Newman Medical MAN-0003-C Vascular Diagnostic System User Manual
- June 15, 2024
- Newman Medical
Table of Contents
- MAN-0003-C Vascular Diagnostic System
- Chapter 1 – About simpleABI
- Chapter 2 – Safety
- Chapter 3 – Setting up your simpleABI
- Chapter 4 – simpleABI Reporting Software
- Chapter 5 – Performing Exams
- Chapter 6 – Detailed Description
- Chapter 7 – Tips & Troubleshooting
- Chapter 8 – Maintenance, Cleaning & Disposal
- Chapter 9 – Specifications
- Chapter 10 – Electromagnetic Compatibility
- Chapter 11 – Warranty, Service, & Support
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Vascular Diagnostic System
User Manual
MAN-0003-C Vascular Diagnostic System
Newman Medical 5350 Vivian Street, Unit C Arvada. CO 80002
1-800-267-5549
info@newman-medical.com
www.newman-medical.com
COMMON TERMINOLOGY
ABI – Ankle Brachial Index – the ratio of the ankle systolic blood pressure
divided by the brachial (arm) systolic blood pressure.
Distal – situated away from the center of the body – i.e. “downstream”. An
example – the ankle is distal to the calf.
Distal Pressure – A pressure taken distal to the method of occlusion. I.e.
taking the brachial pressure below (distal to) the cuff.
DP – (or DPA) — Dorsalis Pedis Artery — the artery on top of the foot. Used in
the ABI exam.
MHz – Megahertz – Millions of cycles per second –the frequency of the sound
transmitted by a vascular probe is measured in megahertz – 8 million cycles
per second, for example.
PAD – Peripheral Artery Disease – Typically atherosclerosis (also known as
hardening of the arteries) in the leg.
PPG – Photoplethysmography –detects the pulse using reflected light from a
small emitter/detector combination.
Proximal – Points situated nearer to the center of the body – i.e. “upstream”.
An example – the calf is proximal to the ankle.
PT – (or PTA) – Posterior Tibial Artery – artery on the inside of the ankle
used in the ABI exam.
PVD – Peripheral Vascular Disease – Includes venous as well as arterial
disease.
PVR – Pulse Volume Recording – a means of evaluating the condition of the
arteries by looking at a waveform of small changes in cuff pressure caused by
pulses in the arteries under the cuff.
Reactive Hyperemia – the transient increase in organ blood flow that occurs
following a brief period of restriction of blood flow. In lower arterial this
is typically done by placing a cuff on the thigh and occluding blood flow for
3-5 minutes, then quickly releasing. An extremely painful exam.
Segmentals – Pressures and PVRs just like the ABI but done in more cuffs on
the leg to assist in localizing the site of any occlusion. The Doppler probe
can remain at the ankle for the pressures.
TBI – Toe Brachial Index – the ratio of the pressure in the toe divided by the
brachial pressure. It is generally measured with a small toe cuff and a PPG
probe
Chapter 1 – About simpleABI
Congratulations on your purchase of a simpleABI System from Newman Medical.
The simpleABI was designed to be the easiest and most intuitive system on the
market for initial vascular assessment.
We hope you find that to be the case. If not, please do not hesitate to call
us and let us know how we can improve the product for you.
What is included with simpleABI PVR Systems
Handheld Doppler with 8MHz probe
PVR Hardware
OPTIONAL – PPG Probe (ABI-300 only)
Vascular Cuffs
USB Cable
OPTIONAL – Laptop PC – simpleABI is designed to work with a computer certified
to IEC 60601-1, IEC-623681 or IEC 60950-1 and running Windows 10 (32 or 64
bit) or Windows 11 (64 bit) in the Operator environment.
simpleABI User Manual
Chapter 2 – Safety
Safety of Equipment
simpleABI Systems are medical equipment. Although designed to withstand
everyday use, care must be taken to ensure safe and continued operation. Do
not abuse equipment. simpleABI Systems were designed according to domestic and
International consensus safety standards. Throughout design of this product,
safety was the paramount concern. In view of that, this product was designed
according to the principle of reducing risk AFAP (As Far As Possible).
Intended Use
This product is intended for use by health care providers for detection of
blood flow in the veins and arteries and as an aid in the diagnosis of
peripheral arterial disease.
CAUTION: Federal law restricts this device to sale by or on the order of
a physician or other licensed practitioner.
Indications
Indications for noninvasive physiologic testing
Common indications for lower extremity arterial physiologic evaluation
include, but are not limited to:
- Peripheral arterial disease (PAD)
- Exercise-related limb pain (claudication symptoms)
- Limb pain at rest
- Extremity ulcer/gangrene
- Assessment of healing potential
- Absent peripheral pulses
- Digital cyanosis
- Cold sensitivity
- Arterial trauma and aneurysms
- Suspected arterial embolization
- Pre-procedure assessment (e.g., angioplasty, stent, bypass graft)
- Follow-up endovascular procedure (e.g., angioplasty, stent)
- Follow-up bypass graft
- Prognosis for healing (e.g., ulcerations, amputations)
Contraindications
Contraindications for noninvasive physiologic testing
Contraindications for lower extremity arterial physiologic evaluation include,
but are not limited to:
- Morbidly obese patients in whom a high-thigh pressure may be not obtainable
- Casts, dressings, staples and/or open wounds
- Suspected or known acute deep venous thrombosis (DVT)
- Incompressible vessels with arterial calcification (TBI may be obtained)
- Post-interventional procedure (e.g., stenting, bypass graft)
- Recent surgery, trauma, ulcers, casts or dressings that should not be compressed by a blood pressure cuff
- Involuntary movement/tremors may render waveform collection suboptimal or unreliable
Contraindications of blood pressure measurements
Contraindications of blood pressure measurements include, but are not limited
to:
- Do not take an arm pressure in an arm with a shunt or dialysis graft.
