CLAS OE 4002 Digital Multimeter Instruction Manual
- May 15, 2024
- CLAS
Table of Contents
CLAS OE 4002 Digital Multimeter
Specifications
- Model: OE 4002 MEGOHMMETRE
- Insulation Test Voltages: 50V, 100V, 250V, 500V, 1000V
- Display Range: 0~5M, 5~50M, 50~100M, 25~250M, 50~500M, 0.5G~2.0M
- Resolution: 0.01M, 0.1M, 1M, 0.1G
- Battery: 4 x 1.5V AA batteries
Product Usage Instructions
Introduction
The OE 4002 Megohmmetre is designed for insulation testing and measurement of various electrical parameters.
Front Panel
The front panel includes an LCD screen and various buttons for operation.
Display Screen
The screen displays insulation test values such as DAR and PI, along with other relevant information.
Buttons
The device features buttons for selecting test modes, functions, and navigating through the menu.
Rotary Selector
The rotary selector allows you to choose between different current measurement modes – DC and AC.
Input Terminal
The input terminal is used for connecting the test leads and is common except for insulation testing.
Functions Overview
The device offers various functions including memory storage, DAR/PI calculation, automatic power-off, and more.
Insulation Testing
For insulation testing, select the appropriate voltage level and measure the insulation resistance as per the display range.
Data Recording in Multimeter Mode
You can record data in multimeter mode for further analysis and reference.
Basic Measurements
- DC Voltage Measurement: Connect the probes and select the DC voltage range for measurement.
- AC Voltage Measurement: Similarly, measure AC voltage by selecting the appropriate range.
- Temperature Measurement: Use the device to measure temperature accurately.
- Resistance and Capacitance Measurement: Measure resistance and capacitance values with ease.
- Continuity and Diode Testing: Verify continuity and test diodes for functionality.
Insulation Measurement Specifications
The insulation test voltage ranges from 50V to 1000V with corresponding display ranges and resolutions for accurate testing.
Battery Replacement
To maintain accurate readings and prevent potential hazards, replace the 4 x 1.5V AA batteries promptly when the battery indicator signals low battery.
Frequently Asked Questions
- Q: Can I perform insulation testing at all voltage levels simultaneously?
- A: No, you need to select the desired test voltage level before conducting an insulation test.
- Q: How do I switch between different measurement modes?
- A: Utilize the rotary selector to switch between DC and AC current measurement modes.
- Q: What should I do if the device shows false readings?
- A: Replace the batteries promptly to avoid false readings that could lead to inaccurate measurements or potential hazards.
“`
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
SECURITY WARNING
This style of digital multimeter is designed and manufactured according to the
safety requirements set out by EN61010-1, EN61010-2-030 standards for
electronic test instruments. Its design and manufacture is strictly based on
the provisions in the 1000V CAT III, 600V CAT IV of IEC61010-1 and the
Stipulation of 2-Pollution Grade.
WARNING
To avoid possible electric shock or personal injury, follow these guidelines:
– Use the Meter only as specified in this manual or the protection provided by
the Meter might be impaired. – Do not use the Meter or test leads if they
appear damaged, or if the Meter is not operating properly. If in doubt, have
the Meter serviced. – Always use the proper terminal, switch position, and
range for measurements before connecting Meter to circuit under test. – Verify
the Meter’s operation by measuring a known voltage. – Do not apply more than
the rated voltage as marked on the Meter, between the terminals or between any
terminal and earth ground. – Use caution with voltages above 30 V ac rms, 42 V
ac peak, or 60 V dc. These voltages pose a shock hazard. – Replace the battery
as soon as the low battery indicator ( ) appears. – Disconnect circuit power
and discharge all high-voltage capacitors before testing resistance,
continuity, diodes, or capacitance. – Do not use the Meter around explosive
gas or vapor. – When using the test leads, keep your fingers behind the finger
guards. – Remove test leads from the Meter before opening the Meter case or
battery door. – Never operate the Meter with the cover removed or the battery
door open. – Use proper protective equipment, as required by local or national
authorities when working in hazardous areas. – Use only the replacement fuse
specified or the protection may be impaired. – Use only the replacement fuse
specified or the protection may be impaired. – If the meter is dirty after
usage, it is advised to clean it by using a humid cloth and mild house hold
detergents. – Never use acid detergent or dissolvants.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
WARRANTY
The meter is warranted to be free from defects in material and workmanship
under normal use and service. The warranty period is one year and begins on
the date of shipment. Parts, product repairs, and services are warranted for
18 months except for misused, altered, neglected, contaminated, or damaged by
accident or abnormal conditions of operation or handling.This warranty does
not apply to fuses, disposable batteries.
