MEAN WELL RPS-30 Series 30W Reliable Green Medical Power Supply User Guide
- August 15, 2024
- MEAN WELL
Table of Contents
30W Reliable Green Medical Power Supply
RPS-30 series
User Guide
RPS-30 Series 30W Reliable Green Medical Power Supply
https://www.meanwell.com/Upload/PDF/Medical_EN.pdf
◼ Features
- 3″x2″ compact size
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Suitable for BF application with appropriate system consideration
- Cooling by free air convection
- EMI class B for class II configuration
- No load power consumption<0.1W
- Extremely low leakage current
- Protections: Short circuit / Overload / Over voltage
- Lifetime > 105K hours
- Operating altitude up to 4000 meters
- 3 years warranty
◼ Applications
- Oral irrigator
- Hemodialysis machine
- Medical computer monitors
- Sleep apnea devices
◼ GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
◼ Description
RPS-30 is a 30W highly reliable green PCB type medical power supply with a high power density on the 3″ by 2″ footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 92% and the extremely low no load power consumption is down below 0.1W. RPS-30 is able to be used for Class II (no FG) system design. The extremely low leakage current is less than 80 µA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.
◼ Model Encoding
SPECIFICATION
NOTE
- All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature.
- 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
- Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1 F & 47 F parallel capacitor.
- Tolerance: includes set up tolerance, line regulation and load regulation.
- Derating may be needed under low input voltages. Please check the derating curve for more details.
- Touch current was measured from primary input to DC output.
- The ambient temperature derating of 3.5°C/1000m with fanless models and of 5°C/1000m with fan models for operating altitude higher than 2000m(6500ft).
- The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf)
※ Product Liability Disclaimer: For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
◼ Block Diagram
fosc: 65KHz
◼ Derating Curve
◼ Static Characteristics
◼ Mechanical Specification
Case No. Unit: mm
AC Input Connector (CN1): JST B3P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | AC/N | JST VHR or equivalent | JST SVH-21T-P1.1 or equivalent |
2 | No Pin | ||
3 | AC/L |
DC Output Connector (CN2): JST B2P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | +V | JST VHR or equivalent | JST SVH-21T-P1 .1 or equivalent |
2 | -V |
◼ Installation Manual
Please refer to: http://www.meanwell.com/manual.html
File Name: -SPECRPS 30 2024-04-15