- Patients with a side affected by stroke, mastectomy or renal fistula should avoid having blood pressure readings taken on this side.
- If a patient has an IV catheter in one arm, use the other arm to take the reading to prevent potential damage to the catheter and interruption of administration of medication.
- Lymphedema
- Paresis or paralysis
- Arterial or venous lines
WARNING: The device is not to be used on or near the eyes.
WARNING: The device is for use only on intact skin.
WARNING: Do not leave a cuff inflated at high pressures for an extended
period of time when on a patient.
WARNING: Probes are not intended to be used on open skin. If there is
evidence of open wound contamination, disinfect the probe before using again
as described Chapter 8 – Maintenance, Cleaning.
WARNING: Any open lesion with the potential for contamination should be
covered with an impermeable dressing.
WARNING: Connect only to a computer certified to IEC/AAMI 60601-1, IEC/UL
62368-1 or IEC/UL 60950-1.
Computer is not intended for patient contact.
WARNING: To avoid safety or compliance issues, no modification of this
equipment is allowed.
WARNING: Components of simpleABI Systems are not user serviceable. To
maintain patient safety, do not maintain or clean equipment while in use.
WARNING: Not intended for use in conjunction with HF Surgical equipment.
WARNING: Doppler probe is not intended for immersion in liquids.
CAUTION: The device is not to be plugged into a telephone or modem
system.
CAUTION: Do not inflate a cuff when it is not on a patient.
AIUM Statements
As Low As Reasonably Achievable (ALARA) Principle
Approved May 19, 2020
The potential benefits and risks of each examination should be considered. The
as low as reasonably achievable (ALARA) principle should be observed when
adjusting controls that affect the acoustic output and by considering both the
transducer dwell time and overall scanning time. Practicing ALARA requires
that users do all of the following:
-
Apply correct examination presets if built into the diagnostic ultrasound device. The review of manufacturer default presets for appropriateness is encouraged.
-
Adjust the power to the lowest available setting that provides diagnostic-quality images. If appropriate, reduce power at the end of each examination so the next user will start with the lowest acoustic output setting.
-
Monitor the mechanical index (MI) and thermal index (TI). Know the recommended upper limit of the MI, TI, and related duration limitations for the type of examination being performed.
-
Move/lift the transducer when stationary imaging is not necessary to reduce the dwell time on a particular anatomic structure. When possible, avoid fields of view that include sensitive tissues such as the eye, gasfilled tissues (lung and ntestines), and
fetal calcified structures(skull and spine) -
Minimize the overall scanning time to that needed to obtain the required diagnostic information.
Prudent Use and Clinical Safety
Approved May 20, 2019
Diagnostic ultrasound has been in use since the late 1950s. Given its known
benefits and recognized efficacy for medical diagnosis, including use during
human pregnancy, the American Institute of Ultrasound in Medicine herein
addresses the clinical safety of such use: No independently confirmed adverse
effects caused by exposure from present diagnostic ultrasound instruments have
been reported in human patients in the absence of contrast agents. Biological
effects (such as localized pulmonary bleeding) have been reported in
experimental mammalian systems at diagnostically relevant exposures, but the
clinical relevance of such effects is either not significant or is not yet
known. Increased outputs and time of exposure can increase the likelihood of
bioeffects. Ultrasound should be used only by qualified health professionals
to provide medical benefit to the patient. Ultrasound exposures during
examinations should be as low as reasonably achievable (ALARA)
Safety in Research Using Diagnostic Ultrasound
Approved May 20, 2019
Diagnostic ultrasound has been in use since the late 1950s. There are no
confirmed adverse biological effects on patients resulting from this use.
Although no hazard has been identified that would preclude the prudent and
conservative use of diagnostic ultrasound in research, experience from normal
diagnostic practice may not be relevant to potential extended exposure times
and altered exposure conditions in research. It is therefore considered
appropriate to make the following recommendation: When examinations are
carried out for purposes of research, ultrasound exposures should be as low as
reasonably achievable (ALARA) within the goals of the study. In addition,
informed consent, using a form approved by an Institutional Review board,
should be obtained from the patient. Informed consent forms should include
information about the anticipated exposure conditions and how these compare
with normal diagnostic practice. Repetitive and prolonged exposures on a
single patient should be justified and consistent with prudent and
conservative use.
Chapter 3 – Setting up your simpleABI
Installation of your simpleABI is straightforward but may require a few common
tools. This installation is typically handled by a trained Newman Medical
salesperson.
Assembly Directions
- Remove all materials from packaging.
- Check that inventory matches packing list.
- Plug PVR hardware into computer using included USB cable.
- Use tubing assembly, as shown below, to attach inflation bulb, PVR box hardware, and cuff tubing.
