PPI Clavex GD Vertical Autoclave Controller with Graphic Display User Manual

June 11, 2024
PPI

PPI Clavex GD Vertical Autoclave Controller with Graphic Display

PPI Clavex GD Vertical Autoclave Controller

FRONT PANEL LAYOUT

Front Panel

PPI Clavex GD Vertical Autoclave Controller 7

Keys Operation

Symbol Key Function
**** Scroll Press to scroll through various Process Information Screens in

Normal Operation Mode.
| Cycle Start & Alarm Acknowledge| Press to Start / Abort Autoclave Cycle. In case of process under alarm condition, press to acknowledge and mute the alarm output.
| DOWN| Press to decrease the parameter value. Pressing once decreases the value by one count; keeping pressed speeds up the change.
| UP| Press to increase the parameter value. Pressing once increases the value by one count; keeping pressed speeds up the change.
| SET-UP| Press to enter or exit set-up mode.
****| ENTER| Press to store the set parameter value and to scroll to the next parameter.

PV Error Indications

Message Error Type Cause
**OPEN** Sensor Open RTD Pt100 / Thermocouple is Broken / Open
**OVR** Over-range Temperature above Max. Specified Range
**UNR** Under-range Temperature below Min. Specified Range

ELECTRICAL CONNECTIONS

PPI Clavex GD Vertical Autoclave Controller 8

Analog Input

  • AI-1: RTD Pt100 Sensor
  • AI-2: Pressure Sensor

Digital Output

  • DO-1: Heater Control
  • DO-2: Air Valve Control
  • DO-3: Event Alarm
  • DO-4: Unused
  • DO-5: Unused
  • DO-6: Unused

Digital Input

  • DI-1: Cycle Start
  • DI-2: Door Open
  • DI-3: Water Level
  • DI-4: Alarm ACK
  • DI-5: Unused
  • DI-6: Unused

Operation Instruction

OPERATOR PARAMETERS

Parameter| Settings (Default Value)
Ster. Temp SP >>| Setpoint LO Limit to Setpoint HI Limit

(Default : 121.0 °C)

Air Outlet SP >>| Setpoint LO Limit to Setpoint HI Limit

(Default : 100.0 °C)

Ster. Time (HH:MM) >>| 0.01 to 99.59 (HH:MM) (Default : 00.10 min)
Change Password >>| 1 to 100

(Default : 0)

SUPERVISORY SERIAL PARAMETERS

Parameter| Settings (Default Value)
Device Id >>| 1 to 127

(Default : 1)

Baud Rate >>| 2400, 4800, 9600

19200, 38400, 57600

(Default : 9600)

Parity >>| None Even Odd

(Default : Even)

Serial Write >>| No Yes

(Default : No)

SUPERVISORY CONTROL

Parameter| Settings (Default Value)
Setpoint LO Limit >>| 0 to Setpoint HI Limit (Default : 90 °C)
Setpoint HI Limit >>| Setpoint LO Limit to 600.0 °C

(Default : 150.0 °C)

Overshoot Temp >>| Sterilization SP to Control SP HI

(Default :131 °C)

Cycle End Strategy >>| Immediate Stop Timer Based Stop Temp Based Stop

(Default : Immediate Stop)

Cycle End Timer >>| 1 to 60 Minutes (Default : 5 Min)
Cycle End Temp >>| Control SP LO to Sterilization SP (Default : 100.0 °C)
Ctrl Band >>| 0.1 to 99.9

(Default : 0.2 °C)

EOP Alarm Time (sec) >>| 0 to 250 Sec. (Default : 10 Sec)
SUPERVISORY STERILIZATION

Parameter| Settings (Default Value)
Steri. Stop Temp >>| Air Valve SP to Sterilization SP (Default : 120.5 °C)
Steri. Reset Temp >>| Air Valve SP to Sterilization SP

(Default : 120.0 °C)

SUPERVISORY PV OFFSET

Parameter| Settings (Default Value)
Temp Zero Offset >>| -50.0 to 50.0

(Default : 0)

Pressure Offset >>| -500 to 500 Counts

(Default : 0.0)

SUPERVISORY > PASSWORD

Parameters| Settings (Default Value)
Change Password >>| 1000 to 1999

(Default : 123)

SUPERVISORY > EXIT

Parameters| Settings (Default Value)
Exit Setup Mode >>| Yes No

(Default : No)

FACTORY TEMPERATURE INPUT

Parameter| Settings (Default Value)
Filter Constant >>| 0.5 to 60.0 Sec. (Default : 0.5 Sec.)
Temperature SP Offset >>| 0.0 to 10.0 ºC (Default : 0.0 ºC)
FACTORY > PRESSURE INPUT

Parameters| Settings (Default Value)
Pressure Indication >>| None Computed Measured

(Default : None)

Units >>| PSI KG/CM2 BAR EU

(Default : PSI)

Resolution >>| 1

0.1

0.01

0.001

(Default : 0.1)

Parameter Settings (Default Value)
Input Type >> 0-20mA

4-20mA

0-5V

0-10V

1-5V

(Default : 0-5V)

Signal LO >>| ****






Default : Signal Low 0.000

Signal High 5.000

Signal HI >>
Range LO >>| 0 to 9999 Counts

(Default : 0.0)

Range HI >>| 0 to 9999 Counts

(Default : 100.0)

FACTORY DIGITAL INPUTS

Parameter| Settings (Default Value)
Remote Cycle Start >>| No Yes

(Default : No)

Water Level Detect >>| No Yes

(Default : No)

Water Level Logic >>| Open : Water Lo Close : Water Lo

(Default : Close : Water Lo)

Door Open Detect >>| Yes No

(Default : No)

Door Open Logic >>| Close : Door Open Open : Door Open

(Default : Close : Door Open)

FACTORY > FACTORY DEFAULT

Parameters| Settings (Default Value)
Set To Default >>| No Yes

(Default : No)

FACTORY > PASSWORD

Parameters| Settings (Default Value)
Change Password >>| 2000 to 2999

(Default : 321)

FACTORY > EXIT

Parameters| Settings (Default Value)
Exit Setup Mode >>| No Yes

(Default : No)

101, Diamond Industrial Estate, Navghar, Vasai Road (E), Dist. Palghar – 401 210.
Sales: 8208199048 / 8208141446
Support: 07498799226 / 08767395333
E: sales@ppiindia.net, support@ppiindia.net

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals