Adapter Kit for use with
Philips Respironics DreamStation 2
INSTRUCTIONS
ADAPTER KIT INCLUDES:
1 – SoClean Express Adapter
2 – Hose Sleeves [For use with 12mm hose]
If you need help connecting your adapter, contact SoClean.
866-501-3705 |
SoClean.com |
info@soclean.com
If you are a new SoClean 2 user:
Detach Ozone Tube from Injection Fitting.
Save Injection Fitting in case you need it in the future.
Move on to step 1.
If you are an existing SoClean 2/DreamStation user:
A. Open the reservoir on the DreamStation.
Remove the SoClean Adapter and Diffuser.
B. Disconnect the Ozone Tube from the SoClean Adapter.
Save the SoClean Adapter and Diffuser in case you need them in the future.
Move on to step 1.
Connecting the SoClean Express Adapter:
with a heated or non-heated hose
- Connect the SoClean Express Adapter to the heated or nonheated hose.
- Connect the Ozone Tube to the SoClean Express Adapter.
- Connect the SoClean Express Adapter with hose and Ozone Tube to the DreamStation 2.
You are now ready to use the SoClean 2.
CAUTION: The Ozone Tube must be connected to the SoClean Express Adapter at all times when operating your PAP or the SoClean 2.
When to use the Hose Sleeve:
The Hose Sleeve is required if you use a 12mm hose.
It is not compatible with 15mm or larger diameter hoses.
An extra sleeve is included in the event of sleeve breakage.
You will only use one Hose Sleeve at a time.
The SoClean requires a snug it around the hose when placed into the hose slot
and the lid is closed. If there is space around the hose and this is not
achieved, the SoClean will not operate.
Determining your hose size:
If you determine that you have a 15mm hose and the SoClean is still not operating, contact Customer Care at 866-501-3705 for assistance.
How to use the Hose Sleeve:
- Position the Hos Sleeve 2 to 3 inches from the connector.
- Position the hose on the Nub side of the Hose Sleeve. Slide the hose until the Sleeve “clicks” into place between two of the hose ribs and close the Sleeve.
- Line up the Hose Sleeve with the gasket, place it in the Hose Slot and shut the lid.
SoClean, Inc.