Dr ger PAC6000 Single Gas Monitors Instruction Manual

October 30, 2023
Dr ger

Dr ger LOGO Instructions for use
Dräger Pac 6×00 / 8×00

MOG 00Dr ger PAC6000 Single Gas Monitors - ICON 1contact@gazdetect.com
+33(0)1.64.09.35.16
Dräger Technology for Life®**

PAC6000 Single Gas Monitors

Dr ger PAC6000 Single Gas Monitors - FIGURE 1

Safety information

These instructions for use may be downloaded in other languages from the database for technical documentation ( www.draeger.com/ifu ) in electronic form, or a free printed copy (order no. 90 33 740) can be obtained from Dräger.
1.1 Basic safety information

  • Before using this product, carefully read these instructions for use and those of the associated products.
  • Strictly follow the instructions for use. The user must fully understand and strictly observe the instructions. Use the product only for the purposes specified in the intended use section of this document.
  • Do not dispose of the instructions for use. Ensure that they are retained and appropriately used by the product user.
  • Only trained and competent users are permitted to use this product.
  • Local and national guidelines that relate to this product (e.g. IEC 60079-14) must be complied with.
  • Only trained and qualified personnel may inspect, repair and maintain the product as described in these instructions for use and the technical manual (order no. 90 33 742) (see section 5 on page 26). Further maintenance work that is not detailed in these instructions for use must only be carried out by Dräger or personnel qualified by Dräger. Dräger recommend a Dräger service contract for all maintenance activities.
  • Use only genuine Dräger spare parts and accessories. Otherwise, the proper functioning of the product may be impaired.
  • Do not use a faulty or incomplete product. Do not modify the product.
  • Notify Dräger in the event of any component fault or failure.
  • Replacing components may impair the intrinsic safety of the device.

Incorrect calibration
An incorrect calibration leads to incorrect measured values.
► The sensitivity must be checked daily before first using the device, with a known concentration of the gas to be measured that corresponds to 25 to 50% of the final concentration. The accuracy must amount to 0 to +20% of the actual value. The accuracy can be corrected by calibration.
1.2 Safety information on explosion protection
Devices or components that are used in potentially explosive atmospheres and which are tested and approved in accordance with national, European or international explosion protection guidelines may only be used under the approved conditions in compliance with the statutory provisions.
Oxygen-enriched atmospheres
Explosion protection is not guaranteed in oxygen-enriched atmospheres (>21 Vol% O2).
► Remove the device from the potentially explosive atmosphere.
Risk of explosion!
► Do not open the gas detector in explosion-hazard areas.
Specific usage conditions

  • Under certain extreme circumstances, exposed plastic and unearthed metal parts of the enclosure may store an ignition-capable level of electrostatic charge.
  • Activities such as placing the device in a pocket or on a belt, operating the keypad or cleaning with a damp cloth, do not present a significant electrostatic risk. However, where a static-generating mechanism is identified, such as repeated brushing against clothing, then suitable precautions shall be taken, e.g. the use of anti-static clothing and footwear.

Conventions in this document

2.1 Meaning of the warning symbols
The following alert messages are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. The meanings of the warning symbols are defined as follows:
WARNING
Indicates a potentially hazardous situation. If not avoided, it could result in death or serious injury.
2.2 Trademarks

Trademark Owner of trademark
Bluetooth Bluetooth SIG, Inc.

2.3 Glossary

Technical term Explanation
Operation signal A periodic optical (green LED) and/or acoustic signal.
D-Light The D-Light allows the user to check and indicate compliance with

certain settings (e.g. bump test interval).
The green LED flashes for a short period and superimposes the optical operation signal.

