TRANSTEK GBF-1319-B6 Body Fat Analyzer User Manual
- June 6, 2024
- TRANSTEK
Table of Contents
User Manual
Body Fat Analyzer
GBF-1319-B6
Thank you very much for selecting the Transtek Body Fat Analyzer GBF-1319-B6.
Please do read the user manual carefully and thoroughly so as to ensure the
safe usage of this product, and keep the manual well for further reference in
case you have problems.
Safety Information
Safety and Usage Information
The warning signs and symbols are essential to ensure your correct and safe
use of this product and protect you and others from injury. Please kindly find
the meanings of the warning signs and symbols, which you may encounter in the
label and user manual, as follows:
| The symbol for “THE OPERATION
GUIDE MUST BE READ”
---|---
| The Bluetooth Combination Mark
| The symbol for “TYPE BF APPLIED
PARTS”
SN| The symbol for “SERIAL NUMBER”
| The symbol for “MANUFACTURE
DATE”
| The symbol for “MANUFACTURER”
| The symbol for “DIRECT CURRENT”
| The symbol for “ENVIRONMENT
PROTECTION – Waste electrical products should not be disposed of with
household waste. Please recycle where
facilities exist. Check with your local authority or retailer for recycling
advice”
CAUTION
The Body Fat Analyzer GBF-1319-B6 offers you a seamless way to manage your
health.
Please be aware that this device is designed for adults’ self-measuring and
self-monitoring body fat level. Any information provided by this device is in
no way meant to treat, cure or prevent any disease or illness from happening.
This device should not be used by anyone who is acutely or chronically ill,
suffering from a disease or taking medications that affect your water levels.
The accuracy of readings for these patients has not been verified. Specific
medical advice should be obtained from a physician.
Before diet and exercise, you must first find a professional medical guide.
The body fat analyzer is equipped with data transmission function. It may emit
electromagnetic energy so as to perform its intended function. Nearby portable
and mobile RF communications equipment can affect the performance of the
device . Portable and mobile RF communications equipment can affect the
measuring accuracy of the body fat analyzer.
Kindly note that the use of accessories, transducers or cables other than
those specified, with the exception of transducers and cables sold by the
manufacturer as replacement parts for internal components, may result in
increased EMISSIONS or decreased IMMUNITY of the device.
Be aware that misuse of electrical equipment can cause electric shock, burns,
fire and other hazards.
Warning that the body fat analyzer should not be used adjacent to or stacked
with other equipment.
The manufacturer will make available on request circuit diagrams, component
parts list, etc.
WARNING : No modifications of this equipment is allowed.
During using the analyzer, the patient will contact the surface and the
electrode of the Analyzer. The surface and the electrode of the analyzer is
type BF Applied part.
The device doesn’t need to be calibrated in five years’ lifetime. Do not touch
the output of batteries when using the analyzer. The patient is an intended
operator. The patient can perform all the operations in the manual, such as
measurement, data transmitting, changing batteries. Please do not use the
device’s inflammable gas environment.
INDICATIONS FOR USE
The Transtek Body Fat Analyzer measures weight and estimates BMI and calories,
and uses bio-electrical impedance analysis (BIA) technology to estimate body
fat, total body water percentage, bone mass, and muscle mass in generally
healthy children 10-17 years old and healthy adults.
It is intended for use in the domestic setting only.
CONTRAINDICATIONS
-
This device is contraindicated for any female subject who may be suspected of, or is pregnant.
Besides providing inaccurate readings, the effects of this device on the fetus are unknown. -
This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument such as a pacemaker or defibrillator.
Your Scale and Its Environment
To ensure your safety as well as the service life of your scale, please
avoid using the scale under the following circumstances:
- Slippery floor such as tile floor
- Jumping onto the platform immediately after bath or with wet feet
- Near a cell phone or microwave oven
Avoid storage in the following locations:
- Where there is water
- Where the device may be exposed to extreme temperatures, humidity, moisture, direct sunlight, dust, or salt air
- Where there is risk of shock or drop Where you store chemicals or full of corrosive gases
- Where in reach of the infants or children
Efficient Use of Your Scale
To ensure the accuracy of measurement, please follow the below instructions
when you start measurement.
- Place the scale on a flat, hard surface. Soft surfaces such as carpets will affect the performance of the scale.
- Step onto the platform with bare feet. Stand still and keep full contact with the scale until the measurement is complete.
- Start measurement at least two hours after Getting up or Dinning.
- Avoid measurement immediately after strenuous exercise, sauna or bath, drinking, and dining.
- Always start measurement in the same time slot and on the same scale located on the same flat, hard surface.
Overview
Device Components
LCD Display
FTA
| Body Fat Analysis Result|
P8
| User ID (Range from P1 to P8)
---|---|---|---
TBW
| Total Body Water Analysis Result| | Male
MUS
| Muscle Mass Analysis Result|
| Male Athlete
BONE
| Bone Mass Analysis Result| | Female
BMR
| Basal Metabolism Rate| | Female Athlete
BMI
| Body Mass Index| | Height
ST
| Stone|
AGE
| Age
ST;IB
| Pound|
| Data hasn’t been sent out
Kg
| Kilogram| ****| Bluetooth Icon
%
| Percentage|
ft
| Foot
cm
| Centimeter|
kcal
| Calorie
Initial Start-Up
General Instructions
Transtek Body Fat Analyzer GBF-1319-B6 applies BIA (Bio-impedance Analysis)
technology. A small amount of weak current flows through the human body so as
to detect
the bio-impedance and estimate calorie, BMI, body fat, body water, muscle mass
and bone mass. The electrical current is small and may not be felt.
This BIA technology is cheap, safe, non-invasive, toxic-free and harmless. It
also possesses the characteristics of simple operation and abundant
information.
The current mentioned above is less than 0.5mA. However, please be aware that
anyone with a wearable or implantable medical electronic instrument, such as a
pacemaker, must avoid using this device.
The intended use of this device is for healthy children 10-17 years old and
healthy adults.
Insert the Batteries
-
Open the battery door in the back of the scale.
-
Insert the batteries (3 x 1.5V AAA) into the battery compartment according to the polarity indications marked inside the compartment.
-
The digits “ 8888 ” will be shown on the LCD.
-
Close the battery door and wait until the digits “
**00**
” are shown on the LCD.
CAUTION
When the symbol “ Lo” appears, the device will power off. Then you shall
replace it with a new set of batteries. Please replace all three batteries at
the same time. Do NOT mix the old batteries with the new ones.
Worn batteries are hazardous waste. Do NOT dispose of them together with the
household garbage. Please refer to the local ordinances and recycling
instructions regarding the disposal of worn batteries and scrapped devices.
If you do not intend to use this unit for a prolonged period of time, it is
advisable to remove the batteries before storing them.
Select Measurement Unit
With batteries correctly installed, press the “UNIT” button in the back of the
scale to select the measurement unit. The default measurement unit is
“kilogram”. You may press the “UNIT” button to choose among kilogram, pound,
and stone.
Note : If there is no operation in the weighing mode, the scale will turn off after a few seconds.
Install APP and Pair-Up
With the advanced Bluetooth 4.0 technology applied, the mobile or portable
equipment, which is equipped with Bluetooth function in line with BLE
Technical Specifications as well as BLP Protocol established by global
organization Bluetooth SIG, are capable to receive your personal health data.
Just simply install the specially-designed app and pair up your scale with
your mobile or portable equipment. Then you may enjoy the comprehensive health
solution provided by Transtek.
Note : It is only available for the device which supports Bluetooth 4.0
technology. If SUCCEED, the symbol “ ****” will be shown on the LCD.
Initial Start-Up
Set Up Your Profile
The body fat analyzer GBF-1319-B6 supports multiple users (Up to 8). You may
follow the below instructions to assign a User ID and set up your own profile,
including Gender, Stature, and Age. 1. Assigning User ID
With batteries correctly installed, press the “SET” key when the LCD displays
0.0 to enter setting mode.
The system will request User ID selection first. As pictured below, “P1”
blinks. The operator may press the function key or to select User ID from P1
to P8.
Press the “SET” key to confirm the User ID.
2. Setting Gender
After confirming the User ID, the system will divert to the Gender setting
automatically.
As pictured below, the portrait blinks. The operator may press the function
key or to select Gender.( Male/Female/Male Athlete/ Female Athlete). Press the
“SET” key to confirm Gender.
Athlete Mode:
An athlete is considered as a person who does 10 hours or more per week of
aerobic activity and has a resting heart rate of 60 beats per minute. These
individuals should select Athlete Mode for the most accurate measurement
results.The body fat analyzer are not calibrated for professional athletes or
bodybuilders.
3. Setting Stature
After confirming Gender, the system will divert to the Stature setting
automatically.
