PPI FrizCon Deep Freezer Controller with Graphic Display Instruction Manual

June 11, 2024
PPI

PPI FrizCon Deep Freezer Controller with Graphic Display

PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-
product

OPERATOR PARAMETERS

OPERATOR PARAMETERS

Parameters| Settings (Default Value)
Filter Repair Ack > >| Yes No

(Default : No)

Cascade HPLP Ack >> Cabinet HPLP Ack >>| Yes No

(Default : No)

Ctrl Setpoint >>| Setpoint Low Limit to Setpoint High Limit

(Default : -80.0)

Ctrl Lo Deviation >>| 0.2 to 99.9

(Default : 2.0)

Ctrl Hi Deviation >>| 0.2 to 99.9

(Default : 2.0)

Ctrl Zero Offset >>| -50.0 to 50.0

(Default : 0.0)

Filter Zero Offset >> or

Coil Zero Offset >>

| -50.0 to 50.0

(Default : 0.0)

Defrost Start >> Defrost Stop >>| Start Stop

(Default : NA)

Change Password >>| 1 to 100

(Default : 0)

SUPERVISORY SERIAL PARAMETERS
Parameters Settings (Default Value)
Device Id >> 1 to 127

(Default : 1)

Baud Rate >>| 2400, 4800,

9600, 19200,

38400, 57600

(Default : 9600)

Parity >>| None Even Odd

(Default : Even)

Serial Write >>| No Yes

(Default : No)

SUPERVISORY CONTROL

Parameters| Settings (Default Value)
Setpoint Lo Limit >>| -100.0 to Setpoint HI Limit (Default : -100.0)
Setpoint Hi Limit >>| ‘Single’ or ‘Single with Filter’ Compressor Mode

Setpoint LO Limit to 100.0

‘Cascade’ or ‘Cascade with Filter’ or ‘Cascade with Coil’ Compressor Mode Setpoint LO Limit to Cascade SP

(Default : -10.0)

Comp On Offset >>| 0.1 to 99.9

(Default : 1.0)

Comp Off Offset >>| 0.1 to 99.9

(Default : 0.1)

PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig
10

FACTORY ALARM PARAMETERS

Parameters| Settings (Default Value)
Low Alarm Hyst >>| 0.1 to 99.9

(Default : 0.2)

Low Alarm Inhibit >>| Disable Enable

(Default : Enable)

High Alarm Hyst >>| 0.1 to 99.9

(Default : 0.2)

High Alarm Inhibit >>| Disable Enable

(Default : Enable)

FACTORY COMPRESSOR OPERATION

Parameters| Settings (Default Value)
Operation Mode >>| Single Single with Filter

Cascade Cascade with Filter Cascade with Coil

(Default : Cascade with Coil)

Cascade Sp >>| -70.0 to 10.0

(Default : -30.0)

Off To On Delay >>| 0 to 1000 Sec (Default : 200 Sec)
Switching Delay >>| 0 to 1000 Sec (Default : 15 Sec)
Cascade Comp Hyst >>| 1 to 50.0

(Default : 20)

FACTORY FILTER OPERATION

Parameters| Settings (Default Value)
Filter Alarm Sp >>| -100.0 to 100.0

(Default : 55.0)

Filter Alarm Hyst >>| 0.1 to 99.9

(Default : 0.2)

Filter Cut-off Sp >>| Alarm SP to 100.0 (Default : 60.0)
FACTORY DEFROST OPERATION

Parameters| Settings (Default Value)
Operation >>| Disable Automatic Manual

(Default : Disable)

Defrost Time [M:S] >>| 1 to 99.59 (Min:Sec) (Default : 5 Sec)
Defrost Interval [H:M] > >| 1 to 99.59 (Hour:Min) (Default : 8 Min)
Parameters| Settings (Default Value)
---|---
Switch Logic >>| Open : Door Open Close : Door Open

(Default : Close : Door Open)

Door Alrm Dly (sec) >>| 0 to 1000 Sec (Default : 60 Sec)
FACTORY CASCADE COMP HPLP

(Available only for Cascade Operation Mode)


Parameters| Settings (Default Value)
Enable >>| No Yes

(Default : No)

Switch Logic >>| Open : HP-LP Close : HP-LP

(Default : Close : HP-LP)

FACTORY CABINET COMP HPLP
Parameters Settings (Default Value)
Enable >> No Yes

(Default : No)

Switch Logic >>| Open : HP-LP Close : HP-LP

(Default : Close : HP-LP)

FACTORY FACTORY DEFAULT
Parameters Settings (Default Value)
Set To Default >> No Yes

(Default : No)

FACTORY PASSWORD

Parameters| Settings (Default Value)
Change Password >>| 2000 to 2999

(Default : 321)

FACTORY EXIT

Parameters| Settings (Default Value)
Exit Setup Mode >>| No Yes

(Default : No)

Table 1

What it means| Range (Min. to Max.)| Resolution
3-wire, RTD Pt100| -199.9 to 600.0°C| Fixed 0.1°C
0 to 20mA DC current| ****

-199.9 to 999.9 units

| ****

Fixed

0.1 unit

4 to 20mA DC current
0 to 5.0V DC voltage
0 to 10.0V DC voltage
1 to 5.0V DC voltage

FRONT PANEL LAYOUT

Keys OperationPPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-
fig 1

PV Error Indications

Message Error Type Cause

PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig
2| Sensor Open| RTD Pt100 Sensor/Cable is Broken / Open
| | | Over-range| Temperature above Max. Specified Range
| Under-range| Temperature below Min. Specified Range

ELECTRICAL CONNECTIONS

PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-
Display-fig 5

CASCADE OPERATION MODE

* For more Information refer User Manual

Cascade Compressors Operation Mode (with Coil Sensor)PPI-FrizCon-Deep-
Freezer-Controller-with-Graphic-Display-fig 6

Cascade Compressors Operation Mode (without Coil Sensor)PPI-FrizCon-Deep-
Freezer-Controller-with-Graphic-Display-fig 6

This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net

101, Diamond Industrial Estate, Navghar, Vasai Road (E), Dist. Palghar – 401 210.
Sales : 8208199048 / 8208141446
Support : 07498799226 / 08767395333
E: sales@ppiindia.net,
support@ppiindia.net