CHAUVIN ARNOUX BC05 Resistance, Capacitance and Inductance Boxes User Guide
- June 3, 2024
- CHAUVIN ARNOUX
Table of Contents
CHAUVIN ARNOUX BC05 Resistance, Capacitance and Inductance Boxes
You have just obtained a resistance, capacity or inductance decade box, and we thank you for your trust in us.
To obtain the best service from your unit:
- read these operating instructions carefully,
- comply with the precautions for use.
GENERAL INSTRUCTIONS
If the device is used in a manner unspecified in these instructions, the protection provided by the device may be compromised.
UNPACKING – PACKING
All instruments are checked mechanically and electronically before shipment. All precautions are taken to be sure you receive an undamaged instrument. If there is damage, notify the carrier immediately.
PRECAUTIONS AND USAGE SAFETY
Before any use
Resistance, condenser and inductance decade boxes are passive elements to be
inserted into trial or development circuits to obtain the desired resistance,
capacity or inductance values by a combination of decades.
These decade boxes are compliant with safety standard IEC 61010 – 1.
For your own safety and to prevent any damage to your equipment, you must
follow the instructions in this manual.
This equipment can be used in category II electrical circuits not exceeding
150 V relative to earth. It must be used indoors in an environment with level
2 pollution and at an altitude below 2,000 m.
Definition of Installation Categories (see IEC 61010-1)
- CAT I: Circuits not directly connected to network and specially protected.
- Example: protected electronic circuits.
- CAT II: Circuits connected directly to low-voltage installation.
- Example: power supply for domestic electrical appliances and portable tools.
- CAT III: Power supply circuits in the building installation. Example: Electrical panel, circuit breakers, fixed industrial machines or devices.
- CAT IV: Source circuits of building low-voltage installation. Example: Power feeders, counters and protection systems.
For safety reasons, you must use only measuring leads, of voltage and category
at least equal to those of the instrument and compliant with standard
IEC 61010.
Do not use damaged equipment.
Before use, always check all the cord insulation.
Meaning of symbols used:
Symbol | Meaning | Symbol | Meaning | |
---|---|---|---|---|
**** | WARNING, risk of DANGER! The operator should refer to this user’s | |||
manual whenever this danger symbol appears. | **** | Earth. | ||
The CE marking indicates compliance with the European Low Voltage Directive |
(2014/35/EU), Electromagnetic Compatibility Directive (2014/30/EU), and Restriction of Hazardous Substances Directive (RoHS, 2011/65/EU and 2015/863/EU).| | ****
The rubbish bin with a line through it indicates that, in the European Union, the product must undergo selective disposal in compliance with Directive WEEE 2012/19/EU. This equipment must not be treated as household waste.
When using the equipment
To prevent electrical discharge, injury or damage to the device, and to
ensure you use the decade box without risk, follow the safety recommendations
below:
- Read the operating instructions fully before using this device and observe all safety instructions.
- This device must be used indoors, up to 2,000 m altitude.
- Always connect the box’s earth terminal to an earth protection.
- Never exceed the protection limit values indicated in the specifications for each type of decade.
- When the decade box is inserted into a circuit, do not touch the unused cords or terminals.
- Every time, after use, disconnect the supplies for the capacity boxes and discharge the condensers.
- Avoid working alone.
- Use the device only as specified in this manual; otherwise, the protection provided by this equipment may be altered.
- Do not use this device if it appears damaged.
- Inspect the integrity of lead insulation. Replace damaged leads.
- Be careful when working with voltages higher than 70 VDC or 33 Vrms and 46.7 Vpp, such voltages can pose a risk of electrocution. Depending on conditions, the use of individual protection is recommended.
- Before opening the equipment, disconnect it from all circuits and cords.
GENERAL PRESENTATION
These decade boxes come in metal casing with a male terminal linked to the casing for earthing and female terminals for a direct connection to each decade.
RESISTANCE BOXES
- 4 modls for values 0 Ω to:
- 11,111 kΩ in 4 decades (BR04)
- 111,111 kΩ in 5 decades (BR05)
- 1,111111 MΩ in 6 decades (BR06)
- 11,111111 MΩ in 7 decades (BR07)
Characteristics:
Uncertainty is ± 1 % of the displayed value.
