LUMEX LF1050 Patient Lift User Manual
- June 5, 2024
- LUMEX
Table of Contents
LUMEX LF1050 Patient Lift
Read this manual before operating your Patient Lift.
Save this manual for future use.
The most current version of this manual can be found online at
www.grahamfield.com.
GF Health Products, Inc. is not responsible for typographical errors. For the most updated and current information on packaging, warranties, products and specifications, including the most current version of these instructions, please visit our website at www.grahamfield.com. Graham-Field and Lumex are registered trademarks of GF Health Products, Inc.
INTRODUCTION
Congratulations on your purchase of the Lumex LF1050 Patient Lift. The
following pages will provide you with important safety and operating
instructions on the use of patient lifts, slings, and accessories, as well as
maintenance and warranty information. Read this manual carefully before
operating your patient lift and refer to it as often as needed. Consult your
authorized distributor and/or healthcare professional with any questions or
concerns regarding safe and effective techniques for operating your patient
lift.
DO NOT attempt any transfer without prior approval of the patient’s healthcare
provider. With their assistance, select the sling that best serves the needs,
abilities, comfort and safety of the person being lifted. It is important that
you use correct slings adapted to the body shape and weight and ability of the
patient and the specific type of transfer needed.
Upon receipt of the patient lift, and prior to each use, inspect it closely to
ensure that nothing
is damaged, there are no loose or broken parts, that all patient lift parts
are in the proper place, etc. Immediately replace any worn or broken
components. Your patient lift should receive maintenance on a regular schedule
and should be inspected daily for proper operation. Contact your Graham-Field
authorized distributor for service and repairs.
Before lifting anyone, thoroughly read the instructions in this manual.
Ideally, observe a team of trained experts performing the lifting procedures.
The operator should then practice the entire lifting procedure several times
with proper supervision and with a capable individual acting as patient. It is
extremely important to know how a patient will feel in the patient lift. We
suggest that the operator try all the various lifting slings and equipment
until the operator is confident with how the patient lift operates. The
lifting procedures should be explained to the patient before the actual
lifting. Remember that safety requires the constant attention of the lift
occupant and the operator. Use extreme care and common sense when operating a
patient lift.
This guide covers the use of the Lumex LF1050 Patient Lift. Keep in mind that
the basic safety procedures included in this manual are to be used as a guide
only. You may find it necessary
to develop your own unique methods for safely utilizing the patient lift.
Again, consult your healthcare professionals for their recommendations and
never hesitate to ask for their assistance.
Info: Before attempting to assemble the patient lift, refer to the section
titled “ASSEMBLY INSTRUCTIONS” contained elsewhere in this manual. Contact
your Graham-Field authorized distributor if additional assistance is needed.
Info: In this manual, the person being lifted is referred to as the “patient”
or “occupant” and the person operating the patient lift is called the
“operator” or “attendant”.
Intended use
The Lumex LF1050 Patient Lift is intended to aid in the transfer of an individual who is unable to independently transfer between bed, chair, bathtub, wheelchair, and/or commode.
WARNING: The Lumex LF1050 Patient Lift is designed for transfer only (to and from bed, chair, bathtub, toilet, wheelchair, commode etc.). This Patient Lift is not intended to be used for transporting or moving a patient from one location to another.
SAFETY PRECAUTIONS
IMPORTANT: Before using patient lift, please read and adhere to the following safety precautions and warnings. Failure to do so could result in serious personal injury or damage to your patient lift.
Always consult your healthcare professional to determine safe methods most
suitable for your individual abilities. Protect yourself, your attendant and
patient lift by having it serviced regularly. If you experience any
malfunction, contact your Graham-Field authorized distributor immediately, as
a hazardous condition could result, causing personal injury or damage to your
patient lift.
Periodic inspection, adjustment and replacement of worn parts are necessary to
provide years of excellent service. Refer to CARE AND MAINTENANCE section of
this manual.
Maintenance MUST be performed by qualified personnel ONLY.
Significance of safety statements
Please note the following special statements, used throughout this manual, and
their significance:
WARNING: Indicates a potential hazard situation or unsafe practice that,
if not avoided, could result in death or serious personal injury.
CAUTION: Indicates a potential hazard situation or unsafe practice that,
if not avoided, could result in minor or moderate personal injury.
NOTICE: Indicates a potential hazard situation or unsafe practice that, if not avoided,could result in product or property damage.
