin with Lock (female)| Type No.| Pin Assignment| None
PIN No.| Output
| R7BF| 1| +Vo
2| -Vo
3| -Vo
4| +Vo
DIN 5 Pin (male)| Type No.| Pin Assignment| None
PIN No.| Output
| R1B| 1| -Vo
2| -Vo
3| +Vo
4| -Vo
5| +Vo
NEUTRIK XLR NC4FX equivalent| Type No.| Pin Assignment| None
PIN No.| Output
| MIC4| 1| +Vo
2| +Vo
3| -Vo
4| -Vo
MOLEX 39-01-2060 (4.2mm) equivalent| Type No.| Pin Assignment| None
PIN No.| Output
FG not connected to output connector| C6P| 1| +Vo
2| +Vo
3| +Vo
4| -Vo
5| -Vo
6| -Vo
AMP 1-480702-0 (6.35mm) equivalent| Type No.| Pin Assignment| Quick Adapter
Accessory
---|---|---|---
PIN No.| Output
__ 4321 FG not connected to
output connector| C4P| 1| +Vo| None
2| +Vo
3| -Vo
4| -Vo
Stripped and tinned leads| Type No.| Pin Assignment| __ None
PIN No.| Output
length, L: 25 mm, L1: 5 mm)|
__ by customer| 1| +Vo
2| __ -Vo
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
Read User Manual Online (PDF format)