CHAUVIN ARNOUX E25 AC DC Current Clamp User Manual
- June 4, 2024
- CHAUVIN ARNOUX
Table of Contents
AC/DC current clamp
Thank you for purchasing an E25 AC/DC current clamp, which we trust will give you satisfaction.
For best results from your instrument:
- read this user manual carefully,
- comply with the precautions for use.
WARNING, risk of DANGER! The operator must refer to this user’s manual
whenever this danger symbol appears.
May be placed on and removed from conductors at dangerous voltages. Type A
current sensor as per IEC/EN 610102-032 or BS EN 61010-2-032.
Equipment is protected by double insulation.
Useful information or tips.
Battery.
USB.
To identify the phase (or the direction) of the primary current.
Chauvin Arnoux has adopted an Eco-Design approach in order to design this
appliance. Analysis of the complete lifecycle has enabled us to control and
optimize the
effects of the product on the environment. In particular, this appliance
exceeds regulation requirements with respect to recycling and reuse.
The CE marking indicates compliance with the European Low Voltage Directive,
2014/35/UE, the Electromagnetic Compatibility Directive, 2014/30/EU, and the
Restriction of Hazardous Substances Directive, (RoHS 2011/65/UE and
2015/863/UE).
The UKCA marking certifies that the product is compliant with the requirements
that apply in the United Kingdom, in particular as regards Low-Voltage Safety,
Electromagnetic Compatibility, and the Restriction of Hazardous Substances.
The rubbish bin with a line through it indicates that, in the European Union,
the product must undergo selective disposal in compliance with Directive WEEE
2012/19/UE: This equipment must not be treated as household waste.
Definition of measurement categories
-
Measurement category IV corresponds to measurements taken at the source of low-voltage installations.
Examples: power feeders, counters, and protection devices. -
Measurement category III corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or fixed industrial devices. -
Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage installations.
Example: power supply to domestic electrical appliances and portable tools.
PRECAUTIONS FOR USE
This device is compliant with safety standard IEC/EN 61010-2-032 or BS EN
61010-2-032, for voltages up to 600V in category III or 300 V in category IV.
Failure to observe the safety instructions may result in electric shock, fire,
explosion, and destruction of the instrument and of the installations.
- The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be taken in use. Sound knowledge and a keen awareness of electrical hazards are essential when using this instrument.
- If you use this instrument other than as specified, the protection it provides may be compromised, thereby endangering you.
- Do not use the instrument on networks of which the voltage or category exceeds those mentioned.
- Do not use the instrument if it seems to be damaged, incomplete, or poorly closed.
- Before each use, check the condition of the insulation on the leads and housing. Any item of which the insulation deteriorates (even partially) must be set aside for repair or scrapping.
- When handling the instrument, keep your fingers behind the guard.
- Do not expose the clamp to sprays of water.
- Use personal protection equipment systematically.
- All troubleshooting and metrological checks must be done by competent, accredited personnel.
PRESENTATION
1.1. DELIVERY CONDITION
The E25 clamp is delivered in a cardboard box with:
- one 9V alkaline battery (type 6LR61 or NEDA 1604A),
- one multilingual getting started guide,
- one multilingual safety data sheet,
- a verification certificate.
1.2. ACCESSORIES
One 5V, 500mA external power supply comprising:
- one mains/USB type A adapter
- one USB type A/micro-USB type B cable
For the accessories and spares, consult our website:
www.chauvin-arnoux.com
1.3. INSERTING THE BATTERY
-
Use a screwdriver to unscrew the captive screw of the battery compartment cover.
-
Slide the battery compartment cover off.
-
Connect the battery to the snap-on connector; take care of the polarity.
You can use a rechargeable Ni-MH battery, but the life will be shorter. The instrument does not recharge rechargeable batteries. -
Place the battery in its compartment.
-
Put the battery compartment cover back in place; make sure that it is completely and correctly closed.
-
Screw the screw back in.
1.4. FUNCTIONS
The E25 clamp is used to measure currents from 5mA to 80ADC OR 60AAC without
opening the circuits in which they flow. It reads out the waveform and
amplitude of the current being measured, in the form of a voltage.
Its shape allows it to reach hard-to-reach places.
This clamp can be used with all measuring instruments having banana inputs. It
can be used with a multimeter, a wattmeter, a recorder, etc.
It can be powered by a battery or at 5VDC via the micro-USB connector.