SYMBOLS
Dangerous Voltage Ground Warning see explain in manual Double insulation Fuse
INTRODUCTION
This meter is a Digital Insulation Multimeter with broad range of measurement,
which can be used to measure DC voltage, AC voltagem, current, AC current,
resistance, capacitance, frequency and temperature.
FRONT PANEL
Ecran Bouton Sélecteur rotatif Terminaux d’entrée
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
DISPLAY SCREEN
Indicator
Description Low battery Indicates when it is time to replace the battery. To
avoid false readings, which could lead to possible electric shock or personal
injury, replace the battery as soon as the low battery indicator appears.
Indicates a test lock will be applied the next time you press Test Button on
the meter, the test lock acts to hold down the button until you press Test
Button again. Hold function, the meter do not update display.
In multimeters function,Indicating the meter working in autorange
function,else the meter work in manual range mode.
Filter: when the rotary switch at ~V Position, the meter active 1KHz low
passed filter; when the rotary switch at Insultation position, the meter
active Smoothing function. Relatvive function.
In insulation test mode, display the DAR value.
In insulation test mode, display the PI value.
Minus, When the measure value lower than 0, display the sign.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
Indicator
Description Greater symbols, In insulation test mode, indicates the measure
value overflow. Unsafe voltage warning, In insulation mode, indicates greater
20V voltage is detected on the input terminals. Indicates the meter working in
Record mode, in this mode,meter can record the last 100 measure value. Display
the Max, Min Avg value.
In Record mode, indicates the counts of the recorded value.
Display the sequence number of the recorded value.
Continuity test function is selected.
Diode test function is selected
DC Voltage or DC Current test function is selected.
AC Voltage or AC Current test function is selected.
Measurements Units.
Insulation test Indicator.when the rotary switch at insulation position, this
sign appers, when the test voltage is present ,the sign alternate on or off.
Source voltage rating for insulation test.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
Indicator
Description The measuring uint of source voltage rating for insulation test.
Insulation Measuring time Unit.
Auxiliary Display :display the output voltage, time.
Primary Display: display the measure value.
DISPLAY MESSAGES
Indicator
Description Appears on Auxiliary display, Indicates the battery too low to
perform Insulation test.
The auto power off function is not actived.
In insulation mode,indicates meter have detected the voltage on the input
terminals.
In insulation mode,indicates the meter performs the auto discharge function;
not to touch any input terminals in this mode.
The Timer function is not actived.
Check the test leads in proper terminals,the rotary switch at current or
insulation position, display the message.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
BUTTONS
Bouton
Description – When rotary switch at the position except insulation, press the
button, the display not updated. – When rotary switch at the insulation
position, press the button toggle the test lock. – When rotary switch at the
insulation position, and the meter in set measure time mode, press the button
to save measure time and exit set-measure time mode.
– Shift:press this button to active the function higher rotary switch. – In
Record mode,function as page up. – In Set-Measureing time mode, Increase the
assign value.
– When rotary switch at the position except insulation:press this button ,the
meter will switch auto range to manual range. Press this button longer than
1S,the meter will switch manual range to autorange. – When rotary switch at
the insulation position, select a source voltage for Test.
– Actived the record function. – In Record mode,function as page down –
Actived the time function (When rotary switch at the insulation position)
– Actived relative function – Active the frequency measure function when
rotary switch at AC voltage;press the button longer than 1S to perform the
frequency measurement
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
Bouton
Display DAR or PI Value.
