eyc-tech FTM06D Thermal Mass Flow Meter
Product Information
The eyc-tech Measuring Specialist is a sensor technology product that enhances your capability to measure various parameters. It is designed to measure air flow, humidity, dew point, differential pressure, liquid flow, temperature, pressure, level, air quality, and signal strength.
Product Usage Instructions
To use the eyc-tech Measuring Specialist, follow the steps below:
- Ensure that the device is properly connected to the power source.
- Turn on the device by pressing the power button.
- Choose the parameter you want to measure by selecting the corresponding option on the device’s interface.
- Position the device in the desired location for measurement.
- Wait for the device to stabilize and display the measurement readings.
- Record or analyze the measurement readings as needed.
- If required, make adjustments to the device settings using the provided controls.
- When finished, turn off the device by pressing the power button again.
- Disconnect the device from the power source.
For more detailed instructions on specific measurements or device features, refer to the complete user manual provided with the product. For any further questions or assistance, please visit our website at www.eyc- tech.com.
Safety Precautions
Please read this Specification carefully, prior to use of this, and keep the
manual properly, for timely reference.
Solemn Statement : This product can not be used for any explosion-proof
area. Do not use this product in a situation where human life may be affected.
eyc-tech will not bear any responsibility for the results produced by the
operators!
Warning!
- Installation and wiring must be performed by qualified personnel in accordance with all applicable safety standards.
- This product must be operated under the operating conditions specified in manual to prevent equipment damages.
- Please using the product under the ordinary pressure, or it will influence safe problem.
- This product must be operated under the operating condition specified in this manual to prevent equipment damages.
- This product must be operated under the normally atmospheric condition to prevent equipment damages.
- To prevent products damage, always disconnect the power supply from the product before performing any wiring and installation.
- All wiring must comply with local codes of indoor wiring and electricalinstallation rules.
- Please use crimp type terminal.
- To prevent personal injury, do not touch the moving part of product in operation. It may cause high humidity atmosphere during the product was breakdown. Please take safety strategy.
Operation Form
Connection Diagram
Installation
- Ducy type
- Flange pipe type Available Options: Installation-Flange Pipe (Unit:mm)
- Inline type
Installation instructions
5. RS-485 and Modbus
FTM06 integrate a RS-485 interface for digital communication as an option
feature. Based on Modbus protocol makes the general convenience on PLC, HMI
and PC connection. For Modus protocol information please download the file
from website. Besides the PLC, HMI application, the user software provide the
device setting and data logging function, it also can free download from
website.
Technical Data
- Max. network size: 32 transmitters
- Communication: with COM-Port (serial interface) of PC
- Max. network expansion: 1200m (3937ft) total length at 9600 baud
- Transmission rate: 9600, 19200, 38400, 57600, 115200 Baud
- Parity: None, Even, Odd
- Data length: 8 bit
- Stop bit: 1 or 2 bit
- Factory default Station address = 1, Data format= 9600, N81
Software and calibration operation step
User may download the configuration software on ec web site. Please decompress the application prior to execute it. Operating System requirements: above Windows XP. Other application program requirements: above Microsoft Office 2003
-
Hardware connection: Connect the FTM06D to PC through USB to R$-485 or RS-232 to RS-485 converter
-
Check the COM port number from Device Manager in Computer Management. e.g. COM1 in illustration
-
Open the FTMO6D UI, go to function ” Interface”, click item “Config” and then setting COM port, BAUD rate, data format and Station ID, pressed “Apply” for connection
-
Scan RS-485 connection Open the FTM06D UI, go to function ” Interface” , click item “Config” and then setting COM port, pressed ” Scan ” bottom for scan devices and pressed ” Close and Export ” when the interested devices found.Pick up the device that you want to connect to and then press “Apply” to go.
-
Setting on Analog Output
In the groupof OUT1, Output tab. The output1 related setting could be found. -
Quantity: Flow Velocity, Flow Volume and Sensor Temperature
-
Response rate : 0… 100 • 100 : filter off › 90 : filter = 60 second › 80 : filter = 120 second, etc.
-
Analog type: 0 .. 20 mA / 4 … 20 mA (if output current) / 0 … 10 V (if output voltage) Range for Upper and Lower
-
Alarm Mode: Check the box if analog output pretend a alarm switch output
-
Alarm Trigger Point: Upper and Lower
-
Alarm Output Level: Upper and Lower
-
Setting on PNP/PNP Output (Optional, Only Available if Function Equipped) In the group of OUT2, Output tab. The PP/PP switch output related setting could be found.
-
Quantity: Flow Velocity, Flow Volume and Sensor Temperature
-
Response rate: 0 … 100 › 100 : filter off • 90 : filter = 60 second • 80 : filter = 120 second, etc.
-
Alarm Mode : High Activate or Low Activate Range for Upper and Lower
-
Alarm Point : Upper and Lower
-
Delay Time: On Delay Time and Off Delay Time (seconds)
-
Setting on Frequency Output (Optional, Only Available if Function Equipped) In the group of OUT2, Output tab. The Frequency output related setting could be found.
-
Quantity: Flow Velocity, Flow Volume and Sensor Temperature
-
Response rate: 0 … 100 • 100 : filter off › 90 : filter = 60 second • 80 : filter = 120 second, etc.
-
Frequency: Upper and Lower
-
Measures: Upper and Lower
-
Alarm Mode: The output would be switch between upper and lower correspond to measures
-
Alarm Point : Upper and Lower
-
Alarm Output : Upper and Lower
-
Setting on Impulse Output (Optional, Only Available if Function Equipped) In the group of OUT2, Output tab. The Impulse output related setting could be found.
-
Quantity: Flow Velocity, Flow Volume and Sensor Temperature
-
Response rate: 0 … 100 › 100 : filter off › 90 : filter = 60 second • 80 : filter = 120 second, etc.
-
Impulse Duration : High and Low
-
Measures : Upper and Lower
-
Impulse Value: The valence of single impulse quantity
-
Impulse Unit : m°, Liter, mL, mm°, ft°, inch°, gal,us (Gallon, us), gal,uk (Gallon, uk)
-
Setting on RS-485, Process Parameters and offset adjustment There are 4 groups in setting tab. The description of each item as below.
Process Parameters : -
Temperature of normal condition
-
Pressure of working condition
-
Relative Humidity of working condition
-
Profile factor of gas speed
Offset Adiustment : -
velocity offset
-
temperature offset
-
velocity cut off
Modbus Protocol : -
station ID
-
Baud Rate
-
Date Frame
-
Flash memory write protect
Miscellaneous : -
Password of keypad menu
-
LED brightness
-
Display Alternate Period: the first field for quantity unit duration, the secondsfield for quantity measure duration (seconds)
-
Data display and logging
-
Assign Temperature Unit: °C / °F
-
Assign Flow Velocity Unit : m/s • ft/s • Flow Volume Unit : Nm°/h, Nm/s, L/min
-
Display Decimals
-
Clear Measure Plot Chart
-
Time scale of plot
-
Vertical scale of plot
-
Export all logging data since device is connected
-
Start/Stop data logging
-
Device Information
-
Model Name of Device
-
Firmware Version of Device
-
Serial Number of Device
-
Firmware Checksum
-
Calibration Date
-
Flow Offset
-
Temperature Offset
-
Flow Calibration Range
-
Temperature Calibration Points
-
Totalizer
-
Flow Velocity
-
Flow Volume Rate
-
Volume Accumulation Totalizer 1
-
Volume Accumulation Totalizer 2
-
unit ofcounter 1
-
unit of counter 2
-
Main switch of totalizer
-
Function switch of totalizer 1
-
Function switch of totalizer 2
-
Ser/Reset Totalizer 1
-
Set/Reset Totalizer 2
Inspection and maintenance
- Maintenance Since this product is inspected and calibrated for high accuracy at the factory before shipment, no calibration on the installation site is necessary when this product is installe For inspection and maintenance follow the instructions below :
- (a) Periodic inspection Periodically inspect this product for its sensing accuracy, and clean the cover Set the period between inspections based on atmospheric dust and other contaminants in the installation environment
- (b) Sensor maintenance Do not damage sensor surface during maintenance process
- (c) Troubleshooting If any problem occurs during operation, refer to the table below for appropriate solutions
- Troubleshooting :
eyc-tech Measuring Specialist enhance your capability with sensor technology
- Air flow
- Humidity
- Dew point
- Differential pressure
- Liquid flow
- Temp
- Pressure
- Level
- Air quality
- Signal meter
CONTACT
- Tel.: 886-2-8221-2958
- Web: www.eyc-tech.com.
- e-mail: info@eyc-tech.com.
- www.eyc-tech.com.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>