apogee MQ-100 Quantum Meter Instructions
- June 5, 2024
- APOGEE
Table of Contents
- apogee MQ-100 Quantum Meter
- INTRODUCTION
- SENSOR MODELS
- SPECIFICATIONS
- DEPLOYMENT AND INSTALLATION
- BATTERY INSTALLATION AND REPLACEMENT
- OPERATION AND MEASUREMENT
- APOGEE AMS SOFTWARE
- MAINTENANCE AND RECALIBRATION
- TROUBLESHOOTING AND CUSTOMER SUPPORT
- Error Codes and Fixes
- RETURN AND WARRANTY POLICY RETURN POLICY
- PRODUCTS BEYOND THE WARRANTY PERIOD
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
apogee MQ-100 Quantum Meter
INTRODUCTION
Radiation that drives photosynthesis is called photosynthetically active
radiation (PAR) and is typically defined as total radiation across a range of
400 to 700 nm. PAR is often expressed as photosynthetic photon flux density
(PPFD): photon flux in units of micromoles per square meter per second (µmol
m-2 s-1, equal to microEinsteins per square meter per second) summed from 400
to 700 nm (total number of photons from 400 to 700 nm). While Einsteins and
micromoles are equal (one Einstein = one mole of photons), the Einstein is not
an SI unit, so expressing PPFD as µmol m-2 s-1 is preferred.
The acronym PPF is also widely used and refers to the photosynthetic photon
flux. The acronyms PPF and PPFD refer to the same parameter. The two terms
have co-evolved because there is not a universal definition of the term
“flux”. Some physicists define flux as per unit area per unit time. Others
define flux only as per unit time. We have used PPFD in this manual because we
feel that it is better to be more complete and possibly redundant.
Sensors that measure PPFD are often called quantum sensors due to the
quantized nature of radiation. A quantum refers to the minimum quantity of
radiation, one photon, involved in physical interactions (e.g., absorption by
photosynthetic pigments). In other words, one photon is a single quantum of
radiation.
Typical applications of quantum sensors include incoming PPFD measurement over
plant canopies in outdoor environments or in greenhouses and growth chambers,
and reflected or under-canopy (transmitted) PPFD measurement in the same
environments.
Apogee Instruments MQ series quantum meters consist of a handheld meter and a
dedicated quantum sensor that is integrated into the top of the meter housing
(MQ-100) or connected by cable to an anodized aluminum housing (MQ-200 and
MQ-300 series). Integrated and separate sensors consist of a cast acrylic
diffuser (filter), photodiode, and are potted solid with no internal air
space. MQ series quantum meters provide a real-time PPFD reading on the LCD
display and offer measurements for both sunlight and electric light
calibrations (menu selectable) that determine the radiation incident on a
planar surface (does not have to be horizontal), where the radiation emanates
from all angles of a hemisphere. MQ series quantum meters include manual and
automatic data logging features for making spot-check measurements or
calculating daily light integral (DLI).
SENSOR MODELS
Apogee MQ series quantum meters covered in this manual are self-contained and come complete with handheld meter and sensor. Line quantum meters, MQ-300 series, provide spatially averaged PPFD measurements. All sensors along the length of the line are connected in parallel, and as a result, Apogee line quantum meters display PPFD values that are averaged from the location of the individual sensors.
Sensor model number and serial number are located on a label on the backside of the handheld meter.
