TRANSTEK TMB-2266 Blood Pressure Monitor
Blood Pressure Monitor TMB-2266 Arm Type
General Description: The TRANSTEK Blood
Pressure Monitor TMB-2266 is a digital monitor designed to measure blood
pressure and heartbeat rate with an arm circumference ranging from 22cm to
32cm, 22cm to 42cm, and 22cm to 45cm. The monitor provides accurate readings
and stores results for easy tracking.
Indications for Use
- The Transtek Blood Pressure Monitor is intended for use by adults only.
- The monitor is designed for indoor use only.
- The arm cuff should fit around an arm circumference ranging from 22cm to 32cm, 22cm to 42cm, and 22cm to 45cm.
Contraindications
The Transtek Blood Pressure Monitor should not be used by individuals with an irregular heartbeat or who have been diagnosed with atrial fibrillation. If you have any concerns about using the monitor, please consult with your healthcare provider before using it.
Measurement Principle
The TMB-2266 uses the oscillometric method to measure blood pressure and heartbeat rate. This method detects the vibrations of the arterial wall caused by blood flow, and then converts the movements into digital readings.
Safety Information
It is important to read the instruction manual thoroughly before using the product to ensure safe usage. The following signs are present in the manual, labeling, or other components:
- “Refer to instruction manual/booklet” to signify that the instruction manual/booklet must be read.
- “Direct Current”
- “Type BF applied part”
- “Serial Number”
- “Date and Country of manufacture”
- “Manufacturer”
- “Recyclable”
- “For indoor use only”
Before You Start
Before using the Transtek Blood Pressure Monitor, follow these steps:
- The monitor can be powered by using batteries or an AC adapter. Choose the power supply that works best for you.
- To install or replace the batteries, follow the instructions in the manual.
- Set the date and time on the monitor.
Measurement
Follow these steps to take a blood pressure reading:
- Apply the cuff to your arm according to the instructions provided.
- Press the “Start” button to begin the measurement.
Data Management
The monitor stores previous blood pressure readings for easy tracking. Follow these steps to recall or delete records:
- To recall previous readings, refer to the instructions in the manual.
- To delete previous readings, refer to the instructions in the manual.
About Blood Pressure
Learn more about blood pressure and how it is measured:
- What are systolic pressure and diastolic pressure?
- What is the standard blood pressure classification?
- Irregular heartbeat detector
- Why does my blood pressure fluctuate throughout the day?
- Why do I get a different blood pressure at home compared to the hospital?
- Is the result the same if measuring on the right arm?
Maintenances
To ensure accurate readings, follow these maintenance tips:
- Clean the monitor and cuff regularly according to the instructions in the manual.
- Store the monitor in a cool, dry place and avoid exposing it to extreme temperatures or moisture.
INTRODUCTION
General Description
Thank you for selecting TRANSTEK blood pressure monitor (TMB-2266). The
monitor features blood pressure measurement, pulse rate measurement and the
result storage. The design provides you with two years of reliable service.
Readings taken by the TMB-2266 are equivalent to those obtained by a trained
observer using the cuff and stethoscope auscultation method.
This manual contains important safety and care information, and provides step
by step instructions for using the product. Read the manual thoroughly before
using the product.
Features:
71mm×82 mm Digital LCD display with white backlight Maximum 199 records per
each user
3rd technonoly: Measuring during inflation
Indications for Use
The Transtek Blood Pressure Monitor is digital monitors intended for use in
measuring blood pressure and heartbeat rate with arm circumference ranging
from 22cm to 32cm, 22cm to 42cm, 22cm to 45cm. It is intended for adult indoor
use only.
Contraindications
- The device is not suitable for use on may be pregnant women or pregnant women.
- The device is not suitable for use on patients with implanted, electrical devices, such as cardiac pacemakers, defibrillators.
Measurement Principle
This product uses the Oscillometric Measuring method to detect blood pressure.
