PPI Clavex GD Vertical Autoclave Controller with Graphic Display User Manual
- June 11, 2024
- PPI
Table of Contents
PPI Clavex GD Vertical Autoclave Controller with Graphic Display
FRONT PANEL LAYOUT
Front Panel
Keys Operation
Symbol | Key | Function |
---|---|---|
**** | Scroll | Press to scroll through various Process Information Screens in |
Normal Operation Mode.
| Cycle Start & Alarm Acknowledge| Press to Start / Abort Autoclave
Cycle. In case of process under alarm condition, press to acknowledge and mute
the alarm output.
| DOWN| Press to decrease the parameter value. Pressing once decreases
the value by one count; keeping pressed speeds up the change.
| UP| Press to increase the parameter value. Pressing once increases the
value by one count; keeping pressed speeds up the change.
| SET-UP| Press to enter or exit set-up mode.
****| ENTER| Press to store the set parameter value and to scroll to the
next parameter.
PV Error Indications
Message | Error Type | Cause |
---|---|---|
**OPEN** | Sensor Open | RTD Pt100 / Thermocouple is Broken / Open |
**OVR** | Over-range | Temperature above Max. Specified Range |
**UNR** | Under-range | Temperature below Min. Specified Range |
ELECTRICAL CONNECTIONS
Analog Input
- AI-1: RTD Pt100 Sensor
- AI-2: Pressure Sensor
Digital Output
- DO-1: Heater Control
- DO-2: Air Valve Control
- DO-3: Event Alarm
- DO-4: Unused
- DO-5: Unused
- DO-6: Unused
Digital Input
- DI-1: Cycle Start
- DI-2: Door Open
- DI-3: Water Level
- DI-4: Alarm ACK
- DI-5: Unused
- DI-6: Unused
Operation Instruction
OPERATOR PARAMETERS
Parameter| Settings (Default Value)
Ster. Temp SP >>| Setpoint LO Limit to Setpoint HI Limit
(Default : 121.0 °C)
Air Outlet SP >>| Setpoint LO Limit to Setpoint HI Limit
(Default : 100.0 °C)
Ster. Time (HH:MM) >>| 0.01 to 99.59 (HH:MM) (Default : 00.10 min)
Change Password >>| 1 to 100
(Default : 0)
SUPERVISORY SERIAL PARAMETERS
Parameter| Settings (Default Value)
Device Id >>| 1 to 127
(Default : 1)
Baud Rate >>| 2400, 4800, 9600
19200, 38400, 57600
(Default : 9600)
Parity >>| None Even Odd
(Default : Even)
Serial Write >>| No Yes
(Default : No)
SUPERVISORY CONTROL
Parameter| Settings (Default Value)
Setpoint LO Limit >>| 0 to Setpoint HI Limit (Default : 90 °C)
Setpoint HI Limit >>| Setpoint LO Limit to 600.0 °C
(Default : 150.0 °C)
Overshoot Temp >>| Sterilization SP to Control SP HI
(Default :131 °C)
Cycle End Strategy >>| Immediate Stop Timer Based Stop Temp Based Stop
(Default : Immediate Stop)
Cycle End Timer >>| 1 to 60 Minutes (Default : 5 Min)
Cycle End Temp >>| Control SP LO to Sterilization SP (Default : 100.0 °C)
Ctrl Band >>| 0.1 to 99.9
(Default : 0.2 °C)
EOP Alarm Time (sec) >>| 0 to 250 Sec. (Default : 10 Sec)
SUPERVISORY STERILIZATION
Parameter| Settings (Default Value)
Steri. Stop Temp >>| Air Valve SP to Sterilization SP (Default : 120.5
°C)
Steri. Reset Temp >>| Air Valve SP to Sterilization SP
(Default : 120.0 °C)
SUPERVISORY PV OFFSET
Parameter| Settings (Default Value)
Temp Zero Offset >>| -50.0 to 50.0
(Default : 0)
Pressure Offset >>| -500 to 500 Counts
(Default : 0.0)
SUPERVISORY > PASSWORD
Parameters| Settings (Default Value)
Change Password >>| 1000 to 1999
(Default : 123)
SUPERVISORY > EXIT
Parameters| Settings (Default Value)
Exit Setup Mode >>| Yes No
(Default : No)
FACTORY TEMPERATURE INPUT
Parameter| Settings (Default Value)
Filter Constant >>| 0.5 to 60.0 Sec. (Default : 0.5 Sec.)
Temperature SP Offset >>| 0.0 to 10.0 ºC (Default : 0.0 ºC)
FACTORY > PRESSURE INPUT
Parameters| Settings (Default Value)
Pressure Indication >>| None Computed Measured
(Default : None)
Units >>| PSI KG/CM2 BAR EU
(Default : PSI)
Resolution >>| 1
0.1
0.01
0.001
(Default : 0.1)
Parameter | Settings (Default Value) |
---|---|
Input Type >> | 0-20mA |
4-20mA
0-5V
0-10V
1-5V
(Default : 0-5V)
Signal LO >>| ****
Default : Signal Low 0.000
Signal High 5.000
Signal HI >>
Range LO >>| 0 to 9999 Counts
(Default : 0.0)
Range HI >>| 0 to 9999 Counts
(Default : 100.0)
FACTORY DIGITAL INPUTS
Parameter| Settings (Default Value)
Remote Cycle Start >>| No Yes
(Default : No)
Water Level Detect >>| No Yes
(Default : No)
Water Level Logic >>| Open : Water Lo Close : Water Lo
(Default : Close : Water Lo)
Door Open Detect >>| Yes No
(Default : No)
Door Open Logic >>| Close : Door Open Open : Door Open
(Default : Close : Door Open)
FACTORY > FACTORY DEFAULT
Parameters| Settings (Default Value)
Set To Default >>| No Yes
(Default : No)
FACTORY > PASSWORD
Parameters| Settings (Default Value)
Change Password >>| 2000 to 2999
(Default : 321)
FACTORY > EXIT
Parameters| Settings (Default Value)
Exit Setup Mode >>| No Yes
(Default : No)