MEAN WELL RPS-60 Series Reliable Green Medical Power Supply Owner’s Manual

June 1, 2024
MEAN WELL

MEAN WELL RPS-60 Series Reliable Green Medical Power Supply

Product Information

  • Specifications:
    • Model: RPS-60 series
    • Power Output: 60W
    • Compliance: EN 60601-1, IEC 60601-1, ANSI/AAMI ES60601-1, BS EN/EN60601-1, IEC60601-1, UL62368-1, BS EN/EN62368-1, IEC62368-1
    • Model Number: TPTC004

Product Usage Instructions

  • Power Supply Connection:
    • Ensure the power supply is compatible with your device’s input requirements. Connect the power supply to a suitable power outlet.
  • Device Connection:
    • Connect the output of the power supply to your medical device following the manufacturer’s guidelines.
  • Power On:
    • After making all necessary connections, switch on the power supply using the provided power button.
  • Monitoring:
    • Monitor the power supply to ensure it is functioning correctly and providing the required power output to your device.
  • Safety Precautions:
    • Always follow the safety guidelines provided in the user manual to prevent any accidents or damage to the equipment.

FAQs

  • Q: Is the RPS-60 series power supply suitable for continuous use in medical applications?
    • A: Yes, the RPS-60 series power supply is designed for reliable and continuous use in medical applications while meeting safety standards.
  • Q: Can I use the power supply with devices that have different power requirements?
    • A: It is essential to match the power supply output with your device’s input requirements to prevent damage. Consult the device manufacturer for compatibility.

Features

  • 4″x2″ compact size
  • IT & Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1, IEC/BS EN/EN60601-1, and lEC/BS EN/EN/UL 62368-1
  • Suitable for BF application with appropriate system consideration
  • Cooling by free air convection
  • EMI Class B for class I configuration
  • No load power consumption<0.75W
  • Protections: Short circuit / Overload / Over voltage
  • Operating altitude up to 3000 meters
  • 3 years warrantyMEAN-WELL-RPS-60-Series-Reliable-Green-Medical-Power-Supply-FIG-1 \(1\)

Applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

GTIN CODE

Description

  • RPS-60 is a 60W highly reliable green PCB-type medical power supply with a high power density on the 4″ by 2″ footprint.
  • It accepts 90~264VAC input and offers various output voltages between 3.3V and 48V.
  • The working efficiency is up to 86% and the extremely low no-load power consumption is down below 0.75W.
  • RPS-60 can be used for Class I (with FG) system design.
  • The extremely low leakage is current EMS BEEN SO 22 SEEN 501, pero in citing all kinds medical red ulation contact medical system equipment.

Model Encoding

SPECIFICATION

MODEL| RPS-60-3.3| RPS-60-5| RPS-60-12| RPS-60-15| RPS-60-24| RPS-60-48
---|---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 3.3V| 5V| 12V| 15V| 24V| 48V
RATED CURRENT| 10A| 10A| 5A| 4A| 2.5A| 1.25A
CURRENT RANGE| 0 ~ 11A| 0 ~ 11A| 0 ~ 5.5A| 0 ~ 4.4A| 0 ~ 2.75A| 0 ~ 1.375A
RATED POWER| 33W| 50W| 60W| 60W| 60W| 60W
PEAK LOAD(10sec.) Note.2| 36.3W| 55W| 66W| 66W| 66W| 66W
RIPPLE & NOISE (max.) **Note.3| 60mVp-p| 60mVp-p| 60mVp-p| 100mVp-p| 100mVp-p| 100mVp-p
VOLTAGE ADJ. RANGE| 3.1 ~ 3.6V| 4.75 ~ 5.5V| 11.4 ~ 13.2V| 13.5 ~ 16.5V| 22.8 ~ 27.6V| 45.6 ~ 52.8V
VOLTAGE TOLERANCE Note.4| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 500ms, 30ms/230VAC            500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)| 60ms/230VAC            12ms/115VAC at full load
INPUT| VOLTAGE RANGE| 90 ~ 264VAC          127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 74%| 79%| 84%| 85%| 87%| 86%
AC CURRENT (Typ.)| 1.8A/115VAC             1 A/230VAC
INRUSH CURRENT (Typ.)| COLD START 60A/230VAC           30A/115VAC
LEAKAGE CURRENT(max.) Note.5| Earth leakage current < 130μA/264VAC , Touch current < 100μA/264VAC
PROTECTION| **

OVERLOAD

| 115 ~ 150% rated output power
Protection type: Hiccup mode, recovers automatically after fault condition is removed


OVER VOLTAGE

| 3.8 ~ 5V| 5.7 ~ 6.8V| 13.8 ~ 16.2V| 17.2 ~ 20.3V| 28.4 ~ 32.4V| 55.2 ~ 64.8V
Protection type: Shut down o/p voltage, re-power on to recover
ENVIRONMENT| WORKING TEMP.| -20 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.6| 3000 meters
SAFETY & EMC

(Note 8)

| ****

SAFETY STANDARDS

| IEC62368-1, UL62368-1, TUV BS EN/EN62368-1, IEC60601-1, TUV BS EN/EN60601-1, UL ANSI / AAMI ES60601-1

(3.1 version), CAN/CSA-C22.2 No. 60601-1:14 – Edition 3, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1

ISOLATION LEVEL| Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC  I/P-FG:2KVAC  O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC/ 25℃/ 70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035 ,BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
END| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m


Voltage dip, interruption

| BS EN/EN61000-4-11| 100% dip 1 period, 30% dip 25 periods, 100% interruptions 250 periods
OTHERS| MTBF| 5153.0K hrs min.       Telcordia SR-332 (Bellcore); 353.6K hrs min.        MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 101.650.829mm or 4″ 2″ 1.141″ inch
PACKING| 0.15Kg; 96pcs/15.4Kg/0.89CUFT
NOTE| 1.  All parameters NOT specially mentioned are measured at 230VAC input, rated load, and 2 ℃ of ambient temperature.

2.  33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.

3.  Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μF & 47μF parallel capacitor.

4.  Tolerance: includes set up tolerance, line regulation, and load regulation.

. Touch current was measured from primary input to DC output.

6. The ambient temperature derating of 3. ℃/1000m with fanless models and of ℃/1000m with fan models for operating altitude higher than 2000m(6 00ft).

7.  Heat Sink HS1,HS2 can not be shorted.

8.  The power supply is considered a component that will be installed into the final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re- confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”

(as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

※ Product Liability Disclaimer: For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

Diagram

Block Diagram

MEAN-WELL-RPS-60-Series-Reliable-Green-Medical-Power-Supply-FIG-1
\(4\)

Output Derating

MEAN-WELL-RPS-60-Series-Reliable-Green-Medical-Power-Supply-FIG-1
\(5\)

Output Derating VS Input Voltage

MEAN-WELL-RPS-60-Series-Reliable-Green-Medical-Power-Supply-FIG-1
\(6\)

Mechanical Specification

MEAN-WELL-RPS-60-Series-Reliable-Green-Medical-Power-Supply-FIG-1
\(7\)

AC Input Connector (CN1) : JST B3P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 AC/N

JST VHR or equivalent

| JST SVH-21T-P1.1 or equivalent
2| No Pin
3| AC/L

DC Output Connector (CN2): JST B4P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1,2 +V JST VHR

or equivalent

| JST SVH-21T-P1.1

or equivalent

3,4| -V

: Grounding Required

  1. HS1, HS2 cannot be shorted.
  2. M1 is safety ground. For better EMC performance, Please secure an electrical connection between M1, M2, and chassis grounding.

Installation Manual

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

MEAN WELL User Manuals

Related Manuals