Description

3.1 Product overview
3.1.1 Gas detector (see page 2, fig. A)

1. Alarm LEDs
2. Operation signal/D-Light
3. Gas inlet
4. Horn
5. Display| 6. [OK] key
7. [▼] key
8. Screw (4x)
9. IR interface
---|---

3.1.2 Display (see page 2, fig. B)

1. Key word symbol
2. Error symbol
3. Information symbol
4. Battery charge status
5. Bluetooth® activated
Bluetooth®  connection established| 6. Measurement unit
7. Span calibration
8. Fresh air adjustment
9. TWA/STEL
10. Peak concentration
---|---

3.2 Intended use
The Dräger Pac 6×00/8×00 is a gas detector and is used to measure and alert the user of gas concentrations in the ambient air.
3.3 Approvals
A copy of the name plate and the declaration of conformity are provided in the enclosed supplementary documentation (order no. 90 33 741).
The rating plate on the gas detector must not be concealed.

Use

4.1 Preparations for use
4.1.1 Initial start-up
The gas detector remains in deep sleep mode upon delivery and must be activated during the initial start-up.

  1. Hold down the [▼] key for approx. 3 s.
    The gas detector is activated.

4.1.2 Switching on the gas detector (see page 2, fig. C)

  1. Hold down the [OK] key for approx. 3 s.

The following is displayed or activated:

  • Display elements, LEDs, alarm signal and vibration alarm
  • Self-test
  • Software version and gas name
  • Alarm thresholds A1 and A2, A3 if applicable
  • Time to next calibration (configurable)
  • Time to expiration of the bump test interval (configurable)
  • Fresh air calibration (configurable)

Before every use, check whether the display elements and information are displayed correctly.
A warm-up phase takes place when first switching on the gas detector (duration depends on the sensor type).
4.1.3 Switching off the gas detector (see page 2, fig. C)

  • Hold down both keys for approx. 3 s, until shut-down is complete.

4.2 Before entering the workplace
WARNING
Serious damage to health!
An incorrect calibration can lead to incorrect measurement results, which may result in serious damage to health.
► Before performing safety measurements, check the calibration by way of a bump test, adjust as necessary, and check all alarm elements. If national regulations exist, the bump test must be performed in accordance with these regulations.
WARNING
Incorrect measurement results!
The gas inlet opening is fitted with a filter that protects against dust and water.
Contamination may change the qualities of the dust and water filter.
► Do not damage the filter. Immediately replace damaged or blocked filters.
For information on calibration, see the technical manual (order no.
90 33 742). The technical manual can be downloaded on the gas detector’s product page under the following internet address: www.draeger.com.
To ensure correct function:

  • Do no cover the gas inlet opening.
  • Place the device on clothing near the mouth.
  • At temperatures lower than -20 °C, the measured value may deviate by > 10 % if the relevant sensor was calibrated at room temperature. Dräger recommends calibration at the primary operating temperature if the measurement will be performed at very low temperatures. Thereby, maximum measurement accuracy is achieved.

After switching on the device, the current measured value is shown in the display.
Check whether the warning [!] appears. If it is displayed, we recommend performing a bump test, as described in Section 4.3.
4.2.1 Performing a manual bump test
WARNING
Danger to health! Do not inhale the test gas.
► Pay attention to the hazard information in the respective safety data sheets.
Two modes can be selected for a bump test. Use the Dräger CC-Vision PC software to set the mode.

  • Quick bump test (test for alarm triggering)
  • Advanced bump test (test for accuracy)

For further information on the 2 bump test modes, see the technical manual (order no. 90 33 742). The technical manual can be downloaded on the gas detector’s product page under the following internet address: www.draeger.com.
The bump test can be performed as follows:

  • Manual bump test
  • Bump test with X-dock (see instructions for use for the Dräger X-dock)
  • Bump test with Bumptest Station (see quick guide to the Bumptest Station)

Prerequisites for the manual bump test:

  • The gas detector is switched on.
  • Suitable test gas cylinder available, e.g. test gas cylinder (order no. 68 11 130) with the following mixed gas ratios: 50 ppm CO, 15 ppm H2S, 2.5 Vol% CH4, 18 Vol% O2

To perform a manual bump test (configurable):

  • See Page 227, fig. E.

The result of the bump test (passed or failed) is saved in the data logger.
4.2.2 Display the peak concentration (Peak), time-weighted average (TWA) and the short-term exposure limit (STEL)

  1. Press the [OK] key in measuring mode.
    This displays the peak concentration and the icon for peak concentration.

  2. Press the [OK] key repeatedly to display the Peak, TWA1) and STEL1) in succession (TWA and STEL only for Pac 6500/8xx0, not for Pac 6000).

