W HLER DC 2000 PRO Pressure Computer User Manual
- June 15, 2024
- W HLER
Table of Contents
- Specifications
- Controls and connections
- Signal menu
- Stress test (DVGW-TRGI)
- Main test menu (DVGW-TRGI)
- Leakage rates and utility life measurements
- Min, Max, AVG menu
- Setup menu for basic settings and company logo
- Logger / data transfer
- Changing the batteries
- Accessories
- Information on disposal
- Declaration of conformity
- Warranty and service
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Pressure Computer Wöhler DC 2000PRO 2023-07-23 Art.-Nr.: 20295
The Measure of Technology
Quick reference
loose leaf at the centre
Specifications
1.1 Important information
Before using this instrument, carefully read and observe all notes contained
in these operating instructions.
Fundamentally, skilled personnel only should use the Wöhler DC 2000PRO for the
purpose that it is intended and within the specified data range. No liability
is accepted under any circumstances or guarantee given for results determined
in conjunction with this
instrument nor for any damage that may arise when using this instrument.
The pressure meter can be used on both gaseous media (air or inert gas) and
liquids, e.g. water or fuel oil. In this case, it is important to observe that
remains of liquid at the pressure connections can falsify the result of the
following gas measurement.
Therefore, the device should no longer be used on gaseous media, after it has
been used to measure liquids.
Manufacturer:
Wöhler Technik GmbH
Wöhler Platz 1
33100 Paderborn
1. 2 Application
The Pressure Computer DC 2000PRO is a high-precision multifunctional meter for
registering differential pressures, flow rates, temperatures, and humidities
(optional). From the basic version on, this device exhibits an extremely wide
dynamic range that not only takes highly sensitive measurements of minimum
draughts and gas pressures in the pascal range, but also lets the user measure
leakage rates and examine sealing properties for main assessments as per DVGW-
TRGI and conduct measurements for load tests during preassessments. A maximum
measuring range of 2 bar and a rupture pressure of 3 bar also provide for
adequate safety at higher pressure ranges. During all measurements the user is
guided by plaintext instructions on the display. This device, which can store
all measured values in a logger, can also be used for measurements of
(ambient) climates. This is made possible by a temperature sensor also
integrated as standard in addition to the pressure sensor. Optionally, the
range of applications can be expanded with an external temperature sensor.
Depending on the selected scan rate, all measured values can be logged for
several years and transferred to a PC via the integrated IrDA interface.
Measurement records can be sent to a thermal printer for printouts with the
company logo. If needed, continuous IrDA transfer can be activated in the
Setup menu so that during measurements all four measured values (pressure,
external and internal temperature, and humidity) and their respective channel
numbers are transferred to a PC every second. The extremely low current
consumption is made possible by an all-new processor technology that
automatically and dynamically adapts the power draw to the measuring task.
Also the mode with the maximum current consumption (6 mA) runs continuously
for over 300 hours on two standard AA batteries (2 Ah). In logger mode the
device can run on the same batteries and without memory overflow for several
years when the scan rate between two measurements is 4 h (4680
measurements x 4 h). The pressure meter can be used on both gaseous media (air
or inert gas) and liquids, e.g. water or fuel oil. In this case, it is
important to observe that remains of liquid at the pressure connections can
falsify the result of the following gas measurement. Therefore, the device
should no longer be used on gaseous media, after it has been used to measure
liquids.
The processor’s arithmetic and logic unit (ALU) can perform simple operations
on the measured values so that, for example, the leakage rate is automatically
displayed in l/h or the flow rate measured with a Prandtl’s tube in m/s.
The German association for safety inspections on gas leak meters, TÜV SÜD
Industrie Service zur Gasleckmengenmessung has certified and approved the DC
2000PRO in accordance with the new DVGW requirements under VP 952 for low-
pressure gas lines complying with DVGW worksheets G 600 and G 624. (DVGW-
Certificate DG-4805BQ0012)
1.3 Measured values
Differential pressure measurements (temperature-compensated piezo bridge)
Measuring range: ±2 bar
1 Pa resolution in measuring range -125.00 hPa to +125.00 hPa, otherwise 10 Pa
Precision: < 3% of measured value, better than ±6 Pa within a range of ±200
Pa.
Internal temperature measurements (NTC)
Measuring range: -20 °C to 60 °C
Precision: < ±2 °C
Resolution: 0.1 °C
External temperature measurements (optional, air temperature sensor order no.
