WASSERMANN Polyquick 260-T105 Pressure Polymerisation Unit and Compactor User Manual
- June 5, 2024
- WASSERMANN
Table of Contents
Polyquick 260-T105 Pressure Polymerisation Unit and Compactor
User Manual
Dear customer,
Thank you for choosing a product from the Wassermann range. Wassermann
DentalMaschinen incorporates the highest standards of quality and the latest
technology. In order to enjoy maximum performance and years of trouble-free
operation, please read this user manual carefully before you connect this
device and start work, and operate the device according to the recommended
guidelines. The operation safety and the functionality of this device can only
be guaranteed if you follow both the general safety guidelines and the
applying laws to prevent accidents as well as the precautions given in this
user manual. We are not liable for any damages which occur due to
inappropriate usage or faulty operation of this device.
Make sure that anyone using this device has read and understood this user
manual.
Keep this user manual in a safe place where it can be referred to as required
at any time.
The unit complies with the relevant EU guidelines.
The unit is subject to the EU guidelines 2012/19/EU (WEEE Directive).
Company address:
Wassermann Dental-Maschinen GmbH
Rudorffweg 15-17
21031 Hamburg, Germany
Phone : +49 (0)40 730 926 -20/ -24 Fax.: +49 (0)40 730 37 24
export@wassermann-dental.com
www.wassermann.hamburg
www.facebook.com/WassermannDentalmaschinen
Features
- Compact table-top unit for pressure polymerisation up to 105°C in a water bath
- Very large capacity Ø 240 mm, usable depth 160 mm, suitable for extra large articulators
- Touch screen for fast and easy operation
- 13 programmes, each with three individually configurable polymerisation stages
- The device can also be operated as a pressure compactor
- Rapid draining: prompt switch between the two modes
- Separate menu for each mode
- Intelligent workflow preset by the software helps the user to ensure operational safety
- Exact remaining time displayed
- Powerful, electronically controlled heating for optimal water temperature
- Pressure relief valve for protection in case of malfunction (7 bar)
- Safety lid with bayonet lock
- Dry-run protection
- Engraving in container indicating max. water level
- Time and cost saving due to thermal insulation
- Made almost entirely of stainless steel
- Low maintenance, easy to service
Safety Symbols used in this Manual
Warning!
This is a warning of risk situations and dangers.
Failure to observe this warning could be life-threatening. These warnings has
to be observed.
Information!
This symbol draws your attention to specific features that has to be observed.
Safety Guidelines
Configuring and operating this equipment requires precise knowledge and observance of the instructions in this user manual. The equipment is designed only for its intended application.
WARNING:
Servicing and repairs should be carried out only by authorised specialists.
Disconnect the power plug before starting any maintenance work.
Make sure that the equipment is connected to the correct power source.
The safety valve on the pressure tank opens automatically if the
operating pressure is more than 7 bar.
The pressure can be activated only if the door is closed completely,
since a safety door contact is installed.
Dry-run protection, which switches off at 112°C (brown button on the left
side of the device).
WARNING: Risk of scalding during the Hot Polymerisation.
Never open the lid while the unit is under pressure!
Open the lid slowly!
Display information have to be followed!
For safety-related reasons (hot water, pressure), only the valves and hoses
meant for this purpose must be used for draining. If parts of different design
are used/installed (self-made), there is a risk of scalding!
For safety reasons, the pressure setting is limited to max. 4 bar (at 20 ° C).
Attention: When working above 100°C, the pressure can increase by approx. 2
bar.
Note: Hence the unit should be filled only with water and should be pre-
heated to the desired working temperature, before the actual work process.
To do this, select the preheating temperature and pressure accordingly.
Responsibility for Operation or Damage
The responsibility for operating the device lies exclusively with the owner or
user if said device is incorrectly serviced, maintained or altered by persons
not employed by an authorised dealer or if the device is used in a manner
contrary to its specified purpose.
The unit has to be maintained and operated in accordance with this user
manual.
Wassermann Dental-Maschinen GmbH is not responsible for damage arising from
the nonobservance of these instructions.
Warranty and responsibility provisions contained in the sales and supply
conditions of Wassermann Dental-Maschinen GmbH are not extended by these
instructions.
Application
The Polyquick 260-T105 is a pressure polymerisation unit for hot and cold
polymerisation of plastics. It can be set to a maximum temperature of 105°C
and the polymerisation pressure can be adjusted to a maximum of 4 bar. In
automatic mode, the polymerisation time can be preselected up to 99 hours. It
has a particularly large pressure pot volume and can hence also accommodate
extra large articulators or up to 6 brass flasks including clamp. A touch
screen and the newly developed control panel allow the user to set the
relevant parameters quickly and easily using self-explanatory symbols.
