LUTRON CL-2006 Free and Total CHLORINE METER User Manual
- June 14, 2024
- Lutron
Table of Contents
LUTRON CL-2006 Free and Total CHLORINE METER
CHLORINE METER
Model : CL-2006
Your purchase of this Free and Total cholesterol meter marks a step forward for you into the field of precision measurement. Although this CHLORINE METER is a complex and delicate instrument, its durable structure developed. Please read the following instructions carefully and always keep this manual within easy reach.
FEATURES
- The meter measures the Free and Total chlorine ( CL ) in the 0.00 to 3.50 ppm ( mg/L ).
- The measuring method is an adaptation of the USEPA Method 330.5 for wastewater and Standard Method 4500-Cl G for drinking water.
- The advanced optical system is based on a special narrow-band LED lamp that allows the most accurate and repeatable reading.
- Friendly and powerful calibration functions are able to validate the good performance of your meter at any time.
- 1.00 ppm Free standard solution and 1.00 ppm Total standard solution are included as the standard accessories.
- The unique optics structure enables the instrument to read with high resolution: 0.01 ppm ( mg/L ).
- Splash waterproof on the front panel.
- Jumbo LCD, easy readout.
- Microprocessor circuit assures maximum possible accuracy and provides special functions and features.
- Battery operated for field and on-site testing convenience.
- Data hold function for freezing the desired value on display.
- Records Maximum and Minimum reading with Recall.
- Heavy duty & compact housing with hard carrying case, designed for easy carryout & operation.
- Auto shut-off is available to save battery life.
- Application: Test swimming pool, municipal water, food and beverage water, or other aqueous solution where fluid clarity is important.
SPECIFICATIONS
Circuit | Custom one-chip of microprocessor LSI circuit. |
---|---|
Display | LCD size : 41 mm x 34 mm |
Range | Free chlorine ( CL ) : |
0.00 to 3.50 ppm ( mg/L ).
Total chlorine ( CL ) :
0.00 to 3.50 ppm ( mg/L ).
Resolution| 0.01 ppm ( mg/L ).
Accuracy| ± 0.02 ppm ( mg/L ).
@ 1.00 ppm ( mg/L )
Light source| LED, 525 nm.
Light detector| Photodiode
Method| The measuring method is an adaptation of the USEPA Method 330.5 and
Standard Method 4500-Cl G.
* The reaction between free ( total )
chlorine and the DPD reagent cause a pink tint in the sample.
Response time| Less than 10 seconds.
Sample volume| 10 mL.
Data Hold| Freeze the display reading.
Memory Recall| Maximum & Minimum value.
Display Sampling Time| Approx. 1 second.
Power off| Auto shut-off saves battery life or manual off by pushing the
button.
Calibration points| Zero chlorine.
1.00 ppm ( Free chlorine ).
1.00 ppm ( Total chlorine ).
Operating Temperature | 0 to 50 ℃. |
---|---|
Operating Humidity | Less than 85% R.H. |
Power Supply | DC 1.5 V battery ( UM4, AAA ) x 6 PCs, or equivalent. |
Power Current | Stand by |
Testing | Approx. DC 12 mA. |
Weight | 320 g/0.70 LB. @ Battery is included. |
Dimension | 155 x 76 x 62 mm |
( 6.1 x 3.0 x 2.4 inch)
Accessories Included| * Instruction manual…………………….. 1 PC
* 1.0 ppm Free Chlorine standard
solution, CF-01…………………………. 1 PC
* 1.0 ppm Total Chlorine standard solution, CT-01…………………………. 1 PC
* Zero Chlorine standard solution,
CL-01 1 PC
* Empty testing bottle…………………… 2 PCs
* Clean cloth………………………………. 1 PC
* Free Chlorine DPD powder…………… 10 PCs
* Total Chlorine DPD powder………….. 10 PCs
Optional Accessories| * Free Chlorine DPD powder ( 10 PCs ), Model: CFP-10
* Total Chlorine DPD powder ( 10 PCs ), Model: CTP-10
* Empty testing bottle, Model: 0601
* 1.0 ppm Free Chlorine standard solution, Model: CF-01
* 1/0 ppm Total Chlorine standard solution, Model: CT-01
* Zero Chlorine standard solution,
Model: CL-01
FRONT PANEL DESCRIPTION
- 3-1 Cover of Testing bottle
- 3-2 Container of Testing bottle
- 3-3 Display
- 3-4 Hold Button ( Esc Button )
- 3-5 TEST/CAL Button
- 3-6 Power Button
- 3-7 ZERO Button
- 3-8 REC Button ( MAX, MIN Button )
- 3-9 Battery Compartment/Cover
- 3-10 Free Chlorine 1.0 ppm standard solution.
- 3-11 Total Chlorine 1.0 ppm standard solution
- 3-12 Zero Chlorine standard solution
- 3-13 Empty testing bottle 1
- 3-14 Empty testing bottle 2
- 3-15 Clean Cloth
- 3-16 Free Chlorine DPD powder ( 10 PCs )
- 3-17 Total Chlorine DPD powder ( 10 PCs )
MEASURING PROCEDURE
Measurement Consideration
- There is a ” White mark ” on the edge of the ” Container ” ( 3-2, Fig. 1 ) and also on the up position of ” Testing Bottle ” ( 3-10, 3-11, 3-12 3- 13, 3-14 Fig. 1 ), refer Fig. 2.
- When making the measurement ( or calibration ), it should keep the ” Container white mark ” face to face together with the ” Testing bottle white mark “
- insert the ” Testing bottle ” to the bottom of ” Container ” ( 3-2, Fig. 1 ) completely.
- Before the measurement, it should be enveloped in the ” Cover ” ( 3-1, Fig. 1 ) completely. Remark: Before the measurement, it should
Measurement
Before the measurement, select the measurement model to ” Free Chlorine ” or ”
Total Chlorine ” First, the setting
ZERO setting for the liquid
-
Power ON the meter by pressing the ” Power Button ” ( 3-6, Fig. 1 ) once.
The Display ( 3-3, Fig. 1 ) will show the text ” Free ( TOTAL ) ” approx. 1 second. then show ” CAL 0 “, now the meter for the ZERO setting procedures. -
Fill the measurement liquid into the ” Testing bottle ” ( 3-11, 3-12, Fig. 1 ) .
-
Insert the ” Testing bottle ” to the bottom of ” Container ” ( 3-2, Fig. 1 ) completely.
Before the measurement, it should be enveloped in the ” Cover ” ( 3-1, Fig. 1 ) completely. -
Press the ” Zero Button ” ( 3-7, Fig. 1 ) once, the display will show the text ” tESt ” ( TEST )” with flashing, then show ” 0.00 “.
Liquid with the DPD powder -
Take the ” Testing bottle ” with the measurement liquid away from the ” Container ” ( 3-2, Fig. 1 ).
Fill the DPD powder into the ” Testing bottle ” with the measurement liquid. It should shake the at least 10 seconds to keep the solution under uniform conditions. Remark : -
For the Free Chlorine measurement, it should use the ” Free Chlorine DPD powder “.
-
For the Total Chlorine measurement, it should use the ” Total Chlorine DPD powder “.
-
Wait one minute approximately, then insert the ” Testing bottle ” to the bottom of the ” Container ” ( 3-2, Fig. 1 ) completely.
Before the measurement, it should be enveloped in the
” Cover ” ( 3-1, Fig. 1 ) completely. -
Press the” Test Button ” ( 3-5, Fig. 1 ) once, the display will show the text ” tESt ” ( TEST )” with flashing, wait approx. 10 seconds, the Display will show the measurement value.
Wash ( Reins ) the Testing bottle
- After the testing, it should wash ( reins ) the Testing Bottle by the Distill Water
Free/Total Chlorine mode selection
-
The meter is under the power-off conditions.
3-8, Fig. 1 ) at the same time ( not release ), then press the ” Hold Button ” ( 3-4, Fig. 1 ) ” REC Button ” ( press the ” Power Button ” ( 3-6, Fig. 1 ) will enter the selection mode to select the function of ” Free Chlorine ” or the ” Total Chlorine “. -
Press the ” TEST Button ” ( 3-5, Fig. 1 ) to select the desiring function ( Free Chlorine or Total Chlorine ), then the ” REC Button ” to confirm and save the selection mode into the memory.
Remark :
- For the swinging pool application, typical to select the ” Free Chlorine ” function.
- For the industrial water pollution application, typical to select the ” Total Chlorine ” function.