9605 or air temperature probe order no. 9611)
Measuring range: -19.9 °C to +99.9 °C
Precision: < ±2 °C
Resolution: 0.1 °C
T98: < 120 s at 1.5 m/s
Humidity measurements (optional, order no.: 7203)
Measuring range: 0% to 100% RH (relative humidity), non-condensing
Precision: < ±2% RH, within a range of 0–90% RH, otherwise < 3% RH
Resolution: 1% RH
1.4 Calculated values
Pressure units – conversion to mbar, hPa, Pa, mm·H2 0, PSI in accordance with
generally applicable conversion rules
Temperature units – conversion from °C to °F in accordance with generally
applicable conversion rules
Flow rates – Prandtl measurements displayed in m/s, automatic continuous
density correction based on temperature signal
Range: 2–150 m/s
Pipeline volumes automatically from 0.0 to 1000.0 l (tested to 250.0 l)
Leakage rate (0.0 to 300.0 l/h) – l/h in accordance with DVGW-TRG worksheet G
624. The physical data needed to convert diverse gas types can be extracted
from a database stored in the device.
Pressure drop – pre-assessments and main assessments in accordance with DVGW-
TRGI worksheet G 600
Statistical characteristics – minimum, average, maximum of all measured and
calculated values in the respective measurement units
Date and time – output to measurement records.
1.3 Logger mode
Scope – 4680 measurements each with measured pressure and humidity values and
two measured temperature values (when external sensor attached), i.e. max
18,720 measured values
These measured values can be stored in memory for over ten years, even without
batteries.
IrDA data transfer, also while data are being registered
User-selectable scan intervals – 30 s, 1 min, 3 min, 10 min, 30 min, 1 h, 3 h,
4 h Control on undervoltage
1.4 Technical data
Current consumption from two AA or dry batteries:
– operating mode: approx. 6 mA
– OFF and logger modes: approx. 16 µA for clock and processor Interfaces
– infrared data transfer to PC
– in situ printout on thermal printer order no. 9130
Storage temperature – -20 °C to +60 °C
Operating temperature – -5 °C to +60 °C in logger mode (excl. display); incl.
display 0–50 °C
Mass – approx. 450 g with protective bag and magnet excl. hose
Dimensions – 54 x 165 x 52 mm
Controls and connections
Figure 1 below shows the indicators and controls on the DC 2000PRO. The
display always shows the trend ▼▲on the left, the numerical value in the
middle, and the corresponding unit of measurement on the right. When a key is
pressed, a cursor ▲ appears in addition at the edge of the display to mark the
subprogram near the display label. This menu is activated when the cursor
flashes.
The functions assigned to the keys resemble those on a mobile phone. In
general, the ± key on the left increments or decrements an entered value or
displaces to the left or right the cursor ▲ (1) at the edge of the display
(see Figure). A short double tap on the ± key switches from increment to
decrement mode or reverses the cursor’s movements. This toggle function is
indicated by a dot at the centre of the display (2). A second double tap
switches back to increment mode or restores the cursor’s original direction.
Example: Pressing the ± key moves the cursor from its position 1 (“Pre-
assessment”) to “Main assessment” on the right. The cursor is returned to
“Pre-assessment” when the key is first double- then single-tapped (see 2, the
dot “•” indicates active reverse mode). Figure 2.1 – Display and controls on
the DC 2000PRO
The ENTER key at the centre of the control panel confirms the entered value or
activates the program selected at the cursor position.
The right C I/O key has two functions. Pressing it once cancels the current
menu option or an incorrectly entered value. Keeping it pressed switches OFF
the device after about three seconds. Figure 2.2 – Connections and sockets
on the DC 2000PRO Pressure Computer
The socket can take a hose with an internal diameter of 5–6 mm or, if the
design of the DC 2000PRO allows this, a DN 2.7-type rapid-action coupling.
Silicone hoses can perforate at overpressures greater than 1 bar, adding to
the leakage rate, so should not be used for these pressure ranges.
In Figure 2.2, the rear of the DC 2000PRO exhibits diffusion apertures for the
internal registration of ambient humidity (optional) and temperature. The
integrated temperature measurement also serves to compensate for temperatures
in the pressure sensor. When the device is used as a precision temperature
sensor with extended measuring range (–19.9 °C to +99.9 °C) the external
combustion air temperature sensor A 500 (order no. 9605) or the external
combustion air probe A 500 (order no. 9611) with a 2 m cable can be used. The
measured data are read out of the logger and transferred to a PC via the IR
(infrared) interface (order no. 9631 serial or 9318 USB or Bluelink 500.)
Signal menu
Before the device is used, a visual check must ensure each and every time that
all functions work properly. When the device is then switched ON it conducts a
self-test. Afterwards the time and date are output. When the device’s logger
mode has been activated, the text “Log” followed by the current measured and
stored values appears instead of the self-test – the device then switches back
OFF. For as long as a flashing cursor points to the subprogram P=0, no hose
should be attached and no differential pressure applied: during this phase the
device is stabilising and determining its zero point.
The subsections under this heading describe the basic functions in the Signal
menu item. Pressing the ± key three times moves the cursor to the subprogram
Signal which is activated at the ENTER key (cursor flashes under Measuring
Mode; see Figure 3.1). Figure 3.1 – Selecting the Signal menu
3.1 Pressure measurements
The subprogram Signal is activated at the ENTER key. The cursor then flashes
and the text “PRESSURE” appears on the display together with the unit of
measurement “mbar”. Now all of the available units can be selected in turn
with the ± key and activated at the ENTER key. In this manner the user can
select one of the following five units for the pressure measurement: Pa, hPa,
mbar, mm·H20, PSI.
