TRANSTEK TMB-2084-T Blood Pressure Monitor User Manual
- June 6, 2024
- TRANSTEK
Table of Contents
TRANSTEK TMB-2084-T Blood Pressure Monitor
INTRODUCTION
General Description
Thank you for selecting TRANSTEK arm type blood pressure monitor (TMB-2084-T).
The monitor features blood pressure measurement, pulse rate measurement and
the result storage. The design provides you with two years of reliable
service. Readings taken by the TMB-2084-T are equivalent to those obtained by
a trained observer using the cuff and stethoscope auscultation method. This
manual contains important safety and care information, and provides step by
step instructions for using the product.
Read the manual thoroughly before using the product.Features:
- 61mm×74 mm Digital LCD display with white backlight
- Maximum 120 records per each user
- 3rd technonoly: Measuring during inflation
Indications for Use
The Transtek Blood Pressure Monitor is digital monitors intended for use in
measuring blood pressure and heartbeat rate with arm circumference ranging
Contraindications
- The device is not suitable for use on may be pregnant women or pregnant women.
- The device is not suitable for use on patients with implanted,electrical devices, such as cardiac pacemakers, defibrillators.
Measurement Principle
This product uses the Oscillometric Measuring method to detect blood
pressure. Before every measurement, the unit establishes a “zero pressure”
equivalent to the atmospheric pressure. Then it starts inflating the arm cuff,
meanwhile, the unit detects pressure oscillations generated by beat-to-beat
pulsatile, which is used to determine the systolic and diastolic pressure, and
also pulse rate.
Safety Information
The signs below might be in the user manual, labeling or other component. They
are the requirement of standard and using.
LCD display signal
Monitor Components
Component list of pressure measuring system
- Cuff
- Air Hose
- PCBA
- Pump
- Valve
List
- Blood Pressure Monitor
- Cuff
- User manual
- 4×AAA batteries
The Choice of Power Supply
- Battery powered mode: 6VDC 4×AAA batteries
- AC adaptor powered mode: 5V 1A
(Please only use the recommended AC Please unplug the adaptor to depart from the using utility power.
Installing and Replacing the Batteries
- Open the battery cover.
- Install the batteries by matching the correct polarity, as shown.
- Replace the battery cover.
CAUTION
- Do not use new and used batteries together.
- Do not use different types of batteries together.
- Do not dispose the batteries in fire. Batteries may explode or leak.
- Remove batteries if the device is not likely to be used for some time.
- Worn batteries are harmful to the environment. Do not dispose with daily garbage.
- Remove the old batteries from the device following your local recycling guidelines.
BEFORE YOU START
Setting Date, Time and Measurement Unit
It is important to set the clock before using your blood pressure monitor, so
that a time stamp can be assigned to each record that is stored in the memory.
(The setting range of the year :2020—2099)
- When the monitor is off, long press “SET” button,it will display [YEAR]. Each press “SET” button will increase the number.
- Press the “MEM” to change the [YEAR]. Each press will decrease the numeral by one in a cycling manner.
- When you get the right year, press “START/STOP” to set down and turn to next step.
- Repeat steps 2 and 3 to set the [MONTH] and [DAY].
- Repeat steps 2 and 3 to set the [HOUR] and [MINUTE].
- Repeat steps 2 and 3 to set the [UNIT].
- After the unit is set,the LCD will display “dOnE” and then it will turn off.
Setting the voice switch
When the monitor is off,press both “STRAT/STOP” and “SET” button, it will
enter voice setting mode. The vocie switch will blink, press “MEM” button to
decrease the volume, press “SET” button to increase the volume, the press
“START/STOP” button to confirm then the monitor will turn off.
Setting the clock
- When the monitor is off, long press both “SET” and “MEM” button, it will enter clock setting mode. Press “MEM” or “SET” button to choose turn off or turn on. Press“START/STOP” button to confirm.
- When you choose turn on, press “START/STOP” button, the LCD will display the time, Choose “MEM” or “SET” button to set the time, press “START/STOP” button to cofirm, the LCD will display“dOnE” and then the monitor will turn off.
Setting the User ID
- When the monitor is off, hold press “SET” button and then the user ID will shows. Press “MEM” or “SET”button to switch the user ID between user 1 and user 2 mode.
- Press “START/STOP” button to confirm user ID , the display will show “User ID + dOnE” and then the monitor will turn off.
Install the App and Pair-Up
Download the Transtek Health app from APP Store or Google Play. Install the
APP, and register an account.
click “My setting”, choose the device and then bind the device and app
Search your test information
After binding the app, back to the beginning page and click “Blood Pressure”
to search your test information.
