Met One 061 Air Temperature Sensor Instruction Manual

June 10, 2024
Met One

061 Air Temperature Sensor
Instruction Manual

MODEL 065
TEMPERATURE SENSOR
OPERATION MANUAL

GENERAL INFORMATION

1.1 Model 065 is an air temperature sensor designed to mount in 073B or 5980 Radiation Shields. The temperature sensor is shielded by concentric enameled aluminum plates to reflect solar energy. The shields are designed to greatly reduce errors due to direct solar  radiation and reflected terrestrial radiation. The 073B  and the 5980 are naturally aspirated. The 065 sensor is a precision thermistor unit that produces a resistance charge inversely proportional to temperature.
1.2 Sensor Cable and Connections
Model 065 is supplied with quick disconnect connector. Mating cables are supplied with several types of termination. Refer to cable drawings at end of this manual  for particular wiring.
1.3 Specifications

Temperature Range -30°C to +50°C
Radiation Shield Mounting 1-inch tubing or ¾ inch IPS pipe
Weight 2 pounds
Size 14-1/2 inches diameter 6-1/2 inches high

INSTALLATAION

2.1 Mount the radiation shield on a ¾ inch horizontal pipe. The unit will slide over the end of the pipe and will lock into place with the self- contained Allen head  set screws. The horizontal pipe should be positioned so that the sensor will not be located too close to a potential heat source. Generally it is recommended that the  unit be on the north side of a tower with the sensing unit at least 3 feet from the tower. These sensors are durable, field proven devices; however,
DO NOT DROP OR EXPOSE THE SENSOR TO HEAVY SHOCK!!!
2.2 Wiring Connections
The output of the thermistor sensor has a relatively high resistance that varies inversely with temperature. It is important not to introduce any parallel resistance  paths. A parallel resistance path may be established by a dirt/moisture build-up between two sensor leads. This may occur in poorly made splices and unprotected  connections. It is advisable to always use a protective coating on exposed sensor connections. Use a coating such as clastic rubber (RTV).

2.3 Direct Wiring to a Met One Instruments Translator
When the sensor is connected directly to a Met One Instruments Translator Module the sensor is loaded with the appropriate resistor to provide a linear output.
2.4 Direct Connection to a Data Logger
When the sensor is connected to a data logger; the data logger must have a terminating resistor to provide a linear output. Refer to Figure 2-1.

OPERATIONAL CHECK-OUT AND CALIBRATION

3.1 Temperature Sensor Check-out
Compare sensor readings against a precision mercury thermometer or other temperature standard. Use either connected logger, translator Module or Ohmmeter to measure and compare readings of temperature vs. resistance.

MAINTENANCE AND TROUBLESHOOTING

**4.1 General Maintenance Schedule***

6 – 12 Month Intervals:
A. Inspect sensor for proper operation per Section 3.1.
*Schedule is based on average to adverse environments.
4.2 Troubleshooting Procedures
Incorrect sensor signal: check sensor input connections; check temperature vs. sensor output signal using Table 3-1. Verify that the sensor has the correct terminating resistor if not used with a Met One Instruments Translator.
4.3 Temperature Sensor Calibration
The sensors are tested for calibration conformity at the factory. It is not possible to make alterations to the sensor’s calibration, as it is fixed.
4.4 Calibration Verification
Install the 065 inside of a Styrofoam container (ice chest) and push a glass thermometer into the container. With the top on the container and no air leaks it is possible to verify the operation of the sensor ±0.5°C within 30 to 60 minutes.

Met One 061 Air Temperature Sensor - fig1

Table 3-1A
Model 065 RESISTANCE CHART

TEMP DEG C RCAL TEMP DEG C                RCAL
-30 110236 10 26155
-29 104464 11 25436
-28 99187 12 24739
-27 94344 13 24064
-26 89882 14 23409
-25 85760 15 22775
-24 81939 16 22159
-23 78388 17 21561
-22 75079 18 20980
-21 71988 19 20416
-20 69094 20 19868
-19 66379 21 19335
-18 63827 22 18816
-17 61424 23 18311
-16 59157 24 17820
-15 57014 25 17342
-14 54986 26 16876
-13 53064 27 16421
-12 51240 28 15979
-11 49506 29 15547
-10 47856 30 15126
-9 46284 31 14715
-8 44785 32 14314
-7 43353 33 13923
-6 41985 34 13541
-5 40675 35 13167
-4 39421 36 12802
-3 38218 37 12446
-2 37065 38 12097
-1 35957 39 11756
0 34892 40 11423
1 33868 41 11097
2 32883 42 10777
3 31934 43 10465
4 31019 44 10159
5 30136 45 9859
6 29284 46 9566
7 28462 47 9279
8 27667 48 8997
9 26899 50 8450

*VALUE WITH 18.7K RESISTOR IN PARALLEL WITH SENSOR
RANGE –30°C TO +50°C
THERMISTOR BEAD 44203
Table 3-1A

Table 3-1B
Model 065 RESISTANCE CHART

TEMP DEG F RCAL TEMP DEG F RCAL
-22 110236 33 34319
-21 106964 34 33757
-20 103855 35 33207
-19 100895 36 32669
-18 98075 37 32141
-17 95385 38 31625
-16 92816 39 31119
-15 90361 40 30622
-14 88011 41 30136
-13 85760 42 29659
-12 83602 43 29192
-11 81532 44 28733
-10 79543 45 28283
-9 77632 46 27841
-8 75794 47 27408
-7 74025 48 26983
-6 72321 49 26565
-5 70678 50 26155
-4 69094 51 25753
-3 67565 52 25357
-2 66088 53 24969
-1 64661 54 24587
0 63281 55 24212
1 61946 56 23843
2 60654 57 23481
3 59402 58 23125
4 58190 59 22775
5 57014 60 22430
6 55874 61 22091
7 54768 62 21758
8 53694 63 21430
9 52651 64 21108
10 51637 65 20790
11 50652 66 20478
12 49695 67 20170
13 48763 68 19868
14 47856 69 19570
15 46974 70 19276
16 46114 71 18987
17 45277 72 18703
18 44461 73 18422
19 43666 74 18146
20 42890 75 17874
21 42134 76 17606
22 41395 77 17342
23 40675 78 17081
24 39972 79 16825
25 39285 80 16572
26 38614 81 16322
27 37958 82 16076
28 37317 83 15834
29 36691 84 15595
30 36078 85 15359
31 35479 86 15126
32 34892 87 14897

Table 3-1B

Table 3-1B
Model 065 RESISTANCE CHART Continued

TEMP DEG F RCAL TEMP DEG F RCAL
88 14670 106 11061
89 14447 107 10883
90 14227 108 10707
91 14009 109 10534
92 13794 110 10362
93 13583 111 10193
94 13374 112 10025
95 13167 113 9859
96 12963 114 9696
97 12762 115 9534
98 12564 116 9374
99 12368 117 9215
100 12174 118 9059
101 11983 119 8904
102 11794 120 8751
103 11607 121 8600
104 11423 122 8450
105 11241

VALUE WITH 18.7K RESISTOR IN PARALLEL WITH SENSOR
RANGE -22˚F TO +122˚F
TERMISTOR BEAD 44203
Tc= -(R
18700/(18700+R)-12175)/127.096
Rt = -(127.096Tc-12175)18700/(127.096*Tc-12175+18700)
Table 3-1B
065-9800 Rev C
Corporate Sales & Service: 1600 NW Washington Blvd. Grants Pass, OR 97526
Tel 541-471-7111 Fax 541-471-7116
www.metone.com[email protected]

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Met One User Manuals

Related Manuals