Description
Back Light on/Off,when the light turn on ,after 10S the meter auto turn off light.
When rotary switch at the insulation position, press the button to perform insulation test.
ROTARY SWITCH
Position
Turn off themeter power.
Fonction
DC Voltage 0.1mV~1000V (Note : mV Range only exist in manual range)
1: AC Voltage 30mV~750V(note:mV Range only exist in manual range)
2: Actived 1KHz low passed filter for AC voltage do not use the Low-Pass
Filter function to verify the presence of hazardous voltages)
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
Position
Fonction Celsius is the default temperature measurement unit press shift
button switch to Fahrenheit measurement unit Temperature from-30°C to1300°C
(-22°F~2372°F)
Ohms : 0.1~60M Capacitance: 0.01nF~60mF
Continuity Diode
DC Current (0.01mA~400mA) AC Current (3.00mA~400mA)
1 : Insulation Test 0.01M~2.0G, Test output Voltage 50V (default)100V 250V,
500V, 1000V, the test output voltage have selected will be saved. 2 : Digit
filter function for Insulation test.
INPUT TERMINAL
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
Borne
Description Input positive terminal except current and inuslation measure.
Common terminal except Inuslation measure.
Input positive terminal for current measure,do not measure any current exceed
400mA (TRMS).
Input positive terminal for Insulation measure.
NOTE 1
To remind you to check that the test leads are in the correct terminals, LEAd
is momentarily displayed when you move the rotary switch to or Insulation
Position
WARNING
To avoid a blown fuse, damage to the Meter,or serious personal injury, never
attempt to make a measurement with a test lead in an incorrect terminal.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
FUNCTION DESCRIPTION
POWER-UP OPTION
Holding a button down while turning the Meter on activates a power-up option.
Power-up options allow you to use additional features and functions of the
Meter. To select a power-up option, hold down the appropriate button indicated
while turning the Meter from OFF to any switch position. Power-up options are
cancelled when the Meter is turned OFF.
Buttons SHIFT
Function Turns on all LCD segments.
MEM
Disables automatic power-off function Display shows PoFF until the button is released.
DAR/PI
Initiates a fully loaded battery test and displays the charge level of the battery until the button is released.
POWER-OFF OPTION
The Meter have automatic power off function (Sleep mode) to conserve battery
power .if there is no function change or button press for 10 minutes. The
Meter comes out of Sleep mode when a key is pressed or when the rotary
switched is changed.
To disable the Sleep mode, hold down MEM button while turning the Meter on.
Sleep mode is always disabled in the recording mode, insulation test active,
or if the auto power off feature has been disabled by pressing MEM button when
the Meter is turned on.
FULLY-LOADED BATTERY TEST
Hold down DAR/PI button while turning the Meter on,the meter display the
battery voltage. Put up the DAR/PI button to exit battery voltage test.
FUNCTION DESCRIPTION : ROTARY SWITCH AT THE POSITION EXCEPT INSULATION
Press Hold button to freeze the displayed value. Press again to release the
display.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
RELATIVE MEASUREMENT
Show the difference between actual value and the relative base. Press REL
Button to enter relative measurement and the meter will record the initial
value when pressing the key. Displayed value = Actual value -Initial value
Press REL Button again to exit relative measurement.
MANUAL RANGING AND AUTORANGING
To enter the Manual Range mode, press Range button and Auto is hide, In the
Manual Range mode, press Range button to increment the range. After the
highest range, the Meter wraps to the lowest range. To exit Manual Range,
press Range button for one second or turn the rotary switch. The Meter returns
to Autorange and Auto is displayed.
DAR AND PI
Sometimes an insulation part with obvious drawbacks (e.g., the insulation part
is broken through under high voltage) is nevertheless with a good absorption
ratio (or polarization index). Therefore, absorption ratio (polarization
index) cannot be used to discover local insulation drawbacks other than
dampness and contamination.
DAR (absorbing ratio) =
DAR (polarization index) =
R10Min= Resistance value measured 10 minutes after applying the test voltage;
R1Min=R60Sec= Resistance value measured 10 minutes after applying the test
voltage; R15Sec= Resistance value measured 10 minutes after applying the test
voltage After performed insulation test, press DAR/PI button, meter display
DAR; press DAR/PI button again, meter display PI. If DAR or PIValue invalid,
themeter display – – – – .
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
INSULATION TEST LOCK
In insulation test mode, press Test button to perform insulation test until
the button is released.when the button is released,the screen display hold
sign. Press Lock Button ,then the screen display Lock sign, press Test Button
,the meter will perform insulation test unitl you press Test button again;The
test lock will unlocked while to cancel insulation test.
TIMER FUNCTION FOR INSULATION
Timer function is only valid in insulation test. In Init-Insulation status
(Init-Insulation status is the rotary switch to Insulation position or have
performed insulation test), press MEM button longer than 2S, the secondary
screen display the remain key time,and primary screen display the `CLOC’.
When the remain key time is zero(as follow figure), Release MEM button
The meter enter timer set-up status, the secondary screen display the preset
time
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
Press MEM button to change positon, press shift button to change value
Press HOLD button will save measuring time,and exit the assign status.
When the measureing time is greater than zero and the test lock is
unlocked,meter will activate the timer function, the meter will be
automatically stop the test when the time is longer than preset time.
NOTE
Press Test button and turn on the meter, the meter display
and the timer function will be invalid.
When the timer function is active,secondary screen display the output volatage,and alternately turn on or off ‘ SEC’
Press shift button ,the secondary screen display the measuring time and alternately turn on or off ‘VDC’ Press test button to stop insulation test.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
DATA RECORDING IN MULTIMETER MODE
Press MEM button to enter recording status, the meter display `MEM’ as
follows:In recording status,meter record the measuring value (when the record
exceed 100 counts, the meter only record the last 100 counts).
In recording status,press MEM button in succession, the meter will display the
Max value,Min value, and Avg value.
In recording status (or when the meter display Max Min value ),press MEM
button longer than 1S,the meter will exit recording status. When meter display
avg value as follows:
Press MEM button longer than 1S,the secondary screen display the remain key
time.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
When the remain key time is zero ,the secondary screen display the counts of
the recorded value as follows
Press MEM button ,the secondary screen display the sequence number of the
recorded value and the Primary display corresponding value.
Press MEM button to page down,and press Shift button to page up; press MEM
button longer than 1S,then the secondary screen display the key remain
time.when the key remain time is zero,the meter will be back to Record status.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
When the key remain time is zero,the meter will be back to Record status.
Data Recording in insulation test mode In insulation test mode, the recording
function will be activated, press MEM button to view the recorded value, the
detailed operation is the same as data-recording in multimeter mode.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
MAKING BASIC MEASUREMENTS
MEASURING DC VOLTAGE
Switch rotary switch to
Position, Input terminals and test leads connecting as follows figure, then connect test
leads to circuit.
mV Range in AutoRangeing is Invalid.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
MEASURING AC VOLTAGE
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
True RMS Meters accurately measure distorted waveforms, but when the input
leads are shorted together in the AC functions, the Meter displays a residual
reading between 1 and 30 counts. When the test leads are open, the display
readings may fluctuate due to interference. These offset readings are normal.
