PPI Clavex Plus 4 Channel Advanced Autoclave Controller

PPI Clavex Plus 4 Channel Advanced Autoclave
Controller

This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net

FRONT PANEL LAYOUT

Front Panel

Front Panel

Keys Operation
Symbol Key Function
Scroll Press to scroll through various Process Information Screens in

Normal Operation Mode.
| Cycle Start & Alarm Acknowledge| Press to Start / Abort Autoclave Cycle. In case of process under alarm condition, press to acknowledge and mute the alarm output.
| DOWN| Press to decrease the parameter value. Pressing once decreases the value by one count; keeping pressed speeds up the change.
| UP| Press to increase the parameter value. Pressing once increases the value by one count; keeping pressed speeds up the change.
| SET-UP| Press to enter or exit set-up mode.
| ENTER| Press to store the set parameter value and to scroll to the next parameter.

PV Error Indications
Message Error Type Cause
Sensor Open Sensor (RTD Pt100) Broken / Open
Over-range Temperature above Max. Specified Range
Under-range Temperature below Min. Specified Range
ELECTRICAL CONNECTIONS

Electrical Connections

OPERATOR PARAMETERS


Parameter| Settings (Default Value)
Select Prnt Option >>| None   Print (New)
Re-Print (Old)
(Default : None)
Ster. Temp SP >>| Setpoint LO Limit to Setpoint HI Limit
(Default : 121.0 °C)
Air Outlet SP >>| Setpoint LO Limit to Setpoint HI Limit
(Default : 100.0 °C)
Ster. Time (min) >>| 1 to 999 (Default : 20 min)
Change Password >>| 1 to 100
(Default : 0)

SUPERVISORY DEVICE


Parameter| Settings (Default Value)
Rec. Interval (sec) >>| 1 to 300 Sec (Default : 60 Sec.)
RTC Time >>| 0 to 23 for Hour
0 to 59 for Min (Default : NA)
RTC Date >>| 1 to 31 (Default : NA)
RTC Month >>| 1 to 12 (Default : NA)
RTC Year >>| 2000 to 2099
(Default : NA)
Device Id >>| 1 to 127
(Default : 1)

SUPERVISORY CONTROL


Parameter| Settings (Default Value)
Setpoint LO Limit >>| 0 to Setpoint HI Limit (Default : 90 °C)
Setpoint HI Limit >>| Setpoint LO Limit to 600 °C
(Default : 150 °C)
Overshoot Temp > >| Ster. Temp SP to Setpoint HI Limit (Default : 131 °C)
Ctrl PV Strategy > >| PV on Map-1 Average PV
PV Map1 Fail-average PV Min Max
(Default : PV on Map-1)
Cycle End Strategy > >| Immediate Stop Timer Based Stop Temp Based Stop
(Default : Immediate Stop)
Cycle End Timer > >| 1 to 60 Minutes (Default : 5 Min)
Cycle End Temp > >| Setpoint LO Limit To Sterilization Temp
(Default : 100 °C)
Ctrl Band > >| 0.1 to 99.9

(Default : 0.2 °C)

EOP Alrm Time(sec) > >| 0 to 250 Sec. (Default : 10 Sec)
Hold Pressure > >| Min. to Max. Pressure Range set for the Pressure Sensor (Default : 15 psi)

SUPERVISORY STERILIZATION


Parameter| Settings (Default Value)
Steri. Stop Temp >>| Air Outlet SP to Sterilization Temp (Default : 120.5 °C)
Steri. Reset Temp >>| Air Outlet SP to Sterilization Temp (Default : 120.0 °C)

SUPERVISORY PV OFFSET


Parameter| Settings (Default Value)
Pressure Offset >>| -50.0 to 50.0
(Default : 0)
Map-1 Offset >>
Map-2 Offset >>
Map-3 Offset >>
Map-4 Offset >>| -50.0 to 50.0
(Default : 0)

SUPERVISORY > PASSWORD


Parameters| Settings (Default Value)
Change Password >>| 1000 to 1999
(Default : 123)

