eyc-tech FDM06S Air Compressor Flow Meter Instruction Manual

June 12, 2024
eyc-tech

eyc-tech FDM06S Air Compressor Flow Meter Instruction Manual
Air Compressor Flow Meter

Security considerations

Please read this Specification carefully, prior to use of this, and keep the manual properly, for timely reference.  Solemn Statement :This product can not be used for any explosion-proof area. Do not use this product in a situation where human life may be affected. eyc-tech will not bear any responsibility for the results produced by the operators !

Warning!

  • installation and wiring must be performed by qualified personnel in accordance with all applicable safety standards.
  • This product must be operated under the operating conditions specified in manual to prevent equipment damages.
  • Please using the product under the ordinary pressure, or it will influence safe problem.
  • This product must be operated under the operating condition specified in this manual to prevent equipment damages.
  • This product must be operated under the normally atmospheric condition to prevent equipment damages.
  • To prevent products damage, always disconnect the power supply from the product before performing any wiring and installation.
  • All wiring must comply with local codes of indoor wiring and electrical installation rules.
  • Please use crimp type terminal.
  • To prevent personal injury, do not touch the moving part of product in operation.
  • it may cause high humidity atmosphere during the product was breakdown. Please take safety strategy.

Operation Form

Differential pressure type air flow transmitter
Operation Form

  • Key Pad Operation Mode

Key Pad
Operation| Display Mode
---|---
Normal| Menu
Click UP| Reserved| Next Option or Increase Value Once
Click OK| Reserved| Submit and Go Next Menu or Return to Normal Mode
Click DOWN| Reserved| Previous Option or Decrease Value Once
Hold UP| Reserved| Speedy Next Option or Increase Value
Hold DOWN| Reserved| Speedly Previous Option or Decrease Value
Hold OK 5 sec| Auto Zero| Reserved
Press UP and DOWN| Go Menu Mode| Reserved

  • Display Symbol
Symbol Description
PASS Password Validation
PF Profile Factor
vo-”c Low Cut-Off
d «5P Display Enable
L,9E Brightness Level
on ON
oFF OFF
—- Submit Setting and Exit Menu Mode

Connection Diagram

Connection Diagram

  • Please make sure the product and the device which connect with RS-485 are on common ground, avoid damaged product.

Installation

The following table specifies the required straight lengths of the pipe sections depending on pipe diameter in case of different disturbances.

<90°) >

Type| Drawing| Upstream straight pipe || Downstream straight pipe
---|---|---|---
Light bend (<90°)| Light Bend| 10xD| 10xD
T-junction| | 15xD| 10xD
Two 90° bends n one plane| | 20xD| 5xD
Two 90° bends with 3- dimensional Two 90° bends with 3- dimensional| | 35xD| 10xD
Shut-off valve| Shut-Off Valve| 45xD| 10xD

 RS-485 and Modbus

FDMO6 integrate a RS-485 interface for digital communication as an option feature. Based on Modbus protocol makes the general convenience on PLC, HMI and PC connection. For Modbus protocol information please download the file from website. Besides the PLC, HMI application, the user software provide the device setting and data logging function, it also can free download from website.

Technical Data :

  1. Max. network size : 32 transmitters
  2. Communication : with COM-Port (serial interface) of PC
  3. Max. network expansion : 1200m (3937ft) total length at 9600 baud
  4. Transmission rate : 9600, 19200, 38400, 57600, 115200 Baud
  5. Parity : None, Even, Odd
  6. Data length : 8 bit
  7. Stop bit : 1 or2 bit
  8. Factory default Station address = 1, Data format= 9600, N81

Autozero

The middle button allows user to set the current pressure to zero point. It is required to press the button about 5 seconds, and user can see A.2ro will be display. Then user can release this button and will see the prompt —-, and the new zero point has been set. Please make sure that the gas is completely still prior to execute this function.

This button also allows user to restore factory default setting. It is required to press the button about 10 seconds, user will first see F.2-o will be display. Then user can release this button and will see the prompt —-, and the new zero point has been set.

Software and configuration step

User may download the configuration software on eyc web site. Please decompress the application prior to execute it. Operating System requirements : above Windows XP. Other application program requirements : above Microsoft Office 2003

  1. Hardware connection : Connect the FDMO06 to PC through USB to RS-485 or RS-232 to RS-485 converter

  2. Check the COM port number from Device Manager in Computer Management. e.g.COM1inillustration
    Configuration

  3. Open the FDMO6C Ul, go to function ” Interface ” ,click item ” Config” and then setting COM port, BAUD rate, data format and Station ID, pressed ” Apply ” for connection
    Configuration

  4. Scan RS-485 connection
    Open the FDMO6 Ul, go to function ” Interface” , clickitem ” Config” and then setting COM port, pressed ” Scan” bottom for scan devices and pressed ” Close and Export ” when the interested devices foundConfiguration
    Pick up the device that you want to connect to and then press “Apply” to go.

  5. Setting on Impulse Output
    In the group of OUTL, Output tab. The output1l related setting could be found.

  6. Quantity : Differential Pressure, Flow Velocity, Air Temperature, Flow Volume and Barometric Pressure

  7. Responserate : 1*order low pass filter, T90 time 0 … 100 » higher value if lower fluctuation but longer response time, lower value if allow higher fluctuation but shorter response time.

  8. Analogtype : 0..20mA/4…20 mA (if output current) / 0 … 10V (if output voltage)

  9. Range for Upper and Lower

  10. Alarm Mode: Check the box if analog output pretend a alarm switch output

  11. Alarm Trigger Point: Upper and Lower

  12. Alarm Output Level: Upper and Lower
    Configuration

  13. Setting on Impulse Output (Optional, Only Available if Function Equipped)
    In the group of OUT2, Output tab. The Impulse output related setting could be found.

  14. Quantity : Differential Pressure, Flow Velocity, Air Temperature, Flow Volume and Barometric Pressure

  15. Response rate : 1st order low pass filter, T90 time 0 … 100 > higher value if lower fluctuation but longer response time, lower value if allow higher fluctuation but shorter response time.

  16. Impulse Value : The valence of single impulse quantity

  17. Upper Range : Span upper point or maximum output. The span lower point always const zero.
    Configuration

  18. Setting on RS-485, Process Parameters and offset adjustment
    There are 4 groups in setting tab. The description of each item as below.

    •   *  **Process Parameters**
      1. Gas molar mass, default air Common gas molar table, the actual molar mass depends on substances molar mass of compounds and mixtures. The table of common molar mass for reference.
        substances| Holar mass (g/mol)
        ---|---
        Air| 28.966
        Argon, Arz| 39.948
        Carbon Dioxide, CO2| 44.01
        Hydrogen, H:| 2.016
        Nitrogen, N>| 28.0134
        Neon, Ne| 20.179
        Oxygen, 0:| 31.9988

      2. Relative Humidity of working condition

      3. Profile factor of gas velocity

      4. Differential Pressure factor, default 1

      5. Pipe shape and dimension: round duct, square duct and any if input area directly.
        Offset  Offset  Adjustment :

      6. velocity offset

      7. temperature offset

      8. differential pressure cut off level

      9. differential pressure offset

      10. atmosphere pressure offset

      11. flow direction: Enable bidirectional flow: considers reserve flow if negative
        differential pressure. Disable bidirectional flow: considers still flow if negative differential pressure.
        Modbus Protocol

      12. station ID

      13. Baud Rate

      14. Date Frame

      15. Flash memory write protect
        Miscellaneous :

      16. Password of keypad menu

      17. LED brightness

      18. Display Alternate Items : Only OUT1, Only OUT2 or display OUT1 and OUT2 in
        turn

      19. Display Alternate Period : the first field for quantity unit duration, the seconds field for quantity measure duration (seconds)
        Configuration

  19. Data display and logging

  20. Assign OUT1 Unit : Differential Pressure : mBar, Pa, hPa, kPa, mmWS, inH20, mmHg ; Flow Velocity : m/s, ft/s ; Air Temperature : °C, °F ; Flow Volume : Kg/h, Nm3/h, L/min, m3/min ; Barometric Pressure : ATM, Bar, hPa, kPa, mmHg

  21. Assign OUT2 Unit : as OUTL’ s

  22. Display Decimals

  23. Clear Measure Plot Chart

  24. Time scale of plot

  25. Vertical scale of plot

  26. Export all logging data since device is connected

  27. Start/Stop data logging
    Configuration

  28. Device Information

  29. Model Name of Device

  30. Firmware Version of Device

  31. Serial Number of Device

  32. Firmware Checksum

  33. Calibration Date

  34. Flow Offset

  35. Temperature Offset

  36. Flow Calibration Range

  37. Temperature Calibration Points
    Configuration

  38. Totalizer

  39. Flow Velocity

  40. Flow Volume Rate Differential pressure type air flow transmitter

  41. Volume Accumulation Totalizer 1. Count positive flow only.

  42. Volume Accumulation Totalizer 2. Count total flow amount without direction if Bidirectional flow mode is enabled. Thus, the reverse flow amount = Totalizer 2

    * Totalizer 1
  5. unit of counter 1
  6. unit of counter 2
  7. Main switch of totalizer
  8. Function switch of totalizer 1
  9. Function switch of totalizer 2
  10. Ser/Reset Totalizer 1
  11. Set/Reset Totalizer 2  

Configuration

Inspection and maintenance

Maintenance
Since this product is inspected and calibrated for high accuracy at the factory before shipment, no calibration on the installation site is necessary when this product is installed For inspection and maintenance follow the instructions below :
(a) Periodic inspection Periodically inspect this product for its sensing accuracy, and clean the cover Set the period between inspections based on atmospheric dust and other contaminants in the installation environment
(b ) Sensor maintenance Do not damage sensor surface during maintenance process
(c ) Troubleshooting If any problem occurs during operation, refer to the table below for appropriate solutions

Troubleshooting :

Problem Cleck items Soluations
  • No output
  • Unstable output

|

  • Disconnected wiring
  • Loose wiring
  • Power supply voltage
  • Sensor damages

|

  • Re-perform wiring

  • Crew on terminal tightly or replace wires

  • Replace the sensor

  • Slow response to output

  • Error in output

|

  • Moisture / Condensation on the product
  • Execute Autozero before measures
  • Check installed location
  • Check installed angle
  • Check dust and contamination on the sensor

|

  • Remove the sensor and filter dry power-off state sensor in clean air seasoning
  • Refer to the section
  • Align measurement head with flow direction
  • Cleaning the filter
  • Changing the filter
  • Calibrate
  • Replace the sensor

eyc-tech Measuring Specialist
enhance your capability with sensor technology
Air flow | Humidity | Dew point | Differential pressure | Liquid flow
Temp. | Pressure | Level | Air quality | Signal meter