APOGEE SQ-622 Extended Range PFD Sensor Owner’s Manual
- June 5, 2024
- APOGEE
Table of Contents
- CERTIFICATE OF COMPLIANCE
- INTRODUCTION
- SENSOR MODELS
- SPECIFICATIONS
- DEPLOYMENT AND INSTALLATION
- CABLE CONNECTORS
- OPERATION AND MEASUREMENT
- MAINTENANCE AND RECALIBRATION
- TROUBLESHOOTING AND CUSTOMER SUPPORT
- RETURN AND WARRANTY POLICY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
OWNER’S MANUAL
EXTENDED RANGE PFD SENSOR
Models SQ-622 and SQ-625
Rev: 28-Oct-2020
CERTIFICATE OF COMPLIANCE
EU Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the
manufacturer:
Apogee Instruments, Inc.
721 W 1800 N
Logan, Utah 84321
USA
for the following product(s):
Models: SQ-622, SQ-625
Type: Extended Range PFD Sensor
The object of the declaration described above is in conformity with the
relevant Union harmonization legislation:
2014/30/EU Electromagnetic Compatibility (EMC) Directive
2011/65/EU Restriction of Hazardous Substances (RoHS 2) Directive
2015/863/EU Amending Annex II to Directive 2011/65/EU (RoHS 3)
Standards referenced during compliance assessment:
EN 61326-1:2013 Electrical equipment for measurement, control, and laboratory
use – EMC requirements
EN 50581:2012 Technical documentation for the assessment of electrical and
electronic products with respect to the restriction of hazardous substances
Please be advised that based on the information available to us from our raw
material suppliers, the products manufactured by us do not contain, as
intentional additives, any of the restricted materials including lead (see
note below), mercury, cadmium, hexavalent chromium, polybrominated biphenyls
(PBB), polybrominated biphenyls (PBDE), bis(2-Ethylhexyl) phthalate (DEHP),
butyl benzyl phthalate (BBP), dibutyl phthalate (DBP), and diisobutyl
phthalate (DIBP). However, please note that articles containing greater than
0.1% lead concentration are RoHS 3 compliant using exemption 6c.
Further note that Apogee Instruments does not specifically run any analysis on our raw materials or end products for the presence of these substances, but rely on the information provided to us by our material suppliers. Signed for and on behalf of Apogee Instruments, January 2021
Bruce Bugbee
President
Apogee Instruments, Inc.
INTRODUCTION
Radiation that drives photosynthesis is called photosynthetically active
radiation (PAR) and is typically defined as total radiation across a range of
400 to 700 nm. PAR is almost universally quantified as photosynthetic photon
flux density (PPFD) in units of micromoles per square meter per second (µmol
m² s², equal to microEinsteins per square meter per second) summed from 400 to
700 nm (total number of photons from 400 to 700 nm). However, ultraviolet and
far-red photons outside the defined PAR range of 400-700 nm can also
contribute to photosynthesis and influence plant responses (e.g., flowering).
Sensors that measure PPFD are often called quantum sensors due to the
quantized nature of radiation. A quantum refers to the minimum quantity of
radiation, one photon, involved in physical interactions (e.g., absorption by
photosynthetic pigments). In other words, one photon is a single quantum of
radiation. Sensors that function like traditional quantum sensors, but measure
a wider range of wavelengths can be thought of as an ‘extended range’ quantum
sensor.
Typical applications of traditional quantum sensors include incoming PPFD
measurement over plant canopies in outdoor environments or in greenhouses and
growth chambers and reflected or under-canopy (transmitted) PPFD measurement
in the same environments. The Extended Range PFD Sensor detailed in this
manual uses a detector that is sensitive to radiation up to about 1100 nm,
well beyond the range of wavelengths that influence photosynthesis and plant
responses. This means this particular sensor should only be used for photon
flux density measurements under LEDs.
Apogee Instruments SQ-600 series Extended Range PFD Sensors consist of a cast
acrylic diffuser (filter),
photodiode, and signal processing circuitry mounted in an anodized aluminum
housing. A cable to connect the sensor to a measurement device is also
included. SQ-600 series Extended Range PFD Sensors are designed for continuous
photon flux density measurements in indoor environments under LEDs. SQ-620
models output an analog signal that is directly proportional to photon flux
density. The analog signal from the sensor is directly proportional to the
radiation incident on a planar surface (does not have to be horizontal), where
the radiation emanates from all angles of a hemisphere.
SENSOR MODELS
This manual covers the amplified voltage output Extended Range PFD Sensors, models SQ-622 and SQ-625 (in bold below). Additional models are covered in their respective manuals.
Model | Signal |
---|---|
SQ-620 | Self-powered |
SQ-622 | 0-2.5 V |
SQ-624 | 4-20 mA |
SQ-625 | 0-5 V |
SQ-626 | USB |
SQ-627 | SDI-12 |
SQ-628 | Modbus |
The sensor model number and serial number are located on the bottom of the sensor. If the manufacturing date of a specific sensor is required, please contact Apogee Instruments with the serial number of the sensor.
SPECIFICATIONS
| SQ-622-SS| SQ-625-SS
---|---|---
Power Supply| 5 to 24 V DC| 5.5 to 24 V DC
Current Draw| 12 V is 57 µA
Sensitivity| 0.625 mV per µmol m-2 s-1| 1.25 mV per µmol m-2 s-1
Calibration Factor| 1.6 µmol m-2 s-1 per mV| 0.8 µmol m-2 s-1 per mV
Calibration Uncertainty| ± 5 % (see Calibration Traceability below)
Output Range| 0 to 2.5 V| 0 to 5 V
Measurement Range| 0 to 4000 µmol m-2 s-1
Measurement Repeatability| Less than 0.5 %
Long-term Drift (Non-stability)| Less than 2 % per year
Non-linearity| Less than 1 % (up to 4000 µmol m-2 s-1)
Response Time| Less than 1 ms
Field of View| 180°
Spectral Range| 340 to 1040 nm ± 5 nm (wavelengths where the response is
greater than 50 %; see Spectral Response below)
Directional (Cosine) Response| ± 2 % at 45° zenith angle, ± 5 % at 75° zenith
angle (see Directional Response below)
Azimuth Error| Less than 0.5 %
Tilt Error| Less than 0.5 %
Temperature Response| -0.11 ± 0.04 % per C
Uncertainty in Daily Total| Less than 5 %
Housing| Anodized aluminum body with acrylic diffuser
IP Rating| IP68
Operating Environment| -40 to 70 C; 0 to 100 % relative humidity; can be
submerged in water up to depths of 30 m
Dimensions| 30.5 mm diameter, 37 mm height
Mass (with 5 m of cable)| 140 g
Cable| 5 m of two-conductor, shielded, twisted-pair wire; TPR jacket; pigtail
lead wires; stainless steel (316), M8 connector
Warranty| 4 years against defects in materials and workmanship
Calibration Traceability
Apogee Instruments SQ-600 series PFD sensors are calibrated through side-by-
side comparison to the mean of four transfer standard PFD sensors under a
reference lamp. The transfer standard PFD sensors are recalibrated with a
quartz halogen lamp traceable to the National Institute of Standards and
Technology (NIST).
Spectral Response
Mean spectral response measurements of six replicate Apogee SQ-600 series Extended Range PFD Sensors. Spectral response measurements were made at 10 nm increments across a wavelength range of 300 to 1100 nm in a monochromator with an attached electric light source. Measured spectral data from each PFD sensor were normalized by the measured spectral response of the monochromator/electric light combination, which was measured with a spectroradiometer
Cosine Response
Directional, or cosine, the response is defined as the measurement error at a specific angle of radiation incidence. Error for Apogee SQ-600 series Extended Range PFD Sensor is approximately ± 2 % and ± 5 % at solar zenith angles of 45° and 75°, respectively.
DEPLOYMENT AND INSTALLATION
Mount the sensor to a solid surface with the nylon mounting screw provided. To accurately measure photon flux density incident on a horizontal surface, the sensor must be level. An Apogee Instruments model AL-100 leveling plate is recommended for this purpose. To facilitate mounting on a cross arm, an Apogee Instruments Model AL120 mounting bracket is recommended.
To minimize azimuth error, the sensor should be mounted with the cable pointing toward true north in the northern hemisphere or true south in the southern hemisphere. Azimuth error is typically less than 0.5 %, but it is easy to minimize by proper cable orientation.
In addition to orienting the cable to point toward the nearest pole, the sensor should also be mounted such that obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor. Once mounted, the blue cap should be removed from the sensor. The blue cap can be used as a protective covering for the sensor when it is not in use.
CABLE CONNECTORS
Apogee started offering cable connectors on some bare-lead sensors in March 2018 to simplify the process of removing sensors from weather stations for calibration (the entire cable does not have to be removed from the station and shipped with the sensor).
The ruggedized M8 connectors are rated IP68, made of corrosion-resistant marine-grade stainless steel, and designed for extended use in harsh environmental conditions.
Instructions
Pins and Wiring Colors: All Apogee connectors have six pins, but not all pins
are used for every sensor. There may also be unused wire colors inside the
cable. To simplify the datalogger connection, we remove the unused pigtail
lead colors at the datalogger end of the cable.
If a replacement cable is required, please contact Apogee directly to ensure ordering the proper pigtail configuration.
Alignment: When reconnecting a sensor, arrows on the connector jacket and
an aligning notch ensure proper orientation.
Disconnection for extended periods: When disconnecting the sensor for an
extended period of time from a station, protect the remaining half of the
connector still on the station from water and dirt with electrical tape or
another method.
Tightening: Connectors are designed
to be firmly finger-tightened only. There is an o-ring inside the connector
that can be overly compressed if a wrench is used. Pay attention to thread
alignment to avoid cross-threading. When fully tightened, 1-2 threads may
still be visible.
WARNING: Do not tighten the connector by twisting the black cable or
sensor head, only twist the metal connector (yellow arrows).
OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying or recording a millivolt signal (an input measurement range of approximately 0-2.5 V (SQ-622) or 0-5 V (SQ-625) is required to cover the entire range of PFD from the sun). To maximize measurement resolution and signal-to-noise ratio, the input range of the measurement device should closely match the output range of the extended range PFD sensor. DO NOT connect the sensor to a power source greater than 24 V DC.
Wiring for SQ-622 and SQ-625
Sensor Calibration
Apogee SQ-622 and SQ-625 Extended Range PFD Sensors have standard calibration
factors of exactly:
SQ-622: 1.6 µmol m-2 s-1 per mV
SQ-625: 0.8 µmol m-2 s-1 per mV
Multiply the calibration factor by the measured mV signal to convert sensor
output to photon flux density in units of µmol m-2 s-1:
Calibration Factor (0.8 µmol m-² s-¹ per mV) * Sensor Output Signal (mV) =
Photon Flux Density (µmol m-² s-¹ )
*0.8 2500 = 2000**
Example of photon flux density measurement with an Apogee model Extended Range
PFD sensor. The LED in this example Full yields a photon flux density on a
horizontal plane of 2000 µmol m -2 -1 s . This yields an output signal of 2500
mV for the 0-5 V option or an output signal of 1250 mV for the 0-2.5 V option.
This signal is converted to photon flux density by multiplying by the
calibration factor.
SQ-622 Sensor Output: 1250 mV Conversion Factor: 1.6 µmol m -2 s -1 per mV
SQ-625 Sensor Output: 2500 mV Conversion Factor: 0.8 µmol m -2 s-1 per mV
MAINTENANCE AND RECALIBRATION
Dust or organic deposits are best removed using water or window cleaner and a soft cloth or cotton swab. Salt deposits should be dissolved with vinegar and removed with an oft cloth or cotton swab. Blocking the optical path between the target and detector can cause low readings. Occasionally, accumulated materials on the diffuser of the upward-looking radiometer and in the apertures of the downward-looking radiometer can block the optical path in three common ways:
- Moisture or debris on the diffuser (upward-looking) or in the apertures (downward-looking).
- Dust during periods of low rainfall.
- Salt deposit accumulation from the evaporation of sea spray or sprinkler irrigation water.
Apogee Instruments’ upward-looking sensors have a domed diffuser and housing for improved self-cleaning from rainfall, but active cleaning may be necessary. Dust or organic deposits are best removed using water, or window cleaner, and a soft cloth or cotton swab. Salt deposits should be dissolved with vinegar and removed with a cloth or cotton swab. Salt deposits cannot be removed with solvents such as alcohol or acetone. Use only gentle pressure when cleaning the diffuser with a cotton swab or soft cloth, to avoid scratching the outer surface. The solvent should be allowed to do the cleaning, not mechanical force. Never use an abrasive material or cleaner on the diffuser.
It is recommended that two-band radiometers be recalibrated every two years. See the Apogee webpage for details regarding the return of sensors for recalibration (http://www.apogeeinstruments.com/tech-supportrecalibration- repairs/).
TROUBLESHOOTING AND CUSTOMER SUPPORT
Independent Verification of Functionality
Apogee SQ-622 and SQ-625 Extended Range PFD Sensors provide an amplified
voltage output that is proportional to incident photon flux density for the
340-1040 nm wavelength range. A quick and easy check of sensor functionality
can be determined using a DC power supply and a voltmeter. Power the sensor
with a DC voltage by connecting the positive voltage signal to the red wire
from the sensor and the negative (or common) to the black wire from the
sensor. Use the voltmeter to measure across the white wire (signal output) and
black wire (signal ground). Direct the sensor head toward a light source and
verify the sensor provides a signal. Increase and decrease the distance from
the sensor head to the light source to verify that the signal changes
proportionally (decreasing signal with increasing distance and increasing
signal with decreasing distance). Blocking all radiation from the sensor
should force the sensor signal to zero.
Compatible Measurement Devices (Dataloggers/Controllers/Meters)
SQ-600 Extended Range PFD Sensors are calibrated with a standard calibration
factor of 1.6 pmol m-2 s-¹ per mV (SQ-622) or 0.8 pmol m-2 s¹ per mV (SQ-625),
yielding a sensitivity of 0.6 mV per pmol m-2 (SQ-622) or 1.3 mV per pmol m-2
(SQ-625). Thus, a compatible measurement device (e.g., datalogger or
controller) should have a resolution of at least 0.6 mV (SQ-622) or 1.3 mV
(SQ-625) in order to provide photon flux density resolution of 1 gm& nr2
An example datalogger program for Campbell Scientific dataloggers can be found
on the Apogee webpage at
https://www.apogeeinstruments.com/downloads/#datalogger.
Cable Length
When the sensor is connected to a measurement device with high input
impedance, sensor output signals are not changed by shortening the cable or
splicing on an additional cable in the field. Tests have shown that if the
input impedance of the measurements device is greater than 1 mega-ohm there is
negligible effect on the calibration, even after adding up to 100 m of cable.
All Apogee sensors use shielded, twisted-pair cables to minimize
electromagnetic interference. For best measurements, the shield wire must be
connected to the earth’s ground. This is particularly important when using the
sensor with long lead lengths in electromagnetically noisy environments.
Modifying Cable Length
See the Apogee webpage for details on how to extend sensor cable length:
(http://www.apogeeinstruments.com/how-to-make-a-weatherproof-cable-splice/).
RETURN AND WARRANTY POLICY
RETURN POLICY
Apogee Instruments will accept returns within 30 days of purchase as long as
the product is in new condition (to be determined by Apogee). Returns are
subject to a 10 % restocking fee.
WARRANTY POLICY
What is Covered
All products manufactured by Apogee Instruments are warranted to be free from
defects in materials and craftsmanship for a period of four (4) years from the
date of shipment from our factory. To be considered for warranty coverage an
item must be evaluated by Apogee.
Products not manufactured by Apogee (spectroradiometers, chlorophyll content
meters, EE08-SS probes) are covered for a period of one (1) year.
What is Not Covered
The customer is responsible for all costs associated with the removal,
reinstallation, and shipping of suspected warranty items to our factory.
The warranty does not cover equipment that has been damaged due to the
following conditions:
- Improper installation or abuse.
- Operation of the instrument outside of its specified operating range.
- Natural occurrences such as lightning, fire, etc.
- Unauthorized modification.
- Improper or unauthorized repair.
Please note that nominal accuracy drift is normal over time. Routine
recalibration of sensors/meters is considered part of proper maintenance and
is not covered under warranty.
Who is Covered
This warranty covers the original purchaser of the product or other party who
may own it during the warranty period.
What Apogee Will Do
At no charge Apogee will:
- Either repair or replace (at our discretion) the item under warranty.
- Ship the item back to the customer by the carrier of our choice. Different or expedited shipping methods will be at the customer’s expense.
How To Return An Item
-
Please do not send any products back to Apogee Instruments until you have received a Return Merchandise Authorization (RMA) number from our technical support department by submitting an online RMA form at www.apogeeinstruments.com/tech-support-recalibration-repairs/. We will use your RMA number for tracking the service item. Call 435-245-8012 or email techsupport@apogeeinstruments.com with questions.
-
For warranty evaluations, send all RMA sensors and meters back in the following condition: Clean the sensor’s exterior and cord. Do not modify the sensors or wires, including splicing, cutting wire leads, etc. If a connector has been attached to the cable end, please include the mating connector – otherwise, the sensor connector will be removed in order to complete the repair/recalibration. Note: When sending back sensors for routine calibration that have Apogee’s standard stainless-steel connectors, you only need to send the sensor with the 30 cm section of cable and one-half of the connector. We have mating connectors at our factory that can be used for calibrating the sensor.
-
Please write the RMA number on the outside of the shipping container.
-
Return the item with freight pre-paid and fully insured to our factory address shown below. We are not responsible for any costs associated with the transportation of products across international borders.
Apogee Instruments, Inc.
721 West 1800 North Logan, UT
84321, USA -
Upon receipt, Apogee Instruments will determine the cause of failure. If the product is found to be defective in terms of operation to the published specifications due to a failure of product materials or craftsmanship, Apogee Instruments will repair or replace the items free of charge. If it is determined that your product is not covered under warranty, you will be informed and given an estimated repair/replacement cost.
PRODUCTS BEYOND THE WARRANTY PERIOD
For issues with sensors beyond the warranty period, please contact Apogee at
techsupport@apogeeinstruments.com
to discuss repair or replacement options.
OTHER TERMS
The available remedy of defects under this warranty is for the repair or
replacement of the original product, and Apogee Instruments is not responsible
for any direct, indirect, incidental, or consequential damages, including but
not limited to loss of income, loss of revenue, loss of profit, loss of data,
loss of wages, loss of time, loss of sales, accruement of debts or expenses,
injury to personal property, or injury to any person or any other type of
damage or loss. This limited warranty and any disputes arising out of or in
connection with this limited warranty (“Disputes”) shall be governed by the
laws of the State of Utah, USA, excluding conflicts of law principles and
excluding the Convention for the International Sale of Goods. The courts
located in the State of Utah, USA, shall have exclusive jurisdiction over any
Disputes.
This limited warranty gives you specific legal rights, and you may also have
other rights, which vary from state to state and jurisdiction to jurisdiction,
and which shall not be affected by this limited warranty. This warranty
extends only to you and cannot by transferred or assigned. If any provision of
this limited warranty is unlawful, void or unenforceable, that provision shall
be deemed severable and shall not affect any remaining provisions. In case of
any inconsistency between the English and other versions of this limited
warranty, the English version shall prevail. This warranty cannot be changed,
assumed, or amended by any other person or agreement
APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA
TEL: 435-792-4700 | FAX:
435-787-8268 | WEB:
APOGEEINSTRUMENTS.COM
Copyright © 2021 Apogee Instruments, Inc.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>