It has:
- a reset pushbutton,
- a range overshoot indicator,
- a power supply indicator,
- automatic standby to prolong battery life.
1.5. E25 CLAMP
2.1. GETTING STARTED
Switch the clamp on by pushing the slide switch to the 1V/A or the 10mV/A
setting.
The 1V/A setting corresponds to the 2A range.
The 10mV/A setting corresponds to the 80A range.
The ON indicator lights green. If it blinks, there remain less than 4 hours of
use. If it fails to light, you must replace the battery (see § 4.2).
2.2. ZERO ADJUSTMENT
- Switch the clamp on.
- Connect the clamp to the measuring instrument. The phase is on the red lead.
- Make sure that the clamp is not on a conductor and that its jaws are correctly closed.
- Place the clamp in the position it will be in during the measurement.
- Press the zero adjustment button.
- The OL indicator lights for approximately three seconds to indicate that the zero adjustments is in progress in the two ranges.
- If the zero has been correctly adjusted, the OL indicator goes off. If it remains on, the zero could not be adjusted.
In this case, check that the clamp is not on any conductor and that its jaws are correctly closed, then press the zero adjustment button again.
Or switch the clamp off and back on, and it is the last adjustment stored that will be used.
2.3. MEASUREMENT
The zero must be adjusted before each measurement.
-
Once the zero has been adjusted, press the moving arm of the clamp to open the jaws.
-
Clamp the cable carrying the current to be measured. Use the centering marks to help you center the cable inside the jaws of the clamp.
The arrow on the clamp must point in the presumed direction of current flow.
-
Release the moving arm gently and make sure that the jaws are correctly closed.
-
The value measured is displayed on the measuring instrument.
If the OL indicator lights, it means that the current is too strong to be measured. If you are in the 1V/A range, switch to the 10mV/A range. -
Apply the conversion factor corresponding to the setting of the switch.
2.4. AUTO OFF
At the end of 10 minutes of operation with no sign of the user’s presence (a
press on the zero adjustment button or an action on the switch), the clamp
switches to standby and, the ON indicator goes off.
To wake up the clamp, press the zero adjustment button or set the switch to a
position other than OFF.
To disable auto-off (operation in permanent mode P ), press the zero
adjustment button when starting up the instrument. The
ON indicator blinks to indicate that the request has been taken into account,
then lights steady orange when you release the zero adjustment button.
When you switch the clamp off by setting the switch to OFF, the auto-off
function is restored.
2.5. INDICATORS
ON indicator
| Off: instrument off
| Lighted green: instrument on
| Blinking green: the batteries will have to be replaced in less than 4h
| Lighted orange: permanent operation P (auto off deactivated)
OL indicator
| Off: the measurement is correct
| Lighted red: The measurement exceeds the range
| Lighted red for 3 seconds: the zero adjustments is in progress.
2.6. MAINS ADAPTER (OPTION)
For long-term measurements, you can connect the clamp to the mains using the
optional line power adapter. You can use any mains/micro-USB adapter that
delivers 50mA or more.
When the instrument is powered via the micro-USB connector, automatic standby
is disabled.
The insulation between the type B micro-USB connector and the measurement
output is 600V CAT III. This makes it possible to connect the clamp, without
risk, to measuring instruments having inputs that are not insulated. The type
B micro-USB connector must not be in contact with conductors or uninsulated
parts at dangerous voltage.
If the external power supply is disconnected, the clamp changes back to
battery-powered operation. The color of the ON indicator indicates whether
automatic standby is enabled (indicator green) or not (indicator orange).
SPECIFICATIONS
3.1. REFERENCE CONDITIONS
Quantity of influence | Reference values |
---|---|
Temperature | 23 ± 5 °C |
Relative humidity | 20 to 75%RH |
Position of the conductor | centered |
Frequency of the measured signal | DC to 65Hz sine wave |
Powered | by battery: 6.5 to 9 V |
external supply: 5 V ± 0.1 V
External electric field| zero
External DC magnetic field (earth’s field)| < 40 A/m
External AC magnetic field| zero
The impedance of the measuring instrument| > 1 MO and 100 pF
The intrinsic uncertainty is the error specified for the reference conditions.
