apogee SB-100 Barometric Pressure Sensor Owner’s Manual
- June 4, 2024
- APOGEE
Table of Contents
apogee SB-100 Barometric Pressure Sensor
INTRODUCTION
- Pressure is defined as force per unit area applied to a surface in a direction perpendicular to the surface. Barometric pressure, or atmospheric pressure, is the force per unit area exerted on Earth’s surface by the mass of air overlying the surface. High pressure indicates more atmospheric air mass over a given area, whereas low pressure indicates less atmospheric air mass. Barometric pressure is strongly dependent on elevation, and decreases as elevation increases, due to less overlying air above the surface (shorter column of air) at higher elevations.
- Barometers are sensors that measure barometric pressure. Aneroid (without liquid) barometers are often electronic and typically use capacitive elements to sense pressure, with the major advantage of capacitive sensing mechanisms being minimal temperature dependence. Capacitive sensing circuits output a voltage that is related to pressure via sensor-specific calibrations. Typical units for barometric pressure are kilopascals [kPa] and millibars [mb] (hectopascals [hPa], pounds per square inch [psi], and millimeters of mercury [mm Hg] or inches of mercury [in Hg] have also been used).
- Barometric pressure is a fundamental weather variable. Typical applications of barometers include barometric pressure measurement in weather networks, often for weather forecasting. Barometric pressure is also an input variable required for calculation of evapotranspiration. Additionally, barometric pressure measurements are used to correct the output of sensors that are sensitive to pressure fluctuations (e.g., Apogee Instruments oxygen sensors).
- Apogee Instruments SB-100 barometric pressure sensors consist of a silicon capacitive sensing element and signal processing circuitry mounted in a compact epoxy plastic/stainless steel housing, and lead wires to connect the sensor to a measurement device. Sensors are weather resistant and are designed for continuous barometric pressure measurement when housed inside the same enclosure as the measurement device (e.g., datalogger or controller). SB-100 pressure sensors output an analog voltage that is directly proportional to barometric pressure.
SENSOR MODELS
The SB-100 barometric pressure sensor is the only pressure sensor model offered by Apogee Instruments.
Sensor model number, serial number, and production date are located on a label between the sensor and pigtail lead wires.
SPECIFICATIONS
| SB-100
---|---
Measurement Range| 15 to 115 kPa (approximate)
Maximum Pressure
Exposure
| 400 kPa (exposure beyond this limit may cause permanent damage to sensor)
Sensitivity| 45.9 mV per kPa; 0.459 mV per 0.01 kPa (approximate)
Calibration Factor| 0.0218 kPa per mV (generic slope; reciprocal of
sensitivity) and 11.4 kPa (generic intercept)
Measurement
Uncertainty
| ± 1.5 kPa (with generic calibration coefficients)
Measurement Repeatability| Less than 0.1 %
Long-term Drift
(Non-stability)
| Less than 1 % per year
Non-linearity| Less than 1 %
Warm-up Time| 20 ms
Response Time| 1 ms
Temperature
Response
| Less than 0.002 % per C for temperatures greater than 0 C, -0.015 % per C for
temperatures less than 0 C
Operating Environment| -40 to 80 C; 0 to 100 % relative humidity (non-
condensing)
Input Voltage
Requirement
| 5 V DC
Output Voltage Requirement| 0 to 5 V DC
Current Draw| 7 mA DC
Dimensions| 1.6 cm diameter
Mass| 5 g
Cable| 12 cm pigtail
DEPLOYMENT AND INSTALLATION
Apogee SB-100 barometric pressures sensors are designed to be mounted inside the datalogger enclosure, where they are protected from the elements, specifically, precipitation, condensation, and dynamic pressure caused by wind. Lead wires are short and allow the sensor to remain near the datalogger wiring panel (as shown below). The datalogger enclosure should not be air tight, as the pressure sensor must be exposed to an environment where the pressure varies with ambient pressure. As a result, vent holes in the enclosure are required and should be in the bottom of the enclosure to minimize the impact of dynamic pressure caused by wind.
OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller) capable of inputting 5 V DC, and measuring and displaying or recording a millivolt (mV) signal (an input measurement range of approximately 100-4800 mV is required to cover the entire pressure range of the sensor). In order to maximize measurement resolution and signal-to-noise ratio, the input range of the measurement device should closely match the output range of the barometric pressure sensor. DO NOT connect the black wire to a power source; applying voltage may damage the sensor.
VERY IMPORTANT: Apogee changed all wiring colors of our bare-lead sensors in March 2018. To ensure proper connection to your data device, please note your serial number and then use the appropriate wiring configuration below. Wiring for SB-100 Serial Numbers 2405 and above
- White: Signal output; connect to single-ended channel
- Red: Power input; connect to 5 V
- Black: Ground for signal output; connect to analog ground
Old Wiring for SB-100 Serial Numbers range 0-2404
- Red: Power input; connect to 5 V
- Black: Signal output; connect to single-ended channel
- Green: Ground for signal output; connect to analog ground
Sensor Calibration
All Apogee SB-100 barometric pressure sensors have a generic calibration
factor (slope): 0.0218 kPa per mV
and generic offset (intercept): 11.4 kPa
Multiply this calibration factor by the measured voltage signal, and then add
the offset, to convert sensor voltage output to barometric pressure (in units
of kilopascals, kPa): Calibration Factor (0.0218 kPa per mV) Sensor Output
Signal (mV) + Offset (kPa) = Barometric Pressure (kPa) 0.0218 4125 + 11.4 =
101.325
The calibration factor and offset are variable from sensor to sensor, and a
sensor-specific calibration factor and offset can be derived by plotting
measured pressure (from a reference) against the measured voltage signal, and
then fitting a linear equation to the results (see Maintenance and
Recalibration section below). The generic calibration coefficients given above
yield barometric pressure in units of kilopascals [kPa], but multiple units
are available and used to report barometric pressure measurements. The
following tables provide calibration factors and offsets in other common
pressure units and conversion of kPa to other common units.
Generic Calibration Factors and Offsets for Common Pressure Units
Units | Calibration Factor (Slope) | Offset (Intercept) |
---|---|---|
kilopascals [kPa] | 0.0218 | 11.4 |
hectopascals [hPa] | 0.218 | 114 |
millibars [mb] | 0.218 | 114 |
pounds per square inch [psi] | 0.00316 | 1.65 |
millimeters of mercury [mm Hg] | 0.164 | 85.5 |
inches of mercury [in Hg] | 0.00643 | 3.36 |
Standard Atmospheric Pressure in Various Units and Conversion Factors for Pressure Units (Relative to Pressure in Kilopascals)
kilopascals [kPa]| hectopascals [hPa]| millibars [mb]| pounds per square inch
[psi]
| millimeters of mercury [mm
Hg]
| inches of mercury [in Hg]
---|---|---|---|---|---
101.325| 1013.25| 1013.25| 14.70| 760.0| 29.92
Conversion Factor| | | | |
1| 10| 10| 0.145| 7.50| 0.295
Normalizing to Sea Level
Before reporting, barometric pressure is often normalized to sea level
(common reference pressure). A simple equation can be used to calculate the
difference in barometric pressure (DP, in kPa) at a given elevation and the
equivalent pressure at sea level:
where E is elevation is meters. To normalize measured barometric pressure to sea level pressure, the pressure difference (DP) from the above equation should be added to the measured pressure.
MAINTENANCE AND RECALIBRATION
- To maintain proper sensor function, the small ports (holes) on each side of the sensor housing (one port in the epoxy plastic and one port in the stainless steel plate) should be kept unobstructed.
- Apogee SB-100 barometric pressure sensors are weather resistant, but not weatherproof. They should be housed in a weatherproof enclosure, where water does not condense. Desiccant can be used to keep water from condensing inside the enclosure. The enclosure must have vents, so the internal air remains in equilibrium with the atmosphere, otherwise the pressure inside the enclosure will not be representative of ambient conditions.
- Apogee SB-100 barometric pressure sensors are not factory calibrated, but come with a generic calibration. A custom calibration can be derived by plotting pressure measurements from a reference barometric pressure sensor versus the measured voltage output from an SB-100, then fitting a linear equation to the data. The slope of the linear equation is the calibration factor and the intercept is the offset.
Example of SB-100 barometric pressure sensor calibration. Barometric pressure measurements from a reference sensor are plotted against the output signal voltage from the SB-100. A linear equation fitted to the data yields the calibration factor and offset, 0.0222 and 9.93, respectively, for this example.
TROUBLESHOOTING AND CUSTOMER SUPPORT
-
Independent Verification of Functionality
Apogee SB-100 barometric pressure sensors require 5 V DC input for operation. A quick and easy check of sensor functionality can be determined using a DC power supply and a voltmeter. Power the sensor with 5 V DC by connecting the positive voltage signal to the red wire from the sensor and the negative (or common) to the black wire from the sensor. Use the voltmeter to measure across the white wire (output signal) and black wire. The sensor should read approximately 4100 mV at sea level, with the voltage output decreasing by approximately 50 mV per 100 meters above sea level. -
Compatible Measurement Devices (Dataloggers/Controllers/Meters)
SB-100 barometric pressure sensors have a calibration factor of approximately 0.0218 kPa per mV, yielding a sensitivity of approximately 0.459 mV per 0.01 kPa. Thus, a compatible measurement device (e.g., datalogger or controller) should have resolution of at least 0.459 mV, in order to provide pressure resolution of 0.01 kPa.
The signal output range of SB-100 sensors is approximately 100 to 4800 mV. A compatible measurement device should also have a full scale range spanning the sensor signal output range.
An example datalogger program for Campbell Scientific dataloggers can be found on the Apogee webpage at http://www.apogeeinstruments.com/content/Barometric- Pressure-Sensor.CR1. -
Modifying Cable Length
When the sensor is connected to a measurement device with high input impedance, sensor output signals are not changed by splicing on additional cable in the field. Tests have shown that if the input impedance of the measurement device is 1 mega-ohm or higher then there is negligible effect on the SB-100 pressure sensor calibration, even after adding up to 100 m of cable. See Apogee webpage for details on how to extend sensor cable length (http://www.apogeeinstruments.com/how-to-make-a-weatherproof-cable-splice/). For cable extensions, shielded, twisted pair cable is recommended, in order to minimize electromagnetic interference. This is particularly important for long lead lengths in electromagnetically noisy environments.
References
- apogeeinstruments.com/content/Barometric-Pressure-Sensor.CR1
- How to Make a Weatherproof Cable Splice
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>