TRANSTEK GBS-2104-G TeleWieght Gen-2 Cellular Body Scale User Manual
- June 6, 2024
- TRANSTEK
Table of Contents
Cellular Body Scale
User Manual
GBS-2104-G
TeleWieght Gen-2
- Thank you for choosing the Transtek Cellular Body Scale!
- Please read this manual thoroughly to know how to use your Cellular Body Scale safely and correctly. Keep the manual well for future reference.
Symbol Meanings
The symbols below may be in the user manual, labeling, or another component. They are the standard requirement of using.
| THE OPERATION GUIDE MUST BE READ| | MANUFACTURER
---|---|---|---
| SERIAL NUMBER| | MANUFACTURE DATE
| DIRECT CURRENT| | RECYCLABLE
| TYPE BF APPLIED PARTS| | TEMPERATURE LIMIT
| HUMIDITY LIMITATION| | ATMOSPHERIC PRESSURE LIMITATION
| Indicates the need for the user to the consul. the instructions for use for
important information such as warnings and precautions that cannot, for a
variety of reasons, be presented on the medical device itself.| | —
ENVIRONMENT PROTECTION-Waste electrical products should not be disposed of
with household waste. Please recycle where facilities exist. Check with your
loc authority or retailer for recycling advice
CAUTION
The Cellular Body Scale GBS-2104-G offers you a way to manage your health.
Please be aware that this device is designed for children (10-17 years old)
and adults’ self-measuring. Any information provided by this device is in no
way meant to treat, cure or prevent any disease or illness from happening. The
Cellular Body Scale is equipped with a data transmission function. It may emit
electromagnetic energy so as to perform its intended function. Nearby portable
and mobile RF communications equipment can affect the performance of the
device. Portable and mobile RF communications equipment can affect the
measuring accuracy of the Cellular Body Scale. Kindly note that the use of
accessories, transducers, or cables other than those specified, with the
exception of transducers and cables sold by the manufacturer as replacement
parts for internal components, may result in increased EMISSIONS or decreased
IMMUNITY of the device. Be aware that misuse of electrical equipment can cause
electric shock, bums, fire, and other hazards. Please do not use the device’s
inflammable gas environment. Warning that the Cellular Body Scale should not
be used adjacent to or stacked with other equipment. The glass surface of the
scale is the type BF Applied part. The device does not need to be calibrated
in five years’ lifetime. Do not touch the output of batteries when using the
scale. You can use this device to take your own measurement, no third-party
operator is required. Please use the device under the environment which is
provided in the user manual. Otherwise, the performance and lifetime of the
device will be impacted and reduced. Store your device in a clean and dry
place. Never place any heavy objects on it. Dispose of accessories, detachable
parts, and the device according to the local guidelines.
Safety Information
INDICATIONS FOR USE
- The Transtek Cellular Body Scale is intended for adults and children (10-17 years old) to measure weight at home.
Your Scale and Its Environment
To ensure your safety as well as the service life of your scale, DO NOT use
the scale:
-
on the slippery floor such as tile floors,
-
immediately after a bath or with wet feet,
-
near a cell phone or microwave oven.
Avoid storage in the following locations: -
Where there is water.
-
Where the device may be exposed to extreme temperatures, humidity, moisture, direct sunlight, dust, or salt air.
-
Where there is a risk of shock or drop.
-
Where you store chemicals or full of corrosive gases.
-
Where in reach of the infants or children.
Efficient Use of Your Scale
To ensure the accuracy of measurement, please follow the below instructions when you start measurement. -
Place the scale on a flat, hard surface. Soft surfaces such as carpets will affect the performance of the scale.
-
Start measurement at least two hours after getting up or dining.
-
Avoid measurement immediately after strenuous exercise, sauna or bath, drinking, and dining.
-
Always start measurement in the same time slot and on the same scale located on the same flat, hard surface.
-
This device is not intended for use in an oxygen-rich environment.
-
Any information provided by this scale is in no way meant to treat, cure or prevent any disease or illness from happening. If in doubt, contact your physician.
-
The device is designed for use in the home healthcare environment.
Overview
Device Components
List
- Body Scale GBS-2104-G
- Four M-size Batteries (1.5V per each)
- User Manual
Overview
LCD Display
| Kilogram| | Signal symbol
---|---|---|---
| Stone| | Data transmitting
| Pound| | Data pending to transmit to a wireless wellness system
| Battery symbol| | No SIM card information
Initial Start-Up
General Instructions
Transtek Cellular Body Scale GBS-2104-G applies SENSE ON technology. It’s not
necessary to turn on or turn off the scale, just stand on the scale for
measurement. It is delicate and trendy, convenient for use.
Insert Batteries
- Open the battery door in the back of the scale.
- Insert the batteries (4 x 1.5V AA) according to the polarity indications marked inside the battery compartment.
- all digits will be shown on the LCD.
CAUTION
- When symbol ‘0,” blinks, the device will power off. Then you shall replace it with a new set of batteries. Please replace all four batteries at the same time. Do NOT mix the old batteries with the new ones.
- Worn batteries are hazardous waste. Do NOT dispose of them together with the household garbage.
- Please refer to the local ordinances and recycling instructions regarding the disposal of worn batteries and scrapped devices.
- Do not mix alkaline, standard (carbon-zinc), or rechargeable (Ni-Cd, Ni-MH, etc) batteries.
- If you do not intend to use this unit for a prolonged period of time, it is advisable to remove the batteries before storing them.
- Use the scale three times a day, and the lifetime of the battery is one year.
Select Measurement Unit
The unit switch buttton “ is on the back of the scale. The default measurement
unit is ” “. You may press the “0” button to choose between “pound”,
“kilogram” and “stone”. Laid down the scale on the ground and it will turn off
in 10s.
How to check and
re-install the SIM card
Flip your scale over to find the SIM card compartment door inside the battery
compartment with the wording “SIM”. Open the SIM card door with a screwdriver
and then you can check and re-install the SIM card. DO NOT enable it when it
is not necessary! DO NOT replace it with another SIM card!
Daily Measurement
-
With original SENSE ON patent technology, the scale will automatically switch on as you step on the platform barefooted.
-
Stand still and keep full contact with the scale, the weight data on the LCD will blink twice, and the weight is locked. Then the scale is automatically connected to the internet.
-
If the scale automatically connected fails, “ ” will be displayed.
-
After connecting to the internet, the weight data will be transmitted. When data is successfully transmitted, “OK” will be displayed and the digit” ” appears.
-
If the data transfer fails, “ will be displayed.
Troubleshooting
Error prompt
Symbol & Code | Description | How to solve |
---|---|---|
Overload. | Retake measurement. |
If the issue persists, check your weight is out of measurement range (11 lb to
440 lb / 5 kg to 200 kg / 0 st :11 lb to 31 st : 1 lb)
| SIM error.| Check and reinstall the SIM card.
If the issue persists, contact customer support.
| No signal.| 1. Move the scale to a strong signal place.
2. Contact customer support.
| Hardware error (such as the communication between the MCU and the module is
abnormal.)| Retake measurement.
If the issue persists, contact customer support.
| Data transmission error| Move to another area, preferably closer to a
window.
Use the device at a location where you get a strong cellular signal with your
mobile phone.
| No cellular coverage or Cellular connection error| Move to another area,
preferably closer to a window.
Use the device at a location where you get a strong cellular signal with your
mobile phone.
| The pre-loaded SIM service SIM card is replaced or the
subscription suspended/terminated.| Make sure the SIM card in the device is
the original one pre-loaded. DO NOT replace it by yourself.
If the issue persists, contact customer support.
| Server connection error| Move to another area, preferably closer to a
window, try again.
Use the device at a location where you get strong cellular sival with your
mobile phone. If the issue persists, contact customer support.
| Certificate error| Contact customer support.
| Hardware error| Retake measurement.
If the issue persists, contact customer support.
When Measuring
Problem | Root Cause | Solution |
---|---|---|
No display on LCD when the device powers on | Batteries not yet installed |
Install the batteries.
(Please refer to Insert the Battery)
Worn batteries| Replace all four batteries at the same time. Please purchase
the authorized batteries for replacement
The device powers off automatically| Worn batteries| Replace all four
batteries at the same time. Please purchase the authorized batteries for
replacement
About the Accuracy of This Product
- This product passes strict inspection before delivery and therefore its accuracy is guaranteed by the manufacturer. Please refer to the above table for the descriptions on accuracy.
- This product is specially designed for Cellular Body scales as well as weight measurement. It should NOT be used by anyone during the process of transaction for verification of goods’ weight.
Specifications
Product Name | Cellular Body Scale |
---|---|
Dimension | Approx 320x320x28.5 mm |
Net Weight | Approximately 2.6 kg (Excluding the dry cells) |
Display | Digital LCD |
Measurement Unit | Pound / Kilogram / Stone |
Measurement Range | 11 lb to 440 lb / 5 kg to 200 kg / 0 st :11 lb to 31 st :1 |
lb
Division| 0.1 kg/ 0.2 lb
Accuracy| 50 ± 0.2 kg; 100 ± 0.3 kg; 150 ± 0.4 kg; 200 ± 0.5 kg;
Working Environment| Temperature: 5°C to 40°C Relative Humidity: 690% RH
Atmospheric pressure: 86 kPa to 106 kPa
Storage Environment| Temperature: -20°C to 60°C Relative Humidity: 690% RH
Atmospheric pressure: 50 kPa to 106 kPa
Power Source| 6V (Four M-size Batteries)
Auto-ON| SENSE ON technology
Auto-OFF| About 10 seconds while showing 0.0;
Accessories| 1. Four AA-size Batteries
2. User Manual
Mode of Operation| Continuous Operation
Protection Against Ingress of Water| 1P21(The first number 2: Protected
against solid foreign objects of 12.5 mm 0 and greater. The second number
1:Protection against vertically falling water drops.)
