LabCon Map PC Recording 4 Channel Mapping PC Software User Manual
- June 11, 2024
- PPI
Table of Contents
LabCon Map PC Recording 4 Channel Mapping PC Software
This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net
OPERATOR PAGE PARAMETERS
Parameters | Settings (Default Value) |
---|
Time Start Command > >
Time Abort Command > >
| Yes No
(Default : No)
Time Interval (H:M) >>| 0.00 to 500.00 (HH:MM)
(Default : 0.10)
Ctrl Set Value >>| Setpoint LO limit to Setpoint HI limit (Resolution 0.1°C for RTD /
DC Linear & 1°C for Thermocouple)
(Default : 25.0)
Ctrl Lo Deviation >>| For RTD & DC Linear : 0.2 to 99.9 For Thermocouple : 2 to 99
(Default : 2.0)
Ctrl Hi Deviation >>| For RTD & DC Linear : 0.2 to 99.9 For Thermocouple : 2 to 99
(Default : 2.0)
Change Password >>| 1 to 100
(Default : 0)
SUPERVISORY
SUPERVISORY > SENSOR INPUT
Parameters | Settings (Default Value) |
---|---|
Ctrl Zero Offset >> | -50 to 50 |
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default : 0.0)
Map-1 Zero Offset >> Map-2 Zero Offset >> Map-3 Zero Offset >> Map-4 Zero Offset >>| -50 to 50
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default : 0.0)
SUPERVISORY > RECORDING
Parameters | Settings (Default Value) |
---|
Rec.
Interval (min) >>
| 0 to 250 Min (Default : 5 Min)
Alarm
Toggle Record >>
| Enable Disable
(Default : Enable)
Del PC Records >>| Yes No
(Default : No)
Del
Display Record >>
| Yes No
(Default : No)
GSM Machine ID >>
(refer note)
| 1 to 128
(Default : 1)
RTC Time >>| 0 to 23 for Hour
0 to 59 for Min
(Default : NA)
RTC Date >>| 1 to 31 (Default : NA)
RTC Month >>| 1 to 12 (Default : NA)
RTC Year >>| 2000 to 2099
(Default : NA)
Device Id >>| 1 to 127
(Default : 1)
SUPERVISORY > CONTROL
Parameters | Settings (Default Value) |
---|---|
Tune > > | Yes No |
(Default : No)
Setpoint LO Limit >>| Min Range for the Selected Input Type to Setpoint HI Limit (Resolution 0.1°C for RTD /
DC Linear & 1°C for Thermocouple)
(Default : 0.0)
Setpoint| Setpoint LO Limit to
HI Limit >>| Max Range for the Selected
Input Type
| (Resolution 0.1°C for RTD /
| DC Linear & 1°C for Thermocouple)
| (Default : 600.0)
Compressor| 0 to 100
Setpoint| (Resolution 0.1°C for RTD / DC Linear & 1°C for
Thermocouple)
| (Default : 45.0)
Compressor Hyst > >| 0.1 to 99.9
(Default : 2.0)
Heat **Ctrl Action >>**| ON-OFF PID
(Default : PID)
Heat Hyst >>| 0.1 to 99.9
(Default : 0.2)
Heat Only Control| Heat + Cool Control Zone : Single| Heat + Cool
Control Zone : Dual
---|---|---
Proportional Band >>
0.1 to 999.9
(Default : 50.0)
| Proportional Band >>
0.1 to 999.9
(Default : 50.0)
| Cz Prop Band >> Proportional Band for Cool Pre-dominant zone 0.1 to 999.9
(Default : 50.0)
Integral Time >> 0 to 3600 sec (Default : 100 sec)| Integral Time >> 0 to 3600 sec (Default : 100 sec)| Cz Integral Time >>
Integral Time for Cool Pre-dominant zone
0 to 3600 sec
(Default : 100 sec)
Derivative Time >>
0 to 600 sec (Default : 16 sec)
| Derivative Time >>
0 to 600 sec (Default : 16 sec)
| Cz Derivative Time >> Derivative Time for Cool Pre-dominant zone
0 to 600 sec (Default : 16 sec)
Cycle Time >>
0.5 to 100.0 sec (Default : 10.0 sec)
| ****
Cycle Time >>
0.5 to 100.0 sec (Default : 10.0 sec)
| Hz Prop Band >> Proportional Band for Heat Pre-dominant zone 0.1 to 999.9
(Default : 50.0)
Overshoot Inhibit > > Enable Disable
(Default : Disable)
| Overshoot Inhibit > > Enable Disable
(Default : Disable)
| Hz Integral Time >>
Integral Time for Heat Pre-dominant zone
0 to 3600 sec
(Default : 100 sec)
Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec)
| Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec)
| Hz Derivative Time >> Derivative Time for Heat Pre-dominant zone
0 to 600 sec (Default : 16 sec)
Cycle Time >>
0.5 to 100.0 sec (Default : 10.0 sec)
Overshoot Inhibit > > Enable Disable
(Default : Disable)
Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec)
SUPERVISORY > PASSWORD
Parameters | Settings (Default Value) |
---|---|
Change Password >> | 1000 to 1999 |
(Default : 123)
SUPERVISORY > EXI
Parameters | Settings (Default Value) |
---|---|
Exit Setup Mode >> | Yes No |
(Default : No)
FACTORY
FACTORY > CONTROL SENSOR INPUT
Parameters | Settings (Default Value) |
---|---|
Input Type >> | Refer Table 1 (Default : RTD Pt100) |
Signal LO >> | |
Signal HI >> | |
Range LO >> | -199.9 to RANGE HI |
(Default : 0.0)
Range HI >>| RANGE LO to 999.9
(Default : 100.0)
FACTORY > MAPPING PARAMETERS
Parameters | Settings (Default Value) |
---|---|
Mapping Points >> | 0 to 4 |
(Default : 0)
Input Type >>| Refer Table 1 (Default : RTD Pt100)
Signal LO >>| | Input Type| Settings| Default
---|---|---
0 to 20mA| 0.00 to Signal High| 0.00
4 to 20mA| 4.00 to Signal High| 4.00
0 to 5V| 0.000 to Signal High| 0.000
0 to 10V| 0.00 to Signal High| 0.00
1 to 5V| 1.000 to Signal High| 1.000
Signal HI >>| | Input Type| Settings| Default
---|---|---
0 to 20mA| Signal Low to 20.00| 20.00
4 to 20mA| Signal Low to 20.00| 20.00
0 to 5V| Signal Low to 5.000| 5.000
0 to 10V| Signal Low to 10.00| 10.00
1 to 5V| Signal Low to 5.000| 5.000
Range LO >>| -199.9 to RANGE HI
(Default : 0.0)
Range HI >>| RANGE LO to 999.9
(Default : 100.0)
FACTORY > ALARM PARAMETERS
Parameters | Settings (Default Value) |
---|---|
Hysteresis >> | 0.1 to 99.9 |
(Default : 0.2)
Inhibit >>| Yes No
(Default : Yes)
FACTORY > HEAT COOL SELECT
Parameters | Settings (Default Value) |
---|---|
Control Strategy >> | Heat Only Cool Only Heat + Cool |
(Default : Heat + Cool)
Control Strategy : Cool Only
Time Delay (sec) >>| 0 to 1000 Sec (Default : 200 Sec)
Control Strategy : Heat + Cool
Compressor Strategy >>| CONT. OFF CONT. ON SP BASED PV BASED
(Default : CONT. ON)
CONT. ON | SP BASED | PV BASED |
---|
Time
Delay (sec) >>
0 to 1000 Sec (Default : 200 Sec)
| Boundary Set Value >>
0 to 100 (Resolution 0.1°C for RTD / DC Linear
& 1°C for Thermocouple)
| Control Zones >> Single Dual
(Default : Single)
Time Delay (sec) >>
0 to 1000 Sec (Default : 200 Sec)
Time Delay (sec) >> 0 to 1000 Sec (Default : 200 Sec)
FACTORY > SOAK TIMER PARAMETERS
Parameters | Settings (Default Value) |
---|---|
Enable >> | No Yes |
(Default : No)
Start Band >>| 0 to 999.9
(Default : 0.5)
Holdback Strategy >>| None Up Down Both
(Default : None)
Hold Band >>| 0.1 to 999.9
(Default : 0.5)
Heat Off >>| No Yes
(Default : No)
Cool Off >>| No Yes
(Default : No)
Power Recovery >>| Abort Restart Continuous
(Default : Restart)
FACTORY > DOOR OPEN
Parameters | Settings (Default Value) |
---|---|
Enable >> | No Yes |
(Default : No)
Start Band >>| 0 to 999.9
(Default : 0.5)
Holdback Strategy >>| None Up Down Both
(Default : None)
Hold Band >>| 0.1 to 999.9
(Default : 0.5)
Heat Off >>| No Yes
(Default : No)
Cool Off >>| No Yes
(Default : No)
Power Recovery >>| Abort Restart Continuous
(Default : Restart)
FACTORY > MAINS FAILURE
Parameters | Settings (Default Value) |
---|---|
Enable >> | No Yes |
(Default : No)
Start Band >>| 0 to 999.9
(Default : 0.5)
Holdback Strategy >>| None Up Down Both
(Default : None)
Hold Band >>| 0.1 to 999.9
(Default : 0.5)
Heat Off >>| No Yes
(Default : No)
Cool Off >>| No Yes
(Default : No)
Power Recovery >>| Abort Restart Continuous
(Default : Restart)
FACTORY > SMS ALERT
Parameters | Settings (Default Value) |
---|---|
Enable >> | No Yes |
(Default : No)
Start Band >>| 0 to 999.9
(Default : 0.5)
Holdback Strategy >>| None Up Down Both
(Default : None)
Hold Band >>| 0.1 to 999.9
(Default : 0.5)
Heat Off >>| No Yes
(Default : No)
Cool Off >>| No Yes
(Default : No)
Power Recovery >>| Abort Restart Continuous
(Default : Restart)
FACTORY > FACTORY DEFAULT
Parameters | Settings (Default Value) |
---|---|
Set To Default >> | Yes No |
(Default : No)
FACTORY > PASSWORD
Parameters | Settings (Default Value) |
---|---|
Set To Default >> | Yes No |
(Default : No)
FACTORY > EXIT
Parameters | Settings (Default Value) |
---|---|
Exit Setup Mode >> | Yes No |
(Default : No)
TABLE 1
TABLE 1
What it means| Range (Min. to Max.)| Resolution
Type J Thermocouple| 0 to +960°C| ****
Fixed 1°C
Type K Thermocouple| -200 to +1376°C
Type T Thermocouple| -200 to +385°C
Type R Thermocouple| 0 to +1770°C
Type S Thermocouple| 0 to +1765°C
Type B Thermocouple| 0 to +1825°C
Type N Thermocouple| 0 to +1300°C
Reserve
| Reserved for customer specific Thermocouple type not listed above. The type
shall be specified in accordance with the ordered (optional on request)
Thermocouple type.
3-wire, RTD Pt100| -199.9 to 600.0°C| Fixed 0.1°C
0 to 20mA DC current| ****
-199.9 to 999.9 units
| ****
Fixed
0.1 unit
4 to 20mA DC current
0 to 5.0V DC voltage
0 to 10.0V DC voltage
1 to 5.0V DC voltage
FRONT PANEL KEYS
Symbol | Key | Function |
---|---|---|
**** |
Scroll
| Press to scroll through various Process Information Screens in Normal
Operation Mode.
| Alarm Acknowledge| ****
Press to acknowledge and mute (if active) alarm output.
| ****
DOWN
| Press to decrease the parameter value. Pressing once decreases the value by
one count; keeping pressed speeds up the change.
| ****
UP
| Press to increase the parameter value. Pressing once increases the value by
one count; keeping pressed speeds up the change.
| ****
SET-UP
| ****
Press to enter or exit set-up mode.
| ****
ENTER
| Press to store the set parameter value and to scroll to the next parameter.
PV ERROR INDICATIONS
Message | Error Type | Cause |
---|---|---|
OPEN | Sensor Open | Sensor (RTD Pt100) Broken / Open |
OVR | Over-range | Temperature above Max. Specified Range |
UNR | Under-range | Temperature below Min. Specified Range |
ELECTRICAL CONNECTIONS
4 Mapping Channels
CONTACT
Address: 101, Diamond Industrial Estate, Navghar, Vasai Road (E), Dist.
Palghar – 401 210.
Sales: 8208199048 / 8208141446
Support: 07498799226 / 08767395333
E: sales@ppiindia.net,
support@ppiindia.net
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>