Banggood HW-300S Coating Thickness Gauge Instruction Manual
- June 5, 2024
- banggood
Table of Contents
Application manual of coating thickness gauge
V1.0
Brief introduction
This product is mainly used to measure the thickness of the non-conductive coating on the metal surfaces, as well as the thickness of the non- ferromagnetic coating on ferromagnetic metal (such as iron, nickel, cobalt, etc.), such as paint thickness on automobile surfaces, the coating thickness of metal parts and so on. This product also has built-in magnetic induction and vortex double principle probe, which can automatically identify the measured metal substrate, only need to be placed on the measured surface, and can automatically calculate the thickness of the coating.
- Probe mode: automatic (AUTO), magnetic induction (F), vortex (NFE)
- Measurement reading
- Statistical display (average, minimum, maximum, number)
- Automatic shutdown instruction
- Low power indicator
- Substrate properties (Fe: iron; NFE: non-iron)
- Unit switching (um micron, mm mm, Mill mils)
- Up key (unit switch)
- Right (zero calibration, switching average/minimum/maximum/number)
- Down key (backlight switch, automatic shutdown switch)
- Left key (probe mode switch, clear statistics)
- Switch machine
- Probe
- Standard diaphragm
- The substrate to be tested
Note: if the symbol is displayed, the battery is normal, and if the symbol is displayed, it means that the battery is almost exhausted and must be exhausted.
Install battery
Please prepare the 2 AAA batteries to be replaced. Disassemble the back of the casing and take out the batteries. Be sure to install new batteries in the direction indicated in the battery compartment.
Measurement step
Step 1. Prepare the parts to be tested.
Step 2. Stay away from the metal object at least 2 cm, and press the switch
key to turn it on.
Note: it is suggested that the instrument should be zeroed with reference
to “zero calibration” before testing.
Step 3. The probe is placed vertically and quickly on the surface to be
measured until the drop is heard, and the measured value is displayed on the
screen, and then the probe is lifted at least 2 cm away from the part to be
measured, and the next measurement can be carried out.
Note: if the automatic shutdown function is turned on and there is no
operation within 3 minutes, the instrument will automatically shut down.
This machine has two calibration methods:
- Basic calibration: the base calibration should be carried out when it is used for the first time or has not been used for a long time, or when replacing the tested substrate material. There are 7 calibration points in the basic calibration, and the calibration unit is mm.
a. Prepare 6 pieces of standard film, the thickness of which is in0. 04 ~ 0.06,0.09 ~ 0. 11, 0.22 ~ 0.28, 0. 45 ~ 0.55,1. 90 ~ 1.05,1.90 ~ 2.00.
b. First hold down the up key to hold, then press the on key, LCD full-screen kettle display, and then hear a BI, LCD display Display 00mm, the lower right corner of the LCD displays the CAL character,ʺRelease the up buttonʺ indicating that the calibration screen is entered.
c. Press the probe lightly on the uncoated aluminum base, at which point the LCD displays 0.00, and then BI-BI calibrates twice.
d. Get rid of the probe and show the number of 0.05mm on the LCD. The second number calibration is carried out, and the probe is gently pressed on the aluminum base where the calibration sheet is placed.
After adjusting the value displayed on the LCD and the thickness of the calibration sheet by pressing the upper and lower keys, remove the probe and Bi-Bi-Sound, the second calibration point has been calibrated.
Zero calibration
In order to improve the test accuracy, it is suggested that the instrument
should be calibrated to zero.
a. Preparation of uncoated substrate to be tested
b. Hold down the right hand until you hear the drop, and the “zero calibration
(ZERO)” on the screen will begin to flicker.
c. The probe is placed vertically and quickly on the surface of the uncoated
substrate to be tested.
After hearing the dripping sound, the screen will display “0”, and the probe
can be lifted at least 2 cm away from the substrate, that is, to return to
zero once.
Unit switching
Press up to switch micron um, millimeter mm, and Mill mi Is units.
Probe mode switching
Press “left” to switch probe mode.
In automatic (AUTO) mode, the instrument can automatically switch the probe
and measure it.
In the magnetic induction (F) mode, the instrument will be measured by
magnetic induction mode. At this time, it is suitable for the measurement of
the ferromagnetic substrates. In Eddy current (N) mode, the instrument will be
measured by Eddy current mode. At this time, it is suitable for the
measurement of the non-ferromagnetic metal substrate.
Statistical display
a. Right-click to switch statistical display (average, minimum, maximum,
number)
b. Hold down the left key for about 3 seconds to clear the statistics and
start the new statistics at the same time. This instrument supports up to 50
data to carry on the statistics, when reaches 50, the latest data will replace
the oldest data, and the statistical value will be updated automatically.
Backlight
Press the “Down key” to turn on and off the backlight.
Automatic shutdown
Hold down the Down key for about 3 seconds to turn on or off the automatic shutdown function. If the automatic shutdown function is turned on and there is no operation within 3 minutes, the instrument will automatically shut down and save batteries.
Troubleshooting
If a significant anomaly is found in the instrument, the user can unplug the battery for 2 seconds, reinstall the battery, and then boot and try again. If the problem cannot be solved, please turn off the machine, hold down the boot button, and do not release until the screen appears “RS”, release the button, at this time the instrument resumes the factory settings successfully. If you are still unable to solve the problem, please contact your dealer in a timely manner.
Technical index
Probe | F probe | N probe |
---|---|---|
Magnetic induction | Measuring principle | Eddy current effect |
Measuring range | 0~2000um | |
Precision | ±(3%+1um) | |
Resolution | Support | |
Zero calibration | Average, minimum, maximum, number | |
Statistics | Diameter 20mm | |
Unit | 0um~99.9um(0.1um) 100um~999um(1um) >1000um(0.01mm) | |
Minimum convex Radian | um,mm,mils | |
Minimum concave Radian | 5mm | |
Minimum measured area | 25mm | |
Minimum substrate thickness | 0.2mm | 0.05mm |
Maximum measuring speed | 2 readings per second | |
Power supply | 2 1.5V 7 battery | |
Operating environment | Temperature:0 ~ 50 °C 20 ~ 90%rh (non-condensed) | |
Preservation environment | Temperature:-10 ^ 60 °C 20 ^ 90%rh (non- |
condensation)
Size/weight/material| 113mmX53mmX24mm/80g/ABS