SUPERVISORY > PRINTER SETTING


Parameters| Settings (Default Value)
Line Per Page >>| 25 to 100
(Default : 60)
Top Margine >>| 1 to 5
(Default : 2)
Bottom Margine >>| 1 to 5
(Default : 2)
Header & Footer >>| No
On First Page Every Page (Default : No)
Enable >>| Yes No
(Default : No)

SUPERVISORY > EXIT


Parameters| Settings (Default Value)
Exit Setup Mode >>| Yes No
(Default : No)

FACTORY > TEMPERATURE INPUT 


Parameters| Settings
(Default Value)
Mapping Channles >>| 1 to 4
(Default : 4)
Input Type >>| RTD Pt100
0-20mA
4-20mA
0-5V
0-10V
1-5V
(Default : RTD Pt100)
Signal LO >>| | Input Type| Settings| Default
---|---|---
0 to 20mA| 0.00 to Signal High| 0.00
4 to 20mA| 4.00 to Signal High| 4.00
0 to 5V| 0.000 to Signal High| 0.000
0 to 10V| 0.00 to Signal High| 0.00
1 to 5V| 1.000 to Signal High| 1.000
Signal HI >>| | Input Type| Settings| Default
---|---|---
0 to 20mA| Signal Low to 20.00| 20.00
4 to 20mA| Signal Low to 20.00| 20.00
0 to 5V| Signal Low to 5.000| 5.000
0 to 10V| Signal Low to 10.00| 10.00
1 to 5V| Signal Low to 5.000| 5.000
Range LO >>| -199.9 to RANGE HI
(Default : 0.0)
Range HI >>| RANGE LO to 999.9
(Default : 100.0)
Temp SP Offset >>| 0.0 to 10.0
(Default : 0.0)

FACTORY > PRESSURE INPUT


Parameters| Settings
(Default Value)
Pressure Enable >>| Yes
No
(Default : No)
Units >>| Psi
Kg/cm²
Bar
Eu
(Default : Psi)
Resolution >>| 0
0.1
0.01
0.001
(Default : 0.1)
Input Type >>| 0-20mA
4-20mA
0-5V
0-10V
1-5V
(Default : 0-5V)
Signal LO >>| | Input Type| Settings| Default
---|---|---
0 to 20mA| 0.00 to Signal High| 0.00
4 to 20mA| 4.00 to Signal High| 4.00
0 to 5V| 0.000 to Signal High| 0.000
0 to 10V| 0.00 to Signal High| 0.00
1 to 5V| 1.000 to Signal High| 1.000
Signal HI >>| | Input Type| Settings| Default
---|---|---
0 to 20mA| Signal Low to 20.00| 20.00
4 to 20mA| Signal Low to 20.00| 20.00
0 to 5V| Signal Low to 5.000| 5.000
0 to 10V| Signal Low to 10.00| 10.00
1 to 5V| Signal Low to 5.000| 5.000
Range LO >>| -1999 to RANGE HI
(Default : 0.0)
Range HI >>| RANGE LO to 9999
(Default : 100.0)

FACTORY DIGITAL INPUTS


Parameter| Settings (Default Value)
Remote Cycle Start >>| Yes No
(Default : No)
Water Level Detect >>| Yes No
(Default : No)
Water Level Logic >>| Open : Water Lo Close : Water Lo
(Default : Close : Water Lo)
Door Open Detect >>| Yes No
(Default : No)
Door Open Logic >>| Close : Door Open Open : Door Open
(Default : Close : Door Open)

FACTORY > FACTORY DEFAULT


Parameters| Settings (Default Value)
Set To Default >>| Yes No
(Default : No)

FACTORY > PASSWORD


Parameters| Settings (Default Value)
Change Password >>| 2000 to 2999
(Default : 321)

FACTORY > EXIT


Parameters| Settings (Default Value)
Exit Setup Mode >>| Yes No
(Default : No)

Customer Support

101, Diamond Industrial Estate, Navghar,
Vasai Road (E), Dist. Palghar – 401 210.
Sales : 8208199048 / 8208141446
Support : 07498799226 / 08767395333
E: sales@ppiindia.net, support@ppiindia.net

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals