MEAN WELL RPS-75 Series 75W Reliable Green Medical Power Supply
Product Information
Specifications
Model | DC Voltage | Rated Current | Rated Power |
---|---|---|---|
RPS-75-3.3 | 3.3V | 15A | 49.5W |
RPS-75-5 | 5V | 14A | 70W |
RPS-75-12 | 12V | 6.3A | 75.6W |
Product Usage Instructions
Features:
- Reliable Green Medical Power Supply
- Complies with EN 60601-1 and IEC 60601-1 standards
Description:
The RPS-75 series is a medical power supply designed for various applications
such as oral irrigators, hemodialysis machines, medical computer monitors, and
sleep apnea devices.
Applications:
- Oral irrigator
- Hemodialysis machine
- Medical computer monitors
- Sleep apnea devices
FAQ
What is the GTIN CODE for the product?
The GTIN CODE can be found on the Mean Well website using the MW Search tool.
Features
- 5″×3″ compact size
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Suitable for BF application with appropriate system consideration
- 75W convection, 100W force air
- EMI Class B for Class I configuration
- No load power consumption <0.75W
- Remote sense function
- Protections: Short circuit/Overload/Over voltage
- Lifetime > 80K hours
- Operating altitude up to 3000 meters
- 3 years warranty
Description
RPS-75 is a 75W highly reliable green PCB-type medical power supply with a high power density on the 5″ by 3″ footprint. It accepts 90~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 86% and the extremely low no-load power consumption is down below 0.75W. RPS-75 can be used for Class I (with FG) system design. The extremely low leakage current is less than 150 MA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF-rated “patient contact” medical system equipment.
Applications
- Oral irrigator
- Hemodialysis machine
- Medical computer monitors
- Sleep apnea devices
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTlN.aspx.
Model Encoding
SPECIFICATION
MODEL| RPS-75-3.3| RPS-75-5| RPS-75-12| RPS-75-15|
RPS-75-24| RPS-75-36| RPS-75-48
---|---|---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 3.3V| 5V| 12V| 15V| 24V| 36V| 48V
RATED CURRENT| 15A| 14A| 6.3A| 5A| 3.2A| 2.1A| 1.6A
CURRENT RANGE| 0 ~ 20A| 0 ~ 18.7A| 0 ~ 8.3A| 0 ~ 6.7A| 0 ~ 4.2A| 0 ~
2.8A| 0 ~ 2.1A
RATED POWER| 49.5W| 70W| 75.6W| 75W| 76.8W| 75.6W| 76.8W
PEAK LOAD (23.5CFM)| 66W| 93.5W| 99.6W| 100.5W| 100.8W| 100.8W| 100.8W
RIPPLE & NOISE (max.) **Note.2| 60mVp-p| 60mVp-p| 60mVp-p| 60mVp-p|
100mVp-p| 100mVp-p| 100mVp-p
VOLTAGE ADJ. RANGE| 2.9 ~ 3.6V| 4.75 ~ 5.5V| 11.4 ~ 13.2V| 13.5 ~ 16.5V|
22.8 ~ 27.6V| 34.2 ~ 39.6V| 45.6 ~ 52.8V
VOLTAGE TOLERANCE Note.3| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%|
±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.5%| ±1.5%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 500ms, 30ms/230VAC 500ms, 30ms/115VAC at full
load
HOLD UP TIME (Typ.)| 90ms/230VAC 20ms/115VAC at full load
INPUT| VOLTAGE RANGE| 90 ~ 264VAC 127 ~370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY(Typ.)| 73%| 78%| 82%| 83%| 85%| 86%| 86%
AC CURRENT (Typ.)| 1.5A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 25A/115VAC 50A/230VAC
LEAKAGE CURRENT(max.) Note.4| Earth leakage current < 150μA/264VAC ,
Touch current < 100μA/264VAC
PROTECTION| OVERLOAD| 140 ~ 180% rated output power
Protection type: Hiccup mode, which recovers automatically after the fault
condition is removed.
OVERVOLTAGE| 3.8 ~ 4.5V| 5.7 ~ 6.8V| 13.8 ~ 16.2V| 17.2 ~ 20.3V| 27.6 ~
32.4V| 41.4 ~ 48.6V| 55.2 ~ 64.8V
Protection type: Shut down o/p voltage, re-power to recover
ENVIRONMENT| WORKING TEMP.| -20 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, and Z axes
OPERATING ALTITUDE Note.5| 3000 meters
SAFETY & EMC**
(Note 7)
| ****
SAFETY STANDARDS
| IEC 60601-1:2005,TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2
CAN/CSA C22.2 No. 60601-1:2014+A2, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1(by request)
ISOLATION LEVEL| Primary-Secondary:2xMOPP, Primary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/
70% RH
EMC EMISSION
| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility
| ****
BS EN/EN61000-4-3
| Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption
| ****
BS EN/EN61000-4-11
| 100% dip 1 period, 30% dip 25 periods,
100% interruptions 250 periods
OTHERS| MTBF| 2837.6K hrs min. Telcordia SR-332 (Bellcore);
446.8K hrs min. MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 12776.231mm or 5″ 3″ 1.22″ inch
PACKING| 0.26Kg; 63pcs/16.3Kg/1.28CUFT
NOTE| 1. All parameters NOT specially mentioned are measured at 230VAC
input, rated load, and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μF & 47μF parallel capacitor.
3. Tolerance: includes set up tolerance, line regulation, and load regulation.
4. Touch current was measured from primary input to DC output.
5. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitudes higher than 2000m(6500ft).
6. Heat Sink HS1,HS2,HS3 can not be shorted.
7. The power supply is considered a component that will be installed into the final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re- confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
※ Product Liability Disclaimer: For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
Block Diagram
- fosc: 65KHz
Derating Curve
Output Derating VS Input Voltage
Mechanical Specification
AC Input Connector (CN1): JST B3P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | AC/N | JST VHR |
or equivalent
| JST SVH-21T-P1.1
or equivalent
2| No Pin
3| AC/L
DC Output Connector (CN2) : JST B8P-VH or equivalent
: Grounding Required
- HS1, HS2, HS3 cannot be shorted.
- M1 is safety ground. For better EMC performance, Please secure an electrical connection between M1, M2, and chassis grounding.
Remote Sense(CN3): JST B2B-XH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | RS+ | JST XHP |
or equivalent
| JST SXH-001T-P0.6
or equivalent
2| RS-
Installation Manual
Please refer to: http://www.meanwell.com/manual.html.
User’s Manual
Documents / Resources
| MEAN
WELL RPS-75 Series 75W Reliable Green Medical Power
Supply
[pdf] Owner's Manual
RPS-75-3.3, RPS-75-5, RPS-75-12, RPS-75-15, RPS-75-24, RPS-75-36, RPS-75-48,
RPS-75 Series 75W Reliable Green Medical Power Supply, RPS-75 Series, 75W
Reliable Green Medical Power Supply, Reliable Green Medical Power Supply,
Green Medical Power Supply, Medical Power Supply, Power Supply, Supply
---|---
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>