Before every measurement, the unit establishes a “zero pressure” equivalent to
the atmospheric pressure. Then it starts inflating the arm cuff, meanwhile,
the unit detects pressure oscillations generated by beat-to-beat pulsatile,
which is used to determine the systolic and diastolic pressure, and also pulse
rate.
Safety Information
The signs below might be in the user manual, labeling or other component. They
are the requirement of standard and using.
| Refer to instruction manual/booklet
To signify that the instruction manual/ booklet must be read.| | Type BF
applied part
---|---|---|---
| Direct Current| | Serial Number
| Date and Country of manufacture| | Manufacturer
| Recyclable| | For indoor use only
| Symbol for “Class II Equipment”
| MR Unsafe
To identify an item which poses unacceptable risks to the patient, medical
staff or other persons within the MR environment.
| Caution
Indicates that caution is necessary when operating the device or control close
to where the symbol is placed, or that the current situation needs operator
awareness or operator action in order to avoid undesirable consequences.
| The symbol indicates that the product should not be discarded as unsorted
waste but must be sent to separate collection facilities for recovery and
recycling.
CAUTION
-
This device is intended for indoor, home use.
-
This device is not intended for public use.
-
This device is portable, but it is not intended for use during patient transport.
-
This device is not suitable for continuous monitoring during medical emergencies or operations.
-
This device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than the arm, or for any purpose other than obtaining a blood pressure measurement.
-
This device is for adults. Do not use this device on neonates or infants. Do not use it on children unless otherwise instructed by a medical professional.
-
Do not use on the women in pregnant, including pre-eclamptic, patients.
-
The device is not suitable for use on patients with implanted, electrical devices, such as cardiac pacemakers, defibrillators.
-
The effectiveness of this device has not been established for use:
- on users with common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation,
- on users with peripheral arterial disease,
- on users undergoing intravascular therapy, or with arteriovenous (AV) shunt.
Consult a medical professional before use.
-
Do not use this device for diagnosis or treatment of any health problem or disease. Contact your physician if you have or suspect any medical problem. Do not change your medications without the advice of your physician or health care professional.
-
If you are taking medication, consult your physician to determine the proper time to measure your blood pressure.
-
This device may be used only for the intended use described in this manual, the manufacturer shall have no liability for any incidental, consequential, or special damages caused by misuse or abuse.
-
Report any unexpected operation or events to the manufacturer.
-
Do not apply the cuff on an arm that has an intravenous drip or a blood transfusion attached.
-
Warning: Do not kink, fold, stretch, compress, or otherwise deform the tube during measuring, as the cuff pressure might continuously increase, which could prevent blood flow and result injury.
-
Warning: Taking blood pressure measurements too frequently could disrupt blood circulation and cause injuries.
-
Warning: Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site frequently for irritation.
-
Warning: Do not place the cuff on the arm of a person whose arteries or veins are undergoing medical treatment, i.e. intra-vascular access or intra-vascular therapy or an arteriovenous (A-V) shunt, which could disrupt blood circulation and cause injuries.
-
Do not place the cuff on the arm on the same side of a mastectomy (especially when lymph nodes have been removed). it is recommended to take measurements on the unaffected side.
-
Do not wrap the cuff on the same arm to which another monitoring device is applied. One or both devices could temporarily stop functioning if you try to use them at the same time.
-
Please check that the operation of the device do not result in prolonged impairment of patient blood circulation.
-
Warning: On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, loosen and remove the cuff immediately. Prolonged high pressure applied to the arm (cuff pressure >300 mmHg or constant pressure >15 mmHg for more than 3 minutes) might lead to bruising and discolored skin.
-
Warning: Do not use this device with high-frequency (HF) surgical equipment at the same time.
-
Warning: This device is not AP/APG equipment. Do not use the device where flammable anesthetic are present, or in environments mixture with air of with oxygen or nitrous oxide.
-
The device contains sensitive electronic components. To avoid measurement errors, avoid taking blood pressure measurements near a strong electromagnetic field radiated interference signal or electrical fast transient/burst signal.