3.1.1 Fast pressure measurements (regulator test)
The unit of measurement hPa features in addition under the name “Fast
pressure” a pressure measurement with instantaneous display. This mode is
particularly suitable for gas regulator tests. The measuring process can be
stopped at the ENTER key, and the device displays the last pressure value
measured. This state is complemented by the trend symbols v . The ± key
initiates a single new measurement. Pressing the ENTER key again returns to
fast measuring mode. This can be ended with the C I/O key. The device
returns to the normal (battery-saving) measuring mode with the unit chosen
last.
Signal AUTO alternately displays pressure, temperature, and air humidity
values in the pressure and temperature units chosen last.
An overpressure at the socket marked with the + sign and an underpressure at
the socket marked with the – sign causes a positive differential pressure to
be displayed. When the differential pressure exceeds 125 hPa, the device
automatically switches to the higher measuring range up to 2000.0 hPa. When
the value falls below 125 hPa in this measuring range, the display switches
back to a resolution of 1 Pa.
3.2 Flow rate measurements based on Prandtl
The flow rate of air in m/s can be measured with a Prandtl’s tube. This
measurement is activated when the ± key is repeatedly pressed in the Measuring
Mode menu until the text “Prandtl” appears with the unit of measurement “m/s”.
The total pressure of the ube is connected to the + overpressure socket and
the static pressure to the – underpressure socket on the DC 2000PRO (see
Figure 3.2).
First of all the device must be “zeroed” in a medium at rest (P=0). Then the
probe is inserted into the gas or air flow, as parallel as possible and with
the tip facing the flow, and the measured values are read off. The current
flow rate v is automatically calculated with Equation (1). According to
Equation (2) the air density ρ in Equation (1) depends in turn on the absolute
air pressure pcur and the current temperature T. whereby:
v | flow rate in m/s |
---|---|
Dp | differential pressure in Pa, measured with the Prandtl’s tube |
ρ | air density in kg/m3 |
pcur | absolute air pressure in hPa, manual entry in the Setup menu item |
(default 1013 hPa)
T| air temperature in °C
The absolute air pressure pcur can be set under the menu item Setup > Ab-
solute pressure. This setting is also used to determine leakage rates under
Section 6. Figure 3.2 – Prandtl’s tube order no. 9487 with combustion air
temperature probe A 500 order no. 9611 for automatic density correction
When the temperature T of the measured air flow deviates from the ambient
temperature of the DC 2000, the combustion air temperature probe can be
inserted over its 2 m cable into the flow parallel to the Prandtl’s tube.
There is then an automatic follow-up of the air density ρ as a function of the
measured temperature T according to Equation (2).
3.3 Temperature measurements
Temperature measurements are activated when the ± key in the Signal menu is
repeatedly pressed until the text “Temperature” appears with the unit of
measurement “°C”. Again pressing the ± key switches to the unit of measurement
“°F”. The ENTER key confirms the selected measuring mode and returns to the
display.
An external temperature sensor can be connected at any time. In this case the
device switches automatically to the external sensor.
For precision measurements, the sensor’s five-digit calibration number (e.g.
cal. no. 10208) should be entered under TLOff in the Setup menu. This
calibration number is printed on a metal foil provided with every temperature
sensor. When no external sensor is
connected, the temperature of an internal sensor is displayed that also serves
to compensate for the temperature of the pressure and the optional humidity
sensor signals. When therefore ambient temperatures and humidities are
constantly measured the housing should not be exposed to direct sunlight or
heat radiation.
3.4 Humidity measurements
Humidity measurements are activated when the ± key in the Singal menu is
repeatedly pressed until the text “Humidity” appears with the unit of
measurement “%”. The ENTER key confirms the selected measuring mode and
returns to the display. The diffusion aperture on the rear of the housing
should not be obstructed. The humidity sensor is a laser-trimmed, capacitive
sensor element with chip-integrated signal processing. It can be retrofitted
or replaced by the user as well (order no. 7203) after the two calibration
values Zero offset (here 0.833 V, Figure 6) and Slope (here 31.31 mV) have
been entered under the Setup menu. Figure 3.3 – Position of the humidity
sensor when the battery compartment cover is removed.
The slope value taken from the provided calibration record must be rounded off
to two decimal places (here 31.311 mV —> 31.31 mV). Figure 3.4 – Calibration
record for a humidity sensor
3.5 Automatically alternating display
The program item Measuring Mode AUTO causes the display to alternate between
pressure, temperature, and air humidity values in the pressure and temperature
units chosen last.
