Dr ger 64 00 000 Accuro Gas Detector Pump Instruction Manual

October 30, 2023
Dr ger

Dr ger 64 00 000 Accuro Gas Detector Pump

Product Information

accuro

The accuro is a manual bellows pump used for taking measurements of gases and vapors in ambient air. It has a stroke counter and an end-of-stroke indicator for accurate and consistent measurements. The pump is designed to be lightweight, compact, and easy to use.

Product Technical Specifications

  • Hub volume: 170 mm
  • Width: 45 mm
  • Height: 85 mm
  • Weight: 250 g
  • Temperature range: 0 to 40°C
  • Humidity range: 0 to 95% non-condensing
  • Ex-approvals: ATEX

Product Preparation

Pump Leak Test

To perform a pump leak test, hold the pump with the end-of-stroke indicator and stroke counter facing the user. Squeeze the pump until it stops, then release it until the bellows are fully expanded. Repeat this process until the end-of-stroke indicator appears. Squeeze the pump completely again and repeat until the number on the stroke counter corresponds to that given in the instructions for use.

Preparing the Tube

To prepare the tube, insert the tip into the opening of the lower pump plate and keep it pressed against the ceramic edge while turning it so that the glass is scored. Pull the tube over at an angle until the tip breaks off. Discard the broken tip. Open the other end of the tube in the same way. Alternatively, turn the handle of the pump to the side and insert the tube tip in the free hole of the plate. Tilt the tube sideways until the tip breaks off. Open the other end of the tube in the same way. Caution: the tips of the tube may have sharp edges, so there is a risk of injury.

Making a Measurement

To take a measurement, hold the pump so that the end-of-stroke indicator and stroke counter are facing the user. Squeeze the pump until it stops, then release it until the bellows are fully expanded. Repeat this process until the end-of-stroke indicator appears. Squeeze the pump completely again and repeat until the number on the stroke counter corresponds to that given in the instructions for use. Evaluate the result in accordance with the instructions for use.

Product Maintenance

To clean or replace the exhaust valve, unscrew the bolt by using the hexagonal screwdriver supplied. Lift off lower pump plate, e.g., by using a coin. Pull valve disc from its seating. Wipe valve seating with a damp cloth. Moisten the stem on a new valve disc and push it into place.

For Your Safety

Strictly follow the Instructions for Use

Any use of the apparatus requires a full understanding and strict observation of these instructions. The apparatus is only to be used for purposes specified here.

Maintenance

  • The apparatus must be inspected and serviced regularly by suitably qualified persons.
  • Repairs to the apparatus may only be carried out by trained service personnel.
  • We recommend that you enter into a service contract with Dräger Safety and that all repairs be conducted by Dräger Safety.
  • Only authentic Dräger spare parts may be used for maintenance.

Accessories

  • Do not use accessory parts other than those listed in the order list.

Use in areas subject to explosion hazards

Equipment or components which have been inspected and approved in accordance with national regulations concerning electrical equipment in explosion-hazard areas (ElexV) or European explosion-protection regulations, may be used only under the conditions explicitly specified in the approval. The equipment or components may not be modified in any manner. The use of faulty or incomplete parts is forbidden. The appropriate regulations must be observed (e.g. EIExV and EN 50014) when carrying out repairs on the equipment or components.

Liability for proper function or damage

The liability for the proper function of the apparatus is irrevocably transferred to the owner or operator, to the extent that the apparatus was serviced or repaired by personnel not employed or authorized by Dräger Safety or when the apparatus was used in a manner not conforming to its intended use. Dräger Safety cannot be held responsible for damage caused by non-compliance with the recommendations given above. The warranty and liability provisions of the terms of sale and delivery of Dräger Safety are likewise not modified by the recommendations given above. Dräger Safety AG & Co. KGaA

Intended Use

  • Gas Detector Pump for all Dräger short term tubes and sampling tubes.

What is What

  1. Tube opener with ceramic cutter
  2. Cover plate
  3. Socket
  4. Rubber bellows
  5. Lower pump plate
  6. Stroke Counter Reset Button
  7. Stroke counter
  8. End-of-Stroke Indicator (End-of-Stroke membrane)
  9. Bolt to secure lower pump plate in place
  10. Hole to break off the tube tip
  11. Hole for attachment of loop handle
  12. Loop handle with snap hook

Preparation

Pump leak test (A)

  • Insert unopened Dräger tube into socket (A1).
  • Squeeze pump completely and release (A2).
  • Pump is adequately leak-proof if the end-of-stroke indicator (B) has not appeared after 15 minutes. If it has, then see chapter “Fault, Cause, Remedy”.
  • Remove unopened Dräger tube from socket.
  • Press button to reset counter to zero (C); button is recessed to prevent accidental resetting.

