apogee INSTRUMENT SQ-500 Full-Spectrum Quantum Sensor Owner’s Manual
- June 3, 2024
- apogee INSTRUMENT
Table of Contents
OWNER’S MANUAL
QUANTUM SENSOR
Model SQ-500
(including SS model, formerly gold)
CERTIFICATE OF COMPLIANCE
EU Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the
manufacturer: Apogee Instruments, Inc.
721 W 1800 N
Logan, Utah 84321
USA
for the following product(s):
Models: SQ-500
Type: Quantum Sensor
The object of the declaration described above is in conformity with the
relevant Union harmonization legislation:
2014/30/EU Electromagnetic Compatibility (EMC) Directive
2011/65/EU Restriction of Hazardous Substances (RoHS 2) Directive
2015/863/EU Amending Annex II to Directive 2011/65/EU (RoHS 3)
Standards referenced during compliance assessment:
EN 61326-1:2013 Electrical equipment for measurement, control and laboratory
use – EMC requirements
EN 50581:2012 Technical documentation for the assessment of electrical and
electronic products with respect to the restriction of hazardous substances
Please be advised that based on the information available to us from our raw
material suppliers, the products manufactured by us do not contain, as
intentional additives, any of the restricted materials including lead (see
note below), mercury, cadmium, hexavalent chromium, polybrominated biphenyls
(PBB), polybrominated diphenyls (PBDE), bis(2-ethylhexyl) phthalate (DEHP),
butyl benzyl phthalate (BBP), dibutyl phthalate (DBP), and diisobutyl
phthalate (DIBP). However, please note that articles containing greater than
0.1% lead concentration are RoHS 3 compliant using exemption 6c.
Further note that Apogee Instruments does not specifically run any analysis on
our raw materials or end products for the presence of these substances, but
rely on the information provided to us by our material suppliers. Signed for
and on behalf of:
Apogee Instruments, June 2020
Bruce Bugbee
President
Apogee Instruments, Inc.
INTRODUCTION
Radiation that drives photosynthesis is called photosynthetically active
radiation (PAR) and is typically defined as total radiation across a range of
400 to 700 nm. PAR is almost universally quantified as photosynthetic photon
flux density (PPFD), the sum of photons from 400 to 700 nm in units of
micromoles per square meter per second (µmol m -2 -1 ). While microEinsteins
and micromoles are equal (one Einstein = one mole of photons), the Einstein is
not an SI unit, so expressing PPFD as µmolm s, equal to microEinsteins m -2 -1
s is preferred. Daily total PPFD is typically reported in units of moles of
photons per square meter per day (mol m -2 -1 s ) and is often called daily
light integral (DLI). -2 -1 d
The acronym PPF is also used and refers to the photosynthetic photon flux. The
acronyms PPF and PPFD refer to the same variable. Both terms are used because
there is not universal definition of the term flux. Flux is sometimes defined
as per unit area per unit time and sometimes defined as per unit time only.
PPFD is used in this manual.
Sensors that measure PPFD are often called quantum sensors due to the
quantized nature of radiation. A quantum refers to the minimum quantity of
radiation, one photon, involved in physical interactions (e.g., absorption by
photosynthetic pigments). In other words, one photon is a single quantum of
radiation.
Typical applications of quantum sensors include measurement of incident PPFD
on plant canopies in outdoor environments or in greenhouses and growth
chambers, and reflected or under-canopy (transmitted) PPFD measurement in the
same environments.
Apogee Instruments SQ-500 series quantum sensors consist of a cast acrylic
diffuser (filter), interference filter, photodiode, and signal processing
circuitry mounted in an anodized aluminum housing, and a cable to connect the
sensor to a measurement device. SQ-500 series quantum sensors are designed for
continuous PPFD measurement
in indoor or outdoor environments. The SQ-500 model quantum sensor outputs a
voltage that is directly proportional to PPFD. The voltage output by the
sensor is directly proportional to the radiation incident on a planar surface
(does not have to be horizontal), where the radiation emanates from all angles
of a hemisphere.
SENSOR MODELS
This manual covers the unamplified analog output quantum sensor, model SQ-500
(listed in bold below).
Additional models are covered in their respective manuals.