-
Wireless communication equipment, such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies may cause interference that may affect the accuracy of measurements. A minimum distance of 1 foot (30 cm) should be kept from such devices during a measurement.
-
You can use this device to take your own measurement, no third-party operator is required.
-
Please use the device under the environment which is provided in the user manual. Otherwise, the performance and lifetime of the device will be impacted and reduced.
-
The device may require up to 30 minutes to warm up / cool down from the minimum / maximum storage temperature before it is ready for use.
-
Warning: Excessive cuff tube lengths could cause strangulation if you don’t manage them properly.
-
Warning: Do not touch output of the batteries/adapter and the user simultaneously.
-
Adapter is specified as a part of ME EQUIPMENT.
-
Warning: The power cord is considered the disconnect device for isolating this equipment from supply mains. Do not position the equipment so that it is difficult to reach or disconnect.
-
The blood pressure monitor, its adapter, and the cuff are suitable for use within the patient environment.
-
Warning: Do not use this device if you are allergic to polyester, nylon, or plastic.
-
Warning: Only use accessories approved by manufacturer. Using unapproved accessories might cause damage to the unit and injure users.
-
Warning: If you experience discomfort during a measurement, such as pain in the arm or other complaints, press the Power button immediately to release the air from the cuff.
-
No calibration is required within two years of reliable service.
-
Do not attempt to repair the unit yourself if it malfunctions. Only have repairs carried out by authorized service centers.
-
At the request of authorized service personnel, circuit diagrams, component part lists, descriptions, and calibration procedures will be made available by the manufacturer or distributor.
-
It is recommended that the performance should be checked after repair, maintenance, and every two years of use, by retesting the requirements in limits of the error of the cuff pressure indication and air leakage (testing at least at 50 mmHg and 200 mmHg).
-
Warning: Do not use the device while under maintenance, or being serviced.
-
Store your device, cuff and adapter in a clean and dry place, protect it against extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on it.
-
Make sure the rubber tube of the cuff is not squeezed, stretched, or kinked during storage.
-
Warning: Keep the device, cuff, and batteries away from children as they may pose a risk of choking or strangulation if used improperly.
-
Clean both device and cuff with a soft, dry cloth. If necessary use a dampened cloth and natural detergent. Do not use alcohol, benzene, or other harsh chemicals.
-
Do not wash the cuff in a washing machine or dishwasher!
-
The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The typical service life is 10000 times.
-
Dispose of accessories, detachable parts, and the device according to the local guidelines.
LCD display signal
Monitor Components
List
-
Blood Pressure Monitor (TMB-2266)
-
Cuff (Type BF applied part) (22cm~32cm or 22cm~42cm or 22cm~45cm)
-
4×AAA batteries
-
User manual
BEFORE YOU START
The Choice of Power Supply
- Battery powered mode: 6VDC 4×AAA batteries
- AC adaptor powered mode: 5V 1A
(Please only use the recommended AC adaptor model).
Please unplug the adaptor to depart from the using utility power.
CAUTION
In order to get the best effect and protect your monitor ,please use the the
right batteries and special power adaptor which complies with local safety
standard.
Installing and Replacing the Batteries
- Open the battery cover.
- Install the batteries by matching the correct polarity, as shown.
- Replace the battery cover.
Replace the batteries whenever the below happen
- The bAt Lo shows
- The display is dim.
- The display does not light up
CAUTION
- Do not use new and used batteries together.
- Do not use different types of batteries together.
- Do not dispose the batteries in fire. Batteries may explode or leak.
- Remove batteries if the device is not likely to be used for some time.
- Worn batteries are harmful to the environment. Do not dispose with daily garbage. Remove the old batteries from the device following your local recycling guidelines.
Setting Date,Time and Voice
It is important to set the clock before using your blood pressure monitor, so
that a time stamp can be assigned to each record that is stored in the memory.
(The setting range of the year :2022—2052 time format: 12H/24H)
-
When the monitor is off, press and hold the “USER” button, it will display the Bluetooth symbol first, then press the “START/STOP” button to enter the
[year] setting, press the “USER” or “MEMORY” button to change the year.
Press and hold the “USER” button to quickly advance the years.
Press and hold the “MEMORY” button to quickly go backwards through the years.
-
Press the “START/STOP” button to confirm the year, then the [date format] will flash. Press the “USER” or “MEMORY” button to switch the date format between [month/day] and [day/month].
-
Press the “START/STOP” button to confirm the date format, then the month will flash. Repeat the same steps to change the month, day, time format, hour and minute.
-
After the minute is set, it will enter the vocie setting. It will speak “ Voice on” or “Voice off”, press the “USER” or “MEMORY” button to change the Vioce volume.
Each press “USER” button will increase the voice volume by one in a cycling manner (voice volume 2 – OFF – 1 – 2 …).
Each press “MEMORY” button will decrease the volume by one in a cycling manner (voice volume 2 -1 – OFF – 2 …).- Voice volume 2, broadcast “Voice on”
- Voice volume 1 broadcast “Voice on”
Voice off
- Press the “START/STOP” button to confirm the voice setting. The LCD will display “do nE” and then turn off after several seconds.
Pair a smart device with the monitor
You are the intended operator of this blood pressure monitor. You can measure
your blood pressure and then save and send measurement data to a smart device
(such as smartphone or tablet) with Bluetooth wireless connectivity and
application.
- Turn on the Bluetooth and the app on your smart device. Make sure both are ON when pair-up is proceeding.
- When the monitor is off, press and hold the “START/STOP” button to start pair-up, the Bluetooth symbol will flash.
- BEFORE YOU START If successful, the Bluetooth symbol will not flash any more and the monitor will automatically shut off after several seconds.
Note
The date and time on your monitor will automatically be set after paired with
your smart device successfully.
If unsuccessful within 60 seconds, it is judged timeout and the monitor will
shut off.
- RF Frequency Range: 2402 MHz to 2480 MHz
- Output Power Range: -1.57dBm
- Supply Voltage: 1.8-3.6 V
- Transmitting Distance: 10 meters
List of compatible devices:
For iOS devices:
The operating system must be iOS 13.0 or more.
For Android devices:
The operating system must be Android 5.0 or more.
CAUTION
- Interference may occur in the vicinity of equipment marked with the following symbol . And TMB-2266 may interfering vicinity electrical equipment. Sensitive people, including pregnant women pre-eclamptic and those who implanted medical electronic instruments, should avoid using the unit whenever possible.
- Keep the monitor at least 20 centimeters away from the human body (especially the head) when the data transmission is proceeding after measurement.
- To enable the data transmission function, this product should be paired to Bluetooth end at 2.4 GHz.
How to mitigate possible interference?
- The range between the device and BT end should be reasonably close, from 1 meter to 10 meters. Please ensure no obstacles between the device and BT end so as to obtain quality connection and to lower the RF output range.
- To avoid interference, other electronic devices (particularly those with wireless transmission / Transmitter) should be kept at least 1 meter away from the monitor.
Setting a user ID
There are 3 user ID available. The user and , each with 199 memory spaces,
are designed for 2 different people to save the measured values separately.
The user , no memory space, is reserved for guest.
- When the monitor is off, press the “USER” button to display the current user ID and flash. Press the “USER” button to switch the user ID between the user , and .
- Press the “START/STOP” button to confirm the selected user ID, the User ID will not flash any more and the monitor will enter the measurement automatically.
MEASUREMENTT
Applying the cuff
-
Remove all jewelry, such as watches and bracelets from your left arm.
Note: If your doctor has diagnosed you with poor circulation in your left arm, use your right arm. -
Roll or push up your sleeve to expose the skin. Make sure your sleeve is not too tight.
-
Hold your arm with your palm facing up and tie the cuff on your upper arm, align the Artery indicator with the main Artery (on the inside of your arm).
Note: Locate the main Artery by pressing with 2 fingers approximately 2 cm above the bend of your elbow on the inside of your left arm. Identify where the pulse can be felt the strongest, that is your main Artery! -
Make sure the bottom edge of the arm cuff 2 to 3 cm above the inside elbow. Then wrap the cuff securely. Note: The cuff should be snug but not too tight. You should be able to insert one finger between the cuff and your arm.
-
Sit upright in a comfortable chair with your back against the backrest of the chair. Keep your feet flat and your legs uncrossed. Place your arm resting comfortably on a flat table. The cuff worn on your arm should be placed at the same level as your right atrium of the heart.
-
Take 5-6 deep breaths and let’s start measuring!
Helpful tips:
Take the measurement in a silent room.
Rest for 5 minutes before a measurement.
Wait at least 3 minutes before another measurement. This allows your blood circulation to recover. Be relax and do not move and talk during the measurement procedure.
For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, on the same arm, or as directed by a physician.
Start the Measurement
A. Single measurement
When the monitor is OFF, press the “START/STOP” button, it will finish the
whole measurement automatically, save and transmit the measurement data
for the selected user. (Take User 1 with voice on for example.)
-
Display the latest measurement record.
-
Adjust the zero. It will read “starting measurement, please remain still and quiet”
-
Cuff wrap detection. symbol “ ” flash
-
Cuff wrap ok. Inflating and measuring.
-
Display, save and read the measurement result “your blood pressure is 116 over 78, pulse is 80”
B. Triple mode measurement
When the monitor is OFF, press the “TRIPLE MODE” button, it will finish the
whole measurement automatically, save and transmit the measurement data
for the selected user. (Take User 1 with voice on for example.)
After the first and second measurements are completed, there will be no
results. After all three measurement is completed, the average results of the
three measurements will be displayed.
-
Display the latest measurement record.
-
Adjust the zero. It will read “starting measurement, please remain still and quiet”
-
Cuff wrap detection. symbol “ ” flash
-
Cuff wrap ok. Inflating and measuring.
The second measurement MEASUREMENT (60 seconds countdown display)
The third measurement MEASUREMENT (60 seconds countdown display)
Display, save and read the measurement result “your blood pressure is 120 over
80, pulse is 90
After the measurement was finished, the Bluetooth symbol “ ” blinks and data
transmission starts. (Please connect the app during the transmission)
If data transmission successful, the symbol “ ” will not blink and “ ” will
disappear, after sevral seconds, “ ” will also disappear, and then the monitor
will turn off automatically after about 1 minute.
If unsuccessful within 60 seconds, it is judged timeout and the monitor will
shut off.
Press “START/ STOP” button to turn off, otherwise it will turn off automatically. Note
- Any time if you want to stop the measurement, you can press the “START/ STOP” button manually.
- When the selected user has history memory record, the monitor will display its last measurement record first before starting measurement.
About the irregular heartbeat and excessive body motion during the measurement.
- During a measurement, If an irregular heartbeat is detected, the symbol will display in the measurement result. See page 26 for more information.
- During a measurement, when the excessive body motion, especially of the arm the cuff is worn is detected, the symbol will flash about 5 seconds and detect again. If it is no longer detected, the symbol will disappear; If still detected, the symbol will final display in the measurement result.
Note
The measured blood pressure reading may not be accurate if this symbol is
displayed.
DATA MANAGEMENT
Recall the Records
-
When the monitor is off, press the “MEMORY” button, the user ID will blink, you can press the “USER” or “MEMORY” button to switch the user ID between user and .
-
Press the “START/STOP” button to confirm the selected user ID.
Then the LCD will display the latest record. (Example shown below for User 1)
When the voice is on, it will read the memory value for you at the same time.
(Take the right memory record 01 for example, it will speak “memory 1, your blood pressure is 120 over 80, pulse is 80”)
Note The most recent record (1) is shown first. Each new measurement is assigned to the first (1) record. All other records are pushed back one digit (e.g., 2 becomes 3, and so on), and the last record (199) is dropped from the list. -
Press the USER or MEM button to display the next record.
Delete the Records
If you did not get the correct measurement, you can delete results by
following steps below.
A: To delete a single measurement:
- Enter the memory recall mode as described in section [Recall the Records]. Press the “USER” or “MEMORY” button to get the measurement you would like to erase.
- Press and hold “MEMORY” button for about 5 seconds, and the display will show a blinking “dEL y”.
- Use the “USER” or “MEMORY” button to switch between “dEL y” and “dEL no”, Press “START/STOP” to confirm the selection.
If “dEL y” is selected, the unit will delete the record, after then it will
return to last memory record and all the records are pushed forward one digit
(e.g., 03 becomes 02, and so on)
If “dEL no” is selected, it will stop the deletion.
B: To delete all measurements:
- Enter the memory recall mode as described in section [Recall the Records].
- Press and hold “USER” and “MEMORY” button button for 5 seconds, and the display will show a blinking “dEL AL” along with the user ID.
- Use the “USER” or “MEMORY” button to switch between “dEL AL” and “dEL no”, Press “START/STOP” to confirm the selection.
If “dEL AL” is selected, the unit will display “dEL do nE” + User ID, and
delete all the record of the current user. Several seconds later, it will
display “—”.
If “dEL no” is selected, it will stop the deletion.
INFORMATION FOR USER
Tips for Measurement
Measurements may be inaccurate if taken in the following circumstances.
Maintenance
In order to get the best performance, please follow the instructions below.
ABOUT BLOOD PRESSURE
What are systolic pressure and diastolic pressure?
When ventricles contract and pump blood out of the heart, the blood pressure
reaches its maximum value in the cycle, which is called systolic pressure.
When the ventricles relax, the blood pressure reaches its minimum value in the
cycle, which is called diastolic pressure.
What is the standard blood pressure classification?
The blood pressure classification published by World Health Organization (WHO)
and International Society of Hypertension (ISH) in 1999 is as follows
:
CAUTION
Only a physician can tell your normal BP range. Please contact a physician if
your measuring result falls out of the range. Please note that only a
physician can tell whether your blood pressure value has reached a dangerous
point.
Level
Blood
Pressure (mm Hg)
|
Optimal
|
Normal
|
High-normal
|
Mild
|
Moderate
|
Severe
---|---|---|---|---|---|---
SYS
|
<120
|
120-129
|
130-139
|
140-159
|
160-179
|
≥180
DIA
|
<80
|
80-84
|
85-89
|
90-99
|
100-109
|
≥110
Irregular Heartbeat Detector
An irregular heartbeat is detected when a heartbeat rhythm varies while the
unit is measuring the systolic and diastolic blood pressure.During each
measurement, the monitor records all the pulse intervals and calculate the
average ; if there are two or more pulse intervals ,the difference between
each interval and the average is more than the average value of ±25% , or
there are four or more pulse intervals ,the difference between each interval
and the average is more than the average value of ±15%,the irregular heartbeat
symbol appears on the display when the measurement results are appeared.
CAUTION
The appearance of the IHB icon indicates that a pulse irregularity consistent
with an irregular heart-beat was detected during measurement. Usually this is
NOT a cause for concern. However, if the symbol appears often, we recommend
you seek medical advice. Please note that the device does not replace a
cardiac examination, but serves to detect pulse irregularities at an early
stage.
Why does my blood pressure fluctuate throughout the day?
- Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions.
- If the person takes medicine, the pressure will vary more.
- Wait at least 3 minutes for another measurement.
What you need to pay attention to when you measure your blood pressure at
home:
If the cuff is tied properly.
If the cuff is too tight or too loose. If the cuff is tied on the upper arm.
If you feel anxious.
Taking 2-3 deep breaths before beginning will be better for measuring.
Advice: Relax for 4-5 minutes until you calm down.
Why do I get a different blood pressure at home compared to the hospital?
The blood pressure is different even throughout the day due to weather,
emotion, exercise etc. Also, there is the “white coat” effect, which means
blood pressure usually increases in clinical settings.
Is the result the same if measuring on the right arm?
It is ok for both arms, but there will be some different results for different
people. We suggest you measure the same arm every time.
TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
PROBLEM| SYMPTOM| CHECK THIS| REMEDY|
---|---|---|---|---
No power
| ****
Display can not light up.
| Batteries are depleted.| Replace with new batteries.|
Batteries are inserted incorrectly.| Insert the batteries correctly.|
Adapter is inserted incorrectly.| Insert the AC adapter correctly.|
High Battery| bAt H shows| The battery is too high.| Replace with
the correct
DC plug.
|
Low Battery| bAt Lo& shows| The battery is too low.| Replace with new
batteries.|
Error message
| ****
E 1 shows
| The cuff is not wrapped or wrapped incorrectly, or the cuff air plug is
loose.| Refasten the cuff and insert air tube plug correctly then measure
again.|
E 2 or
shows
| Excessive body motion (such as shaking of the arm with the cuff on)
or weak Pulse is detected.
| Relax for 5 minutes. and then keep still, measure again.|
E 3 shows| Pulse is not detected during measuring.| Loosen the clothing on the
arm and measure again.|
E 4 shows| The measurement failed.| Relax for 5 minutes and measure again.|
EExx shows| A calibration error occurred (XX can be some digital symbol, such
as 01, 02,etc., if this similar situation appear, all belong to calibration
error.)| Retake the measurement.If the problem persists, contact the retailer
or our customer service department for further assistance. Refer to the
warranty for contact information and return i nstructions.|
Warning message| out shows| Out of measurement range| Relax for a
moment and then measure again. If the problem persists, contact your
physician.|
Power supply| Battery powered mode: 6VDC 4×AAA batteries AC adaptor
powered mode: 5V 1A Please only use the recommended AC adaptor model).
Display mode| Digital LCD V.A.71mm × 82mm
Measurement mode| Oscillographic testing mode
Measurement range
| Rated cuff pressure: 0mmHg~299mmHg Measurement pressure: SYS:
60mmHg~230mmHg DIA: 40mmHg~130mmHg
Pulse value: (40-199)beat/minute
Accuracy| Pressure:5℃-40℃within±3mmHg
Pulse value:±5%
Normal working condition| A temperature range of :+5°C to +40°C A
relative humidity range of 15% to 90%, non-condensing
An atmospheric pressure range of : 700 hPa to 1060 hPa
Storage & transportation condition| Temperature:-20°C to +60°C A relative
humidity range of ≤ 93%, non-condensing
Measurement perimeter of the upper arm| About 22 cm ~ 32 cm, 22cm ~
42cm, 22cm ~ 45cm
Weight| Approx.278g (Excluding the batteries and cuff)
External dimensions| Approx.100mmx138mmx51mm
Attachment| 4×AAA batteries,user manual, Type-C cable
Mode of operation| Continuous operation
Degree of protection| Type BF applied part
Protection against ingress of water| IP21 It means the device could
protected against solid foreign objects of 12.5mm and greater, and protect
against vertically falling water drops.
Device Classification
| Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment
Software Version| A01
NOTE: If the product still does not work, contact Customer Service. Under
no circumstance should you disassemble or attempt to repair the unit by
yourself.
WARNING: No modification of this equipment is allowed.
AUTHORIZED COMPONENT
Authorized Component
please use the TRANSTEK authorized adapter.
Adapter
Type: BLJ06L060100P-U
Input: 100-240V, 50-60Hz, 0.2A max Output: 5V 1000 mA
Contact Information
For more information about our products, please visit
www.transtekcorp.com.