NOTE: Care should be taken when attaching hoses. Do not overtighten
connectors or damage may occur. FIELD INSTALLATION OF SIMPLEABI SOFTWARE
This section applies only to systems where the software is installed in the
field – computers shipped from Newman Medical already have the software loaded
and this section does not apply.
CAUTION: simpleABI is only supported on the following Windows Operating
Systems:
- Windows 10 (32 or 64 bit)
- Windows 11 (64 bit)
Do not install on any other operating systems prior to contacting Newman
Medical Technical Support for information.
CAUTION: Use of other applications (word processing, games, internet) is
not an appropriate use of a medical device and must be avoided to ensure
proper operation of the simpleABI system.
Software Installation Directions
- Plug in included USB Flash Drive into PC
- Copy .msi to C:\ (.msi file will have different names depending on simpleABI system purchased)
- Double Click *.msi file to begin installation
NOTE: DO NOT INSTALL OFF REMOVABLE MEDIA
PVR Driver Installation
- Plug in PVR box. PVR uses Windows default drivers and is ‘plug and play’.
- Driver should be properly installed and can be checked by opening simpleABI Reporting Software, opening an exam, and clicking any PVR button.
a. If installed properly, top of pop-up waveform window will display ‘Connected’ and the following icon. b. If not installed properly, pop-up waveform window will display ‘Not Connected’ and the following icon.
Chapter 4 – simpleABI Reporting Software
File Menu
Beginning a new exam
From the computer Desktop, select the simpleABI icon. The simpleABI start-up
screen with menu items on the top will appear. Select File→New and the exam to
be performed. Depending on the system chosen, you will have different choices
of exams. Choose an exam and it will open on the screen.
Loading an Exam
From the simpleABI start-up screen, Select File→Load and the exam to be
performed. You may select any previous report to populate the loading exam
with data relevant to the exam. I.e., you can load an ABI exam into a
segmental exam; all patient information and arm/ankle pressures and waveforms
from previous exam will be loaded. You would need only to perform upper leg
pressure/waveform measurements. Depending on the system chosen, you will have
different choices of exams. Choose an exam to load, select a previous exam and
it will open on the screen.
Saving an Exam
You can save your finished report using CTRL+S or selecting File→Save. Files
are appended with Patient ID and a timestamp to differentiate exams. You may
also overwrite this default file name. By default, the save command
automatically saves the exam in PDF and XML data files. You may select
additional file types under Edit→Output Formats (TIFF, BMP, JPG, CSV)
Printing an Exam
You can print your exams using CTRL+P or selecting File→Print
Closing an exam
You can close any open exam or editing form by selecting File→Close Form or
CTRL+W
Exiting the software to exit software, select File→Close Form or ALT+F4.
Edit Menu
REPORT INFORMATION You can edit referring doctor, examiner, interpreting
doctor, and practice information and logo under this menu item. Logos and
signatures must be of BMP format to be loaded into the system. There are some
samples pre-filled, and you can add, edit, or delete these as you deem
necessary.
Add/Delete/Edit Doctor/Examiner/Practice Information/Logos
Under the Edit menu there are multiple options for editing the configuration
of your software. You may add, delete, and edit practice information, examiner
information, doctor information, output formats, impressions, and ICD codes.
You can also upload practice logos to customize reports.
To close any editing form, select File→Close Form
OUTPUT FORMATS You can select from different formats to fit your needs.
Options available are PDF, JPG, BMP, TIF, CSV (XML is our data format and
cannot be deselected). By default, PDF is selected.
IMPRESSIONS exam impressions can be pre-filled using this menu item. Use only
numbers for impression numbers.
ICD CODES ICD codes can be edited, added or deleted using this menu item.
There are many pre-populated ICD codes; however, only some relevant codes have
been selected to ‘Show in List’ on exams. You may choose to edit these as you
deem necessary.
CHART OPTIONS chart options allow different scaling or auto-scale for PVR
waveforms. By default, a scaling factor of 1.00 is selected.
DICOM SETUP (OPTIONAL) An optional license key must be purchased to access
DICOM setup options.
simpleABI DICOM provides the ability to connect to a PACS server, query
worklist, and send image or Standard reporting DICOM formats.
Help
Help is found by selecting the Help Menu. You may view system information as
well as, with an internet connection, be directed to Newman Medical’s home
page where you can browse our support pages with helpful hints and videos on
using your simpleABI System.
Chapter 5 – Performing Exams
Interpretation of waveforms
Pulse Volume Recording (PVR) is used to measure the change in pressure in the
cuff caused by the small changes in volume that occur with each pulse. The
analysis of the waveform is based on the shape of the waveform rather than the
amplitude. PVRs are not affected by calcified arterial walls and are
relatively easy to perform.
Normal: Waveforms with a rapid rise and a prominent reflected wave (dicrotic notch) Moderately Abnormal: Waveform starts to broaden and no reflected wave. Abnormal: No reflected wave, slow rise. Potential ‘flatline’
Determining site of occlusion with PVR waveforms:1 A normal waveform at a site
indicates that there is not likely any occlusive disease proximal to that that
cuff.
1 Daigle, Robert J., Techniques in Noninvasive Vascular Diagnosis, Summer
Publishing, www.summerpublishing.com
ABI Examination
Overview: The ABI has become the best known and standardized test for PAD.
Fundamentally it is the systolic blood pressure taken with the cuff at the
ankle divided by the systolic pressure with the cuff on the upper arm.
Have the patient lie down in a supine position relaxed, head and heels
supported in a comfortably warm room. Wrap appropriately sized cuffs (usually
12 cm on the arms and 10 cm on the ankles). On the arms apply the cuffs as you
would for a standard blood pressure. On the legs, have the patient place their
feet flat on the table with knees bent as you apply the cuff above their
ankle. Cuffs should fit snugly so that fingers should slide between the cuff
and limb with difficulty. The lower edge of the ankle cuff should be about an
inch above the top of the medial malleolus (ankle bone).
Although the system will allow you to take the pressures and waveforms in any
order, try to be consistent. The standard software workflow is to perform the
right arm, right ankle, left arm, and finally left ankle.
Arm Pressure
- Connect the hose from the PVR to the first arm cuff. Place the Doppler probe over either the radial (preferred) or the brachial artery. HINT Hold the probe close to the end. It also helps to rest your hand on the patient to keep the probe in place. One of the keys to a successful exam is being able to keep the probe in place as you inflate and deflate. If it moves, you will not be able to hear the Doppler sounds return and you will have to repeat the inflation. Move the probe slowly back and forth laterally until the strongest arterial signal is heard.
- Now inflate the cuff until the Doppler signal disappears and continue to inflate for another 10-20mmHg.
- Slowly (about 2mmHg/second) deflate the cuff.
- Note the pressure from the aneroid at the moment that you hear the first Doppler arterial sound return and enter that pressure into the brachial pressure box. This is the systolic pressure.
Ankle DP Pressure
- Move the tubing to the ankle cuff.
- Obtain the ankle pressure at the Dorsalis Pedis (DP). Place the Doppler over the DP artery on the top of the foot. Use light pressure and move the probe laterally to obtain the best signal. The DP is often harder to find; be sure not to press too hard as it is over boney structures and relatively easy to occlude.
- Now inflate the cuff until the Doppler signal disappears and continue to inflate for another 10-20mmHg.
- Slowly (about 2mmHg/second) deflate the cuff.
- Note the pressure from the aneroid at the moment that you hear the first Doppler arterial sound returnand enter that pressure into the DP pressure box.
Ankle PT Pressure
- Obtain the ankle pressure at the Posterior Tibial (PT). Place the Doppler over the PT artery near the ankle bone. Use light pressure and move the probe laterally to obtain the best signal, making sure to angle the probe proximally. The PT is often found at the midpoint between the ankle bone and the Achilles tendon.
- Now inflate the cuff until the Doppler signal disappears and continue to inflate for another 10-20mmHg.
- Slowly (about 2mmHg/second) deflate the cuff.
- Note the pressure from the aneroid when you hear the first Doppler arterial sound return and enter that pressure into the PT pressure box.
Obtain Waveforms
- On the screen, click on the respective Waveform Field in the report associated with the patient’s right side (the left side of the screen).
- Using the aneroid, inflate the cuff to about 80 mmHg. Deflate the cuff down to about 65mm and stop. You will see a visual guide bar on the screen above the Get Waveform button. The center section of the target pressure guide will turn green (if the guide is red, then pressure is too low or too high). When a constant green, click the Get Waveform button. PVRs require patient cooperation; limb motion affects the waveform. Patients should be instructed not to move or talk during this test. The waveform will appear and when complete the pop-up will disappear and the waveform will be present in the Waveform Field in the exam.
- If you are not satisfied with the waveform, leave the cuff inflated (in the green) and click on the waveform field and then Get Waveform buttons again to overwrite the previous waveform.
- When complete, release the air in the cuff.
Left Side
- Repeat steps above for the left side.
The interpretation of the ABI values is based on recent guidelines.2
HINT If the ankle pressure is high, above 200 mmHg, or the cuff cannot obliterate the Doppler sounds, this indicates that the artery may be incompressible due to calcification.
SAVING/PRINTING REPORT
You can save your finished report using CTRL+S or selecting File→Save. Files
are appended with Patient ID and a timestamp to differentiate exams. You may
also overwrite this default file name. By default, the save command
automatically saves the exam in PDF and XML data files. You may select
additional file types under Edit→Output Formats (TIFF, BMP, JPG, CSV)
The ABI with Toe (TBI) Examination
Background: The ABI with Toe exam is typically performed after an ABI has been
done, especially if the larger ankle arteries appear to be incompressible
(pressure over 200mmHg) or anytime more information is needed about small
vessel disease. The TBI is done using the PPG probe. The patient is supine and
rested in a warm room.
Setting up
On the computer desktop, select the simpleABI icon. To open the ABI with Toe
Report, select File→New→New ABI with Toe Report. The ABI with Toe report form
will open on the screen. Wrap cuffs on the arms (typically 12cm). Wrap cuffs
on the ankles (typically 10cm) and wrap digit cuffs on the great toes as far
down towards the base as possible. Attach the PPG probe and sensor to the
DigiDop, replacing the Doppler probe.
If you have performed a previous ABI Exam, you may load the results into this
exam by selecting File→Load→ABI with Toe Report and selecting the ABI Exam you
previously performed.