If the gas detector is turned off, the values for the peak concentration, TWA and STEL are deleted.
4.3 During operation
WARNING

Danger to life and/or risk of explosion!
The following alarms may indicate a danger to life:

  • A2 alarm (A3 alarm optional)
  • STEL or TWA alarm
  • Device error
    ► Immediately leave the hazard area.

The continuous operation of the gas detector is indicated by an optical and/or acoustic operation signal that is emitted in a 60-second cycle (configurable using the Dräger CC-Vision PC software).
The acoustic operation signal must be switched on for measurements in accordance with EN 45544 (CO, HS) or in accordance with EN 50104 (O).
If the allowable measuring range is exceeded or a negative zero drift occurs, the following message appears in the display: (concentration too high) or (negative drift).
The measuring channels do not need to be checked after a short-term over range (up to one hour) by the EC measuring channels (this does not apply if using DrägerSensor XXS CO H2-CP).
In the event of an alarm, the corresponding displays, the visual, audible and additionally the vibration alarm, are activated, see section 6.2 on page 28.

  1. configurable
  • Push the [▼] button to light up the display.

If the gas detector is used for offshore applications, a distance of 5 m to a compass must be complied with.
4.4 Connecting the gas detector with a smartphone
Only applies to Pac 6500/8000/8500 with Bluetooth® module.
Bluetooth® is not part of the technical suitability test and may only be used in countries for which an approval exists. Contact Dräger for questions on availability.
WARNING
Risk of explosion!
Using an unsuitable smartphone can lead to ignition of combustible or explosive atmospheres.
► The smartphone must be suitable and approved for use in explosionhazard areas.
The gas detector can be connected to a suitable smartphone via Bluetooth®.
The Bluetooth® function and the GATT (Generic Attribute Profile) interface (subject to licensing) can be activated via the PC software CC-Vision. For specific information about the connection via Bluetooth®, see also the instructions for use of the smartphone used.
Prerequisites:

  • Bluetooth® is activated on the gas detector and on the smartphone.
  1. Switch on the gas detector.

  2. On the smartphone, select the gas detector:
    a. Short name of the gas detector in the Adroid/iOS Bluetooth® menu.
    b. Part number and serial number in the optional Dräger app.
    Both the smartphone and the gas detector are showing a 6-digit number.

  3. Check if the numeric codes on the two devices match and if they do confirm on both devices. The connection must be established within 30 s, otherwise the numeric code will become invalid.
    The devices are connected.

A successful connection is saved and in future, the two devices will be connected automatically (configurable via the PC software CC-Vision).
If the connection is lost, the gas detector automatically attempts to establish a new connection.

Maintenance

The device does not require any special maintenance measures.
The following maintenance instructions must be carefully read, understood and followed to prevent flammable or combustible atmospheres from igniting and to ensure that the intrinsic safety of the device is not impaired.
WARNING
Incorrect measurement!
► A bump test and/or calibration must be performed each time the device is opened. This includes any battery change as well as each sensor replacement in the device.
NOTICE
Damage to components!
The gas detector contains components at risk of charging.
► Before opening the gas detector, make sure that the operator is earthed in order to prevent damage to the gas detector. For example, earthing may be ensured by an ESD workplace (electro static discharge).
NOTICE
Damage to the device!
When replacing the battery or the sensor, make sure that no components are damaged or short-circuited.
► Do not use sharp objects to remove the battery or the sensor.

5.1 Replacing the battery
WARNING

Risk of explosion!
► Only the lithium battery type LBT 01** (order no. 83 26 856) may be used.
► Do not remove or replace the batteries in potentially explosive atmospheres. Do no throw used batteries into fire or try to open them by force. Dispose of batteries in accordance with the national provisions.

The battery is part of the Ex approval.

  • Switch off the gas detector. To replace the battery, see Page 227, fig. D.

The battery connector must be snapped in when it is inserted.
5.2 Replacing the dust and water filter

  • See the technical manual to replace the dust and water filter.

For gas detectors with DrägerSensor XXS Ozone (O3) or Phosgene, the front casing must be replaced due to the special diaphragm, see technical manual.
5.3 Cleaning
The gas detector does not require any special care. If it is heavily soiled, wash the gas detector with cold water and use a sponge if necessary. Use a cloth to dry the gas detector.
NOTICE
………..

Damage to the gas detector!
Rough cleaning objects (e.g. brushes), cleaning agents and solvent may destroy the dust and water filter.
► Only clean the gas detector with cold water and, if necessary, with a sponge.

Device settings

Only trained and qualified personnel may change the device settings. For more information on configuration, see the technical manual.
6.1 Factory setting
The factory settings may differ for customised orders.

Bump test mode Quick bump test
Vibration alarm On
Bump test interval Off
Operation signal On
D-Light On
Switch off Always
Data logger interval 1 min
Service life (user) Off

6.2 Alarm settings

Alarm| Display| Latching| Acknowled geable| LED| Horn| Vibration
---|---|---|---|---|---|---
Alarm 1| A1| | | | |
Alarm 2| A2| | | | |
STEL| STEL| | | | |
TWA| TWA| | | | |
 Pre-alarm1)| | | | | |
Main alarm2)| | | | | |
Device error| | | | | |

  1. After the initial battery pre-alarm, the battery’s life span amounts to between 1 day and 2 weeks under normal conditions of use. The life span is shorter at low temperatures and/ or in the event of alarms.
  2. The gas detector automatically switches off after 10 s.

Disposal

**** This product must not be disposed of as municipal waste. It is therefore marked with the adjacent symbol. You can return this product to Dräger free of charge. For information please contact the national marketing organizations or Dräger.
Batteries and rechargeable batteries may not be disposed of as municipal waste. They are therefore marked with the adjacent symbol. Dispose of batteries at battery collection centers as specified by the applicable regulations.

Technical data

8.1 Gas detector
Ambient conditions during operation

Temperature (temperature depending on the sensor)| up to -30 °C … +55 °C
(briefly up to 1 h -40 °C … +55 °C)
---|---
Humidity| 10 … 90 % r. h., non-condensing
Pressure| 700 … 1300 hPa
Ambient conditions during storage:
Temperature| 0 … 40 °C
Humidity| 30 … 80 % r. h., non-condensing
Typ. battery lifetime (under normal conditions):
24 h operation/day,
1 min alarm/day| 24 months
O2 sensor: 10 months
Dual sensors (without O2): 12 months
Typ. battery lifetime (under normal conditions, with Bluetooth module)
24 h operation/day,
1 min alarm/day| 14 months
O2 sensor: 8 months
Dual sensors (without O2): 11 months
Typ. battery lifetime (under normal conditions, with Bluetooth® module)
8 h operation/day,
1 min alarm/day| 30 months
O2 sensor: 12 months
Dual sensors (without O2): 22 months
Alarm volume| approx. 90 dBA at a distance of 30 cm
Dimensions (not including clip)| 64 x 84 x 20 mm
Weight| approx. 106 g (113 g with clip)
Degree of protection| IP 68

Dr ger PAC6000 Single Gas Monitors - FIGURE 2Dr
ger PAC6000 Single Gas Monitors - ICON 11 contact@gazdetect.com
+33(0)1.64.09.35.16
Dräger Safety AG & Co. KGaA
Revalstraße 1
23560 Lübeck, Germany
Tel +49 451 882 0
Fax +49 451 882 20 80
www.draeger.com
90 33 739 – GA 4623.700
© Dräger Safety AG & Co. KGaA
Edition 05 – Juni 2021 (Edition 01 – December 2016)
Subject to alteration

Documents / Resources

| Dr ger PAC6000 Single Gas Monitors [pdf] Instruction Manual
PAC6000, PAC6000 Single Gas Monitors, Single Gas Monitors, Gas Monitors, Monitors, PAC6500, PAC8000, PAC8000V
---|---

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals