TRANSTEK GBF-1257-B Body Fat Analyzer User Manual

June 4, 2024
TRANSTEK

version:A0

User Manual

Body Fat Analyzer 0382
  • Thank you very much for selecting the Greater Goods Body Fat Analyzer 0382.
  • Please do read the user manual carefully and thoroughly so as to ensure the safe usage of this product, and keep the manual well for further reference in case you have problems.

Safety Information

Safety and Usage Information

The warning signs and symbols are essential to ensure your correct and safe use of this product and protect you and others from injury. Please kindly find the meanings of the warning signs and symbols, which you may encounter in the label and user manual, as follows:

| Symbol for “THE OPERATION GUIDE MUST BE READ”| | Symbol for “MANUFACTURER”
---|---|---|---
| Symbol for “SERIAL NUMBER”, includes FCC ID number.| | Symbol for “DIRECT CURRENT”
| Symbol for “TYPE BF APPLIED PARTS”| | Caution: These notes must be abserved to prevent any damage to the device.
| Symbol for “Non-ionizing electromagnetic radiation”| | Symbol for “ENVIRONMENT PROTECTION – Waste electrical products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local authority or retailer for recycling advice”
| Symbol for “MANUFACTURE DATE”

CAUTION

The Body Fat Analyzer 0382 offers you a seamless way to manage your health.
The device is Not to diagnose or treat any medical condition, any information provided by this device is in no way meant to treat, cure or prevent any disease or illness from happening. This device should not be used by anyone who is acutely or chronically ill, suffering from a disease or taking medications that affect your water levels. The accuracy of readings for these patients has not been verified. Specific medical advice should be obtained from a physician.
Before diet and exercise, you must first find a professional medical guide.

The body fat analyzer is equipped with data transmission function. It may emit electromagnetic energy so as to perform its intended function. Nearby portable and mobile RF communications equipment can affect the performance of the device.

Portable and mobile RF communications equipment can affect the measuring accuracy of the body fat analyzer.

Kindly note that the use of accessories, transducers or cables other than those specified, with the exception of transducers and cables sold by the manufacturer as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the device.

Be aware that misuse of electrical equipments can cause electric shock, burns, fire and other hazards. Please do not use the device in flammable gas environment.
Warning that the body fat analyzer should not be used adjacent to or stacked with other equipment.
Manufacturer will make available on request circuit diagrams, component parts list,etc.
WARNING: No modifications of this equipment is allowed.
During using the analyzer, the patient will contact with the surface and the electrode of the Analyzer. The surface and the electrode of the analyzer is type BF Applied part.
The device doesn’t need to be calibrated in five years lifetime.
Do not touch the output of batteries when using the analyzer.
The patient is an intended operator. The patient can perform all the operations in the manual, such as measurement, data transmitting, changing batteries.
Not intended for use with flammable anaesthetics, not intended for use in conjunction with flammable agents, not used in conjunction with oxygen rich environments.
Stop using the analyzer if any discomfort or allergic reaction occurs.
Degeneration or looseness of sensors and electrodes might reduce accuracy.

INDICATIONS FOR USE
  • The Body Fat Analyzer measures weight and uses bioelectrical impedance analysis(BIA) technology to estimate body fat percentage, body water percentage, bone percentage and muscle percentage in 10-17 years old children and less than 85 years old adults.
  • It is intended for use in the domestic setting only.
  • It is not to diagnose or treat any medical condition.
CONTRAINDICATIONS
  1. This device is contraindicated for any female subject who may be suspected of, or is pregnant. Besides provided inaccurate readings, the affects of this device on the fetus are unknown.
  2. This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument such as a pacemaker or defibrillator.
Your Scale and Its Environment

To ensure your safety as well as the service life of your scale, please avoid using the scale under the following circumstances:

  • Slippery floor such as tile floor
  • Jumping onto the platform immediately after bath or with wet feet
  • Near a cell phone or microwave oven

Avoid storage in the following locations:

  • Where there is water
  • Where the device may be exposed to extreme temperatures, humidity, moisture, direct sunlight, dust, or salt air
  • Where there is risk of shock or drop
  • Where you store chemicals or full of corrosive gases
  • Where in reach of the infants or children
Efficient Use of Your Scale

To ensure the accuracy of measurement, please follow below instructions when you start measurement.

  • Place the scale on a flat, hard surface. Soft surface such as carpet will affect the performance of the scale.
  • Step onto the platform with bare feet. Stand still and keep full contact with the scale until the measurement is complete.
  • Start measurement at least two hours after Getting up or Dinning.
  • Avoid measurement immediately after strenuous exercise, sauna or bath, drinking, and dining.
  • Always start measurement in the same time slot and on the same scale located on the same flat, hard surface.

Overview

Device Components
  1. Electrode
  2. Up Key
  3. LCD Display
  4. Down Key
  5. Set Key
  6. Electrode

A UNIT and activation Button
B Battery Compartment

List
  1. Body Fat Analyzer 0382
  2. Four AAA-size Batteries (1.5V per each)
  3. User Manual
LCD Display

TRANSTEK GBF-1257-B Body Fat Analyzer A04

| Body Fat Analysis Result| | Male
---|---|---|---
| Body Water Analysis Result| | Male Athlete
| Muscle Mass Analysis Result| | Female
| Bone Mass Analysis Result| | Female Athlete
| Your Goal Weight| | Kilogram
| Body Mass Index| | Pound
| Age| | Setting Mpde
| Inch| | Percentage
| Centimeter| | User ID (Range form U1 to U8)
| Data transmitting / pending to transmit to wireless wellness system| | Successful Bluetooth Connection
| Low Battery| |

Initial Start-Up

General Instructions

Greater Goods Body Fat Analyzer 0382 applies BIA (Bio-impedance Analysis) technology. A small amount of weak current flows through the human body so as to detect the bio-impedance and estimate BMI, body fat, body water, muscle mass and bone mass. The electrical current is small and may not be felt.
This BIA technology is cheap, safe, non-invasive, toxic-free and harmless. It also possesses the characteristics of simple operation and abundant information.
The current mentioned above is less than 0.5mA. However, please be aware that anyone with an wearable or implantable medical electronic instrument, such as a pacemaker, must avoid using this device.
The intended use of this device is for healthy children 10-17 years old and less than 85 years old healthy adults.
At least 2 hours required for the device to warm / cool from the minimum / maximum storage temperaure between uses until it is ready for intended use when the ambient temperature is 20℃.

Insert the Batteries
  • Open the battery door in the back of the scale.
  • Insert the batteries (4 x 1.5V AAA) into the battery compartment according to the polarity indications marked inside the compartment.
  • The digits “ 8888 ” will be shown on the LCD

TRANSTEK GBF-1257-B Body Fat Analyzer A05

  • Close the battery door and wait until the digits “ ” are shown on the LCD.

TRANSTEK GBF-1257-B Body Fat Analyzer A06

CAUTION

  • When the symbol “ ” appear, the device will power off. Then you shall replace with a new set of batteries. Please replace all four batteries at the same time. Do NOT mix the old batteries with the new one.
  • Worn batteries are hazardous waste. Do NOT dispose of them together with the household garbage.
  • Please refer to the local ordinances and recycling instructions regarding disposal of the worn batteries and scrapped device.
  • If you do not intend to use this unit for a prolonged period of time, it is advisable to remove the batteries before storing.
  • Do not mix alkaline, standard(carbon-zinc), or rechargeable(Ni-Cd, Ni-MH, etc) batteries.
  • Use the scale three times a day, and the lifetime of battery is around 6 months.
Install APP and Pair-Up

With the advanced Bluetooth 4.0 technology applied, the mobile or portable equipments, which are equipped with Bluetooth function in line with BLE Technical Specifications as well as BLP Protocol established by global organization Bluetooth SIG, are capable to receive your personal health data.
Just simply install the specially-designed app and pair up your scale with your mobile or portable equipments. Then you may enjoy the comprehensive health solution provided by Greater Goods.
Note: It is only available for the device which supports Bluetooth 4.0 technology.

  • Turn on Bluetooth and the APP. Make sure both are ON when pair-up is proceeding.
  • Press and hold “UNIT” button in the back of the scale to start pair-up.

Symbol and symbol will be shown on the LCD alternatively, indicating pair-up is proceeding.

TRANSTEK GBF-1257-B Body Fat Analyzer D03

If SUCCEED, symbol “ [ ]” will be shown on the LCD.

TRANSTEK GBF-1257-B Body Fat Analyzer D04

If FAIL, symbol “E1” will be shown on the LCD.

Set Up Your Profile

The body fat analyzer 0382 supports multiple users (Up to 8). You may follow below instructions to assign User ID and set up your own profile, including Gender, Stature, and Age.

1. Assigning User ID

  • With batteries correctly installed, press “SET” key to enter setting.
  • The system will request User ID selection first. As pictured below. The operator may press the function key  or to select User ID among U1 to U8.
  • Press “SET” key to confirm User ID.

TRANSTEK GBF-1257-B Body Fat Analyzer D06

2. Delete Data

  • After confirming User ID, the system will divert to select whether to delete the previous data.
  • As pictured below, the portrait “n” blinks. The operator may press the function key or to select “n” or “y”.
  • Press “SET” key to confirm it.

TRANSTEK GBF-1257-B Body Fat Analyzer D07

3. Setting Gender

  • After confirming Step 2, the system will divert to Gender setting.
  • As pictured below, the portrait blinks. The operator may press the function key or to select Gender.(Male/Female/Male Athlete/ Female Athlete)
  • Press “SET” key to confirm User ID.

TRANSTEK GBF-1257-B Body Fat Analyzer D08

Athlete Mode:
An athlete is considered as a person who does 10 hours or more per week of aerobic activity and has a resting heart rate of 60 beats per minute.These individuals should select Athlete Mode for the most accurate measurement results. The body fat analyzer are not calibrated for professional athletes or body builders.

4. Setting Stature

  • After confirming Gender, the system will divert to Stature setting.
  • As pictured below, the digits “165” blinks. The operator may press the function key or to increase or decrease the numeral. (Range: 100 cm to 220 cm)
  • You may press and hold the function key or for fast changing the numeral.
  • Press “SET” key to confirm Stature.

TRANSTEK GBF-1257-B Body Fat Analyzer D09

5. Setting Age

  • After confirming Stature, the system will divert to Age setting.
  • As pictured below, the digits “30” blinks. The operator may press the function key or to increase or decrease the numeral. (Range: 10 to 85 years old)
  • You may press and hold the function key or for fast changing the numeral.
  • Press “SET” key to confirm Age.

When the LCD displays “STEP ON”, you may start measuring.

TRANSTEK GBF-1257-B Body Fat Analyzer D10

6. Repeat procedure for a second user, or to change user details.
Note: To update or overwrite the memorised data, follow the same procedure, making changes as required.

Select Measurement Unit

With batteries correctly installed, press “UNIT” button in the back of the scale to select measurement unit. The default measurement unit is “lb”. You may press “UNIT” button to choose between pound and kilogram.

TRANSTEK GBF-1257-B Body Fat Analyzer D11

Initialising Your Scale

1. Press the platform centre and remove your foot.
2. “0.0” will be displayed.

TRANSTEK GBF-1257-B Body Fat Analyzer D12
3. The scale will switch off and is now ready for use.

This initialisation process must be repeated if the scale is moved.
At all other times step straight on the scale.

Weight Only Operation

Your Greater Goods Body Fat Analyzer will operate as a conventional weight – reading scale. No special programming steps are required.
Once the scale is initialized, as previously described, you may simply step on the scale to measure your current weight. For only weight reading.

1. Position the scale on a flat, hard surface. Carpeted or uneven floors may affect accuracy.
2. Step onto the scale platform and remain still while the scale computes your weight.
3. The scale will display your weight value.

TRANSTEK GBF-1257-B Body Fat Analyzer D13

4. The scale will automatically turn off after a few seconds.

Note: The weight value won’t be saved in the scale, and won’t be transmitted to the APP.

