LabCon Recording + 4 Channel Mapping + Printer Interface User Manual
- June 11, 2024
- PPI
Table of Contents
LabCon (MAP + Printer)
Recording + 4 Channel Mapping + Printer Interface
User Manual
OPERATOR PAGE PARAMETERS
| Parameters | Settings (Default Value) |
|---|
Time Start Command >>
Time Abort Command >>| Yes No
(Default: No)
Time Interval (H:M) >>| 0.00 to 500.00 (HH:MM)
(Default: 0.10)
Ctrl Set Value >>| Setpoint LO limit to Setpoint HI limit
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default: 25.0)
Ctrl Lo Deviation >>| For RTD & DC Linear: 0.2 to 99.9 For Thermocouple: 2 to
99
(Default: 2.0)
Ctrl Hi Deviation >>| For RTD & DC Linear: 0.2 to 99.9 For Thermocouple: 2 to
99
(Default: 2.0)
Change Password >>| 1 to 100
(Default: 0)
SUPERVISORY > SENSOR INPUT
Parameters| Settings
(Default Value)
---|---
Ctrl Zero Offset >>| -50 to 50
(Resolution 0.1°C for RTD /
DC Linear & 1°C for Thermocouple)
(Default:0.0)
Map-1 Zero Offset >>
Map-2 Zero Offset >>
Map-3 Zero Offset >>
Map-4 Zero Offset >>| -50 to 50
(Resolution 0.1°C for RTD /
DC Linear & 1°C for Thermocouple)
(Default: 0.0)
SUPERVISORY > RECORDING
| Parameters | Settings (Default Value) |
|---|
Rec.
Interval (min) >>| 0 to 250 Min
(Default : 5 Min)
Alarm
Toggle Record >>| Enable Disable
(Default: Enable)
Del Print Records >>| Yes No
(Defaul: No)
Del
Display Record >>| Yes No
(Default: No)
RTC Time >>| 0 to 23 for Hour
0 to 59 for Min
(Default: NA)
RTC Date >>| 1 to 31 (Default: NA)
RTC Month >>| 1 to 12
(Default: NA)
RTC Year >>| 2000 to 2099
(Default: NA)
SUPERVISORY > CONTROL
| Parameters | Settings (Default Value) |
|---|---|
| Tune >> | Yes No |
(Default: No)
Setpoint LO Limit >>| Min Range for the Selected Input Type to Setpoint HI
Limit
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default: 0.0)
Setpoint HI Limit >>| Setpoint LO Limit to Max Range for the Selected
Input Type
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default: 600.0)
Compressor Setpoint >>| 0 to 100
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default: 45.0)
Compressor Hyst >>| 0.1 to 99.9
(Default: 2.0)
Heat Ctrl Action >>| ON-OFF PID
(Default: PID)
Heat Hyst >>| 0.1 to 99.9
(Default: 0.2)
Heat Only Control| Heat + Cool Control Zone: Single| Heat + Cool
Control Zone: Dual
---|---|---
Proportional Band >>
0.1 to 999.9
(Default : 50.0)| Proportional Band >>
0.1 to 999.9
(Default: 50.0)| Cz Prop Band >> Proportional Band for Cool Pre-dominant zone
0.1 to 999.9
(Default: 50.0)
Integral Time >> 0 to 3600 sec (Default : 100 sec)| Integral Time >> 0 to 3600
sec (Default : 100 sec)| Cz Integral Time >>
Integral Time for Cool Pre-dominant zone
0 to 3600 sec
(Default: 100 sec)
Derivative Time >>
0 to 600 sec (Default: 16 sec)| Derivative Time >>
0 to 600 sec (Default: 16 sec)| Cz Derivative Time >> Derivative Time for
Cool Pre-dominant zone
0 to 600 sec (Default : 16 sec)
Cycle Time >>
0.5 to 100.0 sec (Default: 10.0 sec)| Cycle Time >>
0.5 to 100.0 sec (Default : 10.0 sec)| Hz Prop Band >> Proportional Band for
Heat Pre-dominant zone 0.1 to 999.9
(Default : 50.0)
Overshoot Inhibit >> Enable Disable
(Default: Disable)| Overshoot Inhibit >> Enable Disable
(Default: Disable)| Hz Integral Time >>
Integral Time for Heat Pre-dominant zone
0 to 3600 sec (Default : 100 sec)
Cutoff Factor >>
1.0 to 2.0 sec (Default: 1.2 sec)| Cutoff Factor >>
1.0 to 2.0 sec (Default: 1.2 sec)| Hz Derivative Time >> Derivative Time for
Heat Pre-dominant zone
0 to 600 sec (Default: 16 sec)
Cycle Time >>
0.5 to 100.0 sec
(Default: 10.0 sec)
Overshoot Inhibit >> Enable Disable
(Default: Disable)
Cutoff Factor >>
1.0 to 2.0 sec (Default : 1.2 sec)
SUPERVISORY > PRINTER SETTING
| Parameters | Settings (Default Value) |
|---|---|
| Line Per Page >> | 25 to 100 |
(Default: 65)
Top Margine >>| 1 to 5
(Default: 2)
Bottom Margine >>| 1 to 5
(Default: 2)
Header & Footer >>| No
On First Page Every Page
(Default: No)
Enable >>| Yes No
(Default: No)
SUPERVISORY > PRINT EVENT
Parameters| Settings
(Default Value)
---|---
Alarm Toggle >>
RTC Modify >>
Interval Modify >>
Power Up >>
Door Open >>
Mains Fail >>
Setpoint Modify >>| Yes
No
(Default : Yes)
SUPERVISORY > PASSWORD
Parameters| Settings
(Default Value)
---|---
Change
Password >>| 1000 to 1999
(Default: 123)
SUPERVISORY > EXIT
Parameters| Settings
(Default Value)
---|---
Exit Setup Mode >>| Yes
No
(Default: )No
FACTORY > CONTROL SENSOR INPUT
| Parameters | Settings (Default Value) |
|---|---|
| Input Type >> | Refer Table 1 (Default : RTD Pt100) |
| Signal LO >> | |
| --- | --- |
| 0 to 20mA | 0.00 to Signal High |
| 4 to 20mA | 4.00 to Signal High |
| 0 to 5V | 0.000 to Signal High |
| 0 to 10V | 0.00 to Signal High |
| 1 to 5V | 1.000 to Signal High |
| Signal HI >> | |
| --- | --- |
| 0 to 20mA | Signal Low to 20.00 |
| 4 to 20mA | Signal Low to 20.00 |
| 0 to 5V | Signal Low to 5.000 |
| 0 to 10V | Signal Low to 10.00 |
| 1 to 5V | Signal Low to 5.000 |
| Range LO >> | -199.9 to RANGE HI |
(Default : 0.0)
Range HI >>| RANGE LO to 999.9
(Default : 100.0)
FACTORY > ALARM PARAMETERS
| Parameters | Settings (Default Value) |
|---|---|
| Hysteresis >> | 0.1 to 99.9 |
(Default : 0.2)
Inhibit >>| Yes No
(Default : Yes)
FACTORY > HEAT COOL SELECT
| Parameters | Settings (Default Value) |
|---|---|
| Control Strategy >> | Heat Only Cool Only Heat + Cool |
(Default : Heat + Cool)
Control Strategy: Cool Only
Time Delay (sec) >>| 0 to 1000 Sec (Default : 200 Sec)
Control Strategy: Heat + Cool
Compressor Strategy >>| CONT. OFF CONT. ON SP BASED PV BASED
(Default: CONT. ON)
CONT. ON| SP BASED| PV BASED
---|---|---
Time Delay (sec) >>
0 to 1000 Sec
(Default: 200 Sec)| Boundary Set Value >>
0 to 100
(Resolution 0.1°C for RTD / DC Linear & 1°C for Thermocouple)
(Default : 45.0)| Time Delay (sec) >>
0 to 1000 Sec
(Default: 200 Sec)
Control Zones >>
