MEAN WELL RPS-60 Series Reliable Green Medical Power Supply Owner’s Manual
- June 1, 2024
- MEAN WELL
Table of Contents
MEAN WELL RPS-60 Series Reliable Green Medical Power Supply
Product Information
- Specifications:
- Model: RPS-60 series
- Power Output: 60W
- Compliance: EN 60601-1, IEC 60601-1, ANSI/AAMI ES60601-1, BS EN/EN60601-1, IEC60601-1, UL62368-1, BS EN/EN62368-1, IEC62368-1
- Model Number: TPTC004
Product Usage Instructions
- Power Supply Connection:
- Ensure the power supply is compatible with your device’s input requirements. Connect the power supply to a suitable power outlet.
- Device Connection:
- Connect the output of the power supply to your medical device following the manufacturer’s guidelines.
- Power On:
- After making all necessary connections, switch on the power supply using the provided power button.
- Monitoring:
- Monitor the power supply to ensure it is functioning correctly and providing the required power output to your device.
- Safety Precautions:
- Always follow the safety guidelines provided in the user manual to prevent any accidents or damage to the equipment.
FAQs
- Q: Is the RPS-60 series power supply suitable for continuous use in medical applications?
- A: Yes, the RPS-60 series power supply is designed for reliable and continuous use in medical applications while meeting safety standards.
- Q: Can I use the power supply with devices that have different power requirements?
- A: It is essential to match the power supply output with your device’s input requirements to prevent damage. Consult the device manufacturer for compatibility.
Features
- 4″x2″ compact size
- IT & Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1, IEC/BS EN/EN60601-1, and lEC/BS EN/EN/UL 62368-1
- Suitable for BF application with appropriate system consideration
- Cooling by free air convection
- EMI Class B for class I configuration
- No load power consumption<0.75W
- Protections: Short circuit / Overload / Over voltage
- Operating altitude up to 3000 meters
- 3 years warranty
Applications
- Oral irrigator
- Hemodialysis machine
- Medical computer monitors
- Sleep apnea devices
GTIN CODE
- MW Search: https://www.meanwell.com/serviceGTIN.aspx.
Description
- RPS-60 is a 60W highly reliable green PCB-type medical power supply with a high power density on the 4″ by 2″ footprint.
- It accepts 90~264VAC input and offers various output voltages between 3.3V and 48V.
- The working efficiency is up to 86% and the extremely low no-load power consumption is down below 0.75W.
- RPS-60 can be used for Class I (with FG) system design.
- The extremely low leakage is current EMS BEEN SO 22 SEEN 501, pero in citing all kinds medical red ulation contact medical system equipment.
Model Encoding
SPECIFICATION
MODEL| RPS-60-3.3| RPS-60-5| RPS-60-12| RPS-60-15|
RPS-60-24| RPS-60-48
---|---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 3.3V| 5V| 12V| 15V| 24V| 48V
RATED CURRENT| 10A| 10A| 5A| 4A| 2.5A| 1.25A
CURRENT RANGE| 0 ~ 11A| 0 ~ 11A| 0 ~ 5.5A| 0 ~ 4.4A| 0 ~ 2.75A| 0 ~
1.375A
RATED POWER| 33W| 50W| 60W| 60W| 60W| 60W
PEAK LOAD(10sec.) Note.2| 36.3W| 55W| 66W| 66W| 66W| 66W
RIPPLE & NOISE (max.) **Note.3| 60mVp-p| 60mVp-p| 60mVp-p| 100mVp-p|
100mVp-p| 100mVp-p
VOLTAGE ADJ. RANGE| 3.1 ~ 3.6V| 4.75 ~ 5.5V| 11.4 ~ 13.2V| 13.5 ~ 16.5V|
22.8 ~ 27.6V| 45.6 ~ 52.8V
VOLTAGE TOLERANCE Note.4| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±1.0%|
±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 500ms, 30ms/230VAC 500ms, 30ms/115VAC at
full load
HOLD UP TIME (Typ.)| 60ms/230VAC 12ms/115VAC at full load
INPUT| VOLTAGE RANGE| 90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 74%| 79%| 84%| 85%| 87%| 86%
AC CURRENT (Typ.)| 1.8A/115VAC 1 A/230VAC
INRUSH CURRENT (Typ.)| COLD START 60A/230VAC 30A/115VAC
LEAKAGE CURRENT(max.) Note.5| Earth leakage current < 130μA/264VAC ,
Touch current < 100μA/264VAC
PROTECTION| **
OVERLOAD
| 115 ~ 150% rated output power
Protection type: Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE
| 3.8 ~ 5V| 5.7 ~ 6.8V| 13.8 ~ 16.2V| 17.2 ~ 20.3V| 28.4 ~ 32.4V| 55.2 ~
64.8V
Protection type: Shut down o/p voltage, re-power on to recover
ENVIRONMENT| WORKING TEMP.| -20 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, Z axes
OPERATING ALTITUDE Note.6| 3000 meters
SAFETY & EMC
(Note 8)
| ****
SAFETY STANDARDS
| IEC62368-1, UL62368-1, TUV BS EN/EN62368-1, IEC60601-1, TUV BS EN/EN60601-1, UL ANSI / AAMI ES60601-1
(3.1 version), CAN/CSA-C22.2 No. 60601-1:14 – Edition 3, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1
ISOLATION LEVEL| Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP,
Secondary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC/ 25℃/
70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN55035 ,BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
END| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption
| BS EN/EN61000-4-11| 100% dip 1 period, 30% dip 25 periods, 100%
interruptions 250 periods
OTHERS| MTBF| 5153.0K hrs min. Telcordia SR-332 (Bellcore);
353.6K hrs min. MIL-HDBK-217F (25℃)
DIMENSION (LWH)| 101.650.829mm or 4″ 2″ 1.141″ inch
PACKING| 0.15Kg; 96pcs/15.4Kg/0.89CUFT
NOTE| 1. All parameters NOT specially mentioned are measured at 230VAC
input, rated load, and 2 ℃ of ambient temperature.
2. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
3. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μF & 47μF parallel capacitor.
4. Tolerance: includes set up tolerance, line regulation, and load regulation.
. Touch current was measured from primary input to DC output.
6. The ambient temperature derating of 3. ℃/1000m with fanless models and of ℃/1000m with fan models for operating altitude higher than 2000m(6 00ft).
7. Heat Sink HS1,HS2 can not be shorted.
8. The power supply is considered a component that will be installed into the final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re- confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
※ Product Liability Disclaimer: For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
Diagram
Block Diagram
Output Derating
Output Derating VS Input Voltage
Mechanical Specification
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | AC/N |
JST VHR or equivalent
| JST SVH-21T-P1.1 or equivalent
2| No Pin
3| AC/L
DC Output Connector (CN2): JST B4P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1,2 | +V | JST VHR |
or equivalent
| JST SVH-21T-P1.1
or equivalent
3,4| -V
: Grounding Required
- HS1, HS2 cannot be shorted.
- M1 is safety ground. For better EMC performance, Please secure an electrical connection between M1, M2, and chassis grounding.
Installation Manual
- Please refer to: http://www.meanwell.com/manual.html.
- Downloaded from Arrow.com.
- File Name: RPS-60-SPEC 2024-02-23