Model | Decade |
---|---|
X1 W | X10 W |
BR04 | + |
BR05 | + |
BR06 | + |
BR07 | + |
Uncertainty | ±1 % |
I max mADC | 700 |
Recommendations for use:
Use short and strong section cables for low values.
Do not exceed the maximum intensity allowed for each decade. The access
terminals for each decade are used to produce resistant dividers.
CAPACITOR BOXES
1 model (BC05) for a total capacity of 11,1111 µF in 5 decades.
Characteristics:
Residual capacity is around 25 pF.
Uncertainty is ± 2 % of the displayed value (capacity resistance deducted).
Model | Decade |
---|---|
X100 pF | X1 nF |
BC05 | + |
Uncertainty | 2 % |
Recommendations for use:
Use short cables kept away from one another for low values. BR04-05-06-07 –
BC05 – BL07
Do not exceed the maximum tensions: 300 V direct current or 230 V at 50 Hz.
The access terminals for each decade are used to produce capacitive dividers.
INDUCTANCE BOX
1 model (BC07) for a total capacity of 11,111111 H in 7 decades.
Characteristics:
Residual inductance Lr is around 0,15 µH.
Uncertainty is given in the table below for a frequency and a superimposed
direct current under 10 mA (residual self-deducted)
Model | Decade |
---|---|
X1 |
µH
| X10
µH
| X100
µH
| X1
mH
| X10
mH
| X100
mH
| X1 H
BL07| +| +| +| +| +| +| +
Uncertainty| 5 %+/-Lr| 5 %| 5 %| 5 %| 5 %| 5 %| 10 %
Fréquence| 10 kHz| 1 kHz| 100 Hz
I max mADC| 300| 200| 100| 100| 70| 50| 40
Recommendations for use:
Using the twisted cables for low values can help to reduce the effects of the
per unit length.
Do not exceed the maximum intensity allowed for each decade. The access
terminals for each decade are used to produce inductive dividers.
GENERAL SPECIFICATIONS
CONSTRUCTION CHARACTERISTICS
Models | BR04 | BR05 | BC05 | BR06 | BR07 | BL07 |
---|---|---|---|---|---|---|
Weight | 1 kg | 1.4 kg | ||||
Dimensions L (mm) | 310 | 410 | ||||
w (mm) | 86 | 86 | ||||
h (mm) | 76 | 76 |
Protection level: IP 40 (as per EN 60529)
STORAGE AND OPERATING CONDITIONS
Storage conditions:
Temperature: -40°C to +70°C Humidity: 20 % to 96 % Altitude: < 12,000 m
Operating conditions:
- Temperature: 0°C to 40°C
- Humidity: 20 % to 80 %
- Altitude: < 2,000 m
- Pollution level: 2
- Rated voltage: 150 V CAT II relative to earth
CLEANING, MAINTENANCE, WARANTY
Cleaning
Disconnect from all sources of electricity before cleaning, maintenance or
dismantling.
Clean using a soft tissue with a small amount of soap and water, then dry
completely.
Maintenance, metrology:
After a long period of storage, it may be worth turning each of the decade’s
switches before use.
Like all measuring or testing devices, regular metrological verification is
necessary.
To have your devices checked and calibrated, please contact our Cora approved
metrology laboratories or Manumeas reenters.
WARANTY, REPAIRS
WARRANTY
This instrument is guaranteed against any defect in materials or workmanship,
in accordance with the general terms and conditions of sale.
During the guarantee period (1 year), the instrument must be repaired only by
the manufacturer, who reserves the right to repair the instrument or to
exchange all or part of it.
The guarantee is not applicable in the following cases:
- misuse of the instrument or use with incompatible equipment;
- modifications of the instrument without explicit authorization of the manufacturer’s engineering department;
- work carried out on the instrument by a person not approved by the manufacturer;
- adaptation for a specific application, not included in the definition of the instrument or the operating instructions;
- impacts, falls or immersion.
REPAIRS
For all repairs before or after expiry of warranty, please return the device to your distributor.
HOW TO ORDER
- BR04 …………………………………………………………….. P01197401
- BR05 …………………………………………………………….. P01197402
- BR06 …………………………………………………………….. P01197403
- BR07 …………………………………………………………….. P01197404
- BC05 …………………………………………………………….. P01197421
- BL07 …………………………………………………………….. P01197451
Delivered with:
- 1 user’s manual in 5 languages.
ACCESSORIES :
- Cords 25 cm long with rewinder ……P01295056