Preparing the tube (D)

  • Select appropriate Dräger tube for the measurement required. Follow the relevant Instructions for Use!
  • Open tip with Dräger tube opener (item no. 64 01 200).

Or

  • Insert the tip into the opening of the lower pump plate (D1); keep it pressed against ceramic edge while turning it, so that the glass is scored.
  • Pull the tube over at an angle until the tip breaks off.
  • Discard broken tip.
  • Open the other end of tube in the same way.

Or

  • Turn handle of pump to the side and insert tube tip in the free hole of the plate (D2).
  • Tilt tube sideways until the tip breaks off.
  • Open other end of tube in the same way. Be careful when opening the Dräger tube as glass splinters may come off! Keep the pump facing downwards by holding the loop handle, so to prevent glass splinters from entering the pump!
  • Insert Dräger tube (now open at both ends) into pump, the arrow must point towards the pump (E).

Caution: the tips of the tube may have sharp edges; risk of injury!

Making a Measurement

  • Check number of strokes required from the Instructions for Use of the Dräger tube.
  • Hold pump so that the End-of-stroke indicator and the stroke counter are facing the user (F).
  • Squeeze pump until stop.
  • Release pump until bellows are fully expanded.
  • When the end-of-stroke indicator (F1) appears, squeeze pump completely again.
  • Repeat until the number on the stroke counter (F2) corresponds to that given in the Instructions for Use.
  • Evaluation of the result in accordance with the Instructions for Use of the tube in question.
  • Remove used Dräger tube from socket.
  • Flush pump with a few pump strokes in clean air.

Maintenance

To clean or replace the exhaust valve

  • Unscrew bolt by using the hexagonal screwdriver supplied (G1).
  • Lift off lower pump plate, e. g. by using a coin. (G2).
  • Pull valve disc from its seating (H1).
  • Wipe valve seating with a damp cloth.
  • Moisten the stem on a new valve disc and push it firmly into the center hole of the valve seat.
  • Check correct seating by pulling on valve disc lightly.
  • Put back lower pump plate and press it into place (H2).
  • Insert bolt again and screw down handtight. Perform the leak test as described (A).

Replacing the socket/Replacing the membrane

  • Press spring hook down with a suitable tool (e. g. ball pen) (I1).
  • Take out cover plate (I2).
  • Cover plate holds End-of-Stroke Indicator assembly in place.
  • Press spring flanges together and remove socket. (J).
  • Insert new socket.
  • Using a new End-of-Stroke Indicator Membrane, place parts into the housing in the sequence illustrated (K).
  • Observe appropriate position.
  • Insert spring hook of cover plate and push on firmly.
  • Perform the leak test as described (A).

Technical Data

  • Stroke volume: 100 ±5 cm3
  • Length: 170 mm
  • Width: 45 mm
  • Height: 85 mm
  • Weight: 250 g
  • Ambient conditions (Operation and storage)
    • Temperature: –20 oC to +50 oC
    • Humidity; 0 to 95 % non-condensing

Tests and Approvals

Ex-Approvals

  • ATEX: I M1 Ex h I Ma
  • I 1G Ex h IIB T6 Ga
  • II 2G Ex h lIC T6 Gb
  • II 1D Ex h IIIC T85°C Da
    • -20°C ≤ Ta ≤ +50°C
  • BVS 04 ATEX H 068X

ATTENTION

  • Some Dräger tubes must not be used in areas subject to explosion hazards.
  • See the relevant Instructions for Use of the Dräger tubes.

Fault, Cause, Remedy

Fault Cause Remedy
End-of-stroke indicator visible within 15 min Foreign matter in exhaust
valve, Defective valve disc Clean valve and valve seating. Replace valve when

required, (K)
Tube too loose in socket| Insert tube tightly into socket, (E)
Pump does not open| One of the tube ends has not been opened| Open tube at both ends, (D)
Clogged socket| Replace socket, (J)

Order List

Name Item no.
accuro Gas detector pump 64 00 000
Spare part kit 64 00 220
Loop handle, complete 64 00 208
Opener for Dräger tubes 64 01 200
Extension hose 3 m 64 00 077
Gas detector kit 64 00 260
accuro 2000 64 00 200

DGUV Test Certificate

EU-Declaration of Conformity

Contact

Dräger Safety AG & Co. KGaA Revalstraße 1

  • 23560 Lübeck, Germany
  • Tel: +49 451 882-0
  • Fax: +49 451 882-20 80
  • www.draeger.com
  • 90 22 003 – GA 4241.000_MUL041

Official Distributor

© Dräger Safety AG & Co. KGaA
Edition 11 – December 2019

Subject to alteration.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Dr ger User Manuals

Related Manuals