AGILTRON FSCO-500 Series Bench Top Fiber Microscope User Guide

June 13, 2024
AGILTRON

AGILTRON FSCO-500 Series Bench Top Fiber Microscope

AGILTRON FSCO-500 Series Bench Top Fiber Microscope

Thank you

for choosing Eternal Science, please kindly read this manual before use

This manual describes how to use the Microscope (FSCO-500 Series) concise information.

PRODUCT OVERVIEW

Note : Please note that above picture is just the product sketch map, if it doesn’t conform with the real device, please refer the real one.
Product Overview

USER GUIDE

ADAPTOR INSTALLATION

There’re three ways for adaptor installation, please select the corresponding method. For any MF type microscope, please install the MF-L adaptor or MF Cushion block to use firstly. Detail image please refer to adaptors selection.
Adaptor Installation

QUICK START

❶ Connect power port;
❷Turn on the device and make sure a blue light can be seen from adaptor center;
❸ Insert the connector and end face can be seen after device adjusts focal length automatically.

TWO IMAGE SWITCH CHECK

One time focus and touch key to switch to check two images(show as below left image), large rate checks the detail scratch and small rate checks the whole fiber edge.
Two Image Switch Check

BRIGHTNESS CONTROL

The knob on the right side of the scope can adjust the brightness of the light source, which is convenient to observe the details of the end face. It’s only suit for Model A.

CONNECT SECONDARY DISPLAY

Connect the scope with secondary display with the BNC cable, detail refer to the end of user manual.

TWO IMAGE SYNCHRONIZATION CHECK

One time focus to check two images same time, the primary display checks the large magnification and secondary display shows the small magnification.
Wo Image Synchronization Check

EXTENSION CONNECTION (OPTIONAL)

Insert the probe into the Probe in port, press JW1/JW2 same time for 3s to switch the probe view mode, and press again to return. It’s only suit for Model A.
Extension Connection \(Optional\)

USB TO CONNECT PC (OPTIONAL)

End face can be analyzed by company special pass/fail software with USB port⑯ connection to PC.
The software can set IEC/IPC criteria or customize by user.
USB To Connect Pc \(Optional\)

DISPLAY CONTROL

Adjustment key(Primary Display)| Adjustment key(Display with SD slot)| Adjustment key(Secondary Display)
---|---|---
Brightness Control Press JW5 key to enter brightness control,JW3 to reduce brightness, JW4 to increase brightness.| Brightness Control Press one time JW3 key to enter brightness control,JW4 to reduce and JW5 to increase the value.| Brightness Control Press twice Menu key to enter brightness control, VOL- to reduce and VOL+ to increase the value.
Contrast Control Press JW5 key to enter contrast control,JW3 to reduce contrast, JW4 to increase contrast.| Contrast Control Press twice JW3 key to enter contrast control,JW4 to reduce and JW5 to increase the value.| Contrast Control Press one time Menu key to enter contrast control, VOL- to reduce and VOL+ to increase the value.
Sharpness Control Press JW5 key to enter sharpness control,JW3 to reduce sharpness, JW4 to increase sharpness.| Sharpness Control Press three times JW3 key to enter sharpness control,JW4 to reduce and JW5 to increase the value.| Sharpness Control Press four times Menu key to enter sharpness control, VOL- to reduce and VOL+ to increase the value.
Reset Press JW5 key 4 times to select reset, then press JW4 to confirm reset. The above three parameters are all 50 by default| Capture : Press JW6 key to save image data to SD card.
View : Press JW4/JW5 key to view the image forward or backward. Delete : Press JW6 to delete the current image and JW3 to back.|

Note : the function keys are referenced as table, please note that if it doesn’t conform with the real device, please refer to the real one.

PACKAGE DETAILS

FSCO500 Bench Microscope 1 pcs
User Manual 1 pcs
Power Adaptor 1 pcs

PACKAGE,TRANSPORT& STORAGE

☑ Package
Special foam and carton
☑ Transport & Storage
Avoid collision, rain and snow during transportation, do not store with chemical and wet articles.
☑ Environmental Certification
Products components have all been approved by most stringent international RoHS Certification.

OTHER OPERATION NOTICE

☑ When adaptor can not position the fiber firmly, replacement is needed. For a long use of adaptor can cause certain amount of wearing.
☑ Please pay attention to the cleaning and sanitation of lens, reduce the pollution of dust and fiber residue, and prolong the service life.