2 2016 AHA/ACC Guideline on the Management of Patients With Lower Extremity Peripheral Artery Disease, J Am Coll Cardiol. 2017 Mar, 69 (11) e71-e126.
Brachial Pressure
- Begin with the right brachial. Place the Doppler probe at a 45-degree angle to the skin over the radial or brachial artery. Use plenty of gel and slowly move the probe laterally until the best signal is obtained. (You may also use the PPG sensor to listen for the pulse. If using the PPG, place the sensor on the fleshy part of the finger opposite the fingernail. Wait for consistent beeps from the system.)
- Using the handheld aneroid, inflate the cuff until you no longer hear the signal – continue for an additional 10-20 mmHg.
- The thumb button on the aneroid is a variable release mechanism – i.e., the more you depress the button the faster the deflation rate. Using the thumb release, deflate at the suggested rate of 2mmHg/second (10mmHg for every 5 seconds may be slightly easier to monitor)
- When you hear the Doppler signal return, note the pressure from the aneroid. Enter that pressure into the right brachial field on the exam.
Ankle Pressures
- Press tab or use the cursor to move to the Dorsalis Pedis (DP) field. Find the arterial signal using the Doppler probe on the dorsalis pedis artery on top of the foot. Obtain the arterial pressure in the same manner you did on the arm. (Using the aneroid, inflate until occlusion + 10-20mmHg, release at 2mmHg/second, note pressure when Doppler signal returns)
- Press tab or use the cursor to move to the Posterior Tibial (PT) field. Find the Doppler signal on the posterior tibial artery. Obtain the arterial pressure in the same manner as other sites.
Ankle Waveform
- On the screen, click on the respective Waveform Field in the report associated with the patient’s right side (the left side of the screen).
- Using the aneroid, inflate the cuff to about 80 mmHg. Deflate the cuff down to about 65mm and stop. You will see a visual guide bar on the screen above the Get Waveform button. The center section of the target pressure guide will turn green (if the guide is red, then pressure is too low or too high). When a constant green, click the Get Waveform button. PVRs require patient cooperation; limb motion affects the waveform. Patients should be instructed not to move or talk during this test. The waveform will appear and when complete the pop-up will disappear and the waveform will be present in the Waveform Field in the exam.
- If you are not satisfied with the waveform, leave the cuff inflated (in the green) and click on the waveform field and then Get Waveform buttons again to overwrite the previous waveform.
- When complete, release the air in the cuff.
Toe Pressure
- Press tab or use the cursor to move to the toe pressure field. Find the arterial signal using the Doppler or PPG probe on the toe. Obtain the arterial pressure in the same manner you did on the arm. NOTE digit cuffs inflate rapidly. (Using the aneroid, inflate until occlusion +10-20mmHg, release at 2mmHg/second, note pressure when Doppler/PPG signal returns)
Toe Waveform
- On the screen, click on the respective Waveform Field in the report associated with the patient’s toe.
- Using the aneroid, inflate the cuff to about 80 mmHg. Deflate the cuff down to about 65mm and stop. You will see a visual guide bar on the screen above the Get Waveform button. The center section of the target pressure guide will turn green (if the guide is red, then pressure is too low or too high). When a constant green, click the Get Waveform button. PVRs require patient cooperation; limb motion affects the waveform. Patients should be instructed not to move or talk during this test. The waveform will appear and when complete the pop-up will disappear and the waveform will be present in the Waveform Field in the exam.
- If you are not satisfied with the waveform, leave the cuff inflated (in the green) and click on the waveform field and then Get Waveform buttons again to overwrite the previous waveform.
- When complete, release the air in the cuff.
Left Side
- Repeat the above pressures and waveform sequence for the left side of the patient.
Saving/Printing
Save the report by selecting File→Save or pressing CTRL+S. Print the report by
selecting File→Print or pressing CTRL+P.
Exam Helpful Hints
A warm toe is a big help. Keep the toe covered if you can with a towel or warm
towel, if available. Let the patient rest and relax before inflating the
cuffs. Anchor the cord from the PPG sensor to the foot with some adhesive
tape.
Toe pressures are difficult to obtain in patients with small vessel disease.
Pressures can be obtained with the Doppler probe as well, but it is more
difficult as the digit arterial signal is hard to locate.
Chapter 6 – Detailed Description
Operation and Use
The Doppler, probes, and pressure sensors are designed to aid in diagnosis of
vascular conditions including aiding in obtaining systolic pressures during a
vascular examination.
Main Unit
The body of the unit is a one-piece enclosure. The unit was designed for
comfort and ease-of-use during examinations. Each unit is tested to ensure the
highest quality possible.
Turning unit ON/OFF
Power to the unit is controlled by the ON/OFF button located in the center of
the unit. Pressing and releasing the button will cycle the unit through the
power sequence.
The DigiDop automatically turns itself off after 3 minutes of inactivity. This
auto shut off feature preserves battery life and serves to eliminate complete
battery drain in case of accidental failure to shut off the unit.
Battery Monitoring
The DigiDop periodically checks the status of the batteries. When the battery
symbol on the LCD is half full the batteries are getting low and replacement
batteries should be available. When the battery symbol on the LCD flashes, the
battery level is low and batteries should be replaced immediately.
Volume
The DigiDop has multiple volume levels. These are adjusted by sliding the
volume slider on the left-hand side of the main unit.
