MEAN WELL RPS-75 Series 75W Reliable Green Medical Power Supply Owner’s Manual

August 14, 2024
MEAN WELL

MEAN-WELL-Logo

MEAN WELL RPS-75 Series 75W Reliable Green Medical Power Supply

MEAN-WELL-RPS-75-Series-75W-Reliable-Green-Medical-Power-Supply-
Product

Product Information

Specifications

Model DC Voltage Rated Current Rated Power
RPS-75-3.3 3.3V 15A 49.5W
RPS-75-5 5V 14A 70W
RPS-75-12 12V 6.3A 75.6W

Product Usage Instructions

Features:

  • Reliable Green Medical Power Supply
  • Complies with EN 60601-1 and IEC 60601-1 standards

Description:
The RPS-75 series is a medical power supply designed for various applications such as oral irrigators, hemodialysis machines, medical computer monitors, and sleep apnea devices.

Applications:

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

FAQ

What is the GTIN CODE for the product?
The GTIN CODE can be found on the Mean Well website using the MW Search tool.

Features

  • 5″×3″ compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • 75W convection, 100W force air
  • EMI Class B for Class I configuration
  • No load power consumption <0.75W
  • Remote sense function
  • Protections: Short circuit/Overload/Over voltage
  • Lifetime > 80K hours
  • Operating altitude up to 3000 meters
  • 3 years warranty

Description

RPS-75 is a 75W highly reliable green PCB-type medical power supply with a high power density on the 5″ by 3″ footprint. It accepts 90~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 86% and the extremely low no-load power consumption is down below 0.75W. RPS-75 can be used for Class I (with FG) system design. The extremely low leakage current is less than 150 MA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF-rated “patient contact” medical system equipment.

Applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

GTIN CODE

MW Search: https://www.meanwell.com/serviceGTlN.aspx.

Model Encoding

MEAN-WELL-RPS-75-Series-75W-Reliable-Green-Medical-Power-Supply-
Fig-2

SPECIFICATION

MODEL| RPS-75-3.3| RPS-75-5| RPS-75-12| RPS-75-15| RPS-75-24| RPS-75-36| RPS-75-48
---|---|---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 3.3V| 5V| 12V| 15V| 24V| 36V| 48V
RATED CURRENT| 15A| 14A| 6.3A| 5A| 3.2A| 2.1A| 1.6A
CURRENT RANGE| 0 ~ 20A| 0 ~ 18.7A| 0 ~ 8.3A| 0 ~ 6.7A| 0 ~ 4.2A| 0 ~ 2.8A| 0 ~ 2.1A
RATED POWER| 49.5W| 70W| 75.6W| 75W| 76.8W| 75.6W| 76.8W
PEAK LOAD (23.5CFM)| 66W| 93.5W| 99.6W| 100.5W| 100.8W| 100.8W| 100.8W
RIPPLE & NOISE (max.) **Note.2| 60mVp-p| 60mVp-p| 60mVp-p| 60mVp-p| 100mVp-p| 100mVp-p| 100mVp-p
VOLTAGE ADJ. RANGE| 2.9 ~ 3.6V| 4.75 ~ 5.5V| 11.4 ~ 13.2V| 13.5 ~ 16.5V| 22.8 ~ 27.6V| 34.2 ~ 39.6V| 45.6 ~ 52.8V
VOLTAGE TOLERANCE Note.3| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.5%| ±1.5%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 500ms, 30ms/230VAC          500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)| 90ms/230VAC          20ms/115VAC at full load
INPUT| VOLTAGE RANGE| 90 ~ 264VAC         127 ~370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY(Typ.)| 73%| 78%| 82%| 83%| 85%| 86%| 86%
AC CURRENT (Typ.)| 1.5A/115VAC          1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 25A/115VAC          50A/230VAC
LEAKAGE CURRENT(max.) Note.4| Earth leakage current < 150μA/264VAC , Touch current < 100μA/264VAC
PROTECTION| OVERLOAD| 140 ~ 180% rated output power
Protection type: Hiccup mode, which recovers automatically after the fault condition is removed.
OVERVOLTAGE| 3.8 ~ 4.5V| 5.7 ~ 6.8V| 13.8 ~ 16.2V| 17.2 ~ 20.3V| 27.6 ~ 32.4V| 41.4 ~ 48.6V| 55.2 ~ 64.8V
Protection type: Shut down o/p voltage, re-power to recover
ENVIRONMENT| WORKING TEMP.| -20 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, and Z axes
OPERATING ALTITUDE Note.5| 3000 meters
SAFETY & EMC**

