AGILTRON SUVC 4 Port Solid State UV Spot Light User Manual
- June 10, 2024
- AGILTRON
Table of Contents
AGILTRON SUVC 4 Port Solid State UV Spot Light User Manual
Warning:
- UV is harmful to the eye. A shield should be used when operating the tool.
- High-intensity light can ignite a fire. Do not place combusting materials at the light focus spot.
Introduction
The 4-port Solid-State UV Light uses high-power UV LEDs as light sources for
curing UV adhesives. This superior optical design achieves high light coupling
efficiency exceeding 85% and uniform light spot intensity. Each port can be
switched on/off either individually by the button on each head or
simultaneously by an optional foot switch. The four ports are controlled by a
globally same setting of curing time as well as UV output power, values of a
timer and power shown on the front panel are as same settings exerted to every
four ports. Each LED head features an integrated cooling fan.
Figure 1: Overall view of the 4-Port UV lights device
Specifications
Parameter | Value | Units |
---|---|---|
Wavelength | 275 / 365±5 / 385±5 / 405 / 440 | nm |
UV radiant flux | 0 ~1000 | mW |
UV irradiance | 0 ~ 4* | W/cm2 |
UVLED driving current | 0 ~ 1100 | mA |
Curing time setting | 1 second ~ 18 hours | |
UV spot size | Φ5 | mm |
Working distance | 20 | mm |
Cooling method | forced air, fan | |
Operation life | 40,000 | hours |
- Average intensity over the spot size Φ5mm at working distance.
Power Supply
The device requires a 100V~240V AC supply. The power on-off rocker switch is embedded in the AC receptacle that is located on the rear panel.
Functional Descriptions
LCD and its display content on the front pane
Power: UV power setting in percentage 0% ~ 100%.
Timer: Curing time setting to maximum 18H:00M:00S
Rubber buttons on the front panel
POWER: Enters the UV power setting mode
TIMER: Enters the curing time setting mode.
UP: Modify the setting upwards
DOWN: Modify the setting downwards.
Green LED on the front panel
①②③④: Indicates on/off status of each UV port
Ports on the front panel
Four 4-Pin Sockets: Connectors for UV heads
Ports on the rear panel
Synchronizing Foot Switch:
Forces all ports to be synchronously turned on or off.
- USB: Not available, preserved for future.
- Virtual COM Port: Connects this device to a computer with a USB type-B cable, and then the drive can be controlled by the computer via a Virtual COM Port (VCP).
- AC Input: AC power entry and on-off power switch.
Figure 2: Rear panel
Operational Descriptions
Set the UV power
Push the POWER button to let the Power display blink, then the device enters
power setting mode, push UP/DOWN to modify the power setting, setting range is
0% ~ 100% with increasing /decreasing step by 1%, the percentage 100%
corresponds to the maximum UV radiant flux. This global power setting exerts
its same value on every port. If no push happens on any button, the device
automatically exits the power setting mode after 4 seconds.
Set the UV curing time
Push the TIMER button to let the Timer display blinking, then the device
enters curing time setting mode, push UP/DOWN to modify the curring time
setting, setting range is 1 second ~ 18 hours, increasing /decreasing step is
1 of the highest significant digit. This global curring time setting exerts
its same value on every port.
If no push happens on any button, the device automatically exits the curring
time setting mode after 4 seconds.
Independent On/Off mode
The device operates in independent mode when the device is not getting in
synchronous On mode. In independent mode, the display of curring time does not
change; however, if one or more heads are lightening, then the ‘S’ symbol
blinks simultaneously, indicating the curring time of one or more ports is
counted down in the background
Synchronous On/Off mode
Insert the connector of an Agiltron foot switch into the socket of
Synchronizing Foot Switch on the rear panel, pressing the foot switch. Then
all lightening UV heads are immediately turned off. On the contrary, if no UV
head is lightening, pressing the foot switch, then all UV heads are
immediately turned on, and the device enters the synchronous On mode meanwhile
the LCD the curring time in countdown mode.
VCP Remote Control
VCP Settings
Baud rate: 9600 bps, data length: 8 bits, parity: none, stop bit length: 1
bit.
VCP Command Sets
CMD: C0 00 //Check connection status
Echo: C0 01 //UV box is online
CMD: C1 00 //Enable remote control
Echo: C1 01 //Acknowledged
CMD: C2 00 //Disable remote control
Echo: C2 01 //Acknowledged
CMD: C3 0x //Turn on UVx, x: 1, 2, 3, 4
Echo: C3 0x //Acknowledged
C3 00 //Denied
CMD: C3 FF //Turn on all UVs
Echo: C3 FF //Acknowledged
C3 00 //Denied
CMD: C4 0x //Turn off UVx, x: 1, 2, 3, 4
Echo: C4 0x //Acknowledged
C4 00 //Denied
CMD: C4 FF //Turn off all UVs
Echo: C4 FF //Acknowledged
C4 00 //Denied
CMD: C5 xx //Set power percent, Hex data in 01H-64H
Echo: C5 xx //Acknowledged
C5 00 //Denied