Probes:
8 MHz – Vascular. simpleABI Systems come standard with an 8MHz probe designed
for locating shallow lying vessels.
The pen-tip sensor face aids in the location of specific vessels during
examination.
5 MHz – Vascular. An optional 5MHz probe is available for simpleABI Systems
and designed for locating deeper lying vessels. The wider sensors may aid in
maintaining the probe location on the vessel during examination.
PPG – The Photoplethysmograph probe uses the absorption of infrared light by
blood cells in the blood to produce an audible beep each time it detects a
vascular pulse in the near surface vessels.
Chapter 7 – Tips & Troubleshooting
Obtaining Doppler Signals
CAUTION: For any examination utilizing a Doppler, it is essential that an
adequate supply of gel is used to transmit the ultrasound energy from the
probe to the surface of the skin. Re-apply more gel if it starts to dry out or
spread so thinly that an air gap occurs between the probe and the skin. It is
not necessary to cover the entire surface of the probe, only the probe face.
Applying too much gel makes the unit difficult to clean and does not aid in
the performance of the probe. Given the small area of the vascular probes, the
strength of the Doppler signal is highly location specific.
Poor Sound Quality
Inadequate gel use-apply more gel
Probe location-search for vascular sounds as described in Chapter 5 –
Performing Exams
Obtaining Waveforms
- Wrap the cuff snugly.
- While wrapping cuffs, do not let the patient try to help by lifting their leg – as soon as they relax their muscles the cuff will become loose.
- Placing a pillow under the patient’s heels may aid the examination.
- Have the patient remain as still and quiet as possible while taking the waveforms.
- If the patient has tremors that interfere with the waveform, having them do a few dorsiflexions with their toes before taking the waveform can sometimes help.
Battery Symbol on LCD flashing
The voltage of the batteries is low. Charge the batteries as soon as possible.
Chapter 8 – Maintenance, Cleaning & Disposal
CAUTION: Newman Medical recommends leaving your computer on overnight
once a week to allow for system updates to occur while system is not in use.
CAUTION: simpleABI components are not designed for liquid immersion. Do
not soak or drop the main unit or probes in liquid.
CAUTION: The equipment not designed for sterilization processes such as
autoclaving or gamma radiation.
The equipment requires very little maintenance. It is important, however, for
the continued functionality of the unit and the safety of the patients to that
the unit is cleaned and examined regularly as follows:
After every examination
Excess gel should be wiped off after each examination. Unit should be cleaned
with a damp water or alcohol based wipe. Mild soap or detergent can be used.
In particular pay attention to any surface openings on the unit including, but
not limited to, the speaker grill, the battery compartment, the audio output,
and the parting line between the front and back shell.
Practitioners should wash hands and change gloves after every exam. Please
follow local and hospital guidelines for cleaning and disinfection policies.
To disinfect unit, use an appropriate disinfectant spray or wipe and follow
the manufacturer’s instructions.
Store unit in a clean area free from dust and debris in an indoor environment.
If storing the unit for a prolonged period of 90 days or longer without use,
remove the batteries prior to storage.
Periodically
Inspect the unit for signs or cracks or breaks in the surface housing. If any
sign of cracking or damage is evident, use of the unit should be discontinued.
Please contact Newman Medical for service.
Inspect the cuffs for damage. Cuff covers may be washed by removing the
bladders.
Disposal
This device and accessories must be disposed of according to local regulations
after their useful lives. Alternatively, they may be returned to Newman
Medical for recycling and proper disposal.
Battery Replacement
CAUTION: Replace rechargeable batteries only with approved batteries.
Please call customer service or visit our website for further information.
CAUTION: The Doppler unit uses AA batteries. Do not attempt to use any
other size batteries in the unit. Open the battery compartment by pressing the
tab on the battery door and opening the battery compartment.
Remove the drained batteries by pulling one of batteries out of the unit from
the positive (button) end of the battery.
Replace the batteries by paying close attention to the polarity indicators on
the battery and the polarity indicators on the battery door. Align the
batteries according to the symbols located in the battery compartment.
NOTE: If the batteries have been incorrectly inserted, the unit will not
work, but will not be damaged. Please reinsert the batteries correctly.