As pictured below, the digits “ 165 ” blinks. The operator may press the
function key or to increase or decrease the numeral.
You may press and hold the function key or for fast change the numeral. Press
the “SET” key to confirm Stature.
4. Setting Age
- After confirming Stature, the system will divert to Age setting.
- As pictured below, the digits “30” blinks. The operator may press the function key or to increase or decrease the numeral.
- You may press and hold the function key or for fast changing the numeral.
- Press “SET” key to confirm Age.
5. After confirming the Age, the LCD will display “0.0”, then you can start measuring.
6. Repeat the procedure for a second user, or to change user details.
Note: To update or overwrite the memorized data, follow the same procedure,
making changes as required.
Weight Only Operation
Your TRANSTEK Body Fat Analyzer will operate as a conventional weight-reading
scale.
No special programming steps are required.
Once the scale is initialized, as previously described, you may simply step on
the scale to measure your current weight. For only weight reading.
-
Position the scale on a flat, hard surface. Carpeted or uneven floors may affect accuracy.
-
Step onto the scale platform and remain still while the scale computes your weight.
-
The scale will display your weight value.
-
The scale will automatically turn off after a few seconds.
Note: The weight value won’t be saved in the scale, and won’t be transmitted to the APP.
Start Measuring
First Measurement
Please process the user setting before your first measurement. (Refer to Set
Up Your Profile for more details)
STEP 1: Step on the platform barefooted when the LCD display “0.0”.
STEP 2: Stand still and the weight data will be locked with blinking twice when the weight is stable. Keep full contact with the scale until the LCD stops displaying moving “o”.
STEP 3: The measurement results will be displayed followed by Weight, Body Fat, Total Body Water, Muscle Mass, Bone Mass, BMR and BMI. The data will be displayed three times and then turn off.
- If it fails to complete the analysis, the LCD will only display the weight data.
(To find out the solutions, please refer to Troubleshooting for more details.)
When your scale is successfully paired with your smartphone and the Bluetooth is ON, GBF-1319-B6 will process data transmission.
(Please refer to Data Transmission for more details.)
Daily Measurement
With original SENSE ON patent technology, GBF-1319-B6 will switch on as you
step on the platform barefooted.
Stand still and keep full contact with the scale until the LCD stops
displaying moving “o”.
-
After the weight is locked, if identify the stored weight is during this weight within+/-2 kg, it will identify the specific user ID.
-
If it identifies one user, it will skip the step about user selection, then display the Weight, Body Fat, Total Body Water, Muscle Mass, Bone Mass, BMR and BMI data sequentially three times and turn off.
-
If it identifies two or more users,for example,P1, P2 and P3, the LCD will display these users alternately two times, you may press the function key or to choose the User ID. The measuring results will then be stored into the User ID you selected and displayed sequentially three times. If you don’t choose the user, it will turn off.
-
If it fails to identify the possible User ID, the LCD will only display the weight data then shut down.
(To find out the solutions, please refer to Troubleshooting for more details.)
Data Transmission
- When your scale is successfully paired with your smartphone and the Bluetooth is ON, GBF-1319-B6 will process data transmission.
(Please refer to Data Transmission for more details.)
Data Transmission
-
With GBF-1319-B6 successfully pair-up with your smartphone, the measurement data will be transmitted to your mobile via Bluetooth.
-
ONLY when the measuring results are attached to a specific User ID will it be transmitted to your mobile after measurement.
-
The symbol will disappear after successful data transmission, and you may check your personal health data stored in your smartphone.
-
If the data transmission fails, the symbol will remain. The pending measurement data will be temporarily kept in the scale and transmitted to your smartphone when the next measurement is complete.
**** | Successful Bluetooth Connection |
---|---|
Data transmitting to |
wireless wellness system:
-If SUCCEED, the symbol disappears;
-If FAIL, the symbol remains.
CAUTION
Interference may occur in the vicinity of equipment marked with the following
symbol . And the Analyzer may interfere vicinity electrical equipment.
To enable the data transmission function, this product should be paired to a
Bluetooth 4.0 end. How to mitigate possible interference?
- The range between the Analyzer and the Bluetooth end should be reasonably close, from 1 meter to 10 meters. Please ensure no obstacles between the Analyzer and the Bluetooth end so as to obtain quality connection.
- To avoid interference, other electronic devices (particularly those with Bluetooth transmission / Transmitter) should be kept at least 1 meter away from the Analyzer.
Troubleshooting
Error Prompt
Error | Description | Solution |
---|---|---|
Overload. The device will power off. | Stop using this scale for |
measurement.
| Low Battery. The device will power off.| Replace all three batteries in the
same time. Please purchase the authorized batteries for replacement.
| Failure of pairing up your scale with your smartphone.| Please check below
items:
-Bluetooth is ON.
-App Collector is ON.
-Both devices are within the transmission distance of Bluetooth.
| There has data doesn’t upload. (The notebook icon will be displayed.)|
Upload the data.
When Measuring …
Problem | Root Cause | Solution |
---|
Abnormal measuring results:
– Too high; OR
– Too low; OR
– Huge difference between two recent measurements.
| ncorrect posture.| Please step on the platform barefooted and standstill.
The device is located on the soft ground such as a carpet OR on a rugged
surface.| Please place the device on a flat, hard surface.
Problem| Root Cause| Solution
---|---|---
Abnormal measuring results:
– Too high; OR
– Too low; OR
– Huge difference between two recent measurement.
| Your feet are too dry.| Wipe your feet with a damp cloth, keeping them
slightly damp when starting measurement.
No display on LCD when the device powers on.| Batteries not yet installed.|
Install the batteries. (Please refer to Insert the
Batteries )
Worn batteries.| Replace all three batteries at the same time. Please purchase
the authorized batteries for replacement.
CANNOT proceed to analyze calorie, BMI, body fat, total body water, muscle
mass and bone mass.| Step onto the platform wearing socks or shoes.| Please
keep barefooted during the measurement, and keep full contact with the scale
as well.
The system cannot identify the possible User ID with most similar data.|
Please assign a User ID following the instruction in Set Up Your Profile.
The user fails to select the User ID from what the system found.| Please
assign a User ID following the instruction in Set Up Your Profile.
The device powers off.| Low battery.| Replace all three batteries at the same
time. Please purchase the authorized batteries for replacement.
When Data Transmitting …
Problem | Root Cause | Solution |
---|---|---|
Data transmission failed. | Bluetooth is OFF. | **** Turn ON the Bluetooth via |
Setting.
The App is OFF.| Press the icon to turn ON your app.
Out of range of Bluetooth transmission.| Place your smartphone closer to the
scale.
None of the user ID is assigned.| Please assign a User ID following the
instruction in Set Up Your Profile.
About the Accuracy of This Product
- This product passes strict inspection before delivery and therefore its accuracy is guaranteed by the manufacturer. Please refer to the above table for the descriptions on accuracy.
- This product is specially designed for body fat analysis as well as weight measurement. It should NOT be used by anyone during the process of transaction for verification of goods’ weight.
Product Name | Body Fat Analyzer (GBF-1319-B6) |
---|---|
Dimension | Scale: 321x321x23.5mm (Approximately) |
Net Weight | Approximately 1.98kg ( Including the dry cells) |
Display | Digital LCD |
Measurement Unit | Kilogram / Pound / Stone |
Measurement Range | 5kg to 180kg / 11lb to 397lb / 0st: 11lb to 28st: 5lb |
Division | 0.2lb / 0.1kg |
Accuracy | 50kg: ±0.3kg; 100kg: ±0.4kg; |
150kg: ±0.5kg; 180kg: ±0.7kg
Working Environment
| Temperature: 5℃ to 40℃ Relative Humidity: ≤90% RH
Atmospheric pressure: 86kPa to 106kPa
Storage Environment
| Temperature: -20℃ to 60℃ Relative Humidity: ≤90% RH
Atmospheric pressure: 50kPa to 106kPa
Power Source| 4.5V (Three AAA-size Batteries)
Auto-ON| SENSE ON technology
Auto-OFF| The scale will turn off after about 10s if there is no
operation.
Accessories
| 1. Three AAA Size Batteries
2. User Manual
Mode of Operation| Continuous Operation
Bluetooth Frequency Range| 2402MHz-2480MHz
Bluetooth Output Power| 0dBm
Bluetooth Module No| LS8261
Bluetooth Supply Voltage| 3.0-3.6V
Software Version| A12
Degree of protection| Type BF applied part
Protection against ingress of water| IP21
Maintenance
Maintenance
When carrying out usual maintenance, please ensure the practice of the
following Dos and Don’ts:
- DO use a dry soft cloth to wipe the dust.
- DO use a wet soft cloth, dipped into water and wrung out, to wipe the dirt. Then use a dry soft cloth to dry up the device.