Start Measuring

First Measurement
  • Please process the user setting before your first measurement.
    ( Refer to Set Up Your Profile for more details)

STEP 1: Step on the platform barefooted when the LCD display “0.0”.

STEP 2: Keep full contact with the electrodes and the circle of LCD runs clockwise. Stand still and the weight data will be locked with LCD stop blinking when the weight is stable.

TRANSTEK GBF-1257-B Body Fat Analyzer Step 02

STEP 3: After displaying weight data, the circle of LCD runs clockwise again. Then measurement results will be displayed followed by BMI, Body Fat, Muscle Mass, Body Water and Bone Mass. The data will be displayed four times and then turn off.

TRANSTEK GBF-1257-B Body Fat Analyzer Step 03

  • If it fails to complete the analysis, the LCD will only display the weight data.
    ( To find out the solutions, please refer to Troubleshooting for more details.)

    • When your scale is successfully paired with your smartphone and the Bluetooth is ON, 0382 will process data transmission instead.
      ( Please refer to Data Transmission for more details.)
Daily Measurement
  • With original SENSE ON patent technology, 0382 will switch on as you step on the platform barefooted.
  • Stand still and keep full contact with the scale until the LCD stops blinking unit.
  • After the weight is locked, the circle of LCD runs clockwise. If identify the stored weight is during this weight within+/-4.5 kg,it will identify the specific user ID.

TRANSTEK GBF-1257-B Body Fat Analyzer E 01

  • If it identifies one user, it will skip the step about user selection, then display BMI, Body Fat, Muscle Mass, Body Water and Bone Mass. The data will be displayed four times and then turn off.

TRANSTEK GBF-1257-B Body Fat Analyzer E 02

  • If it identifies two or more users,for example, U7 and U8, the LCD will display these users altermately two times, you may press the function key or to choose the User ID. The measuring results will then be sorted into the User ID you selected and displayed sequentially four times. If you don’t choose the user, it will display the weight again and turn off.

TRANSTEK GBF-1257-B Body Fat Analyzer E 03

  • If it fails to identify the possible User ID, the LCD will only display the weight data then shutdown.
    ( To find out the solutions, please refer to Troubleshooting for more details.)

Manage Your Health

  • When your scale is successfully paired with your smartphone and the Bluetooth is ON, 0382 will process data transmission instead.
    ( Please refer to Data Transmission for more details.)
Data Transmission

TRANSTEK GBF-1257-B Body Fat Analyzer E 04

  • With 0382 successfully pair-up with your smartphone, the measurement data will be transmitted to your mobile via Bluetooth.
  • ONLY when the measuring results are attached to a specific User ID will it be transmitted to your mobile after measurement.
  • The symbol will disappear after successful data transmission, and you may check your personal health data stored in your smartphone.
  • If the data transmission fails, the symbol will remain. The pending measurement data will be temporarily kept in the scale and transmitted to your smartphone when next measurement is complete.

| Successful Bluetooth Connection
---|---
| Data transmitting to wireless wellness system:
-If SUCCEED, the symbol disappears;
-If FAIL, the symbol remains.

CAUTION

  • Interference may occur in the vicinity of equipment marked with the following symbol . And the Analyzer may interfere vicinity electrical equipment.
  • To enable the data transmission function, this product should be paired to a Bluetooth 4.0 end.

How to mitigate possible interference?

  1. The range between the Analyzer and the Bluetooth end should be reasonably close, from 1 meter to 10 meters. Please ensure no obstacles between the Analyzer and the Bluetooth end so as to obtain quality connection.
  2. To avoid interference, other electronic devices (particularly those with Bluetooth transmission / Transmitter) should be kept at least 1 meter away from the Analyzer.

Troubleshooting

Error Prompt
Error Description Solution
Overload. The device will power off. Stop using this scale for

measurement.
TRANSTEK GBF-1257-B Body Fat Analyzer F02| Low Battery. The device will power off.| Replace all four batteries in the same time. Please purchase the authorized batteries for replacement.
TRANSTEK GBF-1257-B Body Fat Analyzer F03| Failure of pairing up your scale with your smartphone.| Please check below items: -Bluetooth is ON. -App Collector is ON. -Both devices are within the transmission distance of Bluetooth.
TRANSTEK GBF-1257-B Body Fat Analyzer F04| There has data doesn’t upload. (The notebook icon will be displayed.)| Upload the data.