Chapter 9 – Specifications
Level of Protection against electrical shock
Type B Applied Part; Class II Equipment
Designed to meet the following standards:
IEC60601-1, IEC60601-2, IEC60601-2-37
simpleABI System
Weight (inclusive)| 3.6 kgs (8 pounds)
Transport/Storage Temperature| -20° ~ 50°C (-4° ~ 122°F)
Transport/Storage Humidity| 5% ~ 90%, non-condensing
Doppler Unit
Dimensions-(h x w x d)| 150x65x35 mm (6×2.5×1.25 in)
Weight (with probe & batteries)| 370 grams (13 ounces)
Operating Temperature| 10° ~ 40°C (50° ~ 104°F)
Operating Humidity| 30% ~ 75%
Transport/Storage Temperature| -20° ~ 50°C (-4° ~ 122°F)
Transport/Storage Humidity| 5% ~ 90%, non-condensing
Battery Voltage, type| 3 x 1.5 volts, AA
Battery Life| 800, 1-minute exams
PVR Unit
Dimensions-imperial (h x w x d)| 2.75×2.0x0.8
Weight| 200 grams (8 ounces)
Interface| USB, Type B
Acoustic Properties (DigiDop)
System:| DigiDop
Transducer Model:| 5MHz
Operating Mode:| Continuous Wave (CW)
Application(s):| Peripheral Vascular
Acoustic Output| MI| ISPTA.3
(mW/cm2)| ISPPA.3
(mW/cm2)
---|---|---|---
Global Maximum Value| 0.0223| 86.4| 86.4
Associated Acoustic
Parameters| Pr.3| (Mpa)| 0.041| |
Wo| (mW)| | 9.26| 9.26
fc| (MHz)| 5.61| 5.61| 5.61
zsc.| (cm)| 1.10| 1.10| 1.10
Beam Dimensions| x-6 (cm)| | 0.154| 0.154
y-6 (cm)| | 0.540| 0.540
PD| (usec)| CW| | CW
PRF| (Hz)| n/a| | n/a
EBD| Az. (cm)| | 1.052|
Ele. (cm)| | 0.526|
System:| DigiDop
---|---
Transducer Model:| 8MHz, narrow
Operating Mode:| Continuous Wave (CW)
Application(s):| Peripheral Vascular
Acoustic Output| MI| ISPTA.3
(mW/cm2)| ISPPA.3
(mW/cm2)
---|---|---|---
Global Maximum Value| 0.0495| 555| 555
Associated Acoustic
Parameters| Pr.3| (Mpa)| 0.0923| |
Wo| (mW)| | 9.02| 9.02
fc| (MHz)| 7.84| 7.84| 7.84
Zsp| (cm)| 0.50| 0.50| 0.50
Beam Dimensions| x-6 (cm)| | 0.231| 0.231
y-6 (cm)| | 0.121| 0.121
PD| (usec)| CW| | CW
PRF| (Hz)| n/a| | n/a
EBD| Az. (cm)| | 0.203|
Ele. (cm)| | 0.457|
Measurement Uncertainties:
Total uncertainty for power: | 28.2% |
---|---|
Total uncertainty for ISPTA: | 28.2% |
Total uncertainty for fc: | 2.0% |
Total uncertainty for MI: | 14.1% |
ISPTA.3
derated spatial-peak temporal-average intensity (milliwatts per square
centimeter).
ISPPA.3
derated spatial-peak pulse-average intensity (watts per square centimeter).
The value of IPA.3 at the position of global maximum MI (IPA.3@MI) may be
reported instead of ISPPA.3 if the global maximum MI is reported.
MI
Mechanical Index. The value of MI at the position of ISPPA.3, (MI@ISPPA.3) may
be reported instead of MI (global maximum value) if ISPPA.3 is ≤ 190W/cm2 .
pr.3
derated peak rarefactional pressure (megapascals) associated with the transmit
pattern giving rise to the value reported under MI.
Wo
ultrasonic power (milliwatts). For the operating condition giving rise to
ISPTA.3, Wo is the total time-average power; for the operating condition
subject to reporting under ISPPA.3, Wo is the ultrasonic power associated with
the transmit pattern giving rise to the value reported under ISPPA.3.
fc
center frequency (MHz). For MI and ISPPA.3, fc is the center frequency
associated with the transmit pattern giving rise to the global maximum value
of the respective parameter. For ISPTA.3, for combined modes involving beam
types of unequal center frequency, fc is defined as the overall range of
center frequencies of the respective transmit patterns.
zsp
the axial distance at which the reported parameter is measured (centimeters).
x-6, y-6
are respectively the in-plane (azimuthal) and out-of-plane (elevational) -6 dB
dimensions in the x-y plane where zsp is found (centimeters).
PD
pulse duration (microseconds) associated with the transmit pattern giving rise
to the reported value of the respective parameter.
PRF
the pulse repetition frequency (Hz) associated with the transmit pattern
giving rise to the reported value of the respective parameter.
EBD
the entrance beam dimensions for the azimuthal and elevational planes
(centimeters).
EDS
the entrance dimensions of the scan for the azimuthal and elevational planes
(centimeters).
The reporting values for ultrasonic power, W0, and non-derated spatial average temporal average ISATA required by paragraph 2.1.2 of the FDA Guidance [3] as well as the derated spatial-peak temporal-average intensity, ISPTA.3, provided for reference only, are calculated for all probes as illustrated in the sample calculations below.
For Non-Auto scanning modes reporting parameters are calculated as:
W0 = ISPTA.0 PF
ISATA.0 = W0/(entrance beam area)
ISPTA.3 = ISPTA.0 e -0.069 fc z
Where ISPTA.0, is the non-derated spatial-peak temporal-average intensity, ISATA.0 is the nonderated spatial-average temporal-average intensity at the transducer face and ISPTA.3, is the derated spatial-peak temporal-average intensity, fC, the waveform center frequency, z, the axial distance between the probe and hydrophone, PF, the power factor which is calculated by integrating the normalized cross axis and raster scan data selecting the largest PF value, which is an “effective area” used to calculate W0, the ultrasonic power.
Chapter 10 – Electromagnetic Compatibility
All medical equipment may produce electromagnetic interference or be susceptible to electromagnetic interference. The following are guidance and manufacturer’s declarations regarding EMC for SimpleABI.
- simpleABI Systems need special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the following pages.
- Warning: This equipment is intended for use by healthcare professionals only. As with all electrical medical equipment, this equipment may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures such as re-orienting or relocating the SimpleABI system or shielding the location.
- Portable and Mobile RF communications equipment can affect the performance of SimpleABI. Please use the guidelines and recommendations specified in Tables 15.3.1 and 15.3.2.
- Other Medical Equipment or Systems can produce electromagnetic emissions and therefore can interfere with the functionality of SimpleABI. Care should be used when operating SimpleABI adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, SimpleABI should initially be observed to verify normal operation in the configuration in which it will be used.
- The electrical cables, external power supplies and accessories listed or referenced in this manual have been shown to comply with the test requirements listed in the following tables. Care should be taken to use only manufacturer-recommended cables, power supplies and electrical accessories with SimpleABI. If a third-party supplier offers cables, external power supplies and electrical accessories for use with SimpleABI and they are not listed or referenced in this manual, it is the responsibility of that third-party supplier to determine compliance with the standards and tests in the following tables.
- The use of electrical cables and accessories other than those specified in this manual or referenced documents may result in increased electromagnetic emissions from SimpleABI or decreased electromagnetic immunity of SimpleABI.
Table for Electromagnetic Emissions
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The simpleABI System is intended for use in the electromagnetic environment
specified below. The customer or
the user of the device should assure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment – guidance
RF Emissions – CISPR 11 (Radiated & Conducted)| Group 1| The simpleABI System
uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic
equipment.
RF Emissions – CISPR 11 (Radiated & Conducted)| Class A| The simpleABI System
is suitable for use in all establishments other than domestic establishments
and those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Table for Electromagnetic Immunity
Guidance and Manufacturer’s Declaration — Electromagnetic Immunity
The simpleABl System is intended for use in the electromagnetic environment
specified below. The customer or the end user should assure it is used in such
an environment.
Immunity Test| EN/IEC 60601 Test Level| Compliance Level| Intended
Electromagnetic Environment
Electromagnetic Discharge (ESD)
EN/IEC 61000-4-2| ± 6kV contact
± 8kV air| ± 6kV contact
± 8kV air| Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
EN/IEC 61000-4-4| ± 2kV for power supply
lines
± 1kV for input/output
lines| ± 2kV for power supply
lines
± lkV for input/output
lines| Mains power quality should be that of a typical commercial or hospital
environment.
Power frequency (50/60Hz) magnetic field
EN/IEC 61000-4-8| 3A/m| 3A/m| Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
Guidance and Manufacturers Declaration — Electromagnetic Immunity (continued)
Immunity Test EN/I EC 60601Compliance Level Intended Electromagnetic
Environment Test Level
Conducted RF EN/IEC 61000-4-6
Radiated RF EN/IEC 61000-4-3| 3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz| 3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz| Portable and mobile RF communications
equipment should be used no closer to any part of SimpleABI, including cables,
than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2VP 80MHz to 800 MHz
d = 2.3VP 800MHz to 2.5GHz
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended minimum
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site surveys, should be less than the compliance level in each
frequency range.b
| | | Interference may occur in the vicinity of
equipment marked with the following symbol:
---|---|---|---
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from objects, structures
and people.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the SimpleABl is used exceeds
the applicable RF compliance level above, the SimpleABI should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the SimpleABl.
b Over the frequency range 150kHz to 80MHz, field strengths should be less
than 3V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the SimpleABI.
SimpleABI unit is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of SimpleABI
can help prevent electromagnetic interference by maintaining a minimum
distance between the portable and mobile RF communications equipment
(transmitters) and SimpleABI as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output power of transmitter in watts (W)| Separation distance
according to frequency of transmitter in meters (m)
150kHz to 80MHz
d = 1.2VP| 80MHz to 800MHz
d = 1.2VP| 800MHz to 2.5GHz
d = 2.3VP
0.01| .12| .12| .23
0.1| .38| .38| .73
1.0| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not isted above, the
recommended separation distance (d) in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Chapter 11 – Warranty, Service, & Support
The simpleABI System and components are guaranteed to be free from defects in
material and workmanship for 2 years (5 years for the DigiDop and its probes)
from the original sale of the device. This guarantee includes all parts and
labor required to repair or replace the unit, including shipping the unit back
to the customer. Customer is responsible for the adequate packaging and return
of the unit for servicing. Products will be repaired or replaced in a
reasonable amount of time, to be determined by service personnel.
The manufacturer and distributor of a simpleABI System assume neither
responsibility nor liability for incidental or consequential damages arising
from the purchase of this product.
The manufacturer and distributor of simpleABI Systems are not responsible for
damages occurring from misuse or neglectful handling of the device. Any abuse,
neglect, or alteration of the equipment, including dismantling of the unit
(other than by trained service personnel), from its original specifications
nullify all stated and implied warranties.
To return a unit for servicing:
- Call customer service for a return authorization.
- Clean the product prior to packing and shipping.
- Adequately package and return the unit to:
Newman Medical
Attn: Service 5350 Vivian Street, Unit C Arvada, CO 80002 800-267-5549
MAN-0003-F
References
- medical.com is for sale | www.brandforce.com
- Newman Medical | US manufacturer of the most valuable Dopplers & ABI systems
- Newman Medical | US manufacturer of the most valuable Dopplers & ABI systems
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>