Manufactured by:
Company:
Address:
Guangdong Transtek Medical Electronics Co., Ltd.
Zone A, No.105, Dongli Road, Torch Dev
EMC GUIDANCE
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments.
Warning: Don’t be near the active HF surgical equipment and the RF
shielded room of an ME system for magnetic resonance imaging, where the
intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper operation. If
such use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally.
Warning: Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
Warning: Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the equipment TMB-2266 including cables specified
by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
Technical description:
- All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the excepted service life.
- Guidance and manufacturer’s declaration-electromagnetic emissions and Immunity.
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
Emissions test| Compliance
RF emissions CISPR 11| Group 1
RF emissions CISPR 11| Class [ B ]
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations / flicker emissions
IEC 61000-3-3| Comply
Table 2
Guidance and manufacturer’s declaration – electromagnetic Immunity
Immunity Test| IEC 60601-1-2
Test level|
Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2
| ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air| ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst IEC 61000-4-4| ±2 kV for power supply lines
±1 kV signal input/output 100 kHz repetition frequency| ±2 kV for power supply
lines Not applicable
100 kHz repetition frequency
Surge IEC61000-4-5| ±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV, ±2 kV common mode| ±0.5 kV, ±1 kV differential mode
Voltage dips, short interruptions and voltage variations on power supply input
lines
IEC 61000-4-11
| 0% UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0% UT; 1 cycle and 70% UT; 25/30 cycles; Single phase: at 0°. 0% UT; 250 / 300
cycle| 0% UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0% UT; 1 cycle and 70% UT; 25/30 cycles; Single phase: at 0°. 0% UT; 250 / 300
cycle
Power frequency magnetic field IEC 61000-4-8| 30 A/m
50 Hz / 60 Hz| 30 A/m
50 Hz / 60 Hz
Conduced RF IEC61000-4-6
| 3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between 0,15 MHz and 80 MHz
80% AM at 1 kHz| 3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between 0,15 MHz and 80 MHz
80% AM at 1 kHz
Radiated RF IEC61000-4-3
| 10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz| 10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz
NOTE UT is the a.c. mains voltage prior to application of the test level.
Table 3
Guidance and manufacturer’s declaration – electromagnetic Immunity
Radiated RF IEC61000-4-3
(Test specifications for ENCLOSURE PORT IMMUNITY to
RF wireless communicati- ons equipment)
| Test Frequency (MHz)| Band (MHz)| Service| Modulation| Maximum Power (W)| Distance (m)| IEC 60601-1-2
Test Level
(V/m)
| Compliance level
(V/m)
385| 380-390| TETRA 400| Pulse modulation 18 Hz| 1.8| 0.3| 27| 27
450| 430-470| GMRS 460,
FRS 460
| FM ± 5k Hz
deviation 1 kHz sine
| 2| 0.3| 28| 28
710| 704-787| LTE Band 13,
17
| Pulse modulation 217 Hz| 0.2| 0.3| 9| 9
745
780
810| 800-960| GSM 800/900, TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
| Pulse modulation 18 Hz| 2| 0.3| 28| 28
870
930
1720| 1700-
1990
| GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1,
3, 4,25; UMTS
| Pulse modulation 217 Hz| 2| 0.3| 28| 28
1845
1970
2450| 2400-
2570
| Bluetooth, WLAN, 802.11
b/g/n, RFID 2450, LTE
Band 7
| Pulse modulation 217 Hz| 2| 0.3| 28| 28
5240| 5100-
5800
| WLAN 802.11
a/n
| Pulse modulation 217 Hz| 0.2| 0.3| 9| 9
5500
5785
FCC Statement
FCC ID: OU9-TMB-2266
FCC Regulatory Compliance
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation.
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the
following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and
- This device must accept any interference received, including interference that may cause undesired operation.
RF Exposure Compliance
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. This transmitter must not be co-located or operating
in conjunction with any other antenna or transmitter.