W HLER TI 410 Dew-Point Indicator Instruction Manual
- June 10, 2024
- W HLER
Table of Contents
W HLER TI 410 Dew-Point Indicator Instruction Manual
Preface
Operation Manual Information
This manual enables the user to handle the Wöhler TI 410 safely. Please keep
this manual for your information. In principle, the Wöhler TI 410 Dew-point
Indicator should only be used by skilled personnel for its intended use and
within the specified range of data. Liability is void for any damages caused
by not following this manual.
Notes
Warning! Not following this warning can cause injury or death.
Attention! Not following this note can cause damage to the device.
Note! Tips and other useful information
Intended application
The Wöhler TI 410 Dew-point indicator may only be used for controlling the
build-up of exhaust gas in gas fire places that are dependent on air
condititoning. The device may only be used in interior rooms. Every other
application is not intended and therefore forbidden.
Security advices
WARNING!
Never use the Wöhler TI 410 Dew-point Indicator in any potentially explosive
atmosphere. Never use the unit in any environment where there can be an
explosive gas atmosphere. In such an environment, even batteries should not be
taken out of the unit and/or exchanged.
Transport
Attention!
The device may be damaged by improper transport!
Information on disposal
Electronic equipment does not belong into domestic waste, but must be disposed
in accordance with the applicable statutory provisions. You may hand in any
defective batteries taken out of the unit to our company as well as to
recycling places of public disposal systems or to selling points of new
batteries or storage batteries.
Functional principle
The dew-point indicator shows exhaust gas build-up. Exhaust gas that has been built up is characterised by an increased humidity and temperature relative to the ambient air. A possibly likewise presence of CO content has a strong toxic effect, which is why exhaust gas build-up can be highly dangerous. To provide a sure detection, the measurement is effected via two independent sensors:
- measurement of the surface resistance of the sensor plate
- measurement of the temperature When a pre-determined resistance of the sensor plate is undercut, an optical and an acoustic signal is emitted. The increased temperature of the outflowing exhaust gas is recognized by the fastreacting temperature sensor, which is located in a protective position in a hole at the end of the sensor plate. For to adapt to the different ambient conditions, the operator can adjust the sensitivity in four different steps. To attain a high battery life, the instrument switches off automatically after four minutes, when no key has been pressed.
Technical Data
Description | Specification |
---|---|
Temperature measuring range | -40 to 100°C |
Resolution | 0.1 °C |
Accuracy | ± 1 °C |
Indicators | 1. acoustic through pulsed alarm tone2. optical through |
four-step bar3. increased tem- perature indication relative to the sur-
roundings
Setting of the sensitivi- ty| in four steps
Automatic switch-of| After ten minutes
Power supply:| 9 V block battery (indi- cation if battery volt- age is too
weak)
Operational life| About 100 hours
Sensor plate| Double-sided
Control elements
Legend
- Display
- ON/Off (on the instrument base)
- SENS
- HOLD
- Protective cover for sensor plate
Display options
Fig. 2: Temperature indication sensitivity step 1
Fig. 3: Temperature indication sensitivity step 2
Fig. 4: Bar indication
Fig. 5: Bar indication Hold function
Fig. 6: Temperature unite, Fahrenheit
Fig. 7: Initial picture with battery symbol, bar indication, both temperature units and hold function
Before starting the measurement
Before starting the measurement, test the operational capability of your device:
- Carefully clean the sensor plate with a dry cloth or aceton
- Breathe on the sensor plate
NOTE!
Breathe through your nose (Filter) The device should now respond to
sensitivity step 4 (see chapter 7.2)
NOTE!
Before switching on the device, the temperature of the sensor plate should not
be higher than the room temperature.
Handling
Use of the measuring instrument
The TI 410 is switched on and off by pressing the On/Off button. After
switching on the device, all display possibilities appear in the display.
During this short stabilization phase, the device will redetermine the zero
point for the measurement of the dew point.
NOTE!
Please take care that the sensor is located in dry air during the
stabilization phase.
After that, the measured ambient temperature appears in the display. If there is a rise in humidity, when e.g. the flexarm is passed into an exhaust-gas pipe and the sensor plate is surrounded by exhaust gas flow from both sides, then a oneto-four-step bar is seen in the display, depending on the degree of humidity. As soon as the second bar is shown, a controlled alarm tone additionally sounds. Moreover, the increased temperature of the exhaust gas is shown.
Adjusting the sensitivity
Activating the button ”SENS” a number of times lets you adjust the sensitivity
of the Wöhler TI 410 to an increased humidity in four steps. The respective
current step is shown in the display. Step 4 is the most sensitive step.
Freezing the display indication
By pressing the button ”HOLD”, you have the possibility during the measuring
to freeze the indication, when for example inaccessible positions are to be
measured and subsequently readings are to be comfortably made.
Changeover of the temperature indication
Simultaneous pressing of the ”HOLD” and ”On/Off” buttons; when the meter is
switched on, lets you change between the degree celsius and degree fahrenheit
indications
Trouble Shooting
Indicator | Possible reason | Trouble shooting |
---|---|---|
Battery symbol appears in the display. | The necessary battery volt- age is | |
undercut. | Change Battery. | |
LOW | T < – 40°C | Temperature is out of range. |
HIGH | T > 100°C | Temperature is out of range. |
ERR | Measuring error during the stabilization phase | Switch off the |
device.Wait until the sensor is dry. Switch on the device.
Note of care
The sensitivity of the TI 410 depends very strongly on the condition of the sensor plate. If the sensor plate is dirty, the sensitivity is lowered strongly. Use a cloth dampened with aceton for cleaning. The cleaning must be carefully done so that the temperature sensor is not damaged. Hand contact with the sensor plate should be minimized as much as possible.
Calibration
If the device does not work correctly during the operability test described in chapter 6, a recalibration will be necessary. Before starting the calibration, switch on the device as described in chapter 7.1. Press the buttons “SENS” an “ON/OFF” at the same time, to enter the calibration mode. The icon “C” will appear at the right bottom of the display.
NOTE!
If the calibration mode was activated by mistake, it can be deactivated by
pressing the “SENS” button without changing the calibration.
During the calibration, the TI 410 shows constantly the raw data of the condensation sensor. Hold the sensor above a water quench with a temperature that is approximately 10 °C higher than the room temperature. The measuring signal should increase considerably and stabilize after some seconds. As soon as the signal has stabilized, the measurement must be stopped by pressing “HOLD“. After pressing the buttons “SENS” and “ON/OFF” at the same time, the actual measured value will be assumed as the reference value. This will be shown in the Display by “SAVE”.
Guarantee and Service
Guarantie
In the case of proper usage, the guarantee period for the Wöhler TI 410 Dew-
point indicator amounts to 12 months from the date of sale. Excluded here are
wear-and-tear parts (e.g. batteries). The cost for the transportation and the
packing of the device in the case of repair are not covered by this guarantee.
Service
We attach great importance to SERVICE. Therefore, of course, we are readily
available to assist you after the guarantee period ends.
- Send us the device, and we will repair it and return it to you with our package service.
- Immediate help is provided by our technical staff over the telephone
Declaration of conformity
The product: Wöhler TI 410 Dew-Point Indicator conforms with the essential
protection requirements which are set out in the directives of the European
Council for to adapt the legal provisions of the Member States in respect of
electromagnetic compatibility (2014/30/EU). The following standards were
availed of for to evaluate the product in respect of electromagnetic
compatibility:
EN 61326 : 1997 + A1 : 1998 + A2 : 2001 + A3 : 2003
EN 61000-4-2: 1995+A1:1998+A2 : 2001
EN 61000-4-3 : 2006
Adress Wöhler Technik GmbH
Schützenstr. 41
D-33181 Bad Wünnenberg
Tel.: +49 2953 73-100
Fax: +49 2953 73-250
E-Mail: info@woehler.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>