Stress test (DVGW-TRGI)
Lines with operating pressures up to 100 hPa can be pre-assessed in accordance
with DVGW-TRGI worksheet G 600 and the findings documented very easily with
the DC 2000PRO. First of all, the line must be sealed and a suitable test plug
inserted. The DC 2000PRO must be switched ON before it is connected to the
test plug. After zeroing, press the ± key to activate the menu item stress
test. The DC 2000PRO prompts you to pump to the test pressure (e.g. 1 bar or
1,000 hPa). When the over- and underpressure lines have been swapped over by
accident, the text “Swap” appears on the display. When the pressure reaches
the preset test pressure the stabilising phase is initiated (default Stab.
time = ten minutes). This is marked with the trend symbols ▼▲ . When the
pressure remains within ±10% of the test pressure during this stabilisation
phase the actual pressure loss test is started at the end of the stabilisation
time (default Time = ten minutes). During the stabilisation phase the pressure
loss test can also be initiated manually at the ENTER key. The remaining test
time and the current pressure values are displayed alternately. When the test
time has ended or been cancelled with the C I/O key the cursor flashes at Pre-
assessment on the display and the results can be viewed in turn at each press
of the ± key as follows:
Difference: 54.4 hPa
± key
Start pressure: 1000.3 hPa
± key
Time: 10.00
± key
Stop pressure: 945.9 hPa
± key
Text: “Print …” (activate the printout with the ENTER key)
± key
Difference: 54.4 hPa etc.
These results can also be printed out or transferred to a PC at a later time
under the menu item Print or Log/IR respectively. By default, the print-out
includes a grafic of the pressure progression. You can disable this function
under the item „grafic“.
The record is not deleted until the logger is started or another TRGI test is
conducted.
Main test menu (DVGW-TRGI)
Lines with operating pressures up to 100 hPa can also be main-assessed in
accordance with DVGW-TRGI worksheet G 600 and the findings also documented
very easily with the DC 2000PRO.
First of all, the line must be sealed and a suitable test plug inserted. The
DC 2000PRO must be switched ON before it is connected to the test plug. After
zeroing, press the ± key to activate the menu item Main assessment. The DC
2000PRO prompts you to pump to the test pressure (e.g. 150 hPa). When the
over- and underpressure lines have been swapped over by accident, the text
“Swap” appears on the display. When the pressure reaches the preset test
pressure the stabilising phase is initiated (default Stab. time = ten inutes).
This is marked with the trend symbols ▼▲. When the pressure remains within
±10% of the test pressure during this stabilisation phase the actual pressure
loss test is started at the end of the stabilisation time (default Time = ten
minutes). During the stabilisation phase the pressure loss test can also be
initiated manually at the ENTER key. The remaining test time and the current
pressure values are displayed alternately. When the test time has ended or
been cancelled with the C I/O key the cursor flashes at Main test on the
display and the results can be viewed in turn at each press of the ± key as
follows:
Difference: 17.7 hPa
± key
Start pressure: 110.83 hPa
± key
Time: 10.00 (minutes)
± key
Stop pressure: 93.14 hPa
± key
Text: “Print …” (activate the printout with the ENTER key)
± key
Difference: 17.7 hPa etc.
These results are stored in the DC 2000PRO and can also be printed out or
transferred to a PC at a later time under the menu item Print or Log/IR
respectively. The record is not deleted until the logger is started or another
TRGI test is conducted.
Leakage rates and utility life measurements
6. Leakage rates and utility life measurements as per DVGW worksheet G
624
Leakage rates can be determined in accordance with DVGW worksheet G 624 and
the findings documented very easily with the DC 2000PRO.
First of all, the line must be sealed and a suitable test plug inserted (see
Section 11). The DC 2000PRO must be switched ON and zeroed before it is
connected to the test plug. After the automatic zeroing, press the ± key to
activate the menu item Utility life / Leakage rate. The display now shows the
text “Pipe volume”. This pipe volume can now be determined either
automatically or from a graph.
6.1 Automatically determining pipeline volumes
When the ENTER key is pressed at the same time as the text “Pipe volu- me”
appears, the pipe volume can be determined automatically with the DC 2000PRO.
For a pipe volume up to 100 l a syringe takes a sample volume of 100 ml
(Figure 6.2 on the left). Lines with volumes greater than 100 l can be tested
with a manual test pump (163 ml/stroke, see figure 6.2 on the right). Each
test volume must be entered in the DC 2000PRO with the ± key and confirmed
with the ENTER key. After a repeated zeroing the DC 2000PRO then shows the
text “Pump”. Now the actual sample can be taken with the medical syringe or
the manual test pump up to the set volume. The DC 2000PRO then displays
automatically the pipe volume calculated with Equation (3). As soon as the
display no longer changes this value is confirmed with the ENTER key after
which it is used as V pipe for further determinations of leakage rates in
accordance with G 624.