When Measuring …
Problem Root Cause Solution

Abnormal measuring results:
– Too high; OR
– Too low; OR
– Huge difference between two recent measurement.| Incorrect posture.| Please step on the platform barefooted and stand still.
The device is located on the soft ground such as a carpet OR on a rugged surface.| Please place the device on a flat, hard surface.
Your feet are too dry.| Wipe your feet with a damp cloth, keeping them slightly damp when starting measurement.
No display on LCD when the device powers on.| Batteries not yet installed.| Install the batteries.
(Please refer to Insert the Batteries)
Worn batteries.| Replace all four batteries at the same time. Please purchase the authorized batteries for replacement.
CANNOT proceed to analyze calorie, BMI, body fat, body water, muscle mass and bone mass.| Step onto the platform wearing socks or shoes.| Please keep barefooted during the measurement, and keep full contact with the scale as well.
The system cannot identify the possible User ID with most similar data.| Please assign a User ID following the instruction in Set Up Your Profile.
The user fails to select the User ID from what the system found.| Please assign a User ID following the instruction in Set Up Your Profile.
The device powers off.| Low battery.| Replace all four batteries at the same time. Please purchase the authorized batteries for replacement.

When Data Transmitting …
Problem Root Cause Solution
Data transmission failed. Bluetooth is OFF. Turn ON the Bluetooth via

Setting.
The App is OFF.| Press the icon to turn ON your app.
Out of range of Bluetooth transmission.| Place your smartphone closer to the scale.
None of the user ID is assigned.| Please assign a User ID following the instruction in Set Up Your Profile.

Specifications

Specifications
Product Name Body Fat Analyzer (0382)
Dimension Scale: 321x321x24.5mm (Approximately)
Net Weight Approximately1.98kg
Display Digital LCD
Measurement Unit Kilogram / Pound
Measurement Range 5kg to 180kg / 11lb to 397lb
Division 0.1kg / 0.2lb
Accuracy 50kg: ±0.3kg;                 100kg: ±0.4kg;

150kg: ±0.5kg;               180kg: ±0.7kg
Working Environment| Temperature: 5℃ to 40℃  Relative Humidity: ≤90% RH
Atmospheric pressure: 86kPa to 106kPa
Storage Environment| Temperature: -20℃ to 60℃  Relative Humidity: ≤90% RH
Atmospheric pressure: 50kPa to 106kPa
Power Source| 6V (Four AAA-size Batteries)
Turn on Method| SENSE ON technology
Auto-OFF| The scale will turn off after about a few seconds if there is no operation
Accessories|

  1. Four AAA-size batteries
  2. User Manual

Mode of Operation| Continuous Operation
Degree of protection| Type BF applied part
Software Version| A0
Protection Against Ingress of Water| IP21

About the Accuracy of This Product

  • This product passes strict inspection before delivery and therefore its accuracy is guaranteed by the manufacturer. Please refer to the above table for the descriptions on accuracy.
  • This product is specially designed for body fat analysis as well as weight measurement. It should NOT be used by anyone during the process of transaction for verification of goods’ weight.

Maintenance

Maintenance

When carrying out usual maintenance, please ensure practice of the following Do’s and Don’ts:

  • DO use a dry soft cloth to wipe the dust.
  • DO use a wet soft cloth, dipped into water and wrung out, to wipe the dirt. Then use a dry soft cloth to dry up the device. Which should be done before and after using the device.
  • DON’T wash the device with water or immerse it in water.
  • DON’T use propellant, abrasive or other chemicals to wipe the dirt in avoidance of discolor or malfunction.
  • DON’T maintenance this device when in use.
  • DON’T disassemble this device. If you have any problems, please contact Greater Goods. ( Please refer to Warranty for contact information)
Warranty
  • Greater Goods warrants its products free of defects in materials and workmanship in normal use for a period of FIVE years from the date of retail purchase.

  • This warranty does NOT cover damages caused by misuse or abuse, including but not limited to:
    → Failure caused by unauthorized repairs or modifications;
    → Damage caused by shock or drop during transportation;
    → Failure caused by improper operation inconsistent with the instructions stated in this user manual;
    → Malfunction or damage from failure to provide the recommended maintenance;
    → Damage caused by improper use of power supply.

  • Should this device require maintenance (or replacement at our option) under warranty, please deliver the original package to Greater Goods, LLC. prepaid. Please return the store receipt (with the retail purchase date) and a note with reasons to return on it as well.

Distributed by: Greater Goods, LLC
4427 Chouteau Ave.St.Louis,Missouri,63110,Unite States
TEL:866-991-8494 Website: http://greatergoods.com

Appendix

FCC Regulations
FCC User Guide Information

Radio Frequency Interface Requirements – FCC
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna;
  • Increase the separation between the equipment and receiver;
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected;
  • Consult the dealer or an experienced radio / TV technician for help.

Radio Transmitters (Part 15)
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

FCC RF Exposure Guidelines

Safety Information
Reducing RF Exposure – Use Properly
Only operate the device in accordance with the instructions supplied.
This device complies with FCC radiation exposure limits set forth for an uncontrolled environment.

Health Tips – About Body Fat

Human body is made up of, amongst other things, a percentage of fat. Body fat is vital for a healthy, functioning body, protects vital organs, helps regulate body temperature, stores vitamins and helps the body sustain itself when food is scarce. However, too much body fat or indeed too little body fat will damage to your health. It is difficult to gauge how much body fat we have in our bodies simply by looking at ourselves in the mirror.
This is why it is important to measure and monitor your body fat percentage. Body fat percentage gives you a better measure of fitness than weight alone–the composition of your weight loss could mean you are losing muscle mass rather than fat-you could still have a high percentage of fat even when a scale indicates‘normal weight’.

The table as follows may be used as a guide:
Table of Body Fat Level (Unit: %)

a) The body fat percentage (%): 5%-60%/0.1%

Standard for Men

Rating Age
20-29 30-39
Low <13
Normal 13-20
Moderately High 20.1-23
High >23

Standard for Women

Rating Age
20-29 30-39
Low <19
Normal 19-28
Moderately High 28.1-31
High >31

Source: University of Illinois Department of Food Science and Human Nutrition.
Body Fat Percentage Calculator.

Health Tips – About Body Water

Body water is the single most important component of body weight. It represents over half of your total weight and almost two thirds of your lean body mass (predominantly muscle). Water performs a number of important roles in the body:
All the cells in the body, whether in the skin, glands, muscles, brain or anywhere else, can only function properly if they have enough water. Water also plays a vital part in regulating the body’s temperature balance, particularly through perspiration.
The combination of your weight and fat measurement could appear to be ‘normal’ but your body hydration level could be insufficient for healthy living.

The table as follows may be used as a guide:
Table of Body Water Level (Unit: %)

b) The body water percentage (%): 43%-73%/0.1%

Source: Derived from Wang & Deurenberg: “Hydration of fat-free body mass”. American Journal Clin Nutr 1999,69833-841.

| BF % RANGE| OPTIMAL TBW % RANGE
---|---|---
Men| 4 to 14%| 70 to 63%
15 to 21%| 63 to 57%
22 to 24%| 57 to 55%
25 and over| 55 to 37%
Women| 4 to 20%| 70 to 58%
21 to 29%| 58 to 52%
30 to 32%| 52 to 49%
33 and over| 49 to 37%

Health Tips – About Muscle Mass

According to the American College of Sports Medicine (ACSM), lean muscle mass may decrease by nearly 50 percent between the age of 20 and 90. If you do nothing with it you’re losing muscle and increasing fat. It is also important to know your muscle mass % during weight reduction. At rest, the body burns approximately 110 additional calories for each kilo of muscle gained. Benifits of gaining muscle mass include:

  • Reversing the decline in strength, bone density and muscle mass with aging
  • Maintenance of flexible joints
  • Guide weight reduction when combined with a healthy diet.