SPECIFICATIONS
MQ-100 | MQ-200 | MQ-301 | MQ-303/306 | |
---|---|---|---|---|
Calibration Uncertainty | ± 5 % (see calibration Traceability below) |
Measurement
Repeatability
| Less than 1 %
Long-term Drift (Non-stability)| Less than 2 % per year
Non-linearity| Less than 1 % (up to 3000 µmol m-2 s-1)
Response Time| Less than 1 ms
Field of View| 180°
Spectral Range| 410 to 655 nm (wavelengths where response is greater than 50 %
of maximum; see Spectral Response below)
Directional (Cosine) Response| ± 5 % at 75° zenith angle (see Cosine Response
below)
Temperature
Response
| 0.06 ± 0.06 % per C (see Temperature Response below)
Operating Environment| 0 to 50 C; less than 90 % non-condensing relative
humidity up to 30 C; less than 70 % non-condensing relative humidity from 30
to 50 C; separate sensors can be
submerged in water up to depths of 30 m
Meter Dimensions| 126 mm length; 70 mm width; 24 mm height
Sensor Dimensions
| Integrated w/ Meter| 24 mm diameter;
33 mm height
| 70 mm length; 15 mm width; 15 mm height| 50 mm length; 15 mm width; 15 mm
height
Mass| 150 g| 180 g| 380 g| 300 g
Cable| 2 m of two conductor, shielded, twisted-pair wire; TPR jacket (high
water resistance, high UV stability, flexibility in cold conditions)
Calibration Traceability
Apogee MQ series quantum meters are calibrated through side-by-side comparison
to the mean of four transfer standard quantum sensors under a reference lamp.
The reference quantum sensors are recalibrated with a 200 W quartz halogen
lamp traceable to the National Institute of Standards and Technology (NIST).
Spectral Response
Mean spectral response of six SQ-100 series quantum sensors (error bars
represent two standard deviations above and below mean) compared to defined
plant response to photons. Spectral response measurements were made at 10 nm
increments across a wavelength range of 300 to 800 nm with a monochromator and
an attached electric light source. Measured spectral data from each quantum
sensor were normalized by the measured spectral response of the
monochromator/electric light combination, which was measured with a
spectroradiometer.
Spectral Response for MQ-300 series meters: MQ-301 serial #’s 1297 and above, MQ-303 serial #’s 1097 and above, MQ-306 serial #’s 1171 and above
Spectral response of updated line quantum sensor spectral response graph (in green) vs. the original line quantum sensor response (in blue). For information on effects on accuracy, see the spectral error table on page 15 of this manual.
- MQ-301 serial #’s 1297 and above.
- MQ-303 serial #’s 1097 and above.
- MQ-306 serial #’s 1171 and above.
Temperature Response
Mean temperature response of eight SQ-100 series quantum sensors (errors bars represent two standard deviations above and below mean). Temperature response measurements were made at 10 C intervals across a temperature range of approximately -10 to 40 C in a temperature controlled chamber under a fixed, broad spectrum, electric lamp. At each temperature set point, a spectroradiometer was used to measure light intensity from the lamp and all quantum sensors were compared to the spectroradiometer. The spectroradiometer was mounted external to the temperature control chamber and remained at room temperature during the experiment.
Cosine Response
Directional (cosine) response is defined as the measurement error at a specific angle of radiation incidence. Error for Apogee SQ-100 series quantum sensors is approximately ± 2 % and ± 5 % at solar zenith angles of 45° and 75°, respectively.
Mean directional (cosine) response of six apogee SQ-100 series quantum sensors. Directional response measurements were made on the rooftop of the Apogee building in Logan, Utah. Directional response was calculated as the relative difference of SQ-500 quantum sensors from the mean of replicate reference quantum sensors (LI-COR models LI-190 and LI-190R, Kipp & Zonen model PQS 1). Data were also collected in the laboratory using a reference lamp and positioning the sensor at varying angles.
DEPLOYMENT AND INSTALLATION
Apogee MQ series quantum meters are designed for spot-check measurements, and calculation of daily light integral (DLI; total number of photons incident on a planar surface over the course of a day) through the built-in logging feature. To accurately measure PFFD incident on a horizontal surface, the sensor must be level. For this purpose, each MQ model comes with a different option for mounting the sensor to a horizontal plane.
The AL-210 leveling plate is recommended for use with the MQ-100.The AL-100 leveling plate is recommended for use with the MQ-200. To facilitate mounting to a cross arm, the AL-120 mounting bracket is recommended.MQ-300 series line quantum sensors are leveled using the built-in bubble level located in the handle of the sensor. In addition to leveling, all sensors should also be mounted such that obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor.
BATTERY INSTALLATION AND REPLACEMENT
Use a phillips head screwdriver to remove the screw from the battery cover. Remove the battery cover by slightly lifting and sliding the outer edge of the cover away from the meter. To power the meter, slide the included battery (CR2320) into the battery holder, after removing the battery door from the meter’s back panel.
The positive side (designated by a “+” sign) should be facing out from the meter circuit board.NOTE: The battery cradle can be damaged by using an incorrectly sized battery. If the battery cradle is damaged, the circuit board will need to be replaced. To avoid this costly problem, use only a CR2320 battery.
Battery Removal
Press down on the battery with a screwdriver or similar object. Slide battery
out.
If the battery is difficult to move, turn the meter on its side so that the opening for the battery is facing downward and tap the meter downward against an open palm to dislodge the battery enough so that it can be removed with your thumb to slide the battery out of the battery holder.
OPERATION AND MEASUREMENT
MQ series quantum meters are designed with a user-friendly interface allowing quick and easy measurements. Press the power button to activate the LCD display. After two minutes of non-activity the meter will revert to sleep mode and the display will shut off to conserve battery life.
Press the mode button to access the main menu, where the appropriate calibration (sunlight or electric light) and manual or automatic logging are selected, and where the meter can be reset.
Press the sample button to log a reading while taking manual measurements.
Press the up button to make selections in the main menu. This button is also used to view and scroll through the logged measurements on the LCD display.
Press the down button to make selections in the main menu. This button is also used to view and scroll through the logged measurements on the LCD display.
The LCD display consists of the total number of logged measurements in the
upper right hand corner, the real-time PPFD value in the center, and the
selected menu options along the bottom. Calibration:
To choose between sunlight and electric light calibration, push the mode
button once and use the up/down buttons to make the appropriate selection (SUN
or ELEC). Once the desired mode is blinking, press the mode button three more
times to exit the menu.
Logging:
To choose between manual or automatic logging, push the mode button once and
use the up/down buttons to make the appropriate selection (SMPL or LOG). Once
the desired mode is blinking, press the mode button two more times to exit the
menu. When in SMPL mode press the sample button to record up to 99 manual
measurements (a counter in the upper right hand corner of the LCD display
indicates the total number of saved measurements). When in LOG mode the meter
will power on/off to make a measurement every 30 seconds. Every 30 minutes the
meter will average the sixty 30 second measurements and record the averaged
value to memory. The meter can store up to 99 averages and will start to
overwrite the oldest measurement once there are 99 measurements. Every 48
averaged measurements (making a 24 hour period), the meter will also store an
integrated daily total in moles per meter squared per day (mol m-2 d-1).
Reset:
To reset the meter, in either SMPL or LOG mode, push the mode button three
times (RUN should be blinking), then while pressing the down button, press the
mode button once. This will erase all of the saved measurements in memory, but
only for the selected mode. That is, performing a reset when in SMPL mode will
only erase the manual measurements and performing a reset when in LOG mode
will only erase the automatic measurements.
Review/Download Data:
Each of the logged measurements in either SMPL or LOG mode can be reviewed on
the LCD display by pressing the up/down buttons. To exit and return to the
real-time readings, press the sample button. Note that the integrated daily
total values are not accessible through the LCD and can only be viewed by
downloading to a computer.
Downloading the stored measurements will require the AC-100 communication
cable and software (sold separately). The meter outputs data using the UART
protocol and requires the AC-100 to convert from UART to USB, so standard USB
cables will not work. Set up instructions and software can be downloaded from
the Apogee website (http://www.apogeeinstruments.com/ac-100-communcation-
cable/).
Spectral Error
The combination of diffuser transmittance, interference filter transmittance,
and photodetector sensitivity yields spectral response of a quantum sensor. A
perfect photodetector/filter/diffuser combination would exactly match the
defined plant photosynthetic response to photons (equal weighting to all
photons between 400 and 700 nm, no weighting of photons outside this range),
but this is challenging in practice. Mismatch between the defined plant
photosynthetic response and sensor spectral response results in spectral error
when the sensor is used to measure radiation from sources with a different
spectrum than the radiation source used to calibrate the sensor (Federer and
Tanner, 1966; Ross and Sulev, 2000).