The measuring principle
When a known sample volume Vsample is taken from a pipeline system (e.g. with
a medical syringe or a manual test pump), the total volume Vpipe can be
determined from the resulting pressure change. For small pipe volumes a
medical syringe proves adequate; for larger pipe vo- lumes a manual test pump
can be used. The choice should be such that the pressure change is at least 5
hPa. Boyle’s law returns the following equation for the wanted volume Vpipe:
whereby:
Vpipe | wanted pipe volume, max 1000.0 l |
---|---|
Vsample | sample volume, default 0.100 l, e.g. with 100 ml medical syringe |
Δp | max pressure difference in Pa caused by the taking of the sample |
pabs | absolute air pressure, entered manually under the Setup menu |
Figure 6.2 – Connections for determining the pipe volume V pipe, with a 100 ml
medical syringe (left) or a manual test pump (right)
6.1.1 Determination of the volume of the nozzle
The pressure difference Δp will be measured. It should amount to at least 200
Pa, to get an exact measuring result. Therefore, the volume of the nozzle
VSample must at least amount to 1/500 of the pipe volume. In this case, the
expected error of the result of the volume measurement is equal to the
inaccuracy of the DC 2000PRO which is 3% of the measured value. A higher
pressure will provoque longer stabilization periods during the temperature
compensation and possible leakages will have more influences.
The following table will give you some guiding values for the determination of
Volume of the nozzle | max. pipe volume (DC 2000PRO) |
---|---|
20 ml | 10 l |
50 ml | 25 l |
100 ml | 50 l |
163 ml (1 stroke with the test pump) | 80 l |
489 ml (3 strokes with the test pump) | 240 l |
The entire printout for this determination of leakage rate specifies and also
documents permanently all interim results for determining the volume. One
advantage of this method lies in the insensitiveness to any pipe leaks: the
small sample volume leads to pressure differences only of a few hPa in the
line. Compared with a usual test pressure of 150 hPa, these low test pressures
can scarcely give rise to a perceptible leak.
6.1.2 Determining pipelines from graphs
When the C I/O key is pressed at the same time that the text “Pipe volume”
appears, the pipe volume can then be determined from the pipe lengths and pipe
diameters in figure 6.3 and entered at the ± key. When this entered value is
confirmed with the ENTER key, it is used as V pipe for determining other
leakage rates in accordance with G 624. A blow-up of this diagram can be found
on the front page of the loose leaf inserted at the centre of these operating
instructions.
Figure 6.3 – Aid for determining pipe capacity
For example, a 1/2“ pipe designed with “medium-duty threads in accordance with
DIN 2440” of 10 m length has a pipe volume of 2 l.
6.3.3 Determining the leakage rate
Once the pipeline volume has been measured or entered the DC 2000PRO prompts
the user to enter the test pressure, the test time, and the operating pressure
of the gas line.
Afterwards the line is to be pumped to the test pressure (e.g. 100 hPa). To do
so, replace now the syringe or the manual test pump with the compressed air
pump for applying the test pressure (see Section 11). When the over- and
underpressure lines have been swapped over by accident, the text “Swap”
appears on the display. When the pressure reaches the preset test pressure the
stabilising phase is initiated (default Stab. time = ten minutes). This is
marked with the trend symbols ▼▲. When the pressure remains within ±10% of the
test pressure during this stabilisation phase the actual pressure loss test is
started at the end of the stabilisation time (default Time = ten minutes).
During the stabilisation phase, e.g. when the pressure drops greater than 5
hPa, the pressure loss test can also be initiated manually at the ENTER key.
The remaining test time and the current pressure values are displayed
alternately.
When the test time has ended or been cancelled with the C I/O key, e.g. when
the pressure drop is greater than 15 hPa, the cursor flashes at Utility life /
Leakage rate on the display and the results can be viewed in turn at each
press of the ± key as follows:
Difference: 9.9 hPa
± key
Start pressure: 108.83 hPa
± key
Time: 1 min
± key
Stop pressure: 98.96 hPa
± key
Leakage rate: 8.3 l/h hPa
± key
Text: “Print …” (activate the printout with the ENTER key)
± key
Difference: 9.9 hPa etc.
The leakage rate is calculated automatically with the following equations (4)
and (5) and so complies with the procedure under DVGW-TRGI worksheet G 624:
VB | gas leakage rate in operating mode in l/h |
---|---|
VL | air leakage rate at test pressure |
p Omax | max operating pressure of the gas in situ |
p start | test pressure at start of measurement |
p end | test pressure at end of measurement |
p abs | absolute air pressure, entered manually under Abs. pressure in the menu |
Setup (default 1013 hPa)
f| (absolute air viscosity)/(absolute gas viscosity), selection stored in
Setup under Gas
T meas| measuring time converted to hours (default 1 min)
V pipe| cubic capacity of the test section in litres (see Figure 7
Min, Max, AVG menu
The three cursor positions on the left side of the bottom display edge return the statistical characteristics of all measured and calculated values. The detected minima and maxima and the calculated averages (AVG) can be reset with the P=0 function. The effect of smoothing on calculated averages can be set with the value ALPHA defined in Equation (6). The smaller the value for ALPHA, the greater the effect of smoothing. ALPHA can be varied from 0.01 to 0.99 under the menu item Setup —> ALPHA.