Info: Provides application recommendations or other useful information to ensure that you get the most from your product.
WARNINGS
- Important! Read and understand these instructions before assembling or using the Lumex LF1050 Patient Lift. If you do not understand any part of these warnings, cautions or instructions, contact a healthcare professional for direction in the use of this product. If the Lumex LF1050 Patient Lift is not properly assembled, personal injury and damage to the LF1050 could result.
- If components are damaged or missing, contact your Graham-Field aauthorized distributor immediately. DO NOT use substitute parts. Use only Lumex replacement parts. The use of non-Lumex replacement parts could cause personal injury, property damage, and void the warranty.
- Unauthorized modification of your patient lift could cause personal injury, property damage, and void the warranty.
- The Lumex LF1050 Patient Lift maximum weight capacity is 400 lb, EVENLY DISTRIBUTED. DO NOT exceed the maximum weight capacity.
- GF Health Products, Inc. assumes no responsibility for any damage or injury caused by improper assembly or use of this product.
- Check all parts for shipping damage before using. In case of damage, DO NOT USE the equipment. Contact the carrier or your Graham-Field authorized distributor for further instructions.
- Patient lift may tip over if used incorrectly. Read and adhere to the operating instructions prior to lifting anyone.
- GF Health Products, Inc. strongly recommends that two caregivers take part in the lifting process.
- When transferring a patient, always keep the patient centered over the base.
- Ensure that lifting sling loops are correctly attached to the hooks to prevent the patient from sliding or falling out of the sling, which could result in personal injury. Never use the patient lift to lift or transfer anything other than a person.
- Do not lock the brakes or block the wheels when lifting. The casters must be FREE to roll to allow the patient lift to stabilize itself when the patient is initially lifted from a chair, bed, or any stationary object.
- Do not attempt to maneuver the lift by pushing on the mast, boom, actuator, or patient. Always maneuver the lift with the handle.
- During lifting or lowering, whenever possible, always keep the patient lift legs in the maximum open position.
- To facilitate movement, close patient lift legs before moving the lift.
- Replace any worn or broken parts immediately! Torn, cut, frayed, or broken slings can fail, resulting in serious personal injury. Use only slings that are in good condition.
- Use extreme care, caution, and common sense when utilizing a patient lift. Special attention must be given to patients who are disoriented and cannot cooperate while being lifted, such as comatose, agitated, spastic, or mentally confused persons. Consult their healthcare professional for assistance in lifting these types of patients.
- California Proposition 65 WARNING: This product contains a chemical known to the State of California to cause cancer and reproductive or developmental harm.
- GF Health Products, Inc. specifically disclaims responsibility for any bodily injury or property damage which may occur during any use which does not comply with federal, state or local laws or ordinances.
FEATURES
LIFT COMPONENTS
a. Boom
b. Mast
c. Base
d. Spreader bar
e. Pendant
f. Actuator
g. Control box
h. Caster brake
i. Front caster
j. Emergency Lowering Device
CONTROL BOX COMPONENTS
HANDLING PROCEDURES
SHIPPING AND TRANSPORTATION INSTRUCTIONS
If the patient lift is to be reshipped by common carrier, it should be packed
in the same carton in which it was received.
Unpacking
WARNING: The lift is heavy. To avoid personal injury, DO NOT attempt to remove the lift from the box without assistance.
- Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier or your Graham-Field authorized distributor.
- Remove all loose packing from the carton.
- Carefully remove all the components from the carton.
Info: Unless the patient lift is to be used immediately, retain boxes, containers and packing material for use in storing until use of patient lift is required.
Inspection
- Examine the exterior of the patient lift for nicks, dents, scratches or other damage.
- Inspect all components.
Storage
- Store the repackaged patient lift in a dry area.
- DO NOT place other objects on top of the repackaged patient lift.
ASSEMBLY INSTRUCTIONS
WARNING: The lift is heavy. To avoid personal injury, DO NOT attempt to assemble the lift without assistance.
- Engage the lift brakes, as shown above.
- Remove the bolts from the bottom of the mast and base.
- Pull the mast, boom, and spreader bar to an upright position, as shown above.
- Reinstall the bolts in the holes at bottom of mast. Tighten the nuts.
- Attach the control box hanging bracket to the mast, as shown above. Tighten the nuts.
- Place the control box in the bracket, as shown above.
- Attach the actuator bottom to the mast bracket and insert retainer pin, as shown above. Insert retaining ring through retainer pin hole to secure actuator base to mast bracket.
