PYRAMID HC80 Beampath Helium Flow Controller User Manual
- June 6, 2024
- PYRAMID
Table of Contents
PYRAMID HC80 Beampath Helium Flow Controller
Product Information
Specifications
- Model: HC80
- Document ID: 2765226245
- Version: v1
Product Usage Instructions
HC80 Programming
The HC80 programming concepts and methods are explained in this manual,
focusing on device-specific IO and functionality unique to the HC80.
Remote Sensor IO
The remote sensor IO are associated with the HC80’s remote chamber sensor
module. These IO provide information on chamber oxygen concentration,
pressure, and temperature.
- Path: /hc80/remote_sensor/luminox/oxygen_percent
- Description: READONLY NUMBER – Chamber oxygen concentration as a percentage
- Path: /hc80/remote_sensor/pressure
- Description: READONLY NUMBER – Chamber pressure in PSI
- Path: /hc80/remote_sensor/temperature
Mass Flow Controller IO
The mass flow controller IO relate to the two mass flow controllers inside
the HC80, used for purging and maintaining the chamber at different flow
rates.
-
Path: /hc80/mfc_large/flow_rate
-
Description: READONLY NUMBER – Large flow controller flow rate
in SCCM -
Path: /hc80/mfc_small/flow_rate
-
Description: READONLY NUMBER – Small flow controller flow rate
in SCCM
Relay IO
The relay IO provides information on the state of the HC80
relay, indicating whether it is opened, closed, or locked due to an
interlock.
- Path: /hc80/relay/state
- Description: READONLY STRING – The state of the HC80 relay
FAQ
-
Q: What should I do if the remote sensor cable is disconnected?
A: If the remote sensor cable is disconnected, the IO associated with the remote sensor will no longer accurately reflect the chamber state. It is recommended to reconnect the cable securely to ensure proper functionality. -
Q: How does the HC80 compensate for error sources in reported flow rates?
A: The HC80’s software automatically compensates for error sources in reported flow rates, including latency and potential offsets from zero. This compensation is done internally by the HC80’s software. -
Q: What are the possible states of the HC80 relay?
A: The HC80 relay can be in one of three states: opened, closed, or locked. A locked relay is open but cannot be closed due to an interlock mechanism.
Introduction
Document ID: 2473297357
Author | @ Matthew Nichols |
---|---|
Owner | Project Lead |
Purpose | Explain the programming concepts necessary to use the API and |
extend the product through external applications.
Scope| HC80 related programming concepts.
Intended Audience| Software developers interested in using the
product.
Process| Standard Manual Creation Process
Training| NOT APPLICABLE
References
Document | Document ID | Author | Version |
---|---|---|---|
IGX – Programmer Manual | 2439249921 | @ Matthew Nichols | 2 |
HC80 Programming
The concepts and methods described in this manual build on the concepts established in the IGX – Programmer Manual. Please see that document for explanation and examples of how basic IGX programming and interfaces work. This manual will only cover the device-specific IO and functionality that is unique to the HC80.
Remote Sensor IO
These IO are associated with the HC80s remove chamber sensor module. This
module is separate from the HC80 unit itself and connected by a cable that
goes to the helium chamber. If the cable is disconnected, these IO will no
longer accurately reflect the chamber state.
IO Path | Description |
---|---|
/hc80/remote_sensor/luminox/oxygen_percent | READONLY **NUMBER** |
Chamber oxygen concentration as a percentage.
/hc80/remote_sensor/pressure| READONLY **NUMBER Chamber pressure in
PSI. This is a differential pressure sensor that measures the difference
between the chamber and atmospheric pressure.
/hc80/remote_sensor/temperature| READONLY NUMBER** Chamber temperature
in °C. This temperature is measured from the exhaust gas outlet and may not
exactly reflect the chamber temperature.
Mass Flow Controller IO
These IO relate to the two mass flow controllers inside the HC80. One MFC is
used for purging the chamber at a high flow rate, and the other is used to
maintain the chamber with a very small flow rate. These flow rates are analog
signals and are measured inside the controller. The reported flow rate has a
latency and potentially an offset from zero. The HC80s software automatically
compensates for these error sources internally.
IO Path | Description |
---|---|
/hc80/mfc_large/flow_rate | READONLY **NUMBER** The large flow |
controller flow rate in SCCM.
/hc80/mfc_small/flow_rate| READONLY **NUMBER** The small flow
controller flow rate in SCCM.
Relay IO
IO Path | Description |
---|---|
/hc80/relay/state | READONLY **STRING** The state of the HC80 relay. |
Locked relays are open but cannot be closed due to an interlock.
States: “opened”, “closed”, or “locked”
Version Control
Version| Description| Saved by| Saved on|
Status
---|---|---|---|---
v1| Initial Version| Matthew Nichols| Apr 12, 2024 5:47 PM| APPROVED
Document Control
Current document version: v.1