Model | Signal |
---|---|
SQ-500 | 0-40 mV |
SQ-512 | 0-2.5 V |
SQ-515 | 0-5 V |
SQ-520 | USB |
SQ-521 | SDI-12 |
SQ-522 | Modbus |
Sensor model number and serial number are located near the pigtail leads on the sensor cable. If the manufacturing date of a specific sensor is required, please contact Apogee Instruments with the serial number of the sensor.
SPECIFICATIONS
| SQ-500-SS
---|---
Power Supply| Self-powered
Sensitivity| 0.01 mV per µmol m-2 s-1
Calibration Factor| 100 µmol m-2 s-1 per mV
Calibration Uncertainty| ± 5 % (see Calibration Traceability below)
Measurement Range| 0 to 4000 µmol m-2 s-1
Measurement Repeatability| Less than 0.5 %
Calibrated Output Range| 0 to 40 mV
Long-term Drift (Non-stability)| Less than 2 % per year
Non-linearity| Less than 1 % (up to 4000 µmol m-2 s-1)
Response Time| Less than 1 ms
Field of View| 180°
Spectral Range| 389 to 692 nm ± 5 nm (wavelengths where response is greater
than 50 %)
Spectral Selectivity| Less than 10 % from 412 to 682 ± 5 nm (see Spectral
Response below)
Directional (Cosine) Response| ± 2 % at 45° zenith angle, ± 5 % at 75° zenith
angle (see Directional Response below)
Azimuth Error| Less than 0.5 %
Tilt Error| Less than 0.5 %
Temperature Response| -0.11 ± 0.04 % per C (see Temperature Response below)
Uncertainty in Daily Total| Less than 5 %
Detector| Blue-enhanced silicon photodiode
Housing| Anodized aluminum body with acrylic diffuser
IP Rating| IP68
Operating Environment| -40 to 70 C; 0 to 100 % relative humidity; can be
submerged in water up to depths of 30 m
Dimensions| 24 mm diameter, 37 mm height
Mass (with 5 m of cable)| 100 g
Cable| 5 m of two conductors, shielded, twisted-pair wire; TPR jacket; pigtail
lead wires; stainless steel (316), M8 connector located 25 cm from the sensor
head
Calibration Traceability
Apogee Instruments SQ-500 series quantum sensors are calibrated through side-
by-side comparison to the mean of four transfer standard quantum sensors under
a reference lamp. The reference quantum sensors are recalibrated with a quartz
halogen lamp traceable to the National Institute of Standards and Technology
(NIST).
Spectral Response
Mean spectral response measurements of six replicate Apogee SQ-100 (original)
and SQ-500 (full-spectrum) series quantum sensors. Spectral response
measurements were made at 10 nm increments across a wavelength range of 300 to
800 nm with a monochromator and an attached electric light source. Measured
spectral data from each quantum sensor were normalized by the measured
spectral response of the monochromator/electric light combination, which was
measured with a spectroradiometer.
Temperature Response Mean
temperature response of ten SQ-500 series quantum sensors (errors bars
represent two standard deviations above and below mean). Temperature response
measurements were made at 10 C intervals across a temperature range of
approximately -10 to 40 C in a temperature-controlled chamber under a fixed,
broad-spectrum, electric lamp. At each temperature set point, a
spectroradiometer was used to measure light intensity from the lamp and all
quantum sensors were compared to the spectroradiometer. The spectroradiometer
was mounted external to the temperature control chamber and remained at room
temperature during the experiment.
Cosine Response
Directional (cosine) response is defined as the measurement error at a
specific angle of radiation incidence. Error for Apogee SQ-500 series quantum
sensors is approximately ± 2 % and ± 5 % at solar zenith angles of 45° and
75°, respectively. Mean directional (cosine)
response of seven apogee SQ-500 series quantum sensors. Directional response
measurements were made on the rooftop of the Apogee building in Logan, Utah.
Directional response was calculated as the relative difference of SQ-500
quantum sensors from the mean of replicate reference quantum sensors (LI-COR
models LI-190 and LI-190R, Kipp & Zonen model PQS 1). Data were also collected
in the laboratory using a reference lamp and positioning the sensor at varying
angles.