They do not affect the Meter’s ac measurement accuracy over the specified
measurement ranges. In AC Voltage measuring Mode to actived Low passed filter,
the signal diverts through a filter that blocks unwanted frequencies above 1K
Hz. To actived this funtion, press Shift button,the screen display , to cancel
low passed filter function, press Shift button again ; To avoid possible
electric shock or personalinjury, do not use the Low-Pass Filter function to
verify the presence of hazardous voltages. Voltages greater than what is
indicated may be present. First, make a voltage measurement without the filter
to detect the possible presence of hazardous voltage. Then, select the filter
function.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
MEASURING TEMPERATURE
The Meter measures the temperature of a type-K thermocouple (included). Choose
between degrees Celsius (°C) or degrees Fahrenheit (°F) by pressing Shift
Button. When the type-K thermocouple not connect to meter, the meter display
the ambient temperature the Meter is rated for -30°C to 1300°C, the included
K-type thermocouple is rated for 260°C. For temperatures out of that range,
use a higher rated thermocouple. To avoid risk of shock, do not connect
thermocouple to electrically live circuits.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
MEASURING RESISTANCE AND MEASURING CAPACITANCE
To avoid possible damage to the Meter or to the equipment under test,
disconnect circuit power and discharge all high voltage capacitors before
testing for continuity. Press Shift button switch between Measuring Resistance
and Measuring Capacitance
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
MEASURING CONTINUITY AND MEASURING DIODE
The continuity test and diode test features a beeper that sounds as long as a
circuit is complete. The beeper sounds when a short (<30 )or forward voltage
lower than 300mV. Press Shift button switch between Measuring Continuity and
Measuring Diode.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
MEASURING AC OR DC CURRENT
Check fuse is good before Test; Select proper input terminal,rotary
switch,range ; not to measuring the current exceed the meter current rating
(400mA) Turn power OFF to the circuit under test, break circuit, insert Meter
in series,then turn power on Press Shift button switch between Measuring DC
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
MEASURING FREQUENCY
In Measuring AC Voltage mode, Press Hz button Longer than 1S, the meter switch
measuring voltage to frequency In Frequency mode, the Range button is invalid.
press Hz button, the meter will switch to measuringAC Voltage.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
INSULATION TEST
Insulation tests should only be performed on dead circuits. Check the fuse and test leads before testing.
Switch rotary to Insulation position. if meter displays , please replease battery.
Insert test leads to High/Low terminals. if the meter display Live and , indicator the meter cannot measure on live circuit.please power off Live circuit..
Press Range button to select output voltage;
Press test button to perform insulation test, when the source voltage outputed the screen display .
In insulation measuring ,the screen alternate turn
on or off, the primary screen display the resistance value,
and the secondary screen display the Output voltage.Release the test button then discharges through the
Meter, the meter display DISC.
The secondary Display 0 VDC indicator when the voltage discharge finished.
Disconnect the test leads from circuit..
Active smoothing funciton by press Shift button ,the meter display
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
GENERAL SPECIFICATIONS
