MEAN WELL RPD-75 Series 75W Dual Output Medical Grade Instruction Manual

June 4, 2024
MEAN WELL

MEAN WELL RPD-75 Series 75W Dual Output Medical Grade

Features

  • 5×3 compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration.
  • Free air convection for rated power and 23.5CFM forced air convection for peak load.
  • EMI class B for class Ⅰ configuration.
  • Extremely low leakage current.
  • Protections: Short circuit / Overload / Over voltage
  • Lifetime > 140K hours
  • 3 years warranty.

Applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

GTIN CODE

MW Search: https://www.meanwell.comiserice GTIN asp

Description

RPD-75 is a 75W highly reliable green PCB type medical power supply with a high power density on the 5 by 3 footprint. It accepts 90~264VAC input and offers dual output voltages. RPD-75 is able to be used for Class Ⅰ with FG system design. The extremely low leakage current is less than 150 μA. In addition, it conforms to international medical regulations (2 MOPP) and EMC BS EN/EN55011.

Model Encoding

75 Reliable Dual Output Medical Grade

MODEL RPD-75A RPD-75B






OUTPUT

| OUTPUT NUMBER| CH1| CH2| CH1| CH2
DC VOLTAGE| 5V| 12V| 5V| 24V
RATED CURRENT| 7A| 3A| 5A| 2A
CURRENT RANGE| 1 ~ 9.5A| 0.3 ~ 4A| 1 ~ 6.8A| 0.2 ~ 2.7A
RATED POWER| 71W| 73W
PEAK LOAD (23.5CFM)| 95.5W| 98.8W
RIPPLE & NOISE max **Note.2| 80mVp-p| 120mVp-p| 80mVp-p| 120mVp-p
VOLTAGE ADJ. RANGE| CH1: 4.75 ~ 5.5V| CH1: 4.75 ~ 5.5V
VOLTAGE TOLERANCE Note.3| ±2.0%| ±6.0%| ±2.0%| ±6.0%
LINE REGULATION| ±0.5%| ±1.0%| ±0.5%| ±1.0%
LOAD REGULATION| ±1.5%| ±3.0%| ±1.5%| ±3.0%
SETUP, RISE TIME| 500ms, 30ms/230VAC          500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)**| 90ms/230VAC          20ms/115VAC at full load



INPUT

| VOLTAGE RANGE| 90 ~ 264VAC        127 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 77%| 79%
AC CURRENT (Typ.)| 1.5A/115VAC          1A/230VAC
INRUSH CURRENT (Typ.)| COLD START 25A/115VAC         50A/230VAC
LEAKAGE CURRENT Note.4| Earth leakage current < 150μA/264VAC , Touch current < 100μA/264VAC


PROTECTION

| OVERLOAD| 140 ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE| Ch1: 5.7 ~ 6.8V
Protection type : Shut down o/p voltage, re-power on to recover



ENVIRONMENT

| WORKING TEMP.| -20 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.5| 3000 meters









SAFETY & EMC

(Note 8)

| SAFETY STANDARDS| IEC60601-1, EAC TP TC 004, UL ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved,

TUV BS EN/EN60601-1 approved

ISOLATION LEVEL| Primary-Secondary:2xMOPP,  Primary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC  I/P-FG:2KVAC  O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH


EMC EMISSION

| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–





EMC IMMUNITY

| BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m 80MHz~2.7GHz

Table 9, 9~28V/m 385MHz~5.78GHz

EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG ; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,

100% interruptions 250 periods


OTHERS

| MTBF| 2575.6K hrs min.

Telcordia SR-332 (Bellcore) ; 569.9K hrs min.      MIL-HDBK-217F (25℃)

DIMENSION L WH| 12776.231mm or 5 31.22 nch
PACKING| 0.25Kg; 63pcs/17.3Kg/1.28CUFT


**NOTE**

| All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. Ripple & noise are measured at 20MHz of bandwidth by using a 12 twisted pair-wire terminated with a 0.1af & 47uf parallel capacitor. Tolerance: includes set up tolerance, line regulation and load regulation. Touch Current was measured from primary input to DC output. The ambient temperature derating of 3.5 C/1000m with fan less models and of 5 C/1000m with fan models for operating altitude higher than 2000m(6500ft) Length of set up time is measured at cold first start. Turning ONOFF the power supply may lead to increase of the set up time.Heat Sik HS1,HS2,HS3 can not be shorted. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies as avalilable on http://www.meanwell.com X Product Liability Disclaimer: For detailed information. please refer to https://www.meanwell.com/service Disclaimer.asp

Block Diagram

MEAN WELL RPD-75 Series 75W Dual Output Medical Grade
2

Derating Curve

MEAN WELL RPD-75 Series 75W Dual Output Medical Grade
3

Output Derating VS Input Voltage

Mechanical Specification

Front and Side View

MEAN WELL RPD-75 Series 75W Dual Output Medical Grade
5AC Input Connector CN1 : JST B3P-VH or equivalent

Pin No. Assignment Mating Housing Terminal
1 AC/N JST VHR

or equivalent

| JST SVH-21T-P1.1

or equivalent

2| No Pin
3| AC/L

DC Output Connector (CN2) : JST B8P-VH or equivalent
Pin No. Assignment Mating Housing Terminal
1,2 V1

JST VHR

or equivalent

|

JST  SVH-21T-P1.1

or equivalent

3,4,5| COM
6,7| V2
8| NC

Grounding Required

HS1,HS2,HS3 cannot be shorted. M1 is safety ground. For better EMC performance. Please secure an electrical connection between M1,M2 and chassis grounding.
M 1 M2 Installation Manual Please refer to : http://www.meanwell.com/manual.html