maquira Autoclavable Adult-Universal-Child Cone Positioner Instructions

June 13, 2024
maquira

maquira Autoclavable Adult-Universal-Child Cone Positioner

AUTOCLAVABLE ADULT/UNIVERSAL/CHILD CONE POSITIONER

INDICATION:
The AUTOCLAVABLE ADULT/UNIVERSAL/CHILD CONE POSITIONER assists in guiding the angulation of the dental X-ray equipment, in periodontal procedures, resulting in standardized and clear radiographic images.

COMPOSITION

Thermoplastic, pigment and non-toxic silicone.

PIECES

Positioner for upper and lower incisors and canines.

Positioner for upper right and lower left molars.

Positioner for upper left and lower right molars.
Note: to take lower teeth, the sides of parts 2 and 3 are reversed.

Positioner for radiography bite wings. interproximal

Storage pot (non-autoclavable).

MODE OF USE:
How to place the x-ray film:
Firmly place the radiographic film in the position indicated in the figure, between the holder (for film) and the pressure device.

  • A – Film support.
  • B – Pressure device.
  • C – Bite device.
  • D- Base of the positioner.

Radiography positioner:
For the piece to be positioned in the patient’s mouth, the film and the tooth must be in parallelism. During the positioning of the upper and posterior teeth the archwire must touch the forehead of the patient. In a posterior radiograph the bite is always in the area of the anterior teeth. When you want to dissociate overlapping roots mesialize or distalize the bracket in the mouth, by about 20 degrees horizontally, then you get a distalization or mesialization without distortion.

Radiographic take:
The constituents of this product should be used as shown in the illustrations. The X-ray Angle is correctly matched to the position of the film and tube parallel to the positioner.

  • A – X-ray film
  • B- Base for the film
  • C- Ring
  • D – Tube of the X-ray machine
  • E – Bite device
  • F- Base

STORAGE:
Store the package in a clean place away from direct sunlight. Temperature between 3 °C to 50 °C and humidity up to 75 RH.

WARNINGS AND PRECAUTIONS:
Exclusive use of the professional. Do not use the product without prior cleaning and sterilization.

CLEANING

Before starting the cleaning process, remove the biting device from the parts. Remove organic material using ultrasonic cleaners with enzymatic detergent (minimum 3 enzymes) at dilutions recommended by the detergent manufacturer. Then wash with running water.
Pack the pieces in sterilization wrapping (envelopes of surgical grade paper) and put them in the autoclave with temperature from 121 °C to 134 °C, observing the recommendations of the autoclave manufacturer’s instruction manual. The pot and cap are not autoclavable, they are only used to store the product after sterilization. Remove the bite device from the positioner to autoclave.
Sterilization is valid for one week if the package is not violated. Chemical disinfection can also be performed using, for example,0.2% peracetic acid solution for 1 hour. Repeat the process after each use.

EXPIRATION DATE:
After opened the positioners have an expiration of 50 sterilization cycles.

DISPOSAL MODE:
The product must be disposed of according to the legislation in force in your country.

SYMBOLS

maquira-Autoclavable-Adult-Universal-Child-Cone-
Positioner-9

MADE IN BRAZIL

.,. Fabrlcante I Manufacturer: Maquira lndustria de Produtos Odontol6gicos S.A.
CNPJ: 0S.823.205/0001-90 Endere?o I Direcci6n I Address: Av. Melvin Jones, 773
Pq. Ind. Bandeirantes
Maringa – PR I 87070-030 I Brasil Resp. Tecnico I Technical Resp.: Cristiane Furlan I CRF-PR: 14.038 Reg. ANVISA N”: 80322400059
@SAP:

Atendimento ao cliente Atenci6n al cliente Customer service 0800 726 5848 sap@maquira.com.br

0403006 – rev. 07