AVGnew | ALPHA · current measured value + (1 – ALPHA) · |
---|---|
AVGold | |
AVGnew | mean value at present moment |
AVGold | mean value one second earlier |
ALPHA | weighting factor for current measured value (0.01–0.99) |
---|
Setup menu for basic settings and company logo
In the Setup menu a number of settings can be configured. All configurations
will be kept after switching off the device or changing the battery.
8.1 Basic configurations
When the Setup menu is open, the ± key can be used to configure a number of
settings that are explained in the following. A good overview is also found in
the quick reference in the middle of this manual.
-
Setup — > Rounding
Activating this function rounds off the displayed value to a resolution of five digits. For a display in pascals, for example, this means a resolution reduced from 1 to 5 Pa. The result is a considerably more stabilised display under pressure fluctuations, yet without the delay effects usual for averaging. The device continues calculating values based on the higher resolution (ON/OFF, default OFF). -
Setup — > Medium
This menu selects the gas type and therefore the relative viscosity in Equation (4) (default natural gas).Gas type| f
---|---
natural gas| 1.7
air| 1.0
town gas| 1.3
propane| 2.3
butane| 2.4
hydrogen| 2.0
Table 1 – Relative viscosity of various gases in accordance with Equation (4) based on DVGW-TRGI
-
Setup — > Air pressure
Here you enter the current air pressure pcur in situ (QFE) in hPa for calculating the flow rate with Equation (2) and for determining the leakage rate with Equation (5). This air pressure can range from 800 to 1200 hPa (default 1013 hPa). -
Setup — > Clock
Here you enter the time and date. When the batteries are removed for no longer than a minute, the clock does not have to be corrected. When the batteries are removed for longer, the clock needs only to be put forward by this time. -
Setup — > ALPHA
This sets the weighting factor for the averaging function AVG in the form of Equation (6). ALPHA can accept values between 0.01 and 0.99. The lower the value for ALPHA, the less perceptible are the current fluctuations in the signal (see also Section 7; default 0.90). -
Setup — > Preassessment —> Pressure
Here you can set the test pressure for a pre-assessment (500–1500 hPa, default 1 bar). -
Setup — > Preassessment —> Time
Here you can set the time for a pre-assessment (1–300 min, default 10 min). -
Setup — > Main assessment —> Pressure
Here you can set the test pressure for a main assessment (1–500 hPa, default 150 hPa). -
Setup — > Main assessment —> Time
Here you can set the time for a main assessment (1–300 min, default 10 min). -
Setup — > Stab. time
Here you can set the time for thermal stabilisation before the start of preassessment, main assessment, and the determination of leakage rates. When during this time the current pressure value remains within ±10% of the setpoint pressure, the measurement of the pressure loss starts automatically. The position within this tolerance range and therefore the start of the stabilisation period is marked on the display with the trend symbols. The measurement of the pressure loss can also be initiated manually at any time at the push of a button (1–300 min, default 10 min). -
Setup — > TLOff
Here you can enter the calibration number given on the external temperature sensor’s label (10000–10300, default 10179). -
Setup — > Humidity —> Zero offset
Here you can enter the calibration value Zero offset for the humidity sensor. An example zero offset of 0.833 V is given in Figure 6 under Section 3.4 (0.5–1.0 V, default 0.780 V). -
Setup — > Humidity —> Slope
Here you can enter the calibration value Slope for the humidity sensor. An example slope of 31.31 mV is shown in Figure 6 under Section 3.4. The value taken from the calibration sheet must be rounded to two decimal places (25.00–60.00 mV, default 30.00 V). -
Setup — > Log rate
Here you can set the time between two registrations by the logger. The default is thirty seconds; in other words, the pressure, two temperatures, and the humidity are each measured and the values written to the integrated data memory every thirty seconds (see Section 9.2; default 30 s). -
Setup — > AUTO OFF
This activates and deactivates the AUTO OFF function. This serves to switch OFF the device automatically when no key has been pressed for longer than thirty minutes (default ON). -
Setup — > IrDA
This activates the continuous IrDA data transfer. In this case all four measured values (pressure, 2 x temperature, relative humidity) and their corresponding channel numbers are transferred to a PC every second during the normal measuring process (default OFF). -
Setup — > Fastprinter
If you enter „OFF“, you can print the dates with the Wöhler TD 23 termoprinter. If you enter „On“, you can print faster with the Wöhler TD 600 termofastprinter (default WÖHLER ON). -
Setup- >Graph ON/OFF
Under this item you can select if a graph of the pressure porcess shall be printed out after the pre-assessment menu, main assesment menu and after the leakage rates and utility life measurements. Default is „on“. -
Setup —> Logo Here you can enter the logo text for printouts. This is explained under the following Section 8.2 (default WÖHLER MGKG, DC 2000).
-
Setup —> Default This function restores the device’s setup status on delivery. The logo text is overwritten by the original Wöhler logo. All sensor calibration values remain unchanged. Accidental initiation is prevented by the prompt Sure?.
8.1 Entering a logo The following Tables 2 and 3 are to assist you in entering
a logo into the DC 2000. First enter (preferably with a lead pencil) the text
you want in the upper Table 3. The first two rows take twelve characters that
are printed in bold. The following rows 4 to 6 can take max twenty-four
characters that are printed normally. Finally determine for each row and
column the ASCII codes and enter these in the code fields. These values can
then be entered and stored row by row under Setup —> Logo.
The LOGO converter is also provided in the Excel software that can be
downloaded as freeware from the internet address
www.woehler.de/mgkg. There the text is converted
automatically into ASCII codes.
Table 2 – Conversion example (original logo) Table 3: Tamplate for own
conversions
Logger / data transfer
The menu item Log/IR initiates subprograms that control the long-term data
registration (logging) and its infrared output or transfer.
9.1 Data transfer to a PC
The menu item Log/IR —> IrDA is used to transfer stored measured values and
records to a PC.
The contents stored in the DC 2000PRO are transferred to a PC via the infrared
interface (order no. 9631 serial or 9318 USB or Bluelink 500).
Table 4 – Example receive sequence: channel no. + measured value
303025/01/0213:38:19 | Start: channel no. 3030 + date and time |
---|---|
3031 107.35 | Channel no. 3031 + measured pressure in mbar |
3038 22.0 | Channel no. 3038 + external temperature in °C |
3041 22.0 | Channel no. 3041 + internal temperature in °C |
3040 40 | Channel no. 3040 + relative humidity in % |
303025/01/0213:48:19 | Stop: channel no. 3030 + date and time |
Table 4 shows an excerpt from a transfer sequence read into the Microsoft
HyperTerminal program via the IrDA receiver connected to the serial port COM1
(9600, 8, 1, 0, Xon/Xoff). The data from the last measurement record (leakage
rates, etc.) are transferred after the measured values. This text file in
Table 4 can then be imported into an Excel file where it can be viewed. The
Excel file DC2000.exe can be downloaded as freeware from the internet
(http://mgkg.woehler.de). With this programm the logger data can be imported
directly in the excel-list and it can also depict the pressure loss
measurements in graph form.
9.2 Logging
This menu item starts the logger that can store up to 4680 measurements each
with measured pressure, temperature and (optional) humidity values and two
measured temperature values, i.e. max 18,720 measured values. When the battery
voltage falls below a threshold of 2 V (battery symbol appears on the
display), the logger switches OFF automatically. Before starting the logger
for a longer period you should therefore check the battery status under
Measuring Mode —> V battery. Even without batteries, measured values are
retained in memory for over ten years. When the fastest scan rate of thirty
seconds is chosen, values can then be registered for max one day and fifteen
hours (see Table 5). The following scan intervals are possible and can be
selected in the Setup menu under Setup —> Log rate: 30 s, 1 min, 3 min, 10
min, 30 min, 1 h, 3 h, 4 h.
Table 5: Logger rate and max registration time for 4680 measurements
Log rate | Max possible measuring time |
---|---|
30s | 39 h |
1 min | 3.25 days |
3 min | 9.75 days |
10 min | 32,5 days |
30 min | 13 weeks |
1 h | 27 weeks |
3 h | 19 months |
4 h | approx. 2 years (with good dry bat- teries) |
When the C I/O key is pressed in logger mode the DC 2000PRO only out- puts the
text “Log” followed by the remaining memory and then the current measured
values. The cursor is locked for other commands. Logger mode is ended when the
device is switched OFF at the C I/O button (pressed for three seconds). Data
are transferred to a PC as explained under Section 9.1.
9.3 Data transfer to a pocket PC
The contents stored in the DC 2000PRO can be transferred with the appropriate
software to a pocket PC. This software can be downloaded from the internet
(http:\\mgkg.woehler.de).
9.4 Printing out measurement records
Log/IR —> PRINT – Measurement records can be printed out directly on the TD
600 thermal printer (order no. 4130). When the device is switched OFF the data
are retained until they are overwritten by a new pressure loss measurement or
logger recordings. The text Print appears on the display during the printing
process.
Changing the batteries
When only the text “Self-test …” or the message “Change batteries” appears on the display the batteries are fully depleted. To replace the batteries remove the rear of the device with a slot-head screwdriver. Now you can replace the batteries (IMPORTANT! Note the polarity!) and screw the rear back on. When the batteries are replaced within one minute the clock does not need to be adjusted. Measured data, settings, and calibration values are retained even without batteries for at least ten years.