Degree of protection| Type BF applied part
Maintenance
When carrying out usual maintenance, please ensure the practice of the following Dos and Don’ts:
- Clean the device after use
- DO use a dry soft cloth to wipe the dust.
- DO use a wet soft cloth, dipped into water and wrung out, to wipe the dirt. Then use a dry soft cloth to dry up the device.
- DON’T wash the device with water or immerse it in water.
- DON’T use a propellant, abrasive or other chemicals to wipe the dirt in avoidance of discolor or malfunction.
- Do not attempt to repair the device yourself if it malfunctions. Only have repairs carried out by authorized service centers.
- Do not attempt to modify the device or replace components (e.g., transducers, LCD, etc.) by yourself. If you suspect incorrect operation, contact customer support for further assistance. At the request of authorized service personnel, circuit diagrams, component part lists, descriptions, and calibration procedures will be made available by the manufacturer or distributor.
- Do Not disassemble this device. If you have a problem, contact customer support (Please refer to the section of “Warranty” for customer support information)
- No service and maintenance while in use.
Warranty
-
Transtek warrants its products free of defects in materials and workmanship in normal use for a period of FIVE years from the date of retail purchase.
-
This warranty does NOT cover damages caused by misuse or abuse, including but not limited to:
Failure caused by unauthorized repairs or modifications;
Damage caused by shock or drop during transportation;
Failure caused by improper operation inconsistent with the instructions stated in this user manual;
Malfunction or damage from failure to provide the recommended maintenance;
Damage caused by improper use of power supply. -
Should this device require maintenance (or replacement at our option) under warranty, please deliver the original package to ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD prepaid. Please return the store receipt (with the retail purchase date) and a note with reasons to return it as well.
ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD
No.23, Jin’an Road, Minzhong, Zhongshan, 528441, Guangdong, China Website: www.transtekcorp.com
FCC Regulations
FCC User Guide Information
Radio Frequency Interface Requirements – FCC
Note: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to Part 15 of the FCC rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses,
and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna;
- Increase the separation between the equipment and receiver;
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected;
- Consult the dealer or an experienced radio / TV technician for help.
Radio Transmitters (Part 15)
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
FCC SAR statement
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. End-user must follow the specific operating instructions for satisfying RF exposure compliance. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. The portable device is designed to meet the requirements for exposure to radio waves established by the Federal Communications Commission (USA). These requirements set a SAR limit of 4.0 W/kg averaged over one gram of tissue. The highest SAR value reported under this standard during product certification for use when properly worn Activates on the extremity(hands/feet/ankle/wrist).
Appendix
Warning
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments.
- Warning: Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
- Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Warning: Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.”
- Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment GBS-2104-G, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
EMC Guidance
Guidance and manufacturer’s declaration -electromagnetic emissions and
Immunity
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
Emissions test| Compliance
RF emissions CISPR 11| Group 1
RF emissions CISPR 11| Class B
Harmonic emissions IEC 61000-3-2| Not applicable
Voltage fluctuations/flicker emissions IEC 61000-3-3| Not applicable
Table 2
Guidance and manufacturer’s declaration – electromagnetic
Immunity Test| IEC 60601-1-2 Test level| Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2| ±-8 kV contact
±-2 kV, ±4 kV, -±8 kV, -±15 kV air| ±8 kV contact
±2 kV, -A kV, -±8 kV, ±15 kV air
Electrical fast transient/burst IEC 61000-4-4| Not applicable| Not applicable
Surge IEC 61000-4-5| Not applicable| Not applicable
Voltage dips, short interruptions, and voltage variations on power supply
input lines
IEC 61000-4-11| Not applicable| Not applicable
Power frequency magnetic field| 30 A/m 50Hz/60Hz| 30 A/m 50Hz/60Hz
IEC 61000-4-8| |
Conducted RF IEC 61000-4-6| Not applicable| Not applicable
Radiated RF IEC 61000-4-3| 10 V/m 80 MHz – 2,7 GHz 80%AM at 1 kHz| 10 V/m 80
MHz – 2,7 GHz 80%AM at 1 kHz
NOTE UT is the ac. mains voltage prior to application of the test level
Table 3
ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD
No.23, Jin’an Road, Minzhong, Zhongshan, 528441,
Guangdong, China www.transtekcorp.com