- Attach the actuator top to the boom bracket and insert retainer pin, as shown above. Insert retaining ring through retainer pin hole to secure actuator top to boom bracket.
- Attach the spreader bar bracket to the boom and insert retainer pin, as shown above. Insert retaining ring through retainer pin to secure spreader bar to boom.
- Insert the actuator plug into the control box, as shown above.
- Insert the pendant plug into the control box, as shown above.
Info: Before using the lift, charge the batteries for ten to twelve hours (see instructions on following page).
CHARGING THE BATTERIES
Charge the batteries after assembly.
-
Ensure the battery box power switch is ON (i.e., the red button is up).
-
Insert the charger plug into the battery box charging socket (the top slot).
-
Insert the other charger plug into the power supply.
-
All LED indicator lights should illuminate during charging (see table below).
-
When one green LED indicator light illuminates, the batteries require approximately two to three hours to be fully charged.
-
When only the red LED indicator light illuminates, the batteries require approximately seven to nine hours to be fully charged.
NOTICE: Do not charge the batteries for more than twelve hours. -
Unplug the charger FIRST before using the lift.
Info: The lift will not operate while the batteries are charging.
The lift will not operate while the red reset button is down. -
Replace the batteries when frequent charging is required.
Important Charging Information
Push lift to an appropriate location and charge the batteries with the charger provided.
NOTICE: Avoid unplugging the hand control and motor from the control box. Frequent plugging and unplugging of the hand control and motor to and from the control box may damage the control box.
CHECKING BATTERY CAPACITY
Info: To check remaining battery capacity, press the battery indicator
(blue circle with battery symbol) while the lift is not in motion. Remaining
battery capacity is indicated by LED indicator lights and explained in the
following table.
Red | Amber | Green | Green | Green | Status |
---|---|---|---|---|---|
ON | ON | ON | ON | ON | Fully charged |
ON | ON | ON | OFF | OFF | Needs charge |
ON | ON/OFF | OFF | OFF | OFF | **Charge immediately—DO |
NOT use lift
ON| OFF| OFF| OFF| OFF| Bad batteries or bad
connection**
ALARM BUZZERS
The control box has an alarm buzzer that provides information on the status of
the batteries and the lift. To stop the buzzer, push down the RED RESET button
or press the battery indicator.
Buzzer type | Status | Note |
---|---|---|
2 beeps per second | Low battery (red/amber lights | |
illuminate) | Charge batteries immediately | |
3 beeps per second | Low battery (red light illuminates) | **Charge |
batteries immediately
6 beeps per second| Very low battery (no lights illuminate)|
Charge batteries immediately
3 beeps every other second| Bad battery (no lights illuminate)|
Replace batteries immediately
3 long beeps| Overloaded. Batteries are charged but lift does not
lift| Maximum load capacity exceeded**
OPERATING INSTRUCTIONS
WARNING :
- During lifting and lowering, whenever possible, always keep the base legs of the lift in the widest position.
- Use only Lumex slings specially designed for the Lumex LF1050 Patient Lift.
- Do not put anything (e.g. cushion, pad, etc.) between the user and the sling. This may cause user to slide out of the sling and could cause injury.
- GF Health Products, Inc. strongly recommends that two caregivers take part in the lifting process.
Info: Instructional videos that review the application of several sling
styles and lifting scenarios are available at www.grahamfield.com.
Info: While transferring a patient, the patient lift legs can be in
either open or closed position.
Preparation before lifting
- Turn on the power by turning the RED RESET BUTTON clockwise.
- Press the hand pendant UP or DOWN button once.
- After the battery indicator lights de-illuminate, push the hand pendant UP button and DOWN button to ensure the lift is operating correctly.
- Position the lift base around or under the patient.
- Widen the base.
LIFTING THE PATIENT
WARNING:
- Do not attempt to transfer a patient without prior approval of the patient’s healthcare professional and without having read the instructions and practiced using the patient lift.
- Use special care for those patients who cannot cooperate while being lifted.
- Ensure all bolts, hooks, and loops are securely fastened.
- Use only Lumex slings with the Lumex LF1050 Patient Lift.
- DO NOT use torn, cut, frayed or broken slings as injury could result.
- Ensure the floor area is clear of any obstacle that could cause a caster to stop and possibly tip the patient lift.
Lifting the patient from a bed
- Position patient face-up in center of bed. Roll patient onto their side facing away from you.
