TRANSTEK GBF-2008-B3 Body Fat Analyzer User Manual
- June 6, 2024
- TRANSTEK
Table of Contents
TRANSTEK GBF-2008-B3 Body Fat Analyzer
Safety Information
Safety and Usage Information
The warning signs and symbols are essential to ensure your correct and safe use of this product and protect you and others from injury. Please kindly find the meanings of the warning signs and symbols, which you may encounter in the label and user manual, as follows:
Symbol for “THE OPERATION GUIDE MUST BE READ”|
Symbol for “MANUFACTURER”
---|---
Symbol for “RECYCLE”| Caution: These notes must be abserved to prevent any
damage to the device.
Symbol for “TYPE BF APPLIED PARTS”| Symbol for “COMPLIES WITH EU REQUIREMENTS”
Symbol for “SERIAL NUMBER”| Symbol for “ENVIRONMENT PROTECTION – Waste
electrical products should not be disposed of with household waste. Please
recycle where facilities exist. Check with your local authority or retailer
for recycling advice”
Symbol for “DIRECT CURRENT”
Symbol for “MANUFACTURE DATE”
CAUTION
The Body Fat Analyzer GBF-2008-B3 offers you a seamless way to manage your health. Please be aware that this device is designed for adults’ self- measuring and self-monitoring body fat level. Any information provided by this device is in no way meant to treat, cure or prevent any disease or illness from happening. This device should not be used by anyone who is acutely or chronically ill, suffering from a disease or taking medications that affect your water levels.
The accuracy of readings for these patients has not been verified. Specific medical advice should be obtained from a physician. Before diet and exercise, you must first find a professional medical guide. The body fat analyzer is equipped with data transmission function. It may emit electromagnetic energy so as to perform its intended function. Nearby portable and mobile RF communications equipment can affect the performance of the device . Portable and mobile RF communications equipment can affect the measuring accuracy of the body fat analyzer. Kindly note that the use of accessories, transducers or cables other than those specified, with the exception of transducers and cables sold by the manufacturer as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the device. Be aware that misuse of electrical equipments can cause electric shock, burns, fire and other hazards. Please do not use the device in flammable gas environment. Warning that the body fat analyzer should not be used adjacent to or stacked with other equipment.
Manufacturer will make available on request circuit diagrams, component parts
list,etc. WARNING:No modifications of this equipment is allowed.
During using the analyzer, the patient will contact with the upper surface and
the electrode of the Analyzer. The upper surface of the analyzer is type BF
Applied part.The device doesn’t need to be calibrated in five years lifetime.
Do not touch the output of batteries when using the analyzer.The patient is an
intended operator. The patient can perform all the operations in the manual,
such as measurement, data transmitting, changing batteries.
INDICATIONS FOR USE
The Transtek Body Fat Analyzer measures weight and uses bio-electrical
impedance analysis(BIA) technology to estimate body fat in generally healthy
children 10-17 years old and healthy adults. It is intended for use in the
domestic setting only.
CONTRAINDICATIONS
- This device is contraindicated for any female subject who may be suspected of, or is pregnant. Besides provided inaccurate readings, the affects of this device on the fetus are unknown.
- This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument such as a pacemaker or defibrillator.
Your Scale and Its Environment
To ensure your safety as well as the service life of your scale, please avoid
using the scale under the following circumstances:
- Slippery floor such as tile floor
- Jumping onto the platform immediately after bath or with wet feet
- Near a microwave oven
Avoid storage in the following locations:
- Where there is water
- Where the device may be exposed to extreme temperatures, humidity, moisture, direct sunlight, dust, or salt air
- Where there is risk of shock or drop
- Where you store chemicals or full of corrosive gases
- Where in reach of the infants or children, otherwise there is a risk of swallowing batteries.
Efficient Use of Your Scale
To ensure the accuracy of measurement, please follow below instructions when you start measurement.
- Place the scale on a flat, hard surface. Soft surface such as carpet will affect the performance of the scale.
- Step onto the platform with bare feet. Stand still and keep full contact with the scale until the measurement is complete.
- Start measurement at least two hours after Getting up or Dinning.
- Avoid measurement immediately after strenuous exercise, sauna or bath, drinking, and dining.
- Always start measurement in the same time slot and on the same scale located on the same flat, hard surface.
- This device is not intended for use in an oxygen rich environment.