List of compatible devices:
- For iOS devices:
- The operating system must be iOS 11.0 or more. For Android devices:
- The operating system must be Android 8.0 or more.
Tie the cuff
-
Remove all jewelry, such as watches and bracelets from your left arm. Note: If your doctor has diagnosed you with poor circulation in your left arm, use your right arm.
-
Roll or push up your sleeve to expose the skin. Make sure your sleeve is not too tight.
-
Hold your arm with your palm facing up and tie the cuff on your upper arm, then position the tube off-center toward the inner side of arm in line with the little finger. Or position the artery mark over the main artery (on the artery by pressing with 2 fingers approximately 2 cm above the bend of your elbow on the inside of your left arm. Identify where the pulse can be felt the strongest. This is your main artery.
-
The cuff should be snug but not too tight. You should be able to insert one finger between the cuff and your arm.
-
Sit comfortably with your tested arm resting on a flat surface. Place your elbow on a table so that the cuff is at the same level as your heart. Turn your palm upwards. Sit upright in a chair, and take
5- 6 deep breaths. -
Helpful tips for Patients, especially for Patients with Hypertension:
-
Rest for 5 minutes before first measuring.
-
Wait at least 3 minutes between measurements.
-
This allows your blood circulation to recover.
-
Take the measurement in a silent room.
-
The patient must relax as much as possible and do not move and talk during the measurement procedure. The cuff should maintain at the same level as the right atrium of the heart.
-
Please sit comfortably. Do not cross your legs and keep your feet flat on the ground.
-
Keep your back against the backrest of the chair.
-
For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, on the same arm, or as directed by a physician.
MEASUREMENT
Start the Measurement
Before you start the measurement,Download the Transtek Health app from APP
Store or Google Play,and turn on the Bluetooth. Install the APP, and register
an account. Then set your personal information (Gender, Birthday, Height,
Weight, Name and so on).
- Please switch the User button to select the user between User 1 and User 2. When the monitor is off,press the “START/STOP” button to turn on the monitor, and it will finish the whole measurement,save and transmit the measurement data for the desired user. (Take User 1 for example)
- If the data transmits successful, the symbol and symblo will disappear, and then the monitor will turn off.
- If the data transmition fails, the monitor will turn off automatically.
- Press the “START/STOP” to power off.
Recall the Records
- When the monitor is off, please hold press the “MEM”,it will display the latest record first when the records are less than three groups. When there are three or more than three groups ,it will display the average value of the latest three records first
- Press the “MEM” or “SET” to get the record you want.
Delete the Records
If you did not get the correct measurement, you can delete all results for
the selected user by following steps below.
- Long press “MEM”, when the monitor is in the memory recall mode, the flash display“YES” will show.
- Press “SET” or “MEM”,the flash display “ON” or “YES” will show alternatively,choose “YES” to confirm deleting.
- Press “START/STOP” button to confirm delete the records, the display will show “User ID + dOnE” and then the monitor will turn off.
- If there is no record, press both “START/STOP” and “MEM” button to confirm delete all the records, the following display will show and then the monitor will turn off.
Tips for Measurement
Measurements may be inaccurate if taken in the following circumstances..
Maintenance
In order to get the best performance, please follow the instructions below.
What are systolic pressure and diastolic pressure?
When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.
What is the standard blood pressure classification?
The blood pressure classification published by 1999 is as follows:
Level
Blood
Pressure (mm Hg)
|
Optimal
|
Normal
|
High-normal
|
Mild
|
Moderate
|
Severe
---|---|---|---|---|---|---
SYS| <120| 120-129| 130-139| 140-159| 160-179| 2180
DIA| <80| 80-84| 85-89| 90-99| 100-109| 2110
Irregular Heartbeat Detector
An irregular heartbeat is detected when a heartbeat rhythm varies while the
unit is measuring the systolic and diastolic blood pressure.During each
measurement, the monitor records all the pulse intervals and calculate the
average ; if there are two or more pulse intervals ,the difference between
each interval and the average is more than the average value of ±25% , or
there are four or more pulse intervals ,the difference between each interval
and the average is more than the average value of ±15%,the irregular heartbeat
symbol appears on the display when the measurement results are appeared.
Why does my blood pressure fluctuate throughout the day?
- Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions.
- If the person takes medicine, the pressure will vary more.
- Wait at least 3 minutes for another measurement.
Why do I get a different blood pressure at home compared to the hospital?
The blood pressure is different even throughout the day due to weather,
emotion, exercise etc, Also, there is the “white coat” effect, which means
blood pressure usually increases in clinical settings.
Is the result the same if measuring on the right arm?
It is ok for both arms, but there will be some different results for different
people. We suggest you measure the same arm every time.
TROUBLESHOOTING
This section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If the products not operating as you think it should, check here before arranging for servicing.
PROBLEM | SYMPTOM | CHECK THIS | REMEDY |
---|
No power
| ****
Display will not light up.
| ****
Batteries are exhausted.
| ****
Replace with new batteries
Batteries are inserted incorrectly.| Insert the batteries correctly
AC adaptor is inserted incorrectly.| Insert the AC adaptor tightly
Low batteries| Display is dim or show| Batteries are low.| Replace with
new batteries
Error message
| ****
E01 shows
| ****
The cuff is too tight or too loose.
| Refasten the cuff and then measure again.
E02 shows
| The monitor detected motion or talking during measuring.| Movement can affect the measurement.Relax for a moment and then measure again
E03 shows
| The measurement process does not detect the pulse signal.| Loosen the clothing on the arm and then measure again.
E04 shows
| The treatment of the measurement failed.| Relax for a moment and then measure again.
H bAt shows
| Adaptor voltage is higher than 7.5V| Replace with new adaptor
E+ ,shows on the display.| ****
Matching failure or transmission timeout.
| ****
Reconnect to the bluetooth and re-upload data
EEx,shows on the display.
| ****
Hardware error
| ****
Retake the measurement.If the problem persists, contact the retailer or our customer service department for further assistance. Refer to the warranty for contact information and return instructions.
Warning message| ****
“out ” shows
| ****
Out of measurement range
| Relax for a moment. Refasten the cuff and then measure again. If the problem
persists, contact your physician.
Power supply| Battery powered mode: 6VDC 4×AAA batteries AC adaptor
powered mode: 5V 1A
(Please only use the recommended AC
adaptor model).
---|---
Display mode| Digital LCD V.A.61mm × 74mm
Measurement mode| Oscillographic testing mode
Measurement range
| Rated cuff pressure: 0mmHg~299mmHg(0kPa ~ 39.9kPa) Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa) Pulse value: (40-199)beat/minute
Accuracy| Pressure:
5℃-40℃within±3mmHg(0.4kPa) Pulse value:±5%
Normal working condition
| A temperature range of :+5°C to +40°C A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa
Storage & transportation condition| Temperature:-20°C to +60°C
A relative humidity range of :: 93%, non-condensing,
at a water vapour pressure up to 50hPa
Measurement perimeter
of the upper arm
| About 22 cm ~ 32 cm, 22cm ~ 42cm, 22cm ~ 45cm
Weight| Approx.238g(Excluding the batteries and cuff)
External dimensions| Approx.100.2mm×138.2mm×51mm
Attachment| 4×AAA batteries,user manual,AC adapter
Mode of operation| Continuous operation
Degree of protection| Type BF applied part
Protection against ingress of water| IP21 It means the device could
protected against solid foreign objects of 12.5mm and greater, and protect
against vertically falling water drops.
Device Classification
| Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment
Software Version| A01
AUTHORIZED COMPONENT
Contact Information
For more information about our products, please visit www.transtek.cn.you can
get customer service, usual problems and customer download, transtek will
serve you anytime.
- Authorized European Representative:
- Company: MDSS – Medical Device Safety Service GmbH
- Address: Schiffgraben 41, 30175 Hannover, Germany
COMPLIED STANDARDS LIST
FCC Statement
FCC ID: OU9TMB2080-B
This device complies with Part 15 of the FCC Rules. Operation issubject to the
following two conditions: (1) this device may not cause harmful interference,
and (2) this device must accept any interference received, including
interference that may cause undesired operation.
Caution : The user is cautioned that changes or modifications not
expressly approved by the party responsible for compliance could void the
user’s authority to operate the equipment.
NOTE : This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions,may cause harmful interference to radio
communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
FCC Radiation Exposure Statement:
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. This transmitter must not be co-located or operating
in conjunction with any other antenna or transmitter.
Complied Standards List
Risk management| EN ISO 14971:2012 / ISO 14971:2007 Medical devices –
Application of risk management to medical devices|
---|---|---
Labeling
| EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices.