DEPLOYMENT AND INSTALLATION
Mount the sensor to a solid surface with the nylon mounting screw provided. To accurately measure PPFD incidents on a horizontal surface, the sensor must be level. An Apogee Instruments model AL-100 leveling plate is recommended for this purpose. To facilitate mounting on a cross arm, an Apogee Instruments model AM-110 mounting bracket is recommended. To minimize azimuth error, the sensor should be mounted with the cable pointing toward true north in the northern hemisphere or true south in the southern hemisphere. Azimuth error is typically less than 0.5 %, but it is easy to minimize by proper cable orientation. In addition to orienting the cable to point toward the nearest pole, the sensor should also be mounted such that obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor. Once mounted, the green cap should be removed from the sensor. The green cap can be used as a protective covering for the sensor when it is not in use.
CABLE CONNECTORS
Apogee started offering in-line cable connectors on some bare-lead wire
sensors in March 2018 to simplify the process of removing sensors from
installations for recalibration (the entire cable does not have to be removed
from the station and shipped with the sensor).
The ruggedized M8 connectors are rated IP68, made of corrosion-resistant
marine-grade stainless steel, and designed for extended use in harsh
environmental conditions. Instructions
Pins and Wiring Colors: All Apogee connectors have six pins, but not all pins
are used for every sensor.
There may also be unused wire colors inside the cable. To simplify connection
to a measurement device, the unused pigtail lead wire colors are removed.
If a replacement cable is required, please contact Apogee directly to ensure
ordering the proper pigtail configuration. Alignment: When reconnecting a
sensor, arrows on the connector jacket and an aligning notch ensure proper
orientation.
Disconnection for extended periods: When disconnecting the sensor for an
extended period of time from an installation, protect the remaining half of
the connector still on the station from water and dirt with electrical tape or
other methods. Tightening: Connectors are designed to be firmly finger-
tightened only. There is an O-ring inside the connector that can be overly
compressed if a wrench is used. Pay attention to thread alignment to avoid
cross-threading. When fully tightened, one to two threads may still be
visible.
WARNING: Do not tighten the connector by twisting the black cable, only
twist the metal connector.
OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller)
capable of measuring and displaying or recording a millivolt signal (an input
measurement range of approximately 0 to 25 mV is required to cover the entire
range of PPFD from the sun). In order to maximize measurement resolution and
signal-to-noise ratio, the input range of the measurement device should
closely match the output range of the quantum sensor. DO NOT connect the
sensor to a power source. The sensor is self-powered and applying voltage will
damage the sensor.
VERY IMPORTANT: Apogee changed all wiring colors of our bare-lead sensors
in March 2018 in conjunction with the release of inline cable connectors on
some sensors. To ensure proper connection to your data device, please note
your serial number or if your sensor has a stainless-steel connector 30 cm
from the sensor head then use the appropriate wiring configuration below.
Wiring for SQ-500 Serial Numbers 1559 and above or with a cable connector
Sensor Calibration
Apogee un-amplified full spectrum quantum sensors, model SQ-500, have a
standard PPFD calibration factor of exactly:
100.0 µmol m -2 -1 sper mV
Multiply this calibration factor by the measured voltage to convert sensor
output to PPFD in units of µmol m -2 -1 s : Calibration Factor (100 µmol m -2
-1 s per mV) * Sensor Output Signal (mV) = PPFD (µmol m -2 -1 s )
*100 20 = 2000**
Example of PPFD measurement with an Apogee model SQ500 quantum sensor. Full sunlight yields a PPFD on a horizontal plane at the Earth’s surface of approximately 2000 µmol m -2 -1 s . This yields an output signal of 20 mV. The signal is converted to PPFD by multiplying by the calibration factor of 100.0 µmol m -2 -1 s per mV.
Spectral Error
The combination of diffuser transmittance, interference filter transmittance,
and photodetector sensitivity yields the spectral response of a quantum
sensor. A perfect photodetector/filter/diffuser combination would exactly
match the defined plant photosynthetic response to photons (equal weighting to
all photons between 400 and 700 nm, no weighting of photons outside this
range), but this is challenging in practice. A mismatch between the defined
plant photosynthetic response and sensor spectral response results in a
spectral error when the sensor is used to measure radiation from sources with
a different spectrum than the radiation source used to calibrate the sensor
(Federer and Tanner, 1966; Ross and Sulev, 2000). Spectral errors for PPFD
measurements made under common radiation sources for growing plants were
calculated for Apogee SQ-100 and SQ-500 series quantum sensors using the
method of Federer and Tanner (1966). This method requires PPFD weighting
factors (defined plant photosynthetic response), measured sensor spectral
response (shown in the Spectral Response section on page 7), and radiation
source spectral outputs (measured with a spectroradiometer). Note, this method
calculates spectral error only and does not consider calibration, directional
(cosine), temperature, and stability/drift errors. Spectral error data (listed
in the table below) indicate errors less than 5 % for sunlight in different
conditions (clear, cloudy, reflected from plant canopies, transmitted below
plant canopies) and common broad spectrum electric lamps (cool white
fluorescent, metal halide, high-pressure sodium), but larger errors for
different mixtures of light emitting diodes (LEDs) for the SQ-100 series
sensors. Spectral errors for the SQ-500 series sensors are smaller than those
for SQ-100 series sensors because the spectral response of SQ-500 series
sensors is a closer match to the defined plant photosynthetic response.