AFETER-SALES SERVICE AND PROTECTION

☑ In the guarantee of product quality, a free maintenance service will be provided for any damage caused by the manufacturing process or original components within one year from the date of purchase.
☑ Any man-made damage will not be covered by the free warranty and replacement of materials and accessories will be charged. But the company provide long- life maintenance service for any problem of the product.
☑ For any problem please send us [email protected] for more information.

APPENDIX 1. SECONDARY DISPLAY INSTALLATION GUIDE

①Side mount base
②BNC Video cable
③Power patch cord
④Secondary display
⑤Secondary display card seat
⑥Hand screw
⑦Connecting rod
Secondary Display Installation Guide

Note : Please note that above picture is just the product sketch map, if it doesn’t conform with the real device, please refer the real one.

Horizontal installation

  • Put the connecting rod ⑦ horizontally and screw the side mount base ① corresponding to the left hole of connecting rod 7 with the hand screw 6.
  • Stand up the secondary display card seat⑤ and screw the right side corresponding to the right hole of connecting rod ⑦ with the hand screw 6.
  • Align the back slot of secondary display④ to the card seat⑤ and position it from top to bottom to complete the installation.

Vertical Installation

  • Stand up the connecting rod⑦ and screw the middle of side mount base① with left hole of connecting rod⑦ with the hand screw 6.
  • Stand up the secondary display card seat⑤ and screw the right side corresponding to the right hole of connecting rod⑦ with the hand screw 6.
  • Align the back slot of secondary display④ to the card seat⑤ and position it from top to bottom to complete the installation.

Connection

  • Connect the secondary display with BNC cable②
  • Connect the secondary display with the power patch cord.

APPENDIX 2. AUTO FOCUS USE GUIDE

  1. Connect foot pedal
    As shown in the figure on the right, connect the foot pedal to the interface on the back of the microscope and tighten the nut, insert the fiber and step on the foot pedal to complete focusing.

  2. Adaptor calibration
    When checking different kinds of fiber end face, we need to change different positioning adaptor. At this time, make sure the blue light is emitted from the center of the fixture, insert the connector, and press the JW2 button to automatically find the focus. Then you can see a clear picture.

  3. Single/Multiple fiber mode switch
    The system only remembers the last calibration in both modes. If a single- fiber adaptor and a multi-fiber adaptor needs to be switched frequently, then conversion can be completed only in steps 1 and step 2.
    Single/Multiple fiber mode switch

  4. Other operations
    ❶ Step on the foot pedal to do focus in small adjustments or short click JW1 to same job; Short click JW2 to refocus for large adjustments..
    ❷ Long press JW1 as manual positive forward focus; long press JW2 for manual reverse backward focus.
    ❸ Click JW1 and JW2 same time together to switch the brightness.

APPENDIX 3. COMMON TROUBLE SHOOTING

Exception Causes—>Solutions
Black screen after insertion Without aligning adaptor center hole to the lens

inside device—> Adjust X/Y axis to make sure blue light shoots out of center of adaptor.
Black screen after insertion_two| Mis-operate to close the display—>Press Key 6 in the product image to restart the display or re-start the device.
No end face after focus| Clamp can not position fiber connector due to long wear—> Replace clamp.
Blurred image after the fiber insertion(Model B)| Step on the foot predal to do focus in small adjustments or short click JW1 to same job; Short click JW2 to refocus for large adjustments.
Continuous long focus time or unclear focus(Model B)| Adaptor replacement or check different fibers—> Follow adaptor calibration steps.

APPENDIX 4. OPTIONAL ADAPTERS

Optional Adapters

APPENDIX 5. PC/APC TWO-IN-ONE ADAPTER USE GUIDE

  1. Convert from PC to APC
    ① Loosen the lock nut counterclockwise, ② press down the bottom of the platen, insert the APC jumper, the core should not been seen , clockwise twist ③ Y axis, the core will move up in the screen until the black core totally shows. ④ Finally tighten the lock nut to complete the switch.

  2. Convert from APC to PC
    ① Loosen the lock nut counterclockwise, ② press down the bottom of the platen, insert the PC jumper, the core should not been seen , counterclockwise twist ③ Y axis, the core will move down in the screen until the black core totally shows. ④ Finally tighten the lock nut to complete the switch.
    PC/apc Two-in-one Adapter Use Guide