Symbols to be used with medical device labels, labelling and information to be supplied. Part 1 : General requirements
User manual| EN 1041:2008 +A1:2013 Information supplied by the
manufacturer of medical devices
General Requirements for Safety| EN 60601-1:2006+A1:2013/ IEC
60601-1:2005+A1:2012 Medical
electrical equipment – Part 1: General requirements for basic safety and essential performance
EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
Electromagnetic compatibility| EN 60601-1-2:2015/ IEC 60601-1-2:2014
Medical electrical equipment – Part 1-2: General requirements for basic safety
and essential performance – Collateral standard: Electromagnetic disturbances
– Requirements and tests
Performance requirements| EN ISO 81060-1:2012 Non-invasive
sphygmomanometers – Part 1: Requirements and test methods for non-automated
measurement type
IEC 80601-2-30:2009+A1:2013 Medical electrical equipment- Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
Clinical investigation
| EN 1060-4:2004 Non-invasive sphygmomanometers – Part 4: Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometers
ISO 81060-2:2018 Non-invasive sphygmomanometers – Part 2: Clinical validation of intermittent automated measurement type
Usability
| EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013
Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
IEC 62366-1:2015 Medical devices – Part 1: Application of usability engineering to medical devices
|
Software life-cycle processes| EN 62304:2006/AC: 2008 / IEC 62304:
2006+A1:2015 Medical
device software – Software life-cycle processes
Bio-compatibility
| ISO 10993-1:2018 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
EMC Guidance
- Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
Emissions test| Compliance
RF emissions CISPR 11|
Group 1
RF emissions CISPR 11
|
Class [ B ]
Harmonic emissions IEC 61000-3-2
|
Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3|
Comply
Table 2
Guidance and manufacturer’s declaration – electromagnetic Immunity
Immunity Test| IEC 60601-1-2
Test level
|
Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2
|
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
|
±8 kV contact
±2 kV, ±4kV, ±8 kV,
±15 kV air
Electrical fast transient/burst IEC 61000-4-4|
±2 kV for power supply lines
±1 kV signal input/output 100 kHz repetition frequency
| ±2 kV for power supply lines
±1 kV signal input/output 100 kHz repetition frequency
Surge IEC61000-4-5| ±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
| ±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
| 0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.0 % UT;
1 cycle and 70 % UT; 25/30 cycles;
Single phase: at 0°.0 % UT; 250/300 cycle
| 0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT; 25/30 cycles; Single phase: at 0°. 0 % UT; 250/300 cycle
Power frequency magnetic field IEC 61000-4-8
| 30 A/m
50Hz/60Hz
| 30 A/m
50Hz/60Hz
Conduced RF IEC61000-4-6
| 3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
| 3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
Radiated RF IEC61000-4-3
| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE UT is the a.c. mains voltage prior to application of the test level.
TABLE 3
| Guidance and manufacturer’s declaration – electromagnetic Immunity
---|---
| Immunity Test| IEC 60601-1-2
Test level
|
Compliance level
|
Electrostatic discharge (ESD) IEC 61000-4-2
|
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
|
±8 kV contact
±2 kV, ±4kV, ±8 kV,
±15 kV air
| Electrical fast transient/burst IEC 61000-4-4|
±2 kV for power supply lines
±1 kV signal input/output 100 kHz repetition frequency
| ±2 kV for power supply lines
±1 kV signal input/output 100 kHz repetition frequency
| Surge IEC61000-4-5| ±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
| ±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
| Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
| 0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.0 % UT;
1 cycle and 70 % UT; 25/30 cycles;
Single phase: at 0°.0 % UT; 250/300 cycle
| 0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT; 25/30 cycles; Single phase: at 0°. 0 % UT; 250/300 cycle
|
Power frequency magnetic field IEC 61000-4-8
| 30 A/m
50Hz/60Hz
| 30 A/m
50Hz/60Hz
|
Conduced RF IEC61000-4-6
| 3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
| 3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
|
Radiated RF IEC61000-4-3
| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
| 10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
| NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic Immunity|
Radiated RF IEC61000-4-3
(Test specifications for ENCLOSURE PORT IMMUNITY to
RF wireless communica- tions equipment)
| Test Frequency (MHz)| Band
(MHz)
| Service| Modulation| Modulation
(W)
| Distance (m)| IMMUNITY TEST LEVEL
(V/m)
|
385| 380-390| TETRA 400| Pulse modulation b) 18Hz| 1.8| 0.3| 27|
450
|
430-470
| GMRS 460
FRS 460
| ,FM c) ± 5kHz deviation 1kHz sine|
2
|
0.3
|
28
|
710| 704-787| LTE Band 13,
17
| Pulse modulation b) 217Hz|
0.2
|
0.3
|
9
|
745|
780|
810| 800-960| GSM 800/900, TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
| Pulse modulation b) 18Hz| 2| 0.3| 28|
870|
930|
1720| 1700-
1990
| GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1,
3,
4,25; UMTS
| Pulse modulation b) 217Hz| 2|
0.3
|
28
|
1845|
1970|
2450| 2400-
2570
| Bluetooth, WLAN, 802.11
b/g/n, RFID 2450, LTE
Band 7
| Pulse modulation 217 Hz| 2| 0.3| 28|
5240| 5100-
5800
| WLAN 802.11
a/n
| Pulse modulation 217 Hz| 0.2| 0.3| 9|
5500|
5785|
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>