Overview
Device Components
List
- Body Fat Analyzer GBF-2008-B3
- Four AAA-size Batteries (1.5V per each)
- User Manual
LED Display
General Instructions
Transtek Body Fat Analyzer GBF-2008-B3 applies BIA (Bio-impedance Analysis) technology. A small amount of weak current flows through the human body so as to detect the bio-impedance and estimate body fat. The electrical current is small and may not be felt. This BIA technology is cheap, safe, non-invasive, toxic-free and harmless. It also possesses the characteristics of simple operation and abundant information. The current mentioned above is less than 0.5mA. However, please be aware that anyone with an wearable or implantable medical electronic instrument, such as a pacemaker, must avoid using this device.The intended use of this device is for healthy children 10-17 years old and healthy adults.
Insert the Batteries
- Open the battery door in the back of the scale.
- Insert the batteries (4 x 1.5V AAA) into the battery compartment according to the polarity indications marked inside the compartment.
- All digits will be shown on the LED.
- Close the battery door and wait until the digits “0.00kg” are shown on the LED.
CAUTION
- When the symbol “Lo” first appears, the device can only weigh 10 times and then power off. Then you shall replace with a new set of batteries. Please replace all four batteries at the same time. Do NOT mix the old batteries with the new one.
- Worn batteries are hazardous waste. Do NOT dispose of them together with the household garbage.
- Please refer to the local ordinances and recycling instructions regarding disposal of the worn batteries and scrapped device.
- If you do not intend to use this unit for a prolonged period of time, it is advisable to remove the batteries before storing.
- Do not mix alkaline, standard(carbon-zinc), or rechargeable(Ni-Cd, Ni-MH, etc) batteries.
- If you do not intend to use this unit for a prolonged period of time, it is advisable to remove the batteries before storing.
- Use the scale three times a day, and the lifetime of the battery is around 6 months.
Install APP and Pair-Up
With the advanced Bluetooth 5.0 technology applied, the mobile or portable
equipment, which are equipped with Bluetooth function in line with BLE
Technical Specifications as well as BLP Protocol established by global
organization Bluetooth SIG, are capable to receive your personal health data.
Just simply install the Tuya Smart App and pair up your scale with your mobile
or portable equipment. Then you may enjoy the comprehensive health solution
provided by Tuya.
Note: It is only available for the device which supports Bluetooth 5.0
technology. For the first time, the scale will automatically enter bluetooth
pairing mode.
Turn on bluetooth and the Tuya Smart App. Make sure both are ON when the pair-
up is proceeding.
If SUCCEED, the symbol “ ” will be shown on the LED.
If FAIL, the symbol “ ” will not be shown on the LED.
Select Measurement Unit
With batteries correctly installed, press “UNIT” button in the back of the
scale to select measurement unit. The default measurement unit is “kg”. You
may press “UNIT” button to choose between kilogram and pound.
Daily Measurement
With original SENSE ON patent technology, GBF-2008-B3 will switch on as you
step on the platform barefooted. Stand still and keep full contact with the
electrodes until the LED stops blinking digits.
Data Transmission
- With GBF-2008-B3 successfully pair-up with your smartphone, the measurement data will be transmitted to your mobile via Bluetooth.
- ONLY when the measuring results are attached to a specific User ID will it be transmitted to your mobile after measurement.
- The symbol bluetooth will display after successful data transmission, and you may check your personal health data stored in your smartphone.
- If the data transmission fails, the symbol doesn’t display. The pending measurement data will be temporarily kept in the scale and transmitted to your smartphone when next measurement is complete.
CAUTION
- Interference may occur in the vicinity of equipment marked with the following symbol the Analyzer may interfere vicinity electrical equipment.
- To enable the data transmission function, this product should be paired to a Bluetooth 5.0 or above end.
How to mitigate possible interference?
- The range between the Analyzer and the Bluetooth end should be reasonably close, from 1 meter to 10 meters. Please ensure no obstacles between the Analyzer and the Bluetooth end so as to obtain quality connection.
- To avoid interference, other electronic devices (particularly those with Bluetooth transmission / Transmitter) should be kept at least 1 meter away from the Analyzer.
Troubleshooting
Error | Description | Solution |
---|---|---|
**** | **** |
Overload.
| ****
Stop using this scale for measurement.
|
Low Battery.
| Replace all four batteries in the same time. Please purchase the authorized batteries for replacement.
when Measuring
Problem | Root Cause | Solution |
---|
Abnormal measuring results:
– Too high;
– Too low;
– Huge difference between two recent measurement.
| ****
Incorrect posture
| ****
Please step on the platform barefooted and stand still.
The device is located on the soft ground such as a carpet or on a rugged surface.| ****
Please place the device on a flat, hard surface.
Problem | Root Cause | Solution |
---|
Abnormal measuring results:
– Too high;
– Too low;
– Huge difference between two recent measurement.
|
Your feet are too dry.
|
Wipe your feet with a damp cloth, keeping them slightly damp when starting measurement.
No display on LED when the device powers on.
|
Batteries not yet installed.
|
Install the batteries. (Please refer to Insert the
Batteries )
Worn batteries.
| Replace all four batteries at the same time. Please purchase the authorized batteries for replacement.
CANNOT proceed to analyze body fat.
|
Step onto the platform wearing socks or shoes.
| Please keep barefooted during the measurement, and keep full contact with the electrodes as well.
The system cannot identify the possible User ID with most similar data.
|
Please assign a User ID again on the APP.
The device powers off.
|
Low battery.
| Replace all four batteries at the same time. Please purchase the authorized batteries for replacement.
Error Prompt
Error | Description | Solution |
---|---|---|
**** | **** |
Overload.
| ****
Stop using this scale for measurement.
|
Low Battery.
| Replace all four batteries in the same time. Please purchase the authorized batteries for replacement.
When Measuring
Problem | Root Cause | Solution |
---|
Abnormal measuring results:
– Too high;
– Too low;
– Huge difference between two recent measurement.
| ****
Incorrect posture
| ****
Please step on the platform barefooted and stand still.
The device is located on the soft ground such as a carpet or on a rugged surface.| ****
Please place the device on a flat, hard surface.
Troubleshooting
Problem | Root Cause | Solution |
---|
Abnormal measuring results:
– Too high;
– Too low;
– Huge difference between two recent measurement.
|
Your feet are too dry.
|
Wipe your feet with a damp cloth, keeping them slightly damp when starting measurement.
No display on LED when the device powers on.
|
Batteries not yet installed.
|
Install the batteries. (Please refer to Insert the
Batteries )
Worn batteries.
| Replace all four batteries at the same time. Please purchase the authorized batteries for replacement.
CANNOT proceed to analyze body fat.
|
Step onto the platform wearing socks or shoes.
| Please keep barefooted during the measurement, and keep full contact with the electrodes as well.
The system cannot identify the possible User ID with most similar data.
|
Please assign a User ID again on the APP.
The device powers off.
|
Low battery.
| Replace all four batteries at the same time. Please purchase the authorized batteries for replacement.
When Data Transmitting
Problem | Root Cause | Solution |
---|
Data transmission failed.
| ****
Bluetooth is OFF.
| ****
Turn ON the Bluetooth via Settings.
The APP is OFF.
| ****
Press the icon to turn ON your APP.
Out of range of Bluetooth transmission.
| ****
Place your smartphone closer to the scale.
None of the user ID is assigned.
| ****
Please assign a User ID on the APP.
Specifications
Product Name | Body Fat Analyzer (GBF-2008-B3) |
---|---|
Dimension | Scale: 300x300x27.6mm (Approximately) |
Net Weight | Approximately 1.678kg ( Including the dry cells) |
Display | Digital LED |
Measurement Unit | Pound / Kilogram |
Measurement Range | 11lb to 397lb / 5kg to 180kg |
Division | 5-100kg:0.05kg / 0.2lb;100-180kg:0.1kg / 0.2lb |
Accuracy
|
5-50: ±0.2kg; 50.05-100: ±0.3kg;
100.1-150: ±0.4kg; 150.1-180: ±0.5kg;
Working Environment
| Temperature: 5℃ to 40℃ Relative Humidity: ≤90% RH Atmospheric pressure: 86kPa to 106kPa
Storage Environment
| Temperature: -20℃ to 60℃ Relative Humidity: ≤90% RH Atmospheric pressure:
50kPa to 106kPa
Power Source| 6V (Four AAA-size Batteries)
Auto-ON| SENSE ON technology
Auto-OFF| The scale will turn off after about 30s if there is no
operation.
Accessories| 1. Four AAA-size Batteries
2. User Manual
Mode of Operation| Continuous Operation
Protection Against Ingress of Water
|
IP21(The first number 2: Protected against solid foreign objects of 12.5 mm Ф and greater. The second number 1: Protection against vertically falling water drops.)
Software Version| V1.4
Degree of protection| Type BF applied part
About the Accuracy of This Product
This product passes strict inspection before delivery and therefore its
accuracy is guaranteed by the manufacturer. Please refer to the above table
for the descriptions on accuracy.
This product is specially designed for body fat analysis as well as weight
measurement. It should NOT be used by anyone during the process of transaction
for verification of goods’ weight.
Maintenance
When carrying out usual maintenance, please ensure practice of the following Do’s and Don’ts:
- DO use a dry soft cloth to wipe the dust.
- DO use a wet soft cloth, dipped into water and wrung out, to wipe the dirt. Then use a dry soft cloth to dry up the device.
- DON’T wash the device with water or immerse it in water.
- DON’T use propellant, abrasive or other chemicals to wipe the dirt in avoidance of discolor or malfunction.
- DON’T disassemble this device. If you have any problems, please contact Transtek.
(Please refer to Warranty for contact information)
Warranty
Transtek warrants its products free of defects in materials and workmanship in
normal use for a period of FIVE years from the date of retail purchase.
This warranty does NOT cover damages caused by misuse or abuse, including but
not limited to:
- Failure caused by unauthorized repairs or modifications;
- Damage caused by shock or drop during transportation;
- Failure caused by improper operation inconsistent with the instructions stated in this user manual;
- Malfunction or damage from failure to provide the recommended maintenance; Damage caused by improper use of power supply.
Should this device require maintenance (or replacement at our option) under warranty, please deliver the original package to ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD prepaid. Please return the store receipt (with the retail purchase date) and a note with reasons to return on it as well.
ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD
No.23,Jin’an Road, Minzhong, Zhongshan, 528441, Guangdong, China Tel:
86-760-88282982
Website:http://www.transtek.cn
FCC Regulations
FCC User Guide Information
Radio Frequency Interface Requirements – FCC
Note:
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures:
- Reorient or relocate the receiving antenna;
- Increase the separation between the equipment and receiver;
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected;
- Consult the dealer or an experienced radio / TV technician for help.
Radio Transmitters (Part 15)
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
FCC RF Exposure Guidelines
Safety Information
Reducing RF Exposure – Use Properly
Only operate the device in accordance with the instructions supplied. This
device complies with FCC radiation exposure limits set forth for an
uncontrolled environment.
Warning
- Warning:Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
- Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.”
- Warning:Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment GBF-2008-B3, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Warning
- CAUTION! RISK OF EXPLOSION IF BATTERY IS REPLACED BY AN INCORRECT TYPE.
- DISPOSE OF USED BATTERIES ACCORDING TO THE INSTRUCTIONS.
- Disposal of a battery into fire or a hot oven, or mechanically crushing or cutting of a battery, that can result in an explosion.
- Leaving a battery in an extremely high temperature surrounding environment that can result in an explosion or the leakage of flammable liquid or gas;
- A battery subjected to extremely low air pressure that may result in an explosion or the leakage of flammable liquid or gas.
EMC Guidance
Guidance and manufacturer’s declaration -electromagnetic emissions and
Immunity
Table1
Guidance and manufacturer’s declaration – electromagnetic emissions
Emissions test| Compliance
RF emissions
CISPR 11
| Group 1
RF emissions
CISPR 11
| Class B
Harmonic emissions
IEC 61000-3-2
| Not applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
| Not applicable
Table2
Guidance and manufacturer’s declaration – electromagnetic Immunity
Immunity Test| IEC 60601-1-2
Test level
| Compliance level
Electrostatic discharge (ESD)
IEC 61000-4-2
| ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
| ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst
IEC 61000-4-4
| Not applicable| Not applicable
Surge
IEC 61000-4-5
| Not applicable| Not applicable
Voltage dips, short interruptions and voltage variations on power supply input
lines
IEC 61000-4-11
| Not applicable| Not applicable
Power frequency magnetic
field
| 30 A/m
50Hz/60Hz
| 30 A/m
50Hz/60Hz
IEC 61000-4-8| |
Conducted RF
IEC61000-4-6
| Not applicable| Not applicable
Radiated RF IEC61000-4-3| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE UT is the a.c. mians voltage prior to application of the test level.
Table3
Guidance and manufacturer’s declaration – electromagnetic Immunity
Radiated RF IEC61000-4-3
(Test specifications for ENCLOSURE PORT IMMUNITY to
RF wireless communications equipment)
| Test Frequency (MHz)| Band
(MHz)
| Service| Modulation| Modulation
(W)
| Distance
(m)
| IMMUNITY TEST LEVEL
(V/m)
385| 380 –
390
| TETRA 400| Pulse modulation 18 Hz| 1,8| 0.3| 27
450| 430 –
470
| GMRS 460,
FRS 460
| FM
± 5 kHz deviation 1 kHz sine
| 2| 0.3| 28
710| 704 –
787
| LTE Band 13,
17
| Pulse modulation 217 Hz| 0,2| 0.3| 9
745
780
810| 800 –
960
| GSM 800/900, TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
| Pulse modulation 18 Hz| 2| 0.3| 28
870
930
1720| 1 700 –
1 990
| GSM 1800; CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3,
4, 25;
UMTS
| Pulse modulation 217 Hz| 2| 0.3| 28
1845
1970
2450| 2 400 –
2 570
| Bluetooth, WLAN, 802.11
b/g/n, RFID 2450,
LTE Band 7
| Pulse modulation 217 Hz| 2| 0.3| 28
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>