- DON’T wash the device with water or immerse it in water.
- DON’T use propellant, abrasive or other chemicals to wipe the dirt in avoidance of discolor or malfunction.
- DON’T disassemble this device. If you have any problems, please contact Transtek.
(Please refer to Warranty for contact information)
Warranty
- Transtek warrants its products free of defects in materials and workmanship in normal use for a period of FIVE years from the date of retail purchase.
- This warranty does NOT cover damages caused by misuse or abuse, including but not limited to:
- Failure caused by unauthorized repairs or modifications;
- Damage caused by shock or drop during transportation;
- Failure caused by improper operation inconsistent with the instructions stated in this user manual;
- Malfunction or damage from failure to provide the recommended maintenance;
- Damage caused by improper use of power supply.
- Should this device require maintenance (or replacement at our option) under warranty, please deliver the original package to ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD prepaid. Please return the store receipt (with the retail purchase date) and a note with reasons to return on it as well.
ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD
No.23,Jin’an Road, Minzhong, Zhongshan, 528441, Guangdong, China
Tel: 86-760-88282982
Website: http://www.transtek.cn.
Appendix
FCC Regulations
FCC User Guide Information
Radio Frequency Interface Requirements – FCC
Note: This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the
user is encouraged to try to correct the interference by one or more of the
following measures:
Reorient or relocate the receiving antenna;
Increase the separation between the equipment and receiver;
Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected;
Consult the dealer or an experienced radio / TV technician for help.
Radio Transmitters (Part 15)
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) this device may not cause harmful
interference, and (2) this device
must accept any interference received, including interference that may cause
undesired operation.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
FCC RF Exposure Guidelines
Safety Information
Reducing RF Exposure – Use Properly
Only operate the device in accordance with the instructions supplied.
This device complies with FCC radiation exposure limits set forth for an
uncontrolled environment.
Health Tips – About Body Fat
The human body is made up of, amongst other things, a percentage of fat. Body
fat is vital for a healthy, functioning body, protects vital organs, helps
regulate body temperature,
stores vitamins and helps the body sustain itself when food is scarce.
However, too much body fat or indeed too little body fat will damage to your
health. It is difficult to gauge how much body fat we have in our bodies
simply by looking at ourselves in the mirror. This is why it is important to
measure and monitor your body fat percentage. Body fat
percentage gives you a better measure of fitness than weight alone–the
composition of your weight loss could mean you are losing muscle mass rather
than fat-you could still
have a high percentage of fat even when a scale indicates normal weight’. The
table as follows may be used as a guide: Table of Body Fat Level (Unit: %)
a) The body fat percentage (%): 5%-60%/0.1%
Standard for Men
Standard for Women
Rating | Age | Rating | Age |
---|---|---|---|
20-29 | 30-39 | 40-49 | 50-59 |
low | <13 | <14 | <16 |
Normal | 13-20 | 14-21 | 16-23 |
23-32
Moderately High| 20.1-23| 21.1-24| 23.1-26| 24.1-27| 25.1-28| Moderately High|
28.1-31| 29.1-32| 30.1-33| 31.1-34| 32.1-35
High| >23| >24| >26| >27| >28| High| >31| >32| >33| >34| >35
Source: University of Illinois Department of Food Science and Human Nutrition.
Body Fat Percentage Calculator.
www.ag.uiuc.edu/~food-lab/ai/bfc.html
Health Tips – About Body Water
Body water is the single most important component of body weight. It
represents over half of your total weight and almost two thirds of your lean
body mass (predominantly
muscle). Water performs a number of important roles in the body: All the cells
in the body, whether in the skin, glands, muscles, brain or anywhere else, can
only function properly if they have enough water. Water also plays a vital
part in regulating the body’s temperature balance, particularly through
perspiration. The combination of your weight and fat measurement could appear
to be ‘normal’ but your body hydration level could be insufficient for healthy
living. The table as follows may be used as a guide:
Table of Body Water Level (Unit: %)
b) The body water percentage (%): 43%-73%/0.1%
Source: Derived from Wang & Deurenberg: “Hydration of fat-free body mass”.