Accessories
-battery (1.5 V AA) | order no. 2999 |
---|---|
-combustion air sensor, plug typ | order no. 9605 |
-combustion air temperature sensor, 280 mm | order no. 9611 |
-combustion air temperature sensor, 100mm | order no. 9651 |
-magnet for securing sensors | order no. 6142 |
-holding strap with spring hook | order no. 9805 |
-USB IR interface for PC | order no. 9318 |
-thermal quickprinter TD 600 | order no. 4130 |
-thermal paper, 10 rolls | order no. 9145 |
-measuring hose, single | order no. 2338 |
-measuring hose with brass coupling DN 10 | order no. 7209 |
-retrofit for humidity measurements | order no. 7203 |
-disposable medical syringe 100 ml | order no. 53196 |
-manual test pump | order no. 2412 |
-syringe connecting hose | order no. 30545 |
Information on disposal
You may hand in any defective batteries taken out of the unit to our company
as well as to recycling places of public disposal systems or to selling points
of new batteries or storage batteries.
In the European Union, electronic equipment does not belong into domestic
waste but – in accordance with Directive 2002/96/EC of the European Parliament
and of the Council of 27 January 2003 on Waste Electrical and Electronic
Equipment – must be disposed of in an expert manner. If you do no longer need
this unit, please dispose of it in accordance with the applicable statutory
provisions.
Declaration of conformity
The manufacturer: | WÖHLER Messgeräte Kehrgeräte GmbH |
---|---|
declares that the product: | Schützenstr. 38, 33181 Bad Wünnenberg |
product name: | Pressure Computer |
model number: | Wöhler DC 2000, version 05 and later |
device class: | D, in accordance with VP 952 under DVGW |
has been certified and approved by TÜV SÜD Industrie Service zur
Gasleckmengenmessung in accordance with the DVGW requirements under VP 952 for
low-pressure gas lines complying with DVGW worksheets G 600 and G 624 (DVGW
type approval certificate DG-4805BQ0012).
This type approval also included tests by TÜV SÜD Industrie Service zur
Gasleckmengenmessung on the device’s compliance with the requirements for
conformity and electromagnetic compatibility:
This declaration is submitted for the above manufacturer by:
Dr Stephan Ester, Managing Director
Bad Wünnenberg, 19 October 2004 DVGW type examination certificate
DG-4805BQ001 2
Registration Number
Field of Application | products of gas supply |
---|---|
Owner of Certificate | WOHLER Technik GmbH |
WOhler-Platz 1, D-33181 Bad WOnnenberg
Distributor| WOHLER Technik GmbH
WOhler-Platz 1, D-33181 Bad WOnnenberg
Product Category| mobile metering appliances: Leakage gauge for low pressure
gas pipes (4805)
Product Description| battery-powered leakage gauge for low pressure gas pipes
with digital display
Model| DC 2000 PRO
Test Reports| type testing: M-G 1204-00/17 from 25.01.2017 (TSG) laboratory
control test: U 2739-00/21 from 03.03.2021 (TSG)
Test Basis| DVGW G 5952 (01.04.2014)
Date of Expiry / File No.
01/21/2025 / 22-0140-GNV
Expiry date/file number
DVGW CERT GmbH
Certification Authority
Josef-Wirmer-Str. 1-3
53123 Bonn
Tel. +49 228 91 88 – 888
Fax +49 228 91 88 – 993
www.dvgw-cert.com
info@dvgw-cert.com
Warranty and service
14.1 Warranty
Every WÖHLER Pressure Computer DC 2000PRO has been subjected to full
functional tests and does not leave our production plant until after extensive
quality control. The details of the final inspection are set down in an
inspection report and stored on our premises. When the device is used for the
intended purpose, the warranty period for the device is twelve months
following the date of sale. This warranty does not extend to wearing parts
(e.g. batteries) or damage to the pressure sensor caused by excess loading.
This warranty does not cover the costs of transport and packaging when the
device is sent to us for repairs.
This warranty becomes void when the device is repaired or otherwise modified
by unauthorised parties.
We attach great importance to our services not only during the warranty
period. We are naturally always there to help you.
- Immediate service work performed directly on our premises when you bring the device to us in Bad Wünnenberg.
- You send us the device, and you will get it back repaired within an average of only five days, delivered by our parcel service.
- You can also obtain immediate help by telephone from our technical personnel.
14.2 Calibration
Although the device has no wearing parts, we recommend an annual inspec- tion
by the manufacturer or an authorised service agency.
Points of sale and service
Germany
Wöhler Technik GmbH
Wöhler-Platz 1
33181 Bad Wünnenberg
Tel.: +49 2953 73-100
info@woehler.de
www.woehler.de| Austria
Wöhler GmbH
Burgerfeld 16
3150 Wilhelmsburg
Tel.: +43 2746 313 13 10
www.woehler.at
---|---
USA
Wohler USA Inc.
208 S Main Street
Middleton, MA 01949
Tel.: +1 978 750 9876
www.wohlerusa.com| France
Wöhler France SARL
17 A impasse de Grousset
31590 Lavalette
Tel.: +33 5 61 52 40 39
info@woehler.fr
www.woehler.fr
Your contact:
www.woehler-international.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>