Quantum sensors are the most common instrument for measuring PPFD because they
are about an order of magnitude lower cost the spectroradiometers, but
spectral errors must be considered. The spectral errors in the table below can
be used as correction factors for individual radiation sources.
Spectral Errors for PPFD Measurements with Apogee SQ-100 and SQ-500 Series
Quantum Sensors
Radiation Source (Error Calculated Relative to Sun, Clear Sky)| SQ-100
Series PPFD Error [%]| SQ-500 Series PPFD Error [%]
---|---|---
Sun (Clear Sky)| 0.0| 0.0
Sun (Cloudy Sky)| 0.2| 0.1
Reflected from Grass Canopy| 3.8| -0.3
Transmitted below Wheat Canopy| 4.5| 0.1
Cool White Fluorescent (T5)| 0.0| 0.1
Metal Halide| -2.8| 0.9
Ceramic Metal Halide| -16.1| 0.3
High Pressure Sodium| 0.2| 0.1
Blue LED (448 nm peak, 20 nm full-width half-maximum)| -10.5| -0.7
Green LED (524 nm peak, 30 nm full-width half-maximum)| 8.8| 3.2
Red LED (635 nm peak, 20 nm full-width half-maximum)| 2.6| 0.8
Red LED (667 nm peak, 20 nm full-width half-maximum)| -62.1| 2.8
Red, Blue LED Mixture (80 % Red, 20 % Blue)| -72.8| -3.9
Red, Blue, White LED Mixture (60 % Red, 25 % White, 15 % Blue)| -35.5| -2.0
Cool White LED| -3.3| 0.5
Warm White LED| -8.9| 0.2
Federer, C.A., and C.B. Tanner, 1966. Sensors for measuring light available
for photosynthesis. Ecology 47:654-657.
Ross, J., and M. Sulev, 2000. Sources of errors in measurements of PAR.
Agricultural and Forest Meteorology 100:103-125.
Yield Photon Flux Density (YPFD) Measurements
Photosynthesis in plants does not respond equally to all photons. Relative
quantum yield (plant photosynthetic efficiency) is dependent on wavelength
(green line in figure below) (McCree, 1972a; Inada, 1976). This is due to the
combination of spectral absorptivity of plant leaves (absorptivity is higher
for blue and red photons than green photons) and absorption by non-
photosynthetic pigments. As a result, photons in the wavelength range of
approximately 600-630 nm are the most efficient.
Defined plant response to photons (black line, weighting factors used to
calculate PPFD), measured plant response to photons (green line, weighting
factors used to calculate YPFD), and SQ-500 series quantum sensor response to
photons (sensor spectral response).
One potential definition of PAR is weighting photon flux density in units of
final m at each wavelength between 300 and 800 nm by measuring relative
quantum yield and summing the result. This is defined as yield photon flux
density (PFD, units of final m -2 -1 s ) (Sager et al., 1988). There are
uncertainties and challenges associated with this definition of PAR.
Measurements used to generate the relative quantum yield data were made on
single leaves under low radiation levels and at short time scales (McCree,
1972a; Inada, 1976). Whole plants and plant canopies typically have multiple
leaf layers and are generally grown in the field or greenhouse over the course
of an entire growing season. Thus, the actual conditions plants are subject to
are likely different than those the single leaves were in when measurements
were made by McCree (1972a) and Inada (1976). In addition, the relative
quantum yield shown in the figure above is the mean from twenty-two species
grown in the field (McCree, 1972a).