American Journal Clin Nutr 1999,69833-841.
| BF % RANGE| OPTIMAL TBW % RANGE
---|---|---
Men
| 4 to 14%| 70 to 63%
15 to 21%| 63 to 57%
22 to 24%| 57 to 55%
25 and over| 55 to 37%
Women
| 4 to 20%| 70 to 58%
21 to 29%| 58 to 52%
30 to 32%| 52 to 49%
33 and over| 49 to 37%
Health Tips – About Muscle Mass
According to the American College of Sports Medicine (ACSM), lean muscle mass
may decrease by nearly 50 percent between the age of 20 and 90. If you do
nothing with it you’re losing muscle and increasing fat. It is also important
to know your muscle mass % during weight reduction. At rest, the body burns
approximately 110 additional calories for each kilo of muscle gained. Benifits
of gaining muscle mass include:
- Reversing the decline in strength, bone density and muscle mass with aging
- Maintenance of flexible joints
- Guide weight reduction when combined with a healthy diet.
The table as follows may be used as a guide: Muscle Mass Percentage
(Source: International Commission on Radiological Protection,1975)
Men | Approximately 40% of total body weight |
---|---|
Women | Approximately 30% of total body weight |
Health Tips – About Bone Mass
Regular exercise and a balanced diet can help maintain healthy bones. Like
muscle, bone is a living tissue that can respond to exercise by becoming
stronger. For most
people, bone mass peaks in their thirties. Then people begin to lose bone.
Regular exercise can help prevent that loss. Calcium and vitamin D, good
sources of which are dairy products, green leafy vegetables and fish,
contribute to healthy bones. The bone mass readings given by this product are
anestimation of the amount of bone in
your body. Those with osteoporosis or low bone densities may not get accurate
estimations. If you have any concern regarding your bones please consult your
doctor timely. The table as follows may be used as a guide:
Bone Mass Percentage
(Source: Rico et al.1993)
The average bone mass percentage for both men and women is between 4 to 5%.
Health Tips – About BMI
Maintaining proper weight is essential to healthy living. Obesity has been
linked to a higher risk of heart disease, diabetes and some forms of cancer.
The Body Mass Index (BMI) is the most widely accepted measurement of weight
according to health professionals. The Analyzer will calculate your BMI on
your height and weight.
The table as follows may be used as a guide:
| NORMAL| OVERWEIGHT| OBESE
---|---|---|---
| | Body wight (pounds)|
58| | |
59
60
61
62
63
64
65
67
68
69
70
71
72
73
74
75
| | |
Guidance and manufacturer’s declaration – electromagnetic Immunity
Immunity Test| IEC 60601-1-2
Test level
| Compliance level
Electrostatic discharge (ESD)
IEC 61000-4-2
| ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
| ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst
IEC 61000-4-4
| Not applicable| Not applicable
Surge
IEC 61000-4-5
| Not applicable| Not applicable
Voltage dips, short interruptions and voltage variations on power supply input
lines
IEC 61000-4-11
| Not applicable| Not applicable
Power frequency magnetic
field
| 30 A/m
50Hz/60Hz
| 30 A/m
50Hz/60Hz
IEC 61000-4-8| |
Conducted RF
IEC61000-4-6
| Not applicable| Not applicable
Radiated RF IEC61000-4-3| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE UT is the a.c. mians voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic Immunity
Radiated RF IEC61000-4-3
(Test specifications for ENCLOSURE PORT IMMUNITY to
RF wireless communications equipment)
| Test Frequency (MHz)| Band
(MHz)
| Service| Modulation| Modulation
(W)
| Distance
(m)
| IMMUNITY TEST LEVEL
(V/m)
385| 380 –
390
| TETRA 400| Pulse modulation 18 Hz| 1,8| 0.3| 27
450| 430 –
470
| GMRS 460,
FRS 460
| FM
± 5 kHz deviation 1 kHz sine
| 2| 0.3| 28
710| 704 –
787
| LTE Band 13,
17
| Pulse modulation 217 Hz| 0,2| 0.3| 9
745
780
810| 800 –
960
| GSM 800/900, TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
| Pulse modulation 18 Hz| 2| 0.3| 28
870
930
1720| 1 700 –
1 990
| GSM 1800; CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3,
4, 25;
UMTS
| Pulse modulation 217 Hz| 2| 0.3| 28
1845
1970
2450| 2 400 –
2 570
| Bluetooth, WLAN, 802.11
b/g/n, RFID 2450,
LTE Band 7
| Pulse modulation 217 Hz| 2| 0.3| 28
ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD
No.23,Jin’an Road, Minzhong, Zhongshan, 528441, Guangdong, China
TEL:86-760-88282982 www.transfer.cn