- Fold sling lengthwise, outside out, and place it halfway under patient with lower edge just beneath knees and sling handle on the outside, facing away from patient. If the sling has a head support, ensure that it supports the patient’s neck and head.
- Roll patient onto their side facing you. Pull sling out beneath patient inside up, smooth it, and center patient on sling.
- Pull the leg loops forward and under the thighs. Cross the loops, one through the other.
- Roll the patient lift under the bed, locating the spreader bar over the patient. Be careful not to lower the spreader bar onto the patient. Center the spreader bar over the patient before lifting. WARNING: Do not lock or block the patient lift casters when lifting. The casters must be free to roll so that the patient lift can stabilize as the patient is lifted from the bed. WARNING: To ensure stability while lifting or lowering a patient, the patient lift legs should be in the maximum open position.
- Hook the sling loops onto the spreader bar.
- If the patient is in a hospital bed, raise the bed’s head section to elevate the patient.
- Use the hand pendant to carefully and slowly raise the patient. If possible, lower the bed. Reposition the patient as needed, using the handle on the back of the sling, to ensure a safe and comfortable sitting position as the patient is raised.
- Raise the patient until the buttocks are just above the mattress. Lift the patient’s legs and turn the patient to face the lifting column. Move the patient lift away from the bed. Ensure that the patient’s arms and legs clear the patient lift, bed, or any other obstacles.
Lowering the patient onto a bed
- Raise the patient lift until the patient’s buttocks are above the surface of the bed. Ensure that there are no obstacles underneath or around the bed. Move the patient lift under the bed. Ensure the patient does not hit any portion of the bed and that the patient’s legs have cleared the bed.
- Center the patient over the bed, and rotate the patient to face the foot section of the bed. Use the hand pendant to SLOWLY lower the patient onto the bed. Support the patient’s head, if necessary, while being lowered. Remove the sling by reversing the procedures used when lifting patient from bed.
Lifting the patient from the floor
- Position the patient lift behind the patient, placing a pillow over the base to protect and support the patient’s head and neck. Place the patient’s head carefully on the pillow. Position the patient lift with the spreader bar centered over the patient.
- Lower the spreader bar. Attach the sling to the spreader bar hooks.
- Lift the patient up above the floor. Position the patient in the sling by pulling the handle on the back of the sling.
- Turn the patient to face the lifting column. Remove pillow from base and place the patient’s feet on the base, straddling the column for extra stability.
Lowering the patient to the floor
- Place a pillow on the base to protect the patient’s head and neck (this is unnecessary when using a sling with a head support).
- Reverse the above procedures to lower.
Transferring the patient from a chair to a wheelchair or commode chair
WARNING: GF Health Products, Inc. recommends that two attendants be used when transferring a patient to and from a wheelchair.
-
Fold the sling.
-
While maintaining support, gently lean the patient forward. Place the folded sling behind the patient’s back and push it down until it touches the seat of the chair.
-
Use the loops to pull the leg sections of the sling forward and beneath the patient’s thighs. Then cross the loops, one through the other.
-
Move the patient lift around to the front of the chair.
-
Hook all sling loops onto the spreader bar.
-
Carefully raise the patient.
-
The second attendant should now move the wheelchair into position.
-
Engage the wheelchair wheel locks to prevent movement.
WARNING: Before transfer, ensure wheelchair wheel locks are in locked position. Wheelchair wheel locks must be in LOCKED POSITION before lowering the patient into the wheelchair, or unexpected wheelchair movement could result, which may result in serious injury to both the patient and the attendants. -
Position the patient over the wheelchair with their back against the back of the wheelchair.
-
SLOWLY lower the patient into the wheelchair.
-
With one attendant behind the wheelchair and the other operating the lift, the attendant behind the chair should pull back on the handle or sides of the sling to place the patient into the back of the chair. This will maintain a good center of balance and prevent the chair from tipping forward.
Removing the sling
Gently pull the leg sections to the side, out from under the patient’s thighs.
Standing by the patient’s side, lean the patient forward while supporting the
patient with one hand. Pull the sling up from behind the patient’s back and
position the patient in the chair.
WARNING: DO NOT pull sharply on the sling. Pulling sharply on the sling may cause the patient to fall forward, resulting in personal injury. Always support the patient by keeping one arm around the patient’s shoulders when placing or removing the sling.