The mean relative quantum yield for the same species grown in growth chambers
was similar, but there were differences, particularly at shorter wavelengths
(less than 450 nm). There was also some variability between species (McCree,
1972a; Inada, 1976). -2 -1 s McCree (1972b) found that equally weighting all
photons between 400 and 700 nm and summing the result, defined as
photosynthetic photon flux density (PPFD, in units of final m ), was well
correlated to photosynthesis, and very similar to the correlation between YPFD
and photosynthesis. As a matter of practicality, PPFD is a simpler definition
of PAR. At the same time as McCree’s work, others had proposed PPFD as an
accurate measure of PAR and built sensors that approximated the PPFD weighting
factors (Biggs et al., 1971; Federer and Tanner, 1966). The correlation
between PPFD and PFD measurements for several radiation sources is very high
(figure below), as an approximation, YPFD = 0.9PPFD. As a result, almost
universally PAR is defined as PPFD rather than YPFD, although YPFD has been
used in some studies. The only radiation sources shown (figure below) that
don’t fall on the regression line are the high-pressure sodium (HPS) lamp,
reflection from a plant canopy, and transmission below a plant canopy. A large
fraction of radiation from HPS lamps is in the red range of wavelengths where
the YPFD weighting factors (measured relative quantum yield) are at or near
one. The factor for converting PPFD to YPFD for HPS lamps is 0.95, rather than
0.90. The factor for converting PPFD to YPFD for reflected and transmitted
photons is 1.00. -2 -1 s Correlation between photosynthetic photon flux
density (PPFD) and yield photon flux density (PFD) for multiple different
radiation sources. PDF is approximately 90 % of PPFD. Measurements were made
with a spectroradiometer (Apogee Instruments model PS-200) and weighting
factors shown in the previous figure were used to calculate PPFD and PFD.
Biggs, W., A.R. Edison, J.D. Eastin, K.W. Brown, J.W. Maranville, and M.D.
Clegg, 1971. Photosynthesis light sensor and meter. Ecology 52:125-131.
Federer, C.A., and C.B. Tanner, 1966. Sensors for measuring light available
for photosynthesis. Ecology 47:654-657. Inada, K., 1976. Action spectra for
photosynthesis in higher plants. Plant and Cell Physiology 17:355-365. McCree,
K.J., 1972a. The action spectrum, absorptance, and quantum yield of
photosynthesis in crop plants. Agricultural Meteorology 9:191-216. McCree,
K.J., 1972b. Test of current definitions of photosynthetically active
radiation against leaf photosynthesis data. Agricultural Meteorology
10:443-453. Underwater Measurements and Immersion Effect Sager, J.C., W.O.
Smith, J.L. Edwards, and K.L. Cyr, 1988. Photosynthetic efficiency and
phytochrome photoequilibria determination using spectral data. Transactions of
the ASAE 31:1882-1889. When a quantum sensor that was calibrated in air is
used to make underwater measurements, the sensor reads
Immersion Effect Correction Factor
When a radiation sensor is submerged in water, more of the incident radiation
is backscattered out of the diffuser than when the sensor is in air (Smith,
1969; Tyler and Smith, 1970). This phenomenon is caused by the difference in
the refractive index for air (1.00) and water (1.33), and is called the
immersion effect. Without correction for the
immersion effect, radiation sensors calibrated in air can only provide
relative values underwater (Smith, 1969; Tyler and Smith, 1970). Immersion
effect correction factors can be derived by making measurements in air and at
multiple water depths at a constant distance from a lamp in a controlled
laboratory setting.
Apogee SQ-500 series quantum sensors have an immersion effect correction
factor of 1.32. This correction factor should be multiplied by PPFD
measurements made underwater to yield accurate PPFD.
Further information on underwater measurements and the immersion effect can be
found on the Apogee webpage
(http://www.apogeeinstruments.com/underwater-par-measurements/).
Smith, R.C., 1969. An underwater spectral irradiance collector. Journal of
Marine Research 27:341-351. Tyler, J.E., and R.C. Smith, 1970. Measurements of
Spectral Irradiance Underwater. Gordon and Breach, New York, New York. 103
pages
MAINTENANCE AND RECALIBRATION
Blocking of the optical path between the target and detector can cause low
readings. Occasionally, accumulated
materials on the diffuser of the upward-looking sensor can block the optical
path in three common ways:
- Moisture or debris on the diffuser.
- Dust during periods of low rainfall.
- Salt deposit accumulation from the evaporation of sea spray or sprinkler irrigation water.
Apogee Instruments’ upward-looking sensors have a domed diffuser and housing
for improved self-cleaning from rainfall, but active cleaning may be
necessary. Dust or organic deposits are best removed using water, or window
cleaner, and a soft cloth or cotton swab. Salt deposits should be dissolved
with vinegar and removed with a cloth or cotton swab. Salt deposits cannot be
removed with solvents such as alcohol or acetone. Use only gentle pressure
when cleaning the diffuser with a cotton swab or soft cloth to avoid
scratching the outer surface. The solvent should be allowed to do the
cleaning, not mechanical force. Never use abrasive material or cleaner on the
diffuser.
Although Apogee sensors are very stable, nominal accuracy drift is normal for
all research-grade sensors. To ensure maximum accuracy, we generally recommend
sensors are sent in for recalibration every two years, although you can often
wait longer according to your particular tolerances.
To determine if a specific sensor needs recalibration, the Clear Sky
Calculator (www.clearskycalculator.com)
website and/or smartphone app can be used to indicate PPFD incidents on a
horizontal surface at any time of day at any location in the world. It is most
accurate when used near solar noon in spring and summer months, where accuracy
over multiple clear and unpolluted days is estimated to be ± 4 % in all
climates and locations around the world. For best accuracy, the sky must be
completely clear, as reflected radiation from clouds causes incoming radiation
to increase above the value predicted by the clear sky calculator. Measured
PPFD can exceed PPFD predicted by the Clear Sky Calculator due to reflection
from thin, high clouds and edges of clouds, which enhances incident PPFD. The
influence of high clouds typically shows up as spikes above clear sky values,
not a constant offset greater than clear sky values.
To determine recalibration need, input site conditions into the calculator and
compare PPFD measurements to calculated PPFD for a clear sky. If sensor PPFD
measurements over multiple days near solar noon are consistently different
than calculated PPFD (by more than 6 %), the sensor should be cleaned and re-
leveled. If measurements are still different after a second test, email
calibration@apogeeinstruments.com
to discuss test results and the possible return of sensor(s).
This calculator determines the intensity of radiation falling on a horizontal
surface at any time of the day in any location in the world. The primary use
of this calculator is to determine the need for the recalibration of radiation
sensors. It is most accurate when used near solar noon in the summer months.
This site was developed and maintained by:
Homepage of the Clear Sky Calculator. Two calculators are available: one for quantum sensors (PPFD) and one for pyranometers (total shortwave radiation).
**| FOR QUANTUM SENSORS| Input Parameters for Estimating
Photosynthetic Photon Flux (PPF):| Output from Model:| HOME**
---|---|---|---|---
1. For best accuracy, comparison should be made on clear, non-polluted,
summer days within one hour of solar noon.| Latitude =
Longitude =
Longitudetz =?| 41.7″| Model Estimated PPF = 1994 mol ms?c
Measured PPE = 9990 mol ms?
|
2 Enter input parameters in the blue cells at right. Definitions are shown
below.| Elevation =?
Day of Year =?| 111.8| DIFFERENCE FROM MODEL= 0.2 %|
3 Sensor must be level and perfectly dean. Enter your measured solar radiation
in the blue “Measured WE’ cell at far right.| Time of Day = (6 min = 0.1 hr)
Daylight Savings = +| 105
1400m
172
12.9| CONTAC! APOGEE FOR RECALIBRATION|
4 Difference between the model and your sensor is shown in the yellow
“DIFFERENCE FROM MODEL” cell at right.| Air Temperature =| 1hr
25c| Name:
E-mail:|
5 Run the model on replicate days. Contact Apogee for recalibration if the
measured value is more than 5 % different than the estimated value. You will
be contacted within two business days.| Relative Humidity =| 30%| Phone:
Serial #:
Comments:|
For a discussion on model accuracy and sensitivity of input parameters,
CLICK HERE.| RECALCULATE MODEL| Please include all requested
information.
SEND INFE TO APOGEE
---|---|---
• INPUT AND OUTPUT DEFINITIONS
Latitude =latitude of the measurement site [degrees]; for the southern
hemisphere, insert as a negative number; Info may be obtained from
http://itouchmap.com/lationg.html
Longitude = longitude of the measurement site [degrees]; expressed as positive
degrees west of the standard meridian in Greenwich, England (e.g. 74° for New
York, 260° for Bangkok, Thailand, and 358° for Paris, France).
Longitude,, =longitude of the center of your local time zone [degrees];
expressed as positive degrees
This site Is developed and maintained by: calibrationeapogee-inst.com
Clear Sky Calculator for quantum sensors. Site data are input in blue cells in
the middle of the page and an estimate of PPFD is returned on the right-hand
side of the page.
TROUBLESHOOTING AND CUSTOMER SUPPORT
Independent Verification of Functionality
Apogee SQ-500 series quantum sensors are self-powered devices and output an
analog signal proportional to incident PPFD. A quick and easy check of sensor
functionality can be determined using a voltmeter with a millivolt resolution.
Connect the positive lead wire from the voltmeter to the white wire from the
sensor and the negative (or common) lead wire from the voltmeter to the black
wire from the sensor. Direct the sensor head toward a light source and verify
the sensor provides a signal. Increase and decrease the distance from the
sensor head to the light source to verify that the signal changes
proportionally (decreasing signal with increasing distance and increasing
signal with decreasing distance). Blocking all radiation from the sensor
should force the sensor signal to zero.
Compatible Measurement Devices (Dataloggers/Controllers/Meters)
Model SQ-500 quantum sensors are calibrated with a standard calibration factor
of 100.0 µmol m per mV, yielding a sensitivity of 0.01 mV per µmol m -2 -1 s .
Thus, a compatible measurement device (e.g., datalogger or controller) should
have a resolution of at least 0.01 mV in order to provide a PPFD resolution of
1 µmol m -2 -1 s and at least 0.001 mV in order to provide PPFD resolution of
0.1 µmol m -2 -1 s -2 -1 s . An example datalogger program for Campbell
Scientific dataloggers can be found on the Apogee webpage at
http://www.apogeeinstruments.com/content/Full-Spectrum-Quantum-Sensor-
Unamplified.CR1.
Cable Length
When the sensor is connected to a measurement device with high input
impedance, sensor output signals are not changed by shortening the cable or
splicing on additional cable in the field. Tests have shown that if the input
impedance of the measurement device is greater than 1 mega-ohm there is
negligible effect on the calibration, even after adding up to 100 m of cable.
All Apogee sensors use shielded, twisted pair cables to minimize
electromagnetic interference. For best measurements, the shield wire must be
connected to earth’s ground. This is particularly important when using the
sensor with long lead lengths in electromagnetically noisy environments.
Modifying Cable Length
See the Apogee webpage for details on how to extend sensor cable length:
(http://www.apogeeinstruments.com/how-to-make-a-weatherproof-cable-splice/).
Unit Conversion Charts
Units other than photon flux density (e.g., energy flux density, illuminance)
may be required for certain applications. It is possible to convert PPFD from
a quantum sensor to other units, but it requires the spectral output of the
radiation source of interest. Conversion factors for common radiation sources
can be found in the Knowledge Base on the Apogee website
(http://www.apogeeinstruments.com/knowledge-base/; scroll down to the
Quantum Sensors section). A spreadsheet to convert PPFD to energy flux density
or illuminance is also provided in the Knowledge Base on the Apogee website
(http://www.apogeeinstruments.com/content/PPFD-to-Illuminance-
Calculator.xls). Apogee SQ-500 series quantum sensors are calibrated to
measure PPFD in units of µmol m -2 -1 s APOGEE INSTRUMENTS, INC.
721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA
TEL: 435-792-4700
FAX: 435-787-8268 |
WEB: APOGEEINSTRUMENTS.COM
Copyright © 2020 Apogee Instruments, Inc.
APOGEE INSTRUMENTS, INC.
721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA
TEL: 435-792-4700
FAX: 435-787-8268 |
WEB: APOGEEINSTRUMENTS.COM Copyright © 2020
Apogee Instruments, Inc.
References
- apogeeinstruments.com/content/Full-Spectrum-Quantum-Sensor-Unamplified.CR1
- How to Make a Weatherproof Cable Splice
- Support | Apogee Instruments
- Underwater PAR Measurements | Apogee Instruments
- Clear Sky Calculator | Apogee Instruments Inc.
- Clear Sky Calculator | Apogee Instruments Inc.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>