Complies with IEC/ EN 61010-1 1000V CAT III, 600 V CAT IV. 1000V DC Voltage, 750VAC Voltage (TRMS). 6600 Count (except for capacitance, temperature,insulation). When the input terminals is V and COM, The overload protective voltage is up to AC 250 V (TRMS) ; When the input terminals is mA and COM, protective Current is 0.4A; when the input terminals is Hi and Lo, The overload protective voltage is 600V. Batteries: Four AA batteries (NEDA 15A or IEC LR6) Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 M with a duty cycle of 5 seconds on and 25 seconds off. Insulation Measurement Range: 0.01 M to 2 G. Insulation Test Voltages: 50V, 100V, 250V, 500V, 1000V. Insulation Source Voltage: + 20%, – 0%. Insulation Short-Circuit Test Current : 1 .5mA nominal. Insulation Auto Discharge : Discharge time <1 second for C = 1F or less. Insulation Live Circuit Detection : Inhibit test if terminal voltage > 20V prior to initialization of test. Insulation Maximum Capacitive Load: Operable with up to 1F load. Storage Temperature : -40 °C to 60 °C. Operating Temperature: 0°C to 40 °C . Storage Altitude: 12000m. Operating Altitude: 2000m 1000V CAT III, 3000m 1000V CAT II. Temperature Coefficient: 0.05 × (specified accuracy) per °C for temperatures < 18 °C or > 28 °C. Relative Humidity: 40%~75%. Size: 205(L)×102(W)× 58(H)mm. Weight: approx 390g.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
ACCURACY
DC VOLTAGE
Range
660mV 6.6V 66V 660V 1000V
Resolution
0.1mV 1mV 10mV 100mV 1V
Accuracy 0.1mV
±(0.5%+5) ±(0.5%+5) ±(0.5%+5) ±(0.5%+5)
CAPACITANCE
Range
660mV 6.6V 66V 660V 750V
Resolution
0.1mV 1mV 10mV 100mV 1V
Accuracy – – – – – –
±(1.5%+30) ±(1.5%+30) ±(1.5%+30) ±(1.5%+30)
TEMPERATURE
Range
Resolution
-30°C~1300°C
1°C
-22°F~2372°F
1°F
Accuracy ±(1.0%+2) ±(1.0%+4)
Accuracies apply following 90 minutes settling time after a change in the ambient temperature of the instrument.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
RESISTANCE
Range
660 6.6K 66K 660K 6.6M 66M
CAPACITANCE
Range
66nF 660nF 6.6F 66F 660F 6.6mF
66mF
Resolution
0.1 1 10 100 1K 10K
Resolution
10pF 0.1nF 1nF 10nF 0.1F 1F 10F
DC CURRENT
Range
66mA 400mA
Resolution
0.01mA 0.1mA
AC CURRENT
Range
66mA 400mA
Resolution
0.01mA 0.1mA
Accuracy ±(1.2%+5) ±(1.2%+5) ±(1.2%+5) ±(1.2%+5) ±(2.0%+20) ±(2.0%+20)
Accuracy ±(5.0%+20) ±(5.0%+20) ±(5.0%+20) ±(5.0%+20) ±(5.0%+20) ±(5.0%+20)
±(5.0%+20)
Accuracy ±(1%+5) ±(1%+5)
Accuracy ±(1.5%+30) ±(1.5%+30)
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
FREQUENCY
Range 660.0Hz 6.600Hz 66.00Hz >10kHz
Resolution
0.1Hz 1mV 10mV
Accuracy ±(1.5%+5) ±(1.5%+5) ±(1.5%+5) – – – – – – –
INSULATION
Output Voltage 50V(0~20%)
100V(0~20%)
250V(0~20%) 500V(0~20%)
1000V(0~20%)
Display Range 0~5M 5~50M 0~5M 5~50M
50~100M 0~25M
25~250M 0~50M
50~500M 0~5M 500M
0.5G~2.0M
Resolution 0.01M 0.1M 0.01M 0.1M 1M 0.1M 1M 0.1M 1M 0.1M 1M 0.1G
Accuracy ±(3%+5) ±(3%+5) ±(3%+5) ±(3%+5) ±(3%+5) ±(3%+5) ±(3%+5) ±(3%+5) ±(3%+5) ±(3%+5) ±(3%+5) ±(3%+5)
REPLACING BATTERY AND FUSE
Only qualified service personnel are required to repair it.
To avoid false readings, which could lead to possible electric shock or
personal injury, replace the batteries (4 x 1.5V AA batteries) as soon as the
battery indicator appears.
Use only fuses with the amperage interrupt voltage and speed ratings specified
(F 400mA, 1000V).
Turn the rotary switch to OFF and remove the test leads from the terminals.
OE 4002 MEGOHMMETER 1000V CAT. III & 600V CAT. IV
ACCESSORIES
Item
Quantity
Test Leads
2
Clips
2
K Type Thermocouple
1
Battery AA LR6
4
Manual
1
Multi-Function Socket
1
CLAS Equipements 83 chemin de la CROUZA
73800 CHIGNIN FRANCE
Tél. +33 (0)4 79 72 62 22 Fax. +33 (0)4 79 72 52 86
OE 4002
MEGOHMMETRE 1000V CAT. III & 600V CAT. IV
MEGOHMMETER 1000V CAT. III & 600V CAT. IV
Si vous avez besoin de composants ou de pièces, contactez le revendeur En cas
de problème veuillez contacter le technicien de votre distributeur agréé
If you need components or parts, please contact the reseller. In case of
problems, please contact your authorized technician.