MEAN WELL GSM40B series Medical Grade Power Supplies Installation Guide

August 21, 2024
MEAN WELL

MEAN WELL GSM40B series Medical Grade Power Supplies

Features

  • 2 pole AC inlet IEC320-C8, Class 11 power unit

  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1/-1-11 and IEC/BS EN/EN60601-1/-1-11

  • Extremely low leakage current

  • No load power consumption<0.1W

  • Energy efficiency level VI and meet CoC Version 5 (Except 5~9V for Level V)

  • Comply with Korea K-MEPS(only GSM40B48-P1J)

  • -30~+70°C wide range working temperature

  • Protections: Short circuit / Overload / Over voltage

  • LED indicator for power on

  • Lifetime > 90 K hours

  • Various DC plug quick adapter accessory available
    (Plug kit sold separately, please refer to: https://www.meanwell.com/upload/pdf/DC_plug.pdf)

  • 3 years warranty

Applications

  • Mobile clinical workstation
  • Oral irrigator
  • Portable hemodialysis machine
  • Breath Machine
  • Medical computer monitor

GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx

Description

GSM40B is a highly reliable, 40W desktop style single-output green medical adaptor series. This product is equipped with a 2-pin (no FG) standard IEC320-C8 power plug, adopting the input range from 80 VAC to 264VAC. The entire series supplies different output voltages between 5VDC and 48VDC that can satisfy the demands for various kinds of medical electrical devices. The circuitry design meets the international medical standards (2*MOPP), having an ultra low leakage current (<50 MA), fitting the medical devices in direct electrical contact with the patients.

With the efficiency up to 91% and the extremely low no-load power consumption below 0.1W, GSM40B is compliant with USA EISA 2007/DoE, Canada NRCan, Australia and New Zealand MEPS, EU ErP, and meet Code of Conduct (CoC) Version 5. The supreme feature allows the adaptor to save the energy when it is either under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical shock. GSM40B is approved with the international medical safety certificates.

Model Encoding

SPECIFICATION

ORDER NO.| GSM40B05- P1J| GSM40B07- P1J| GSM40B09- P1J| GSM40B12-P1J| GSM40B15-P1J| GSM40B18- P1J| GSM40B24- P1J| GSM40B48- P1J
---|---|---|---|---|---|---|---|---








OUTPUT

| SAFETY MODEL NO.| GSM40B05| GSM40B07| GSM40B09| GSM40B12| GSM40B15| GSM40B18| GSM40B24| GSM40B48
DC VOLTAGE Note.2| 5V| 7.5V| 9V| 12V| 15V| 18V| 24V| 48V
RATED CURRENT| 5A| 5.34A| 4.45A| 3.34A| 2.67A| 2.22A| 1.67A| 0.84A
CURRENT RANGE| 0 ~ 5A| 0 ~ 5.34A| 0 ~ 4.45A| 0 ~ 3.34A| 0 ~ 2.67A| 0 ~ 2.22A| 0 ~ 1.67A| 0 ~ 0.84A
RATED POWER (max.)| 25W| 40W| 40W| 40W| 40W| 40W| 40W| 40W
RIPPLE & NOISE (max.) **Note.3| 80mVp-p| 80mVp-p| 100mVp-p| 100mVp-p| 100mVp-p| 120mVp-p| 150mVp-p| 150mVp-p
VOLTAGE TOLERANCE Note.4| ±5.0%| ±5.0%| ±5.0%| ±3.0%| ±3.0%| ±3.0%| ±2.5%| ±2.5%
LINE REGULATION Note.5| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
LOAD REGULATION| ±5.0%| ±5.0%| ±5.0%| ±3.0%| ±3.0%| ±3.0%| ±2.5%| ±2.5%
SETUP, RISE TIME Note.6| 1000ms, 30ms / 230VAC          1500ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.)**| 50ms / 230VAC         24ms / 115VAC at full load




INPUT

| VOLTAGE RANGE Note.7| 80 ~ 264VAC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 81%| 85.5%| 86%| 88%| 88.5%| 89%| 90%| 91%
AC CURRENT (Typ.)| 1A / 115VAC         0.5A / 230VAC
INRUSH CURRENT (Typ.)| Cold start 30A / 115VAC         60A / 230VAC
LEAKAGE CURRENT(max.)| Touch current < 50μA/264VAC



PROTECTION

| ****

OVERLOAD

| 105 ~ 160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed


OVER VOLTAGE

| 5.2 ~ 7,0V| 7.8 ~ 10.2V| 9.4 ~ 12.2V| 12.6 ~ 16.2V| 15.7 ~ 20.3V| 18.9 ~ 24.3V| 25.2 ~ 32.4V| 50.4 ~ 64.8V
Protection type : Shut down o/p voltage, re-power on to recover




ENVIRONMENT

| WORKING TEMP.| -30 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03% / ℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.8| 3000 meters















SAFETY & EMC

(Note 9)

| SAFETY STANDARDS| IEC60601-1/IEC60601-1-11, BS EN/EN60601-1/ BS EN/EN60601-1-11, ANSI/AAMI ES60601-1 / ES60601-1-11(3.1 version),

PSE J60950-1(48V only),KC K60950-1(48V only), CAN/CSA-C22.2 No. 60601-1:14 – Edition 3, EAC TP TC 004 approved

ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH





EMC EMISSION

| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22, CAN ICES-3(B)/NMB-3(B),MSIP KN32| Class B
Radiated emission| BS EN/EN55011 (CISPR11), FCC PART 15 /

CISPR22, CAN ICES-3(B)/NMB-3(B),MSIP KN32

| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–









EMC IMMUNITY

| BS EN/EN60601-1-2, BS EN/EN61204-3
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact


RF field susceptibility

| ****

BS EN/EN61000-4-3

| Level 3, 10V/m( 80MHz~2.7GHz )

Table 9, 9~28V/m( 385MHz~5.78GHz )

EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m


Voltage dip, interruption

| ****

BS EN/EN61000-4-11

| 100% dip 1 periods, 30% dip 25 periods,

100% interruptions 250 periods


OTHERS

| MTBF| 3505.7K hrs min.      Telcordia SR-332 (Bellcore) ; 719.4K hrs min.      MIL-HDBK-217F (25℃)
DIMENSION| 1255031.5mm (LWH)
PACKING| 0.29Kg; 40pcs/12.6Kg/1.04CUFT


CONNECTOR

| PLUG| See page 4~5 ; Other type available by customer requested
CABLE| See page 4~5 ; Other type available by customer requested






NOTE

| 1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.

2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.

3. Ripple & noise are measured at 20MHz by using a 12″ twisted pair terminated with a 0.1μF & 47μF capacitor.

4. Tolerance: includes set up tolerance, line regulation, load regulation.

5. Line regulation is measured from low line to high line at rated load.

6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.

7. Derating may be needed under low input voltages. Pleas check the derating curve for more details.

8. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).

9. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”

(as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

DC plug changeable through:

  1. Customization of the standard part with an optional DC plug according to the table (MOQ applicable)
  2. Quick adapter accessory (sold separately without MOQ) Please refer to below table and online selection guide: https://www.meanwell.com/upload/pdf/DC_plug.pdf

Example quick adapter accessory:

Optional DC plug: (Available in customized cable or quick adapter)

Installation Manual
Please refer to: http://www.meanwell.com/manual.html

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

MEAN WELL User Manuals

Related Manuals