Data Hold
During the measurement, press the ” Hold Button ” ( 3-4, Fig. 1 ) once will
hold the me
-
- Press the ” Hold Button ” once again will release the data hold function.
Data Record ( Max., Min. reading )
-
The data record function records the maximum and minimum readings. Press the ” REC Button ” ( 3-8, Fig. 1 ) once to start the Data Record function and there will be a ” REC. ” symbol on the display.
-
With the ” REC. ” symbol on the display :
- a) Press the ” REC Button ” ( 3-8, Fig. 1 ) once, the ” REC. MAX. ” symbol along with the maximum value will appear on the display.
- b) If intend to delete the maximum value, just press the ” Hold Button ” ( 3-4, Fig. 1 ) once, then the display will show the ” REC. ” symbol only & execute the memory function continuously.
Press the ” REC Button ” ( 3-8, Fig. 1 ) again, the
-
REC. MIN. ” symbol along with the minimum value will appear on the display. If intend to delete the minimum value, just press the ” Hold Button ” ( 3-4, Fig. 1 ) once, then the display will show the ” REC. ” symbol only & execute the memory function continuously.
CALIBRATION PROCEDURE
- The meter can be calibrated under the following calibration points:
Free chlorine
* Zero
-
1.00 ppm
Total chlorine
* Zero -
1.00 ppm 2. The meter ship along with
-
Zero Chlorine standard solution ( CL-01 ) X 1 PC,
-
1.0 ppm Free Chlorine standard solution ( CF-01 ) X 1 PC.
-
1.0 ppm Total Chlorine standard solution ( CT-01) X 1 PC. as the standard accessories. 3. The complete calibration should be executed by following two solutions:
-
Zero Chlorine standard solution
-
1.0 ppm standard solution ( Free or Total ). 4. Zero chlorine calibration
-
Insert the ” Zero Chlorine standard solution into the ” Container ” ( 3-2, Fig. 1) and envelope it in the ” Cover ” ( 3-1, Fig. 1 ) completely, other procedures please refer to Chapter 4-1, 4-2.
-
Press the ” CAL Button ” ( 3-5, Fig. 1 ) continuously until the Display shows the text ” CAL ” and then releases the button. * During the display show text ” CAL “, press the ” CAL Button ” ( 3-5, Fig. 1 ), the display will show Now the meter is ready for the ” Zero Chlorine ” calibration
Press ” CAL Button ” ( 3-5, Fig. 1 ) once, the Display will show the following text with flashing ( approx. 10 seconds ). Then the Display will show: Now the meter is finished the ” Zero Chlorine ” calibration procedures and is ready for ” 1.00 ppm ” calibration procedures.- 1.00 ppm calibration
Refer to page 12, when finish the ” Zero Chlorine ” calibration procedures and the Display show The meter is ready for ” 1.0 ppm ” calibration.
Insert the ” 1.0 ppm standard solution into the ” Container ” ( 3-2 , Fig. 1) and envelope in the
” Cover ” ( 3-1, Fig. 1 ) completely, other procedures please refer to chapters 4-1, 4-2.
Press the ” CAL Button ” ( 3-5, Fig. 1 ) once, the Display will show the following text with flashing ( approx. 10 seconds ). Then the LCD display will return to the normal measurement screen, now the meter is finished the calibration procedures ( Zero, 1.0 ppm calibration ) completely and ready for the measurement.
- 1.00 ppm calibration
Remark:
The calibration procedures for the function of ” Total chlorine ” and the ”
Free chlorine ” are independent.
The complete calibration procedures for the function of ” Free chlorine ”
measurement should be executed by
- Zero Chlorine standard solution
- 1.0 ppm Total standard solution
BATTERY REPLACEMENT
-
When the left corner of the LCD display shows” “, it is necessary to replace the battery. However, in-spec. measurement may still be made for several hours after low battery indicator appears before the instrument become inaccurate.
-
Loss the ” Battery Cover Screws ” and slide the ” Battery Cover ” ( 3-9, Fig. 1 ) away from the instrument and remove the battery.
-
Replace with DC 1.5 V battery ( UM4, AAA,
Alkaline/heavy duty ) x 6 PCs, and reinstate the cover. -
Make sure the battery cover is secured after changing the battery.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>