The table as follows may be used as a guide:
Muscle Mass Percentage

(Source: International Commission on Radiological Protection,1975)

Men Approximately 40% of total body weight
Women Approximately 30% of total body weight
Health Tips – About Bone Mass

Regular exercise and a balanced diet can help maintain healthy bones. Like muscle, bone is a living tissue that can respond to exercise by becoming stronger. For most people, bone mass peaks in their thirties. Then people begin to lose bone. Regular exercise can help prevent that loss.
Calcium and vitamin D, good sources of which are dairy products, green leafy vegetables and fish, contribute to healthy bones.
The bone mass readings given by this product are anestimation of the amount of bone in your body. Those with osteoporosis or low bone densities may not get accurate estimations. If you have any concern regarding your bones please consult your doctor timely.

The table as follows may be used as a guide:

Bone Mass Percentage
(Source: Rico et al.1993)

The average bone mass percentage for both men and women is between 4 to 5%.

Health Tips – About BMI

Maintaining proper weight is essential to healthy living. Obesity has been linked to a higher risk of heart disease, diabetes and some forms of cancer. The Body Mass Index (BMI) is the most widely accepted measurement of weight according to health professionals. The Analyzer will calculate your BMI on your height and weight.

The table as follows may be used as a guide:

TRANSTEK GBF-1257-B Body Fat Analyzer table

Warning

Warning: Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.”
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment 0382, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

EMC Guidance

Guidance and manufacturer’s declaration -electromagnetic emissions and Immunity

Table 1

Guidance and manufacturer’s declaration – electromagnetic emissions

Emissions test| Compliance
RF emissions CISPR 11| Group 1
RF emissions CISPR 11| Class B
Harmonic emissions IEC 61000-3-2| Not applicable
Voltage fluctuations/ flicker emissions IEC 61000-3-3| Not applicable

Table 2

Guidance and manufacturer’s declaration – electromagnetic Immunity

Immunity Test| IEC 60601-1-2
Test level| Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2| ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air| ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst IEC 61000-4-4| Not applicable| Not applicable
Surge IEC 61000-4-5| Not applicable| Not applicable
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11| Not applicable| Not applicable
Power frequency magnetic field| 30 A/m
50Hz/60Hz| 30 A/m
50Hz/60Hz
IEC 61000-4-8| |
Conducted RF IEC61000-4-6| Not applicable| Not applicable
Radiated RF IEC61000-4-3| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE UT is the a.c. mians voltage prior to application of the test level.

Table 3

Guidance and manufacturer’s declaration – electromagnetic Immunity

Radiated RF IEC61000-4-3 (Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment)| Test Frequency (MHz)| Band (MHz)| Service| Modulation| Maximum Power (W)| Distance (m)| IEC 60601-1-2 Test Level (V/m)| Compliance level (V/m)
385| 380 – 390| TETRA 400| Pulse modulation 18 Hz| 1,8| 0.3| 27| 27
450| 430 – 470| GMRS 460, FRS 460| FM ± 5 kHz deviation 1 kHz sine| 2| 0.3| 28| 28
710| 704 – 787| LTE Band 13, 17| Pulse modulation 217 Hz| 0,2| 0.3| 9| 9
745
780
810| 800 – 960| GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5| Pulse modulation 18 Hz| 2| 0.3| 28| 28
870
930
1720| 1700 – 1990| GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS| Pulse modulation 217 Hz| 2| 0.3| 28| 28
1845
1970
2450| 2400 – 2570| Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7| Pulse modulation 217 Hz| 2| 0.3| 28| 28
5240| 5100- 5800| WLAN, 802.11 a/n| Pulse modulation 217Hz| 0.2| 0.3| 9| 9
5500
5785

Distributed by: Greater Goods, LLC
4427 Chouteau Ave.St.Louis,Missouri,63110,Unite States

TEL:866-991-8494 Website: http://greatergoods.com

ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD
23,Jin’an Road, Minzhong, Zhongshan, 528441, Guangdong, China

TEL:86-760-88282982 www.transtek.cn

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals