Banggood FS8233Pro Touch Screen Intelligent Digital Multimeter User Manual
- June 4, 2024
- banggood
Table of Contents
Banggood FS8233Pro Touch Screen Intelligent Digital Multimeter
Safety information
To fully understand the function of the meter and ensure safe operation,
please read and follow it carefully Follow the use method of this manual.
Special care should be taken when using this meter, improper use may cause
electric shock or damage to the meter. In use, it should follow the usual
safety regulations and take effective safety precautions.
The instrument meets the safety requirements of IEC-61010-1, IEC-61010-2-030,
IEC-61010-2-032) electronic measuring instrument, belongs to secondary
pollution, and the overvoltage standard is CAT III 600V
Please follow the safety operation guide and use the meter safely.
Safety instructions
-
When using the meter, users must follow standard safety rules ‘ -Universal protection against electric shock
-
Prevent misuse of meter
-
After receiving the meter, check for damage in transit.
-
After storage and shipment in poor condition, check and confirm whether the meter is damaged.
-
The pen must be in good condition. Before use, check whether the insulation of the pen is damaged and whether the metal wire of the wire is exposed
Electrical symbols -
Note (For important safety information, please refer to the instruction manual)
-
Can be used on dangerous charged objects
-
Double insulation protection (Class II)
-
In accordance with iec-61010-1 overvoltage (installation) Class II, contamination level 2 refers to the level of voltage protection provided to withstand pulses.
maintenance -
Do not attempt to open the bottom case to adjust or repair the instrument. Such operations should only be performed by technicians who are fully aware of the instrument and the risks of electric shock.
-
The pen should be removed from the line under test before
opening the instrument backing. -
To avoid electric shocks that may be caused by incorrect
readings, charge the meter immediately when it shows the “ ” symbol. -
Clean the meter with a damp cloth and mild detergent. Do not use abrasives or solvents.
-
Turn off the power when the meter is not in use.
-
The meter cannot be charged for a long time. After being fully charged, unplug the Type-c line charging cable.
2describe
Part name
- Non-contact voltage detection sensing area
- Type-c charging cable port
- LCD touch button
- Display
- Input socket
- The power button
LCD touch display
Technical indicators
Gauges shall be specified for a period of one year, between 18°C and 28°C, with relative humidity less than Recalibration in 75% condition.
Comprehensive Indicators
- Automatic selection of measuring function and range.
- Full range overload protection.
- The maximum voltage between the measuring end and the ground is 1000V DC or 750V AC
- Working height: maximum 2000m
- Display: LCD
- The maximum display value is 9999 digits.
- Polarity indication: automatic indication, ‘-‘ indicates negative polarity.
- Superview display: ‘OL’ or ‘-OL’.
- Sampling time: about 3 times/second
- Unit display: with function, electric quantity unit display.
- Automatic shutdown time: 15 minutes
- Power supply: 3.7V/600mAh lithium battery
- Battery undervoltage indicator: LCD symbol.
- Temperature coefficient: less than 0.1X accuracy /°C
- Operating temperature: 0° C-40 ° C
- Storage temperature: -10°C- 50°C
Technical indicators
Alternating current
Range | Resolution | Accuracy |
---|---|---|
999.9mA | **** | ± (1.0% reading + 8 digits) |
10A | **** |
- Minimum input current 20mA AC current
- Maximum input current 10A AC current
- Frequency range: 40HZ-1000Hz;
Dc current
Range | Resolution | Accuracy |
---|---|---|
999.9mA | ± (1.0% reading +5 digits) | |
10A |
- Minimum input current 20mA DC current
- Maximum input current 10A DC current
Dc voltage
Range | Resolution | Accuracy |
---|---|---|
999.9mV | **** | **** |
± (0.5% reading + 3 digits)
9.999V|
99.99V|
999.9V| ****
- In AUTO mode, the input voltage ranges from 0.5V to 1000V DC
- The input voltage ranges from 0.0mV to 999.9mV DC in mV mode
Note:
In the mV range, with the input open, the meter may have a pulsating reading,
This is normal because of the high sensitivity of the meter when connected to
the pen When you measure the circuit, you get a true measurement.
