Dr ger X-node Indoor and Outdoor Gas Monitoring Instruction Manual
- June 9, 2024
- Dr ger
Table of Contents
X-node Indoor and Outdoor Gas Monitoring
Instruction Manual
Safety-related information
1.1 Basic safety information
- Before using this product, carefully read these instructions for use and those of the associated products.
- Strictly follow the instructions for use. The user must fully understand and strictly observe the instructions. Use the product only for the purposes specified in the intended use section of this document.
- Do not dispose of the instructions for use. Ensure that they are retained and appropriately used by the product user.
- Comply with all local and national rules and regulations associated with this product.
- Maintenance work may only be carried out by Dräger or by qualified personnel trained by Dräger MSI. Otherwise, Dräger MSI will reject any responsibility for the product working properly after repair and for the validity of approvals.
- Use only genuine Dräger spare parts and accessories for maintenance. Otherwise, the functional integrity of the product may be impaired.
- Do not use faulty or incomplete products. Do not modify the product.
- Notify Dräger in the event of any component fault or failure.
- Only store the product at room temperature and without exposure to solvents, plasticizes, exhaust gases or combustibles.
- Do not use the product in a potentially explosive atmosphere.
- Do not open rechargeable batteries or throw them into a fire.
Conventions in this document
2.1 Meaning of the warning notices
The following warning notices are used in this document to alert the user to
potential hazards. The meanings of the warning notices are defined as follows:
Warning sign | Signal word | Consequences of non-compliance |
---|---|---|
CAUTION | Indicates a potentially hazardous situation. If not avoided, |
it could result in physical injury. It may also be used to alert against unsafe practices.
2.2 Trademarks
Trademark | Owner of trademark |
---|---|
DrägerSensor ® | Dräger |
Bluetooth ® | Bluetooth SIG, Inc. |
Description
3.1 Product overview
52533
- Mounting lug
- Status LED
- Housing of the Dräger X-node sensing head
- Sensor
- Bayonet ring
- USB port/USB port seal
Function description
The Dräger X-node sensing head contains the signal processing electronics of a
sensor in the DrägerSensor ® format. The sensing head and sensor together
constitute the measurement system. The Vol% (volume percent), ppb (part per
billion) or ppm (parts per million) measuring units are volume concentrations.
Irrespective of the sensor used, the ambient pressure, temperature, relative
humidity, acceleration and the alignment of the device are measured.
4.1 Status visualization
The status LED is controlled by the set alarm thresholds. The alarm thresholds
are preconfigured specifically for each sensor (see sensor technical data) and
can be changed individually.
Table 1
Status LEDs flashing | Meaning |
---|---|
Flashing green | The measured value is below the first alarm threshold. |
Flashing yellow | The measured value is between the first and second alarm |
threshold.
Flashing red| The measured value is above the second alarm threshold.
Flashing blue| The sensor is not accepted by X node (incorrect sensor).
Other status messages can be found in the device’s technical manual.
Intended use
The measurement system constantly measures the concentration of the harmful gas, depending on the sensor used, in the ambient air under atmospheric conditions.
Usage
6.1 Prerequisites for use
CAUTION
Risk of malfunction
The function of the sensing head is tested prior to delivery, but a functional
test must still be carried out after installation (see chapter “Testing the
sensitivity”).
► A functional test must be carried out by the user.
6.2 Installing the sensor
- Twist off the bayonet ring (5).
- Remove the sensor from its packaging.
- Place the sensor (4) in the opening.
- Secure the sensor with the bayonet ring.
Dräger X-node app
To configure the sensing head, install the Dräger X-node app on a Bluetooth-
enabled iOS or Android device.
The Dräger X-node app is used for data exchange with Dräger Xnode gas sensing
heads via Bluetooth. The initial connection to the device via Bluetooth can
only be established with the help of an Owner Key for reasons of safety. The
Owner Key is enclosed in the package. Additional information can be found in
the device’s technical manual.
X-node 1
Owner Key: 2M5AF9V23H6R
Part Number: 5601008
Serial Number: KSPD-0177
LW:D0:0000000000000055:FC824A9A324C0E54:04BC0001:O2M5AF9V23H6R:SKSPD0177:P722A53C10E545601008:C4B90
Select the sensing head from the menu. The following functions are supported, among others:
- Defining the PIN
- LoRa provisioning
- Querying user authorization
- Displaying device data
- Displaying measured values
- Configuring the sensing head
- Calibrating the sensor
- Defining the alarm thresholds
- Carrying out firmware updates
- Logger of the last 7 days
Further operating instructions are provided in the Dräger X-node app.
Getting started
8.1 Mounting Dräger X-node
CAUTION
The sensing head can be damaged.
To ensure product safety even when an external power supply is connected to
the micro USB socket, only an approved, commercially available USB power
supply unit may be used that does not exceed the parameters 5.25 V / 3 A / 15
W.
► The sensing head may only be used with USB power supply units that comply
with the USB 3.1 (Gen2) specification or lower.
CAUTION
The sensing head can be damaged.
No IP rating applies during the charging process.
► The sensing head must be protected against spray water and dust while the
rechargeable battery is charging.
The internal rechargeable battery of the sensing head must be charged as
follows prior to the installation of the sensor in the sensing head (3):
- Remove the seal for the micro USB port (6).
- Connect the plug of a micro USB cable to the micro USB port (6) of the sensing head.
- Connect the micro USB cable to a compatible power supply unit.
- Connect the power supply unit to the mains power supply.
- When the charging process is complete, the micro USB port must be resealed with the seal.
For optimum measurements the sensing head must be installed at the centre of
the area to be monitored. The sensing head must be installed so that it is
accessible for maintenance work and suspended from the mounting lug (1) or
installed upright on the bayonet ring (5).
The sensing head must be mounted vertically with the sensor (5) pointing down,
±30 ° at a location with low vibration.
8.2 LoRa gateway connection
For successful data exchange between sensing head and gateway,the gateway must
be within range. Additional information on LoRa can be found in the device’s
technical manual.
8.3 Checking the sensitivity and the zero-point
Zero-point
- Start the Dräger X-node app.
- Press the “Calibrate zero-point” button in the “Calibration” menu item.
- Expose the sensor for zero calibration to an appropriate zero point gas (see app).
Sensitivity
- Start the Dräger X-node app.
- Press the “Span calibration” button in the “Calibration” menu item.
- Follow the instructions.
- Apply the gas from the test gas cylinder (1) with the gas supply adapter (2) to the sensor with a minimum of 0.5 L/min.
✓ The calibration is performed as soon as the sensor signal is stable or the maximum calibration time has expired. The maximum calibration time depends on the sensor type used (e.g. CO 6809605 3 min., NH3 6813735 5 min.).
Maintenance
The sensing head and the LoRa gateway do not require maintenance. The service
life of the used sensor depends on the sensor type. When the sensor is taken
out of the sensing head, the measurement system switches into the power-save
mode. Further information about the sensor can be found in the respective
sensor data sheet.
9.1 Maintenance
Observe the charge status of the internal rechargeable battery. Carry out a
daily visual inspection to check the operational readiness. Have specialists
carry out an annual device inspection and sensor calibration.
CAUTION
Health hazard
The dust filter in front of the sensor must not be contaminated. A
contaminated filter prolongs the response time of the sensor. This affects the
sensitivity.
► Replace the contaminated filter.
9.2 Cleaning
Clean the sensing head surface with lukewarm water. Check for mechanical
damage.
CAUTION
The sensing head may be damaged
The sensing head must not be present when the area of use is cleaned and
disinfected.
► Remove the sensing head from the area of use.
Transport
CAUTION
Risk of damage to the sensor
The sensor must be packed in pressure-tight packaging for shipping via air
transport. The sensor can lose its measuring sensitivity.
► Only transport the sensor via air transport in the appropriate packaging.
Storage
The storage temperature may be between 0 °C and +40 °C. Note the maximum storage period as detailed on the packaging label.
Disposal
This product must not be disposed of as household waste. This is indicated by
the adjacent symbol.
You can return this product to Dräger free of charge. For information please
contact the national marketing organizations or Dräger.
Technical operating data
Messgase:| CI2, CO, CO2, H2, H2 O2, H2 S, NH3, NO, NO 2, O 2, O3, other gases
on request
---|---
Measuring range:| See sensor data sheet
Transmission frequency:| LoRa = 868 MHz, Bluetooth = 2,4 GHz LoRa = EU 863 –
870 MHz, AU 915 – 928 MHz, US 902 – 928 MHz, AS 920 – 923 MHz Bluetooth = 2,4
GHz
Memory function:| Real time clock with backup, data logger
Electrical data|
Battery type:| Lithium-Polymer-Akku
Battery life:| up to 24 months
Battery voltage:| 3.6V
Advertisement:| 3 color LED
Interface:| USB interface
Housing:|
Case material:| Polycarbonat
Housing protection class:| IP65
Dimensions:| ca. 170 mm x 90 mm
Weight:| ca. 300 g
Environmental conditions|
Temperature:| -20ºC to +50ºC
Humidity:| 5% to 95% r. f
Print:| 700 hPa bis 1300 hPa
Electric security:| ****
Radio approval:| EN 62311:2008
EN 300 220-1 V3.1.1
EN 300 220-2 V3.1.1
EN300 328 V2.2.2: (2019-07)
47 CFR Part 15.247
AS/NZS 4268:2017
13.1 Technical data when operated with CO2 sensor 6800482
Operating pressure range: | 700 hPa to 1300 hPa |
---|---|
Measuring range: | 0 to 10000 ppm |
Accuracy at 25 °C
±30 ppm, +3 % of measured value
Alarm threshold A1:| 1000 ppm
Alarm threshold A2:| 2000 ppm
13.2 Technical data when operated with NH3 AL sensor 6813735
Operating pressure range: | 700 hPa to 1300 hPa |
---|---|
Measuring range: | 0 to 100 ppm |
Resolution: | 0.5 ppm |
Accuracy: | 1.5 ppm or ± 10 % of the measured value (larger value applies) |
Response time: | T50 < 30 seconds |
Long-term drift: | < ± 10 % of measured value/year |
13.3 Technical data of ambient sensors
Ambient pressure:| Accuracy
700 hPa to 1100 hPa = ± 1.0 hPa,
1100 hPa to 1300 hPa = ± 1.5 hPa
---|---
Temperature:| Accuracy
0 °C to 50 °C = ± 1.0 °C
-20 °C to 0 °C = ± 1.25 °C
Humidity:| 5 % to 95 % RH: Accuracy ± 5 % RH
Manufacturer
Dräger MSI GmbH
Rohrstraße 32
D-58093 Hagen
Germany
+49 2331 9584-0
9300763 – 4675.160 me
© Dräger Safety AG & Co. KGaA
Edition: 03 – 2023-02
Subject to alterations
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>