Spectral errors for PPFD measurements made under common radiation sources for
growing plants were calculated for Apogee SQ-100/300 and SQ-500 series quantum
sensors using the method of Federer and Tanner (1966). This method requires
PPFD weighting factors (defined plant photosynthetic response), measured
sensor spectral response (shown in Spectral Response section on page 7), and
radiation source spectral outputs (measured with a spectroradiometer). Note,
this method calculates spectral error only and does not consider calibration,
directional (cosine), temperature, and stability/drift errors. Spectral error
data (listed in table below) indicate errors less than 5 % for sunlight in
different conditions (clear, cloudy, reflected from plant canopies,
transmitted below plant canopies) and common broad spectrum electric lamps
(cool white fluorescent, metal halide, high pressure sodium), but larger
errors for different mixtures of light emitting diodes (LEDs) for the SQ-100
series sensors. Spectral errors for the SQ-500 series sensors are smaller than
those for SQ-100 series sensors because the spectral response of SQ-500 series
sensors is a closer match to the defined plant photosynthetic response.
Quantum sensors are the most common instrument for measuring PPFD, because
they are about an order of magnitude lower cost the spectroradiometers, but
spectral errors must be considered. The spectral errors in the table below can
be used as correction factors for individual radiation sources. Spectral
Errors for PPFD Measurements with Apogee SQ-100 and SQ-500 Series Quantum
Sensors
Radiation Source (Error Calculated Relative to Sun, Clear Sky)
| SQ-100/300 Series
PPFD Error [%]
| SQ-500 Series
PPFD Error [%]
---|---|---
Sun (Clear Sky)| 0.0| 0.0
Sun (Cloudy Sky)| 0.2| 0.1
Reflected from Grass Canopy| 3.8| -0.3
Transmitted below Wheat Canopy| 4.5| 0.1
Cool White Fluorescent (T5)| 0.0| 0.1
Metal Halide| -2.8| 0.9
Ceramic Metal Halide| -16.1| 0.3
High Pressure Sodium| 0.2| 0.1
Blue LED (448 nm peak, 20 nm full-width half-maximum)| -10.5| -0.7
Green LED (524 nm peak, 30 nm full-width half-maximum)| 8.8| 3.2
Red LED (635 nm peak, 20 nm full-width half-maximum)| 2.6| 0.8
Red LED (667 nm peak, 20 nm full-width half-maximum)| -62.1| 2.8
Red, Blue LED Mixture (80 % Red, 20 % Blue)| -72.8| -3.9
Red, Blue, White LED Mixture (60 % Red, 25 % White, 15 % Blue)| -35.5| -2.0
Cool White LED| -3.3| 0.5
Warm White LED| -8.9| 0.2
Federer, C.A., and C.B. Tanner, 1966. Sensors for measuring light available for photosynthesis. Ecology 47:654-657. Ross, J., and M. Sulev, 2000. Sources of errors in measurements of PAR. Agricultural and Forest Meteorology 100:103-125.Spectral Errors for PPFD and YPFD Measurements with Apogee MQ Series Quantum Meters
Radiation Source (Error Calculated Relative to Sun, Clear Sky)| *MQ-300
Series
| PPFD Error
[%]
| YPFD Error
[%]
---|---|---|---
Sun (Clear Sky)| 0.0| 0.0| 0.0
Sun (Cloudy Sky)| 0.1| 1.4| 1.6
Reflected from Grass Canopy| 7.4| 5.7| -6.3
Reflected from Deciduous Canopy| 4.6| 4.9| -7.0
Reflected from Conifer Canopy| 5.7| 5.5| -6.8
Transmitted below Grass Canopy| 4.8| 6.4| -4.5
Transmitted below Deciduous Canopy| 4.2| 6.8| -5.4
Transmitted below Conifer Canopy| 3.8| 5.3| 2.6
Radiation Source (Error Calculated Relative to Cool White
Fluorescent, T5)
Cool White Fluorescent (T5)| –| 0.0| 0.0
Cool White Fluorescent (T8)| –| -0.3| -1.2
Cool White Fluorescent (T12)| –| -1.4| -2.0
Compact Fluorescent| –| -0.5| -5.3
Metal Halide| –| -3.7| -3.7
Ceramic Metal Halide| –| -6.0| -6.4
High Pressure Sodium| –| 0.8| -7.2
Blue LED (448 nm peak, 20 nm full-width half-maximum)| –| -12.7| 8.0
Green LED (524 nm peak, 30 nm full-width half-maximum)| –| 8.0| 26.2
Red LED (635 nm peak, 20 nm full-width half-maximum)| –| 4.8| -6.2
Red, Blue LED Mixture (85 % Red, 15 % Blue)| –| 2.4| -4.4
Red, Green, Blue LED Mixture (72 % Red, 16 % Green, 12 % Blue)| –| 3.4| 0.2
Cool White Fluorescent LED| –| -4.6| -0.6
Neutral White Fluorescent LED| –| -6.7| -5.2
Warm White Fluorescent LED| –| -10.9| -13.0
Quantum sensors can be a very practical means of measuring PPFD and YPFD from
multiple radiation sources, but spectral errors must be considered. The
spectral errors in the table above can be used as correction factors for
individual radiation sources.
PPFD Error [%] for MQ-301 serial #’s 1297 and above, MQ-303 serial #’s 1097
and above, and MQ-306 serial #’s 1171 and above.
Yield Photon Flux Density (YPFD) Measurements
Photosynthesis in plants does not respond equally to all photons. Relative
quantum yield (plant photosynthetic efficiency) is dependent on wavelength
(green line in figure below) (McCree, 1972a; Inada, 1976). This is due to the
combination of spectral absorptivity of plant leaves (absorptivity is higher
for blue and red photons than green photons) and absorption by non-
photosynthetic pigments. As a result, photons in the wavelength range of
approximately 600-630 nm are the most efficient.
Defined plant response to photons (black line, weighting factors used to calculate PPFD), measured plant response to photons (green line, weighting factors used to calculate YPFD), and SQ-100 series and SQ-300 series quantum sensor response to photons (sensor spectral response).One potential definition of PAR is weighting photon flux density in units of mol m-2 s-1 at each wavelength between 300 and 800 nm by measured relative quantum yield and summing the result. This is defined as yield photon flux density (YPFD, units of mol m-2 s-1) (Sager et al., 1988). There are uncertainties and challenges associated with this definition of PAR. Measurements used to generate the relative quantum yield data were made on single leaves under low radiation levels and at short time scales (McCree, 1972a; Inada, 1976). Whole plants and plant canopies typically have multiple leaf layers and are generally grown in the field or greenhouse over the course of an entire growing season. Thus, actual conditions plants are subject to are likely different than those the single leaves were in when measurements were made by McCree (1972a) and Inada (1976). In addition, relative quantum yield shown in the figure above is the mean from twenty-two species grown in the field (McCree, 1972a). Mean relative quantum yield for the same species grown in growth chambers was similar, but there were differences, particularly at shorter wavelengths (less than 450 nm). There was also some variability between species (McCree, 1972a; Inada, 1976).
McCree (1972b) found that equally weighting all photons between 400 and 700 nm and summing the result, defined as photosynthetic photon flux density (PPFD, in units of mol m-2 s-1), was well correlated to photosynthesis, and very similar to correlation between YPFD and photosynthesis. As a matter of practicality, PPFD is a simpler definition of PAR. At the same time as McCree’s work, others had proposed PPFD as an accurate measure of PAR and built sensors that approximated the PPFD weighting factors (Biggs et al., 1971; Federer and Tanner, 1966). Correlation between PPFD and YPFD measurements for several radiation sources is very high (figure below), as an approximation, YPFD = 0.9PPFD. As a result, almost universally PAR is defined as PPFD rather than YPFD, although YPFD has been used in some studies. The only radiation sources shown (figure below) that don’t fall on the regression line are the high pressure sodium (HPS) lamp, reflection from a plant canopy, and transmission below a plant canopy. A large fraction of radiation from HPS lamps is in the red range of wavelengths where the YPFD weighting factors (measured relative quantum yield) are at or near one. The factor for converting PPFD to YPFD for HPS lamps is 0.95, rather than 0.90. The factor for converting PPFD to YPFD for reflected and transmitted photons is 1.00.
Correlation between photosynthetic photon flux density (PPFD) and yield photon
flux density (YPFD) for multiple different radiation sources. YPFD is
approximately 90 % of PPFD. Measurements were made with a spectroradiometer
(Apogee Instruments model PS-200) and weighting factors shown in the previous
figure were used to calculate PPFD and YPFD.Biggs, W., A.R. Edison, J.D.
Eastin, K.W. Brown, J.W. Maranville, and M.D. Clegg, 1971. Photosynthesis
light sensor and meter. Ecology 52:125-131.
Federer, C.A., and C.B. Tanner, 1966. Sensors for measuring light available
for photosynthesis. Ecology 47:654-657. Inada, K., 1976. Action spectra for
photosynthesis in higher plants. Plant and Cell Physiology 17:355-365. McCree,
K.J., 1972a. The action spectrum, absorptance and quantum yield of
photosynthesis in crop plants. Agricultural Meteorology 9:191-216.
Immersion Effect Correction Factor
When a radiation sensor is submerged in water, more of the incident radiation
is backscattered out of the diffuser than when the sensor is in air (Smith,
1969; Tyler and Smith, 1970). This phenomenon is caused by the difference in
the refractive index for air (1.00) and water (1.33), and is called the
immersion effect. Without correction for the immersion effect, radiation
sensors calibrated in air can only provide relative values underwater (Smith,
1969; Tyler and Smith, 1970). Immersion effect correction factors can be
derived by making measurements in air and at multiple water depths at a
constant distance from a lamp in a controlled laboratory setting. Apogee
SQ-100 series and SQ-300 series quantum sensors have an immersion effect
correction factor of 1.08. This correction factor should be multiplied by PPFD
measurements made underwater to yield accurate PPFD. Further information on
underwater measurements and the immersion effect can be found on the Apogee
webpage (http://www.apogeeinstruments.com/underwater-par-
measurements/).
Smith, R.C., 1969. An underwater spectral irradiance collector. Journal of
Marine Research 27:341-351. Tyler, J.E., and R.C. Smith, 1970. Measurements of
Spectral Irradiance Underwater. Gordon and Breach, New York, New York. 103
pages
APOGEE AMS SOFTWARE
Downloading data to a computer requires the AC-100 communication cable and the
free ApogeeAMS software. The meter outputs data using the UART protocol and
requires the AC-100 to convert from UART to USB, so standard USB cables will
not work. The most recent version of ApogeeAMS software can be downloaded at
http://www.apogeeinstruments.com/downloads/.
When the ApogeeAMS software is first opened, it will show a blank screen until
communication with the meter is established. If you click “Open Port” it will
say
“connection failed.”To establish communication, make sure the meter is plugged
into your computer using the AC-100 communication cable. To connect click the
dropdown menu button and “COM#” options will appear. For more details on how
to figure out which COM is the right one, watch our video.When you have
connected to the correct COM#, the software will say “Connected”. Click
“Sample Data” to view saved sample readings.“Daily Totals” shows all of the
saved Daily Light Integral (DLI) totals per day.
Click “30 Min Avg” to see the meter’s 99, 30-minute averages.To analyze the data, click on “File” and “Save As” to save the data as a .csv file.Or, you can highlight the numbers, copy, and paste them into a blank Excel spreadsheet. Data will need to be comma delimited.
MAINTENANCE AND RECALIBRATION
Blocking of the optical path between the target and detector can cause low readings. Occasionally, accumulated materials on the diffuser of the upward- looking sensor can block the optical path in three common ways:
- Moisture or debris on the diffuser.
- Dust during periods of low rainfall.
- Salt deposit accumulation from evaporation of sea spray or sprinkler irrigation water.
Apogee Instruments upward-looking sensors have a domed diffuser and housing
for improved self-cleaning from rainfall, but active cleaning may be
necessary. Dust or organic deposits are best removed using water, or window
cleaner, and a soft cloth or cotton swab. Salt deposits should be dissolved
with vinegar and removed with a cloth or cotton swab. Salt deposits cannot be
removed with solvents such as alcohol or acetone. Use only gentle pressure
when cleaning the diffuser with a cotton swab or soft cloth to avoid
scratching the outer surface. The solvent should be allowed to do the
cleaning, not mechanical force. Never use abrasive material or cleaner on the
diffuser.
Although Apogee sensors are very stable, nominal calibration drift is normal
for all research-grade sensors. To ensure maximum accuracy, recalibration
every two years is recommended. Longer time periods between recalibration may
be warranted depending on tolerances. See the Apogee webpage for details
regarding return of sensors for recalibration
(http://www.apogeeinstruments.com/.
To determine if your sensor needs recalibration, the Clear Sky Calculator
(www.clearskycalculator.com) website
and/or smartphone app can be used to indicate the total shortwave radiation
incident on a horizontal surface at any time of day at any location in the
world. It is most accurate when used near solar noon in spring and summer
months, where accuracy over multiple clear and unpolluted days is estimated to
be ± 4 % in all climates and locations around the world. For best accuracy,
the sky must be completely clear, as reflected radiation from clouds causes
incoming radiation to increase above the value predicted by the clear sky
calculator. Measured values of total shortwave radiation can exceed values
predicted by the Clear Sky Calculator due to reflection from thin, high clouds
and edges of clouds, which enhances incoming shortwave radiation. The
influence of high clouds typically shows up as spikes above clear sky values,
not a constant offset greater than clear sky values.
To determine recalibration need, input site conditions into the calculator and
compare total shortwave radiation measurements to calculated values for a
clear sky. If sensor shortwave radiation measurements over multiple days near
solar noon are consistently different than calculated values (by more than 6
%), the sensor should be cleaned and re-leveled. If measurements are still
different after a second test, email
calibration@apogeeinstruments.com to
discuss test results and possible return of sensor(s).
Homepage of the Clear Sky Calculator. Two calculators are available: one for quantum sensors (PPFD) and one for pyranometers (total shortwave radiation).
Clear Sky Calculator for quantum sensors. Site data are input in blue cells in middle of page and an estimate of PPFD is returned on right-hand side of page.
TROUBLESHOOTING AND CUSTOMER SUPPORT
Verify Functionality
Pressing the power button should activate the LCD and provide a real-time PPFD
reading. Direct the sensor head toward a light source and verify the PPFD
reading responds. Increase and decrease the distance from the sensor to the
light source to verify that the reading changes proportionally (decreasing
PPFD with increasing distance and increasing PPFD with decreasing distance).
Blocking all radiation from the sensor should force the PPFD reading to zero.
Battery Life
When the meter is maintained properly the coin cell battery (CR2320) should
last for many months, even after continuous use. The low battery indicator
will appear in the upper left hand corner of the LCD display when the battery
voltage drops below 2.8 V DC. The meter will still function correctly for some
time, but once the battery is drained the pushbuttons will no longer respond
and any logged measurements will be lost. Pressing the power button to turn
off the meter will actually put it in sleep mode, where there is still a
slight amount of current draw. This is necessary to maintain the logged
measurements in memory. Therefore, it is recommended to remove the battery
when storing the meter for many months at a time, in order to preserve battery
life.
Low-Battery Error after Battery Replacement
A master reset will usually correct this error, please see the master reset
section for details and cautions. If a master reset does not remove the low
battery indicator, please double check that the voltage of your new battery is
above 2.8 V, this is the threshold for the indicator to turn on.
Master Reset
If a meter ever becomes non-responsive or experiences anomalies, such as a low
battery indicator even after replacing the old battery, a master reset can be
performed that may correct the problem. Note that a master reset will erase
all logged measurements from memory.
Step 1: press the power button so that the LCD display is activated.
Step 2: Slide the battery out of the holder, which will cause the LCD
display to fade out.
Step 3: After a few seconds, slide the battery back into the holder. The
LCD display will flash all segments and then show a revision number (e.g.
“R1.0”). This indicates the master reset was performed and the display should
return to normal.
Error Codes and Fixes
Error codes will appear in place of the real-time reading on the LCD display
and will continue to flash until the problem is corrected. Contact Apogee if
the following fixes do not rectify the problem.
Err 1: battery voltage out of range. Fix: replace CR2320 battery and
perform master reset.
Err 2: sensor voltage out of range. Fix: perform master reset.
Err 3: not calibrated. Fix: perform master reset.
Err 4: CPU voltage below minimum. Fix: replace CR2320 battery and perform
master reset.
Modifying Cable Length
Although it is possible to splice additional cable to the separate sensor of
the appropriate MQ model, note that the cable wires are soldered directly into
the circuit board of the meter. Care should be taken to remove the back panel
of the meter in order to access the board and splice on the additional cable,
otherwise two splices would need to be made between the meter and sensor head.
See Apogee webpage for further details on how to extend sensor cable length:
(http://www.apogeeinstruments.com/.
Unit Conversion Charts
Apogee MQ series quantum sensors are calibrated to measure PPFD in units of
µmol m-2 s-1. Units other than photon flux density (e.g., energy flux density,
illuminance) may be required for certain applications. It is possible to
convert the PPFD value from a quantum sensor to other units, but it requires
spectral output of the radiation source of interest. Conversion factors for
common radiation sources can be found on the Unit Conversions page in the
Support Center on the Apogee website (http://www.apogeeinstruments.com/unit-
conversions/).
A spreadsheet to convert PPFD to energy flux density or illuminance is also
provided on the Unit Conversions page in the Support Center on the Apogee
website
(http://www.apogeeinstruments.com/content/.
RETURN AND WARRANTY POLICY RETURN POLICY
Apogee Instruments will accept returns within 30 days of purchase as long as the product is in new condition (to be determined by Apogee). Returns are subject to a 10 % restocking fee.
WARRANTY POLICY
What is Covered
All products manufactured by Apogee Instruments are warranted to be free from
defects in materials and craftsmanship for a period of four (4) years from the
date of shipment from our factory. To be considered for warranty coverage an
item must be evaluated by Apogee. Products not manufactured by Apogee
(spectroradiometers, chlorophyll content meters, EE08-SS probes) are covered
for a period of one (1) year.
What is Not Covered
The customer is responsible for all costs associated with the removal,
reinstallation, and shipping of suspected warranty items to our factory. The
warranty does not cover equipment that has been damaged due to the following
conditions:
- Improper installation or abuse.
- Operation of the instrument outside of its specified operating range.
- Natural occurrences such as lightning, fire, etc.
- Unauthorized modification.
- Improper or unauthorized repair.
Please note that nominal accuracy drift is normal over time. Routine recalibration of sensors/meters is considered part of proper maintenance and is not covered under warranty.
Who is Covered
This warranty covers the original purchaser of the product or other party who
may own it during the warranty period.
What Apogee Will Do
At no charge Apogee will:
- Either repair or replace (at our discretion) the item under warranty.
- Ship the item back to the customer by the carrier of our choice.
PRODUCTS BEYOND THE WARRANTY PERIOD
For issues with sensors beyond the warranty period, please contact Apogee at
techsupport@apogeeinstruments.com to
discuss repair or replacement options.
OTHER TERMS
The available remedy of defects under this warranty is for the repair or
replacement of the original product, and Apogee Instruments is not responsible
for any direct, indirect, incidental, or consequential damages, including but
not limited to loss of income, loss of revenue, loss of profit, loss of data,
loss of wages, loss of time, loss of sales, accruement of debts or expenses,
injury to personal property, or injury to any person or any other type of
damage or loss. This limited warranty and any disputes arising out of or in
connection with this limited warranty (“Disputes”) shall be governed by the
laws of the State of Utah, USA, excluding conflicts of law principles and
excluding the Convention for the International Sale of Goods. The courts
located in the State of Utah, USA, shall have exclusive jurisdiction over any
Disputes.
This limited warranty gives you specific legal rights, and you may also have
other rights, which vary from state to state and jurisdiction to jurisdiction,
and which shall not be affected by this limited warranty. This warranty
extends only to you and cannot by transferred or assigned. If any provision of
this limited warranty is unlawful, void or unenforceable, that provision shall
be deemed severable and shall not affect any remaining provisions. In case of
any inconsistency between the English and other versions of this limited
warranty, the English version shall prevail. This warranty cannot be changed,
assumed, or amended by any other person or agreement.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>