Info: If possible, the patient should be encouraged to take an active part in removing the sling. The patient can lean forward, lift their thighs, and help in removing the leg sections.
Transferring the patient from a bed to a wheelchair or commode chair
-
For smooth, easy lifting, have the wheelchair or commode chair in ready position with wheel locks in locked position.
-
With patient face-up in center of bed, roll patient onto their side away from attendant.
-
Place sling, folded lengthwise, halfway under patient with lower edge just beneath knees.
-
Roll patient to side toward attendant, pull sling to smooth it, and center patient on sling.
-
Pull the leg loops forward and under the thighs.
-
Cross the loops, one through the other.
-
With base of lift under bed, use the hand pendant to carefully lower spreader bar.
-
Attach sling to spreader bar.
-
Use the hand pendant to slowly lift patient, maintaining patient’s position facing attendant, until patient’s feet swing easily off the bed.
-
Position patient above wheelchair or commode chair. Lock lift brakes.
WARNING: Before transfer, ensure wheelchair or commode chair wheel locks are in locked position. Wheelchair wheel locks must be in LOCKED POSITION before lowering the patient into the wheelchair, or unexpected wheelchair movement could result, which may result in serious injury to both the patient and the attendants. -
Use the hand pendant to SLOWLY lower patient. During descent, assist patient in attaining correct sitting posture.
-
When patient is seated, push down on boom to slacken sling loops. Patient can remain seated on sling.
-
For commode chair use, adjust clothing before moving lift to straddle commode chair. Keep sling loops taut and ensure that patient is in a comfortable position.
EMERGENCY RELEASE
In case of lift failure while using the Lumex LF1050 Patient Lift, follow the
procedure below to safely lower the user. The Emergency Lowering Device is
located at the top of the actuator shaft. It is intended for use if the
actuator fails to operate while a patient is suspended.
The device consists of a plastic collar ring that should be turned clockwise
continually until the patient has been lowered.
Contact your dealer immediately if standard troubleshooting techniques do not
correct the failure. Do not attempt to lift patients until all failure issues
have been safely resolved.
Info: The emergency lowering device is intended for use during lift failure. This device will allow lowering of patients only. Please contact your Graham-Field authorized distributor immediately in case of failure.
CARE AND MAINTENANCE
Proper care and maintenance are essential to keeping your Lumex LF1050 Patient Lift in a safe operating condition. In addition to inspecting the unit before each use, periodic maintenance checks should be done. We further recommend that this unit be serviced at least once annually by a Graham-Field authorized distributor.
WARNING:
- It is extremely important that the patient lift be inspected before each use. Ensure that all hardware and accessories are secure and that the actuator is functioning properly. Failure to do so could result in patient/attendant injury or damage to your patient lift. Torn, cut, frayed or broken slings can fail, resulting in serious injury. Only use slings in good condition. Inspect before each use. Destroy and discard old worn and unusable slings.
- Service and repair of the Lumex LF1050 Patient Lift should be performed only by a Graham-Field authorized distributor. Graham-Field is not responsible for any consequence resulting from any unauthorized service or repair.
- Unauthorized modification of the patient lift or the use of non-Lumex replacement parts may change the structure of the lift and could create a hazardous condition, which may result in serious injury and will void the warranty.
- The electronics and actuator contain no serviceable components. If service is required, the unit must be returned to the factory for repair. DO NOT attempt to open the electronics or actuator or obtain local service; this will void the warranty and may result in damage to the unit. Consult your Graham-Field authorized distributor for further information.
When you believe a component or part is not functioning properly, immediately contact your Graham-Field authorized distributor, as a potentially hazardous condition could exist.
BEFORE EACH USE
Before each use of the patient lift, ensure that:
- All nuts and bolts are tight
- The base can be easily widened
- All lift parts are in place
- All casters turn freely
- Caster brakes can be engaged
- All necessary items (i.e. slings, wheelchairs, etc.) are accessible and ready for use
- Check battery capacity
AT LEAST ONCE A MONTH
The lift should be thoroughly inspected by a person qualified to recognize any
signs of wear, and looseness of bolts or parts. Replace any worn parts
immediately.
LUBRICATION
When the lift is serviced, and every month thereafter, place a drop of oil on
the following points:
- Top of mast
- Spreader hinge
- Caster axles
CLEANING
Regular cleaning is recommended every three months; clean the lift more often
as needed. A soft cloth, dampened with water and a mild NON-ABRASIVE detergent
(household cleaner, soapy water, or hospital-grade disinfectant) is all that
is needed to clean your patient lift. Automobile wax or furniture polish will
help maintain the finish over a long period of time.
NOTICE: DO NOT wash patient lift underwater pressure or steam clean.
CARE OF SLINGS
Lumex slings are washable at 150° F and autoclavable at 185° F for 30 minutes.
The head support’s plastic insert should be removed before washing. The slings
can be air dried or dried by using a dryer on LOW heat.
NOTICE: DO NOT bleach slings!
MAINTENANCE SCHEDULE
Info: Only qualified persons should service and repair your Lumex LF1050
Patient Lift. Regular maintenance of your patient lift is necessary to ensure
continuing proper and safe operation. Read and observe the following
recommended maintenance schedule:
Item | Inspect | Inspect / adjust |
---|---|---|
initially | monthly | |
Boom & spreader bar |
Check connections between 1) boom and spreader bar and 2) boom and mast for
improper connection, looseness, or wear| X| X|
Inspect boom for bending or deflection| X| X|
Mast| | |
Inspect mast for bending or deflection| X| X|
Inspect steering bar for damage or loose hardware| X| X|
Base and foot pedal| | |
Ensure all hardware is tight| X| X|
Ensure casters and axle bolts are tight| X| X|
Inspect rubber parts of casters for deflection| X| | X
Apply grease to caster ball bearings if needed| | | X
Actuator and control box| | |
Ensure control box is firmly affixed to the mast| X| X|
Ensure pins that connect actuator to boom and mast are firmly affixed|
X| X|
Ensure cable that connects actuator and control box is firmly affixed|
X| X|
Ensure battery indicator has at least 2 green LEDs illuminated before
use| X| X|
Sling and sling hardware| | |
Check sling for wear; discard if worn| X| X|
Ensure sling hardware is in good condition before each use| X|
X|
TROUBLESHOOTING
The following list of encountered problems and their solutions will assist you in determining what may be causing your Lumex lift not to function as designed. If you have a problem occurring which is not listed below, please contact your distributor or GF Health Products, Inc. Technical Support for help. Do not attempt to repair components or parts on your lift, as this may invalidate your warranty or cause further problems that may result in patient injury. Stop using your lift immediately if it is not functioning correctly or any warning beeps are heard.
SYMPTOM | POSSIBLE CAUSE & SOLUTION |
---|
I need to recharge my batteries often or they fail to hold a charge when I
charge them| Replace the two internal batteries as they are at the end of
their life cycle
Batteries should be changed every 18-24 months depending on usage
When I plug my charger into the control box I see no lights on the battery
indicator| Check to make sure the red reset button is in the up
position
Check to make sure the charger is firmly in the control box and plugged
into a wall socket
Replace the charger as it is at the end of its life cycle
If the charger is transmitting current, then you need to replace the
control box PCB
My actuator makes a clicking sound when I attempt to use my lift|
Check to make sure the actuator plug is pushed into the control box so the
O-ring is not visible
The actuator needs to be replaced as it is at the end of its life cycle
The control box PCB may be malfunctioning
The actuator will either go up or down but not both| Replace the hand
control as it is at the end of its life cycle
The control box PCB may be malfunctioning
There is a grinding sound inside the actuator when lifting| Replace
the actuator as the internal gears are stripped
The actuator stops and starts while lifting or lowering| Replace the
batteries as they are at the end of their life cycle
My lift will not operate even though it shows the batteries are charged|
Disconnect the charger from the control box as the lift will not work when
the charger is attached
Check to make sure the hand control is properly inserted to control box
Check to make sure the actuator is properly connected to the control box
The hand control needs to be replaced as it is at the end of its life
cycle
My lift will not operate and battery indicator has no lights illuminated|
Check to make sure the red re-set button is in the up position
The internal batteries may not be connected together properly with the
connection wires
Batteries will no longer hold a charge and must be replaced
The control box PCB is damaged and must be replaced
When lifting I hear 3 long beeps from the control box| You are
attempting to lift someone heavier than the safe working load
Stop the lift and lower safely to bed or chair immediately
When lifting I hear short beeping sounds from the control box| Charge
the batteries
Replace the batteries if they fail to hold a charge
WARNING: If you experience a problem with your lift and are unable to service it yourself, contact your Graham-Field authorized distributor.
SPECIFICATIONS
Info: All dimensions are given in inches unless otherwise specified.
Base width | Open: 37.5″ |
---|
Closed: 26″
Base height clearance| 4.75″
Base length| 44″
Lifting height range
Info: As measured from sling hooks.
| Minimum: 27″
Maximum: 69″
Range: 42″
Mast height| 50″
Maximum weight capacity| 400 lb (181 kg), EVENLY DISTRIBUTED
Caster size| 4″ diameter locking; 3″ diameter non-locking
Lifting mechanism| 24 DC actuator
Battery charger| AC 110V / 220V to DC 24V 1 amp
LIMITED WARRANTY
GF Health Products Inc. (“Graham-Field”) warrants this product to be free from
manufacturing defects in material and workmanship for its period of warranty
that is in accordance with industry standards. This warranty is extended only
to the original purchaser/consumer or dealer/non-consumer of this new product
and to no other purchaser or transferee.
GF Health Products, Inc. warrants the Lumex LF1050 Patient Lift and its
components to be free from defects in workmanship and materials as listed
below:
Lift frame and spreader bar: ……………………………………………………3 years
Actuator: ………………………………………………………………………………..2 years
Parts, including casters, control box, PCB, pendant, charger: …2 years
Batteries: ………………………………………………………………………………..1 year
The Warranty period for the consumer commences on the first date a product is
delivered to consumer by seller/dealer. If the product is rented or leased,
the warranty period commences on the invoice date from GF Health Products,
Inc. A copy of the invoice showing date of purchase must be provided when
submitting warranty claims. When proof of purchase date is not provided,
warranty coverage shall commence upon GF Health Products, Inc.’s invoice date
to the dealer/purchaser.
If within the warranty period, the product or component part is proven to GF
Health Products, Inc.’s satisfaction to be defective, GF Health Products, Inc.
shall provide, at its option, one of the following: (1) repair or replacement
of any defective or nonconforming part or product or (2) a credit and/or
refund of the original selling price. GF HEALTH PRODUCTS, INC.’S SOLE
OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO
SUCH REPAIR, REPLACEMENT, CREDIT AND/OR REFUND. This warranty does not include
any labor charges incurred in replacement part(s) installation or any
associated freight or shipping charges to the manufacturer.
For warranty service, please contact the authorized distributor from whom you
acquired your GF Health Products, Inc. product. Upon receiving notice of an
alleged defect in a product, GF Health Products, Inc. will issue a return
authorization. The defective product or part(s) must then be returned, at the
purchaser’s cost, for warranty inspection using the serial number as
identification (or, if the product is not serialized, lot number and date
code) within thirty (30) days of return authorization issue date. In the event
you do not receive satisfactory warranty service, please contact GF Health
Products, Inc. at the address on the following page. DO NOT return products to
our factory without prior authorization.
LIMITATIONS AND EXCLUSIONS: The foregoing warranty shall not apply to serial numbered products if the serial number has been removed or defaced. Products subjected to negligence, abuse, misuse, improper operation, improper maintenance, improper cleaning, improper storage, or damages beyond GF Health Products, Inc.’s control are not covered by this warranty, and that evaluation shall be solely determined by GF Health Products, Inc. This warranty shall not apply to problems arising from normal wear and tear or failure to follow instructions. The warranty shall also not apply to products modified without GF Health Products, Inc.’s express written consent; nor shall it apply if parts not manufactured by GF Health Products, Inc., or if parts not complying with original equipment specifications are added to GF Health Products, Inc. products, or if the product or part is serviced by an entity not authorized by GF Health Products, Inc.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES AND IMPLIED WARRANTIES, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN, AND THE REMEDY FOR VIOLATIONS OF ANY IMPLIED WARRANTY SHALL BE LIMITED TO THE REPAIR, REPLACEMENT, CREDIT AND/OR REFUND OF THE DEFECTIVE PRODUCT OR PART PURSUANT TO THE TERMS CONTAINED HEREIN. GF HEALTH PRODUCTS, INC. SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.
This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state (province to province). Some states (provinces) do not allow the exclusion or limitation of incidental or consequential damage, or limitation on how long an implied warranty lasts, so the above exclusion and limitations may not apply to you.
The warranties contained herein contain all the representations and warranties
with respect to the subject matter of this document, and supersede all prior
negotiations, agreements and understandings with respect thereto. The
recipient of this document hereby acknowledges and represents that it has not
relied on any representation, assertion, guarantee, warranty, collateral
contract or other assurance, except those set out in this document.