The Polyquick 260-T105 can also be used as a pressure compactor without
heating.
Only use the device for this type of application.
Before Starting
Transport
Before transporting the unit, ensure that it has been unplugged from the power
socket.
Make sure that it is packed correctly in order to avoid accidental damage.
Be sure to check for any transport damage when unpacking the goods. Note down
any damage if found.
Installation
Open the box, remove the packing materials, and carefully lift out the device
and accessories. Check the included accessories.
The device has to stand horizontally on a steady and even surface.
Install the device in a place where it will not block the working area and the
functionality (take the dimensions into account).
Do not install the unit outdoors or in places without proper ventilation.
Before start-up, be sure the device reaches room temperature.
It is advisable to install the unit firmly on the drainage system, since very
hot water can drain off from it. This also simplifies the process of complete
emptying, e.g. for cleaning.
Storage
The unit location is very important when it comes to workplace safety, even if it is only to be set up there for a short period. The room should be dry, well ventilated and vibrationfree. An even temperature and wooden supports also help. The unit should not be stored or set up outdoors. If the unit is to be stored for an extended period, protect it from moisture and dust.
Installation/ Start-up
Before commissioning, please install the following connections: Install the pressure hose included in the scope of supply and provide for an appropriate outlet.
After setting the workpiece, fill the device with tap water, at most up to the
engraving in the tank. Establish a pneumatic connection of minimum 2 bar. If
the pneumatic pipe is long, pay attention to the hose cross-section (internal
diameter min. 6 mm). Every pneumatic pipe must always have a maintenance unit.
Ensure a constant pressure by using the pressure reducer.
Finally, insert the plug into the socket – make sure that the voltage
specification of the device and the mains network is identical.
Never operate the device without water, since the heating unit immediately
heats up to 60°C or to the preheating temperature when the power switch is
activated. If required, the heating unit can be switched off via the display.
Setting the Correct Pressure
Before filling the tank, remove the accessories and set the correct
polymerisation pressure. Proceed as follows (image, see 7.2):
Temperatures higher than 100 °C: Working is only possible under pressure.
If no pressure is set, the heating switches off at 100 °C.
Hence working without pressure is only possible at temperatures below 100 °C.
- Activate main switch (green rocker switch).
- Close lid.
- Touch the Pressure key
5
(image, see 7.2). (3 bar default). - If the pressure is very less (min. 2 bar), increase the pressure (max.4 bar at 20°C) using the black regulator on the back side. Release the regulator by pulling to do the setting and press it in once again.
- If the pressure is too high, switch off the pressure
5
(image, see 7.2) and release the pressure. Turn the pressure regulator on the back side in the “minus” direction several times; then, apply pressure to the tank. You can set the desired pressure by turning the regulator in the “plus” direction (Pressure regulator at 20 °C: max. 4 bar)
For safety reasons, the pressure setting is limited to max. 4 bar (at 20 ° C).
Attention: When working above 100°C, the pressure can increase by approx. 2
bar.
Note: Hence the unit should be filled only with water and should be pre-heated
to the desired working temperature, before the actual work process.
To do this, select the preheating temperature and pressure accordingly.
Start-up
Define the unit’s main parameters before switching it on. These settings are
performed in the Set-Up Menu 3. Familiarise yourself with the system’s
various options before starting the polymerisation process. This particularly
applies to functions 7 and 14.
(Please see „setup menu page 1“ on page 10 of this manual).
Preheating temperatures above 100 °C are not stored. When restarting, the
preheating temperature is 60 °C.
The Start menu in Polymerisation mode
*Inverted (background is black) = ON
The Start menu in Compactor mode (without heater)
Setup menu page 1
In the Setup menu the following settings can be made:
Setup menu page 2
Description of features 5 – 14
Pressure ON/OFF
Bevore starting the programme you have to choose the pressure.
Temperatures higher than 100 °C: Working is only possible under pressure.
If no pressure is set, the heating switches off at 100 °C.
Hence working without pressure is only possible at temperatures below 100 °C.
Programme parameter und Polymerisation
Defines the parameters for the programme temperature and time.
You can store 13 programmes counter (1-13).
In every counter up to three polymerisation stages 1-3 can be mapped.
Touch the numbers to change the values.
Polymerisation level 1 (Programme 1):
Examples of level 2 and level 3 :
Touch the arrow (Start) to start the process of polymerisation.
Pause or cancel the programme by pressing the “Stop/Pause”symbol.
In programme 2-13 the practice is the same.
The Polymerisation
Preheating temperature
Defines the target temperature once the main switch has been activated. The
pre-heat temperature also applies to waiting times and rest periods. Default:
60°C
If the programme temperature selected is higher, the temperature reverts back
to this value when the programme ends.
After defining the desired working mode 14 , it may make sense to adjust
the pre-heat temperature of the programme temperature. This avoids long heat-
up or cool-down phases.
Preheating temperatures above 100 °C are not stored. When restarting, the
preheating temperature is 60 °C.
Clock timer
This defines the weekday activation and deactivation times.
Programmes running currently have a higher priority; they are not switched off until complete.
Sleep button display
When you press the Sleep button, the device goes into the Sleep mode (all functions except the Timer function are switched off). The device gets actuated when you press the touch display or with the timer and heats up or cools down to the preset pre-heating temperature.
Heater ON/OFF
Choosing the pressurepolymerisation mode (pressure and temperature) or
Compactor mode (pressure without heater).
Selection of the temperature unit (Fahrenheit or Celsius) Pictograms for the time and date function cycle.
Selection of the working mode
Defines whether the polymerisation process should start immediately or only
after the heating value has been reached.
(a) (Default) The device heats to the temperature programmed under 6.
After reaching this value, the result measured is validated by a control
cycle. If the value is confirmed, the wax polymerisation process will start.
(b) The programmed polymerization begins immediately.
Information
Here the parameters of the touch screen, sensors, hardware and software are
displayed.
Operation
General Operating Instructions
All instructions for using the unit, whether in verbal or written form, are
based on our own experience and experimentation and can only be regarded as
guidelines.
Please put on heat-resistant gloves
During longer periods of non-use please disconnect the power plug.
Operation
Prevent locking problems:
The sliding parts of the lid/ container (bayonet lock) must be greased with
Vasiline if the lid is difficult to move.
The O-ring has to be used with special lubricant Super O-Lube or similar.
- Open the lid.
- First set the workpieces, then fill the water till all the polymerisation parts are covered (max. up to the engraving = Marking in the tank) and close the lid up to the stop.
- Switch on the main switch, the device will automatically heat up to 60°C or up to the value selected as the pre-heating temperature. If warning A appears, close the lid.
- Activate the pressure 5.
- The process of polymerisation starts when you touch the symbol 6. After the process of polymerisation has ended, the temperature falls/rises to the preset temperature again.
Troubleshooting
Fault | Cause | Solution |
---|---|---|
Polyquick 260- heat T105 does not | Lid is open (Fig. 1) | Close lid against |
safety contact.
Temperatur-programme- parameter is to low| Adapt the parameters.
Preheating temperature is to low| Adapt the parameters.
Heater faulty| Please contact our service department.
Pressure does not increase in chamber| Lid seal 0-ring faulty| Check 0-ring
and replace if necessary.
Pressure regulator set to 0 (pressure gauge).| Check pressure and adjust if
necessary.
Lid not fully closed| Close lid against safety contact or contact faulty.
Temperature monitor off or faulty| Press in safety knob (left side of unit)
and/or replace knob. Please contact our service department.
Display blank| Temperature monitor off or faulty| Press in safety knob (left
side of unit) and/or replace knob. Please contact our service department.
Touchscreen malfunction| Calibrate the touch display| Switch on the unit.
While the logo of
Wassermann appears please contact the touch display two times.
Then you will reach the self-explanatory touch panel-calibration*
(Page 17, Fig. 2-3).
Touchscreen under pressure| Please contact our service department.
| Screen protector damaged?| Please contact our service department.
Unit does nothing| Blown fuse| Check the two fuses seperately.
Disconnect the power plug before replacing blown fuse at the rear of unit.
Item no. 582013
Fuse 6,3 AM/250 V D 5 x 20
If the above recommendations do not solve the problem, contact your dental depot or our service department.
* VIDEO: CALIBRATION
Care and Maintenance
Cleaning
Disconnect the power plug before starting any maintenance work.
The identification plate has always to be kept in easily legible condition and
has not to be removed.
Remove external dirt from time to time with some form of cold cleaner.
Use only cold cleaners to avoid damaging the paintwork or removing the
lettering.
Maintenance
At regular intervals, check the cover gasket (item no. 380063) for damages;
for this purpose, remove the gasket ring (reddish brown silicon ring) from the
ring groove of the pressure tank using a pair of blunt tweezers. The ring
groove for holding the gasket ring must always be kept clean.
The O-ring must be used with special lubricant Super O-Lube or similar.
The sliding parts of the lid/ container (bayonet lock) must be greased with
Vasiline if the lid is difficult to move.
Check the functioning of the lid safety switch by opening and closing the lid
multiple times; a soft click must be audible while closing or opening.
Change the sealing ring every 18 months.
Repairs
Servicing or repairs to the unit has only to be carried out by qualified
technicians.
Only original spare parts are to be used. Responsibility for the product is
voided if unauthorised persons alter it or if inappropriate components are
installed.
Spare Parts
If necessary please contact our service hotline phone.
Service Hotline 0049 (0)40 / 730 926 -20/ -24
Scope of Delivery/ Accessories
Included parts | Item no. |
---|---|
Flask bracket support | 170414 |
Compressed air hose, 2 m | 320026 |
Discharge hose, 2 m | 320031 |
Fuse 6,3 AM/250 V D 5 x 20, 2 pcs. | — |
Accessories/ Spare parts | Item no. |
--- | --- |
Flask bracket support | 170414 |
Compressed air hose per meter | 320026 |
Discharge hose per meter | 320031 |
On request: flask basket as special accessory | — |
Fuse 6,3 AM/250 V D 5 x 20 | 582013 |
Warranty
The warranty period for our equipment is 12 months. If faults occur within the
warranty period, contact your dental depot or get in touch directly with our
service department.
Your equipment should only be operated in perfect condition. If faults occur
which could harm operators or third parties, the unit should not be used until
it has been fixed.
This warranty does not cover damage caused by improper use, external
mechanical causes, transport damage or interference with the unit by
unauthorized persons.
Technical Data
Polyquick 260-T105 | Item no.: 170993 |
---|---|
Voltage | 220–240 V / 50/60 Hz |
Power consumption | 4.3 A |
Output | 1100 W |
W x H x D | 420 x 450 (open=490) x 510 (+20 for air socket) mm |
Weight | 41 kg |
Compressed air connection | min. 2 bar |
Operating pressure | max. 4 bar (limiting at 20°C) |
Pressure chamber | Ø 240 mm, depth 175 mm (usable depth 160 mm) |
The noise level of the unit amounts to ≤ 70 dB (A).
Other voltages on request.
Technical changes reserved.
Disposing of the Unit
The unit has to be disposed by an authorized recycling operation. The selected
company
has to be informed of all possibly health-hazardous residues in the unit.
Information on Disposal for Countries within the EU
To conserve and protect the environment, prevent environmental pollution and
improve the recycling of raw materials, the European Commission adopted a
directive that requires the manufacturer to accept the return of electrical
and electronic units for proper disposal or recycling.
Within the European Union units with this symbol should not therefore be
disposed of in unsorted domestic waste.
For more information regarding proper disposal please apply at your local
authoritie.
EU Declaration of Conformity
in accordance with 2014/35/EU (Low Voltage Directive) and 2014/30/EU (Electromagnetic Compatibility Directive) and 2006/42/EC (Machinery Directive) and 2011/65/EU (RoHS Directive) and 2014/68/EU (Pressure Units Directive)
Manufacturer:| WASSERMANN
Dental-Maschinen GmbH
Rudorffweg 15-17
21031 Hamburg
Germany
---|---
Model:| Polyquick 260-T105
Applicable standards:| DIN EN 61010-1
DIN EN 61000-6-3
DIN EN 61000-6-1
DIN 45635-1
DIN EN 60335-1
Product description:| Polimerisation
Unit and Compactor for dental applications
Item no. 170993
Hiermit wird bestätigt, dass die oben bezeichnete Maschine den genannten EU-
Richtlinien entspricht. Diese Erklärung wird ungültig, falls die Maschine ohne
unsere Zustimmung verändert wird.
This is to confirm that the above mentioned machine complies with the
described EU guidelines. This declaration becomes invalid if the machine is
modified without our approval.
https://www.wassermann.hamburg/
Wassermann Dental-Maschinen GmbH Rudorffweg 15-17 • 21031 Hamburg – Germany
Phone: +49 (0)40 730 926-20/ -24
Fax: +49 (0)40 730 37 24 export@wassermann-dental.com •
www.wassermann.hamburg
/WassermannDentalmaschinen
Documents / Resources
|
WASSERMANN Polyquick 260-T105 Pressure Polymerisation Unit and
Compactor
[pdf] User Manual
Polyquick 260-T105, Pressure Polymerisation Unit and Compactor
---|---
References
- Dental.com | Dental Photo Assessments Powered by AI
- Home
- Index of /fileadmin/Redaktion/Service_Kunden/Touchscreen_Kalibrierung