Single
Dual
(Default: Single)
Time Delay (sec) >>
0 to 1000 Sec
(Default : 200 Sec)
FACTORY > SOAK TIMER PARAMETERS
| Parameters | Settings (Default Value) |
|---|---|
| Enable >> | No Yes |
(Default: No)
Start Band >>| 0 to 999.9
(Default: 0.5)
Holdback Strategy >>| None Up Down Both
(Default : None)
Hold Band >>| 0.1 to 999.9
(Default : 0.5)
Heat Off >>| No Yes
(Default : No)
Cool Off >>| No Yes
(Default: No)
Power Recovery >>| Abort Restart Continuous
(Default: Restart)
FACTORY > DOOR OPEN
| Parameters | Settings (Default Value) |
|---|---|
| Enable >> | Yes No |
(Default : No)
Switch Logic >>| Close: Door Open Open: Door Open
(Default: Close: Door Open)
Door
Alrm Dly (sec) >>| 0 to 1000 Sec
(Default : 60 Sec)
FACTORY > MAINS FAILURE
| Parameters | Settings (Default Value) |
|---|---|
| Enable >> | Yes No |
(Default: No)
Switch Logic >>| Close: Mains Fail Open: Mains Fail
(Default: Close: Mains Fail)
FACTORY > PASSWORD
| Parameters | Settings (Default Value) |
|---|---|
| Change Password >> | 2000 to 2999 |
(Default: 321)
FACTORY > FACTORY DEFAULT
| Parameters | Settings (Default Value) |
|---|---|
| Set To Default >> | Yes No |
(Default: No)
FACTORY > EXIT
| Parameters | Settings (Default Value) |
|---|---|
| Exit Setup Mode >> | Yes No |
(Default: No)
TABLE 1
| What it means | Range (Min. to Max.) | Resolution |
|---|---|---|
| Type J Thermocouple | 0 to +960°C | Fixed 1°C |
| Type K Thermocouple | -200 to +1376°C | |
| Type T Thermocouple | -200 to +385°C | |
| Type R Thermocouple | 0 to +1770°C | |
| Type S Thermocouple | 0 to +1765°C | |
| Type B Thermocouple | 0 to +1825°C | |
| Type N Thermocouple | 0 to +1300°C | |
| Reserve | Reserved for customer specific Thermocouple type not listed above. |
The type shall be specified in accordance with the ordered (optional on
request) Thermocouple type.
3-wire, RTD Pt100| -199.9 to 600.0°C| Fixed
0.1°C
0 to 20mA DC current| -199.9 to 999.9 units| Fixed
0.1 unit
4 to 20mA DC current
0 to 5.0V DC voltage
0 to 10.0V DC voltage
1 to 5.0V DC voltage
FRONT PANEL KEYS
| Symbol | Key | Function |
|---|---|---|
| **** | Scroll | Press to scroll through various Process Information Screens in |
Normal Operation Mode.
| Alarm Acknowledge| Press to acknowledge and mute (if active) alarm
output.
| DOWN| Press to decrease the parameter value. Pressing once decreases
the value by one count; keeping pressed speeds up the change.
| UP| Press to increase the parameter value. Pressing once increases the
value by one count; keeping pressed speeds up the change.
| SET-UP| Press to enter or exit set-up mode.
****| ENTER| Press to store the set parameter value and to scroll to the
next parameter.
PV ERROR INDICATIONS
| Message | Error Type | Cause |
|---|
|
Sensor Open| Sensor (RTD Pt100) Broken / Open
| Over-range| Temperature above Max. Specified Range
| Under-range| Temperature below Min. Specified Range
ELECTRICAL CONNECTIONS