(Note 7)

| ****

SAFETY STANDARDS

| IEC 60601-1:2005,TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2

CAN/CSA C22.2 No. 60601-1:2014+A2, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1(by request)

ISOLATION LEVEL| Primary-Secondary:2xMOPP,  Primary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC  I/P-FG:2KVAC  O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH



EMC EMISSION

| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact


RF field susceptibility

| ****

BS EN/EN61000-4-3

| Level 3, 10V/m( 80MHz~2.7GHz )

Table 9, 9~28V/m( 385MHz~5.78GHz )

EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m


Voltage dip, interruption

| ****

BS EN/EN61000-4-11

| 100% dip 1 period, 30% dip 25 periods,

100% interruptions 250 periods

OTHERS| MTBF| 2837.6K hrs min.      Telcordia SR-332 (Bellcore); 446.8K hrs min.       MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 12776.231mm or 5″ 3″ 1.22″ inch
PACKING| 0.26Kg; 63pcs/16.3Kg/1.28CUFT
NOTE| 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load, and 25℃ of ambient temperature.

2. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μF & 47μF parallel capacitor.

3. Tolerance: includes set up tolerance, line regulation, and load regulation.

4. Touch current was measured from primary input to DC output.

5. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitudes higher than 2000m(6500ft).

6. Heat Sink HS1,HS2,HS3 can not be shorted.

7. The power supply is considered a component that will be installed into the final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re- confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

※ Product Liability Disclaimer: For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

Block Diagram

  • fosc: 65KHz

MEAN-WELL-RPS-75-Series-75W-Reliable-Green-Medical-Power-Supply-
Fig-3

Derating Curve

MEAN-WELL-RPS-75-Series-75W-Reliable-Green-Medical-Power-Supply-
Fig-4

Output Derating VS Input Voltage

MEAN-WELL-RPS-75-Series-75W-Reliable-Green-Medical-Power-Supply-
Fig-5

Mechanical Specification

MEAN-WELL-RPS-75-Series-75W-Reliable-Green-Medical-Power-Supply-
Fig-6

AC Input Connector (CN1): JST B3P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 AC/N JST VHR

or equivalent

| JST SVH-21T-P1.1

or equivalent

2| No Pin
3| AC/L

DC Output Connector (CN2) : JST B8P-VH or equivalent

: Grounding Required

  1. HS1, HS2, HS3 cannot be shorted.
  2. M1 is safety ground. For better EMC performance, Please secure an electrical connection between M1, M2, and chassis grounding.

Remote Sense(CN3): JST B2B-XH or equivalent

Pin No. Assignment Mating Housing Terminal
1 RS+ JST XHP

or equivalent

| JST SXH-001T-P0.6

or equivalent

2| RS-

Installation Manual

Please refer to: http://www.meanwell.com/manual.html.

User’s Manual

Documents / Resources

| MEAN WELL RPS-75 Series 75W Reliable Green Medical Power Supply [pdf] Owner's Manual
RPS-75-3.3, RPS-75-5, RPS-75-12, RPS-75-15, RPS-75-24, RPS-75-36, RPS-75-48, RPS-75 Series 75W Reliable Green Medical Power Supply, RPS-75 Series, 75W Reliable Green Medical Power Supply, Reliable Green Medical Power Supply, Green Medical Power Supply, Medical Power Supply, Power Supply, Supply
---|---

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

MEAN WELL User Manuals

Related Manuals