.Ac voltage
Range | Resolution | Accuracy |
---|---|---|
999.9mV | **** | **** |
±( 0.8% reading + 5 digits)
9.999V|
99.99V|
999.9V| ****
- In AUTO mode, the input voltage ranges from 0.5V to 1000V AC
- In mV mode, the input voltage ranges from 0.0mV to 999.9mV AC
- Frequency rate pattern: 40HZ-1000Hz
Line connection test
Range | Resolution | explain |
---|---|---|
**** | **** |
1.0Ω
| If resistance of the line being measured is less than 50 Ω , buzzer in the instrument
will make continuous alarming sounds.
Overload protection: 600V DC or AC (valid value)
Resistance
Range | Resolution | Accuracy |
---|---|---|
999.9Ω | **** | **** |
± ( 0.8% reading + 3 digits)
9.999KΩ|
99.99KΩ|
999.9KΩ|
9.999MΩ|
99.99MΩ| ****
Input protection: maximum 600V DC or AC valid value.
Capacitance
Range | Resolution | Accuracy |
---|---|---|
9.999nF | reading digits | |
999.9nF | **** |
±( 2.5%reading+20 digits)
9.999uF|
99.99uF|
999.9uF|
9.999mF|
99.99mF|
Input protection: the maximum value is 600V DC or AC.
The temperature
Range | Resolution | Accuracy |
---|---|---|
**** | **** | **** |
| ± (1% reading +3℃)
| | ****
| ± (1% reading +3℉)
- Accuracy does not include error of the thermocouple probe.
- Overload protection 250V DC or AC (valid value)
Frequency
Pass grade A:
- Measuring range: 40HZ-1000Hz
- Input signal pattern: > 3A AC current (Effective Value)
Through V gear:
- Measuring range: 40HZ-1000Hz
- Input signal pattern: > 0.8V AC current (Effective value)
Operational guidelines
Reading hold
In the process of measurement, if you need to hold the reading, touch H the key, the display value will be locked, touch H the key again, can remove the reading hold.
Flashlight
Touch the button to turn on the lighting function of the flashlight barrel,
which will be automatically turned off after about 1 minute.。
Automatic shutdown
1If there is no operation within 15 minutes after startup, the meter will
enter a hibernation state and automatically shut down to save electric energy.
The buzzer has one sound one minute before the shutdown. 2After automatic
shutdown, press the power button to wake up the instrument to work.
Preparation before measurement
Tap the power source button to power on. If the battery voltage is low (about
2.8V), the monitor will display the” ” symbol, then should charge. Press the
power button again to turn it off.
When the instrument is not measured, the instrument enters the automatic
scanning state and the instrument displays
Non-contact Voltage Detection (NCV)
Touch the NCV key to enable the NCV function and the instrument displays NCV
The NCV sensor is placed close to the measured wire, and the instrument can
detect whether the measured wire is gt. 90V ac voltage. When the meter detects
the ac power, the meter buzzer alarms and the LCD analog bar displays the
induction intensity.
Note:
- Voltage may still exist even if there is no alarm indication. Do not rely on contactless voltage detectors To see if there’s a voltage on the wire. Probe operation may depend on socket design, and insulation thickness Different types of factors such as the impact.
- In NCV mode, the meter does not measure voltage, resistance, and current at the same tim
Ac-dc current measurement
Insert the stylus into the 10A current input hole, and the current input terminal is inserted into the stylus automatic identification function Automatically enters the default DC current gear of the current gear system. Remove the 10A current input hole and the stylus automatically jumps out of the current gear and enters the AUTO measurement mode
Note:
When the meter is in diode gear, the pen is inserted into the 10A input hole
and the meter will not enter the current gear.
Connect the pen into the measured signal, when the measured signal > The
meter will display the current measured value at 20mA. When measured ac signal
> At 3000rnA, the V Ω instrument pair displays the AC current frequency.
Dc voltage measurement
Insert the black stylus into the COM jack and the red stylus into the V jack. Connect the markers together The voltage source or load ends are measured and read on the LCD.
AUTO automatic measurement mode:
when the measured signal > At 0.5V, the meter will display the current
measured DC voltage value. When the signal being measured. At 0.5V, the meter
will default to the resistance value and display the internal resistance value
of the measured signal. Touch the ” ” key to switch to the DC mV voltage
level. Connect the pen to both ends of the voltage source or load for
measurement, reading on the LCD.
DC mV measurement mode:
when the measured signal < At 600.0mV, the meter will display the
current measured DC voltage
value. When the measured signal > The meter will display “OL” at 999.9mV.
Ac voltage measurement
Insert the black stylus into the COM jack and the red stylus into the V jack. Connect the pen to both ends of the voltage source or load for measurement, reading on the LCD.
AUTO automatic measurement mode:
when the measured signal > At 0.5V, the meter will display the current
measured AC voltage value. When the signal being measured. 0.5 VInstrument
will
default to the resistance value, display the measured signal internal
resistance value.
Touch the “ ” key to switch to ac mV voltage. Connect the pen to both ends of
the voltage source or load for measurement, reading on the LCD.
AC mV measurement mode:
when the measured signal < At 600.0mV, the meter will display the V Ω
current measured AC voltage value.
When the signal is measured; The meter will display “OL” at 999.9mV.”
Resistance measurement
warning
When measuring the impedance on the line, make sure that the circuit is
disconnected and the capacitor on the circuit discharges completely.
-
Insert the black stylus into the COM jack and the red stylus into the “Ω” jack.
-
AUTO AUTO range tap. At this time, the meter is in an automatic scanning state.
-
Connect the pen to the measured resistance or both ends of the line for measurement, reading in the LCD display.
Note:- When the input is open, the LCD will display the “OL” overrange status.
- If the resistance being measured is higher than 10MΩ, the meter may take several seconds to stabilize the reading, which is normal for high resistance readings.
The line of fire test
Touch the “Live” key to switch to LIVE gear, the instrument shows Live
Connect the red test pen to the Live input socket. Insert the single pen into
the power socket L jack or close to the live wire. If the instrument detects
ac voltage, it will display voltage signal strength according to the detected
signal strength, and the buzzer will give alarm sound of different intensity.
Capacitance measurement
Touch the “ ” key to switch the capacitor gear.
Use the other end of the test pen to measure the capacitance value of the
capacitor to be measured and read the measured value from the LCD.
Note
- When measuring large capacitors, a fixed time is required to stabilize readings.
- When measuring polar capacitors, pay attention to the corresponding polarity to avoid damaging the instrument.
Temperature measurement
Touch “SEL” key to switch TEMP file, thermocouple red plug into the end of °C
/°F, black plug into the COM jack. When the reading is stable, the temperature
can be read directly from the display screen.
Note:
The maximum measurement temperature of type K thermocouple is 250°, with
instantaneous measurement up to 300°C.
Diode test
- Insert the black stylus into the COM jack and the red stylus into the “ ” jack.
- Touch SEL key to switch to diode position.
- Connect the red stylus to the diode anode and the black stylus to the diode cathode for testing.
- Read on the LCD.
Note:
The meter displays an approximation of the forward voltage drop of the diode.
If the pen is connected inversely or the pen is open, the LCD displays “OL”.
Line connection test
warning
Danger of electric shock. When testing the circuit on and off, make sure the
circuit is disconnected and the capacitor on the circuit is completely
discharged
- Insert the black marker into the COM jack and the red marker into the ” ” jack.
- Touch the “SEL” key to switch to the line connection test state.
- Connect the pen to both ends V Ω of the line for measurement.
If the resistance of the measured line is less than 50 Ω, the buzzer inside the meter will sound.
Note: If the meter is open or the resistance of the measured line is greater than 999.9Ω, “OL” is displayed.
Maintenance
Type – c line charging
warning
During the charging process, the meter pen should be removed from the
measuring circuit to avoid the danger of lectric shock.
If the ” ” symbol appears, it means to charge.
Replace the test pen
warning When replacing the stylus, you must replace the same or the
same grade of the table. Pen meter must be in good condition, pen grade: 1000V
10A.
Note:
If the insulation layer of the pen is damaged, such as the exposed wire of the
wire, the pen must be replaced.
The attachment
- Test the pen Grade: 1000V 10A A pair of
- Directions for use A copy of the
- thermocouple a
- Charging line Type-c a