It is expressed as a percentage of the output signal (R) plus an offset in mV:
± (a% R + b)
3.2. ELECTRICAL CHARACTERISTICS
Range | 1V/A (2A) | 10mV/A (80A) |
---|---|---|
Specified measurement range | 0.005 to 2 Adc | |
0.005 to 1.5 Aac | 0.05 to 50 Adc | |
0.05 to 40 Aac | 50 to 80 Adc |
40 to 60 Aac
Intrinsic uncertainty| ≤ ± (2%R + 5 mV)| ≤ ± (4%R + 0,5 mV)| ≤ ± 12%R
Phase shift (DC to 65Hz)| ≤ 1°| ≤ 1°| ≤ 1°
3.2.1. TYPICAL CURVES
Typical amplitude error curve for a DC current, 10mV/A range
Typical amplitude error curve for a 60Hz AC current, 10mV/A range
3.2.2. NOISE
Typical noise level at outpthe ut | DC | AC |
---|---|---|
10mV/A range | ± 120 µVdc | 180 µVrms |
Caliber 1V/A | ± 8 mVdc | 4 mVrms |
3.2.3. ZERO ADJUSTMENT
Minimum zero adjustment range: ±1.5 Adc in steps of approximately 1.2 mA.
3.2.4. FREQUENCY RESPONSE
Typical amplitude error curve as a function of frequency, I = 1A, 10mV/A range
Typical amplitude error curve as a function of frequency, I = 0,5A, 1V/A range
Typical amplitude error curve as a function of frequency, I = 1A, 10mV/A range
Typical amplitude error curve as a function of frequency, I = 0,5 A, 1 V/A range
3.2.5. PULSE RESPONSE
Pulse response at ±2 Apeak at the frequency of 1 kHz in the 10mV/A range
Pulse response at ±0.5 Apeak at the frequency of 1 kHz in the 1V/A range
3.3. OPERATING LIMITS
- Conductor temperature: ≤90°C, 110°C peak
- The temperature of the jaws: ≤ 80°C
- The curve of derating versus frequency
3.4. VARIATIONS IN THE RANGE OF USE
Quantity of influence| Range of influence| Error in % of
reading
---|---|---
Typical| Maximum
Temperature| -10 to + 50 .0| The drift of the zero
± 10 mA/°C|
| The drift of the gain
± 800 ppm/°C
Relative humidity| 0 to 85%HR| | < 0.5 %
Frequency| DC at 20 kHz| | see curves
Position of the conductor| | | ± 0.5 %
Adjacent conductor| carrying a current of 10 A at
60 Hz| | ± 4 mA/A
Load| RL=10 k0| -2.%|
Common mode AC| the voltage at 50 Hz| | ± 1 mA/100 V
Radiated fields| 10V/m
80 MHz to 1 GHz| | ± 4 A
Remanence| for 80 ADC| ± 370 mAoc|
3.5. POWER SUPPLY
The instrument is powered by a 9V battery (type 6LR61 or NEDA 1604A).
The typical battery life is 80h with an alkaline battery.
The instrument can be powered by an external supply (5Vdc, 50mA) via the type
B micro-USB connector
3.6. ENVIRONMENTAL CONDITIONS
The device must be used in the following conditions:
Indoor use.
Degree of pollution | 2 |
---|---|
Altitude | < 2000 m |
Transport altitude | ≤ 12,000 m |
3.7. CONSTRUCTION SPECIFICATIONS
Dimensions (L x W x H) | 231 x 36 x 67 mm |
---|---|
Weight | approximately 330 g |
Measurement lead | 1.5 m long |
USB cord | 15 cm long |
Clamping capacity: 11.8mm in diameter
Protection by the housing
- IP 20 per IEC 60529
- Resistance of the jaws per IEC/EN 61010-2-032 or BS EN 61010-2-032
3.8. CONFORMITY TO INTERNATIONAL STANDARDS
The instrument is compliant with IEC/EN 61010-2-032 or BS EN 61010-2-032, 600
V in category III.
Double or reinforced insulation .
Type of current sensor per IEC/EN 61010-2-032 or BS EN 61010-2-032: type .
3.9. ELECTROMAGNETIC COMPATIBILITY
The device is in conformity with standard IEC/EN 61326-1 or BS EN 61326-1.
MAINTENANCE
Except for the battery, the instrument contains no parts that can be replaced by personnel who have not been specially trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent” may gravely impair safety.
4.1. CLEANING
Disconnect the instrument completely and turn the rotary switch to OFF. Also,
make sure that no cable is clamped.
Use a damp soft cloth and dry rapidly with a dry cloth or forced air. Do not
use alcohol, solvents, or hydrocarbons.
The air gaps of the clamp must always be kept clean.
Do not leave the clamp in very damp places, or exposed to splashes.
4.2. REPLACING THE BATTERY
The battery must be replaced when the On indicator remains off when the device
is switched on in the absence of an external supply.
- Withdraw the conductor from the clamp and disconnect it. Set the switch to OFF.
- use a screwdriver to unscrew the captive screw of the battery compartment cover then pull the cover out from the end of the mobile arm.
- Replace the spent battery with a new battery.
Spent primary and storage batteries must not be treated as ordinary household waste. Take them to the appropriate collection point for recycling.
- Place the battery in its compartment, with the correct polarity.
- Close the compartment and check that it is completely and correctly closed.
- Screw the screw back in.
4.3. MANUAL ADJUSTMENT
The manual adjustment serves to adjust the gain of the clamp without using a
PC. To maintain good measurement accuracy, we recommend checking the clamp
once a year.
4.3.1. EQUIPMENT NECESSARY
- A 200 Aac, 40 to 60Hz current generator
- A 10 Aac, 60Hz current generator, accuracy class ≤0.2%
- A 1 Aac, 60Hz current generator, accuracy class ≤0.2%
- A voltmeter, accuracy class ≤0.2%
4.3.2. ADJUSTMENT PROCEDURE
-
First, demagnetize the clamp by clamping a conductor carrying an AC current of at least 200 Arms at a frequency between 40 and 60Hz. Then gently withdraw the clamp from the conductor, in which the current is still flowing.
-
Place the clamp at an ambient temperature of 23±2°C for one hour. It must not be on a conductor, and the jaws must be correctly closed. Connect the voltmeter in VAC mode to the output of the clamp.
-
To enter the adjustment mode, hold the DC Zero button down and shift the switch from the OFF setting to the range to be ad- justed (10mV/A or 1 V/A). Keep the DC Zero button pressed for 30 seconds, until the ON indicator blinks orange, then green.
Release the DC Zero button. The clamp is in adjustment mode. -
The clamp then performs an adjustment of the zero. The OL indicator goes off when the adjustment has been made successfully.
-
Clamp a conductor carrying a current of:
• 10Aac 60Hz for the 10mV/A range
• 1Aac 60Hz for the 1V/A range -
Then press the DC Zero button. The first press substantially lowers the polarization adjustment of Hall effect sensors. Subsequent presses increase this adjustment by one step. So press the DC Zero button until the correct output voltage is reached.
• 100mVrms for the 10mV/A range.
• 1Vrms for the 1V/A range. |
If you overshoot this value, continue to press the DC Zero button until the output signal falls below the desired value, then restart the adjustment. -
Once this adjustment is done, press the DC Zero button again for 30 seconds, until the ON indicator blinks orange, then green.
You can then release the DC Zero button. The adjustment is recorded and the clamp exits from the adjustment mode.
Remarks
When the clamp is in the adjustment mode (in other words, from step 3), any
change of the switch setting causes an exit from
the adjustment mode with no modification. The clamp will then use the previous
adjustments.
To adjust both ranges, you must switch the clamp off, then repeat the
adjustment procedure from step 3.
WARRANTY
Except as otherwise stated, our warranty is valid for 24 months starting from the date on which the equipment was sold. Extract from our General Conditions of Sale provided on request.
The warranty does not apply in the following cases:
- Inappropriate use of the equipment or use of incompatible equipment.
- Modifications were made to the equipment without the explicit permission of the manufacturer’s technical staff.
- Work done on the device by a person not approved by the manufacturer.
- Adaptation to a particular application is not anticipated in the definition of the equipment or not indicated in the user’s manual.
- Damage caused by shocks falls, or floods.
FRANCE
Chauvin Arnoux
12-16 rue Sarah Bernhardt
92600 Asnières-sur-Seine
Tél : +33 1 44 85 44 85
Fax : +33 1 46 27 73 89
info@chauvin-arnoux.com
www.chauvin-arnoux.com
INTERNATIONAL
Chauvin Arnoux
Tél : +33 1 44 85 44 38
Fax : +33 1 46 27 95 